Recall of Device Recall Trabecular Metal ( TM)

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Zimmer, Inc..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    66968
  • Event Risk Class
    Class 2
  • Event Number
    Z-0609-2014
  • Event Initiated Date
    2013-12-10
  • Event Date Posted
    2013-12-30
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2014-09-11
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Prosthesis, shoulder, non-constrained, metal/polymer cemented - Product Code KWT
  • Reason
    Zimmer is initiating a lot specific recall of the cpt hip stem, versys build-up block, and trabecular metal reverse glenosphere medical devices due to reports of the products being packaged in the incorrect outer carton. product labels and patient labels are correct.
  • Action
    Zimmer sent an Urgent Medical Device Recall letter dated December 10, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to review the notification and ensure affected personnel were aware of the contents. If affected product was found, it should be quarantines and their Zimmer representative notified. The Zimmer sales representative will remove the recalled product from their facility. Customers with questions were instructed to call 1-877-946-2761. .

Device

  • Model / Serial
    Part 00-4349-036-11, Lot 62393502
  • Product Classification
  • Device Class
    2
  • Implanted device?
    Yes
  • Distribution
    Worldwide Distribution -USA including OH, FL, WI, MI NJ, SC, CA, NC, PA, OR, WA, TN, KY, OK, KS, AZ, TX, CO, NV and Internationally to Germany. There have been 71 distribution events. Note that this product is lot controlled and may be distributed multiple times if it was returned and returned to the shelf. 7 units have been returned, including three through the complaint process, which leaves 64 affected units in the field subject to removal.
  • Product Description
    Trabecular Metal (TM) Reverse Glenosphere 36mm || Part 00-4349-036-11 || The reverse shoulder system is used in the treatment of the following: severe pain or significant disability in degenerative, rheumatoid, or traumatic disease of the glenohumeral joint; un-united humeral head fractures of long duration; irreducible 3 and 4-part proximal humeral fractures; avascular necrosis of the humeral head, or other difficult clinical management problems (such as a failed total shoulder arthropolastey or grossly rotator cuff deficient joint) where arthrodesis or resectional arthroplasty is not acceptable. 510(k) K130661
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Zimmer, Inc., 345 E Main St, Warsaw IN 46580-2746
  • Manufacturer Parent Company (2017)
  • Manufacturer comment
    “The safety of patients has always been, and continues to be, Zimmer Biomet’s top priority and it is our honor to be a leader in this industry for the past 90 years,” Zimmer Biomet told ICIJ in a statement. “We adhere to strict regulatory standards, and work closely with the FDA and all applicable regulatory agencies in each of our regions as part of our commitment to operating a first-rate quality management system across our global manufacturing network. The company added that it is focused on staying at the forefront of innovation and doing right by the millions of patients who rely on the company’s products.
  • Source
    USFDA