Recall of Device Recall The CARESCAPE" Monitor B850 version 1.0.1.13 and 1.0.2.1

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Ge Healthcare, Llc.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    55352
  • Event Risk Class
    Class 2
  • Event Number
    Z-1801-2010
  • Event Initiated Date
    2010-05-06
  • Event Date Posted
    2010-06-11
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2012-06-05
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Monitor, physiological, patient(with arrhythmia detection or alarms) - Product Code MHX
  • Reason
    Ge healthcare has become aware of a safety issue associated with the carescape" monitor b850 when monitoring ecg. the safety issue includes the potential for a delay in treatment.
  • Action
    The customers were sent on May 6, 2010, a GE Healthcare "Urgent Medical Device Correction" letter dated April 22, 2010. The letter was addressed to Healthcare Administrator / Risk Manager, Chief of Nursing, Director of Biomedical Engineering. The letter listed the Safety Issue, Affected Product Details, Product Correction, Safety Instructions and Contact Information. The customers were instructed to immediately discontinue use of the "Update Lead Set" function until the device software has been upgraded. If you have any questions regarding this medical device correction or identification of affected items, please contact Technical Support at 1-800-558-7044, or your local service representative.

Device

  • Model / Serial
    Parent SN : SED09451341GA SED09471459GA SED09471460GA SED09471461GA SED09471463GA SED09471464GA SED09471467GA SED09471468GA SED09471469GA SED09471470GA SED09471472GA SED09471474GA SED09471476GA SED09471480GA SED09471481GA SED09471486GA SED09471488GA SED09471491GA SED09471493GA SED09471496GA SED09471478GA SED09491583GA SED09491587GA SED09491595GA SED09491596GA SED09491597GA SED09491599GA SED09511674GA SED09511676GA SED10021713GA SED09310446GA SED09310448GA SED09310452GA SED09310453GA SED09310454GA SED09310456GA SED09310458GA SED09310461GA SED09310462GA SED09310468GA SED09310470GA SED09310477GA SED09310496GA SED09310498GA SED09310524GA SED09310525GA SED09320526GA SED09320527GA SED09320529GA SED09320530GA SED09320531GA SED09320532GA SED09320533GA SED09320534GA SED09320535GA SED09320536GA SED09320537GA SED09320539GA SED09320541GA SED09320542GA SED09320543GA SED09320545GA SED09320546GA SED09320547GA SED09320548GA SED09320549GA SED09320550GA SED09330585GA SED09340662GA SED09340663GA SED09370799GA SED09370802GA SED09370803GA SED09370804GA SED09370805GA SED09370806GA SED09370809GA SED09370811GA SED09370812GA SED09370813GA SED09370814GA SED09370815GA SED09370817GA SED09370819GA SED09370820GA SED09370821GA SED09370822GA SED09370825GA SED09370826GA SED09370827GA SED09370828GA SED09370829GA SED09370832GA SED09370833GA SED09390834GA SED09390835GA SED09390836GA SED09390838GA SED09390839GA SED09390840GA SED09390841GA SED09390842GA SED09390843GA SED09390847GA SED09390856GA SED09390857GA SED09390860GA SED09390874GA SED09390876GA SED09390877GA SED09390879GA SED09390882GA SED09390883GA SED09390890GA SED09400894GA SED09400896GA SED09400941GA SED09400945GA SED09400948GA SED09410954GA SED09420980GA SED09420990GA SED09420992GA SED09421001GA SED09451339GA SED09451340GA SED09451342GA SED09451344GA SED09451345GA SED09451346GA SED09451348GA SED09451369GA SED09471432GA SED09471462GA SED09471477GA SED09471494GA SED09491589GA SED09491592GA SED09491600GA SED09491601GA SED09491603GA SED09511667GA SED09511675GA SED10021683GA SED10021684GA SED10021707GA SED09370830GA SED09390846GA SED09390850GA SED09390852GA SED09390862GA SED09390863GA SED09390864GA SED09390865GA SED09390866GA SED09390869GA SED09390870GA SED09390873GA SED09390875GA SED09390878GA SED09390880GA SED09390881GA SED09390884GA SED09390885GA SED09390886GA SED09390887GA SED09390888GA SED09390889GA SED09390893GA SED09400897GA SED09400898GA SED09400899GA SED09400900GA SED09400901GA SED09400902GA SED09400903GA SED09400904GA SED09400905GA SED09400907GA SED09400909GA SED09400910GA SED09400911GA SED09400912GA SED09400913GA SED09400914GA SED09400915GA SED09400916GA SED09400917GA SED09400919GA SED09400920GA SED09400921GA SED09400922GA SED09400925GA SED09400943GA SED09400946GA SED09400949GA SED09400951GA SED09471490GA SED09451303GA SED09461399GA SED09461402GA SED09461406GA SED09461411GA SED09491519GA SED09491520GA SED09491522GA SED09491523GA SED09491524GA SED09491526GA SED09491527GA SED09491528GA SED09491534GA SED09491536GA SED09491537GA SED09491540GA SED09491543GA SED09491544GA SED09491545GA SED09491547GA SED09491548GA SED09491564GA SED09491565GA SED09491567GA SED09491569GA SED09491572GA SED09491575GA SED09491576GA SED09491577GA SED09491578GA SED09310466GA SED09310479GA SED09310482GA SED09310517GA SED09340628GA SED09340637GA SED09340647GA SED09370723GA SED09370725GA SED09370726GA SED09370727GA SED09370729GA SED09370730GA SED09370738GA SED09370739GA SED09370741GA SED09370744GA SED09370746GA SED09370747GA SED09370748GA SED09370749GA SED09370750GA SED09370752GA SED09370753GA SED09370798GA SED09421004GA SED09421006GA SED09421007GA SED09421009GA SED09421010GA SED09421011GA SED09431059GA SED09431063GA SED09431064GA SED09431065GA SED09431067GA SED09431068GA SED09431070GA SED09431071GA SED09431074GA SED09441078GA SED09441157GA SED09441158GA SED09441159GA SED09441160GA SED09441161GA SED09441162GA SED09441165GA SED09441168GA SED09441172GA SED09441173GA SED09441175GA SED09441177GA SED09441178GA SED09441180GA SED09441186GA SED09441189GA SED09441195GA SED09441198GA SED09441199GA SED09441200GA SED09441201GA SED09441202GA SED09441203GA SED09441204GA SED09441206GA SED09441207GA SED09441208GA SED09441209GA SED09441210GA SED09441211GA SED09441212GA SED09441277GA SED09441286GA SED09451311GA SED09451312GA SED09451313GA SED09451314GA SED09451315GA SED09451316GA SED09451318GA SED09451319GA SED09451322GA SED09451351GA SED09451366GA SED09451367GA SED09451368GA SED09451370GA SED09451371GA SED09451372GA SED09451373GA SED09451374GA SED09451375GA SED09451377GA SED09451378GA SED09461379GA SED09461380GA SED09461381GA SED09461382GA SED09461383GA SED09461384GA SED09461385GA SED09461386GA SED09461387GA SED09461388GA SED09461389GA SED09461390GA SED09461391GA SED09461392GA SED09461393GA SED09461394GA SED09461396GA SED09461397GA SED09461398GA SED09461400GA SED09461401GA SED09461403GA SED09461404GA SED09461405GA SED09461407GA SED09461408GA SED09461409GA SED09461410GA SED09461412GA SED09461414GA SED09461416GA SED09471431GA SED09471433GA SED09471440GA SED09471441GA SED09471442GA SED09471448GA SED09471450GA SED09471452GA SED09471456GA SED09471457GA SED09471479GA SED09471483GA SED09491579GA SED09491590GA SED09491594GA SED09491614GA SED09491615GA SED09491617GA SED09491619GA SED10021697GA SED10021714GA SED09451353GA SED09390837GA SED09390851GA SED09390855GA SED09390858GA SED09441081GA SED09310497GA SED09340645GA SED09340648GA SED09451310GA SED09471449GA SED09471454GA SED09471455GA SED09491530GA SED09491531GA SED09491532GA SED09340651GA SED09400927GA SED09400934GA SED09390849GA SED09390868GA SED09390871GA SED09370824GA SED09390845GA SED09431042GA SED09400926GA SED09400935GA SED09400936GA SED09400940GA SED09421012GA SED09421019GA SED09431025GA SED09431034GA SED09431036GA SED09431038GA SED09431039GA SED09431040GA SED09431043GA SED09431044GA SED09431056GA SED09431057GA SED09451347GA SED09340653GA SED09340661GA SED09390844GA SED09390853GA SED09390854GA SED09390891GA SED09431054GA SED09431055GA SED09431058GA SED09431060GA SED09441255GA SED09441257GA SED09441259GA SED09441261GA SED09441264GA SED09441265GA SED09441267GA SED09441268GA SED09441272GA SED09441273GA SED09441275GA SED09441280GA SED09441282GA SED09441288GA SED09441291GA SED09441294GA SED09451302GA SED09491535GA SED09491538GA SED09390861GA SED09400931GA SED09400933GA SED09400937GA SED09400953GA SED09431033GA SED09451323GA SED09451326GA SED09451330GA SED09451331GA SED09451336GA SED09451338GA SED09340664GA SED09451333GA SED09280427GA SED09310463GA SED09340606GA SED09340643GA SED09340652GA SED09400918GA SED09441260GA SED09461395GA SED09461417GA SED09461418GA SED09471420GA SED09471421GA SED09471423GA SED09471426GA SED09471430GA SED09471435GA SED09471443GA SED09471444GA SED09471445GA SED09471446GA SED09471447GA SED09441155GA SED09441156GA SED09441163GA SED09441164GA SED09441166GA SED09441167GA SED09441169GA SED09441170GA SED09441171GA SED09441174GA SED09441176GA SED09441184GA SED09441185GA SED09441187GA SED09441188GA SED09441190GA SED09441191GA SED09491521GA SED09491525GA SED09491539GA SED09491566GA SED09491568GA SED09491570GA SED09491571GA SED09491573GA SED09491574GA SED09390867GA SED09390872GA SED09400906GA SED09370743GA SED09370818GA SED09370823GA SED09370831GA SED09390848GA SED09400929GA SED09400930GA SED09400932GA SED09400938GA SED09400939GA SED09400950GA SED09400952GA SED09421013GA SED09421015GA SED09421016GA SED09421017GA SED09421021GA SED09431022GA SED09431023GA SED09431024GA SED09431027GA SED09431028GA SED09431029GA SED09431031GA SED09431037GA SED09431045GA SED09431046GA SED09431047GA SED09431048GA SED09431049GA SED09431050GA SED09431051GA SED09431052GA SED09431053GA SED09431062GA SED09441256GA SED09441258GA SED09441262GA SED09441263GA SED09441266GA SED09441269GA SED09441270GA SED09441271GA SED09441274GA SED09441276GA SED09441278GA SED09441279GA SED09441281GA SED09441283GA SED09441284GA SED09441285GA SED09441287GA SED09441289GA SED09441290GA SED09441292GA SED09441293GA SED09451299GA SED09451301GA SED09451306GA SED09451307GA SED09451308GA SED09451320GA SED09451321GA SED09451325GA SED09451327GA SED09451328GA SED09451329GA SED09451332GA SED09451334GA SED09451335GA SED09451337GA SED09461413GA SED09461415GA SED09471419GA SED09471422GA SED09471424GA SED09471425GA SED09471427GA SED09471428GA SED09471429GA SED09471434GA SED09471436GA SED09471437GA SED09471438GA SED09471439GA SED09471485GA SED09471497GA SED09491586GA SED09511672GA SED10021685GA SED10021686GA SED10021689GA SED10021693GA SED10021695GA SED09280439GA SED09310459GA SED09310460GA SED09310465GA SED09310485GA SED09310506GA SED09310519GA SED09330569GA SED09330579GA SED09330583GA SED09330599GA SED09330600GA SED09330601GA SED09330602GA SED09330603GA SED09330604GA SED09340607GA SED09340608GA SED09340612GA SED09340613GA SED09340615GA SED09340624GA SED09340625GA SED09340626GA SED09340630GA SED09340633GA SED09340634GA SED09340636GA SED09340639GA SED09350666GA SED09350667GA SED09350674GA SED09350675GA SED09350691GA SED09350693GA SED09350694GA SED09350695GA SED09350696GA SED09350699GA SED09350700GA SED09350715GA SED09350716GA SED09350717GA SED09350718GA SED09350721GA SED09370722GA SED09370724GA SED09370728GA SED09370731GA SED09370732GA SED09370733GA SED09370734GA SED09370735GA SED09370737GA SED09370740GA SED09370745GA SED09370751GA SED09370794GA SED09370795GA SED09370796GA SED09370797GA SED09370800GA SED09370801GA SED09280429GA SED09330605GA SED09340614GA SED09340623GA SED09340632GA SED09340635GA SED09340638GA SED09340641GA SED09340644GA SED09340656GA SED09370742GA SED09370807GA SED09370808GA SED09370810GA SED09370816GA SED09390892GA SED09400895GA SED09400908GA SED09400924GA SED09410955GA SED09410956GA SED09410959GA SED09410960GA SED09410961GA SED09410962GA SED09410963GA SED09410964GA SED09410965GA SED09410966GA SED09410967GA SED09410968GA SED09410969GA SED09410970GA SED09410971GA SED09410972GA SED09410973GA SED09410974GA SED09410975GA SED09410976GA SED09410977GA SED09410979GA SED09420988GA SED09421014GA SED09421018GA SED09421020GA SED09431026GA SED09431030GA SED09431032GA SED09431035GA SED09441196GA SED09451300GA SED09451304GA SED09451305GA SED09451309GA SED09451317GA SED09451324GA SED09471451GA SED09471453GA SED09471458GA SED09471482GA SED09471489GA SED09491580GA SED09491581GA SED09491588GA SED09491591GA SED09491606GA SED09491620GA SED09511664GA SED09511665GA SED09511669GA SED09511670GA SED09511671GA SED09511673GA SED09511677GA SED09511678GA SED10021690GA SED10021692GA SED09340642GA SED09340660GA SED09340665GA SED09340659GA SED09350668GA SED09350669GA SED09350670GA SED09350671GA SED09350672GA SED09350673GA SED09350676GA SED09350677GA SED09350690GA SED09350692GA SED09350697GA SED09350698GA SED09350701GA SED09350712GA SED09350713GA SED09350714GA SED09350719GA SED09350720GA SED09370736GA SED09340611GA SED09340616GA SED09340618GA SED09340619GA SED09340620GA SED09340622GA SED09340646GA SED09400923GA SED09400942GA SED09400944GA SED09400947GA SED09340621GA SED09340629GA SED09340640GA SED09340654GA SED09340655GA SED09491626G SED09491635G SED09491638G SED09511642G SED09511644G SED09511645G SED09511650G SED09511653G SED09511655G SED09511656G SED09511657G SED09491533GA SED09491541GA SED09491546GA
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    Worldwide distribution: USA including CO, MA, NC, WI, and countries of UNITED KINGDOM, SWITZERLAND, SWEDEN, SPAIN, SINGAPORE, PORTUGAL, NORWAY NETHERLANDS, ITALY, INDIA, GERMANY, FRANCE, FINLAND, BELGIUM, and AUSTRALIA.
  • Product Description
    GE Healthcare, CARESCAPE" Monitor B850, Software version 1.0.1.13 and 1.0.2.1 || The CARESCAPE " Monitor B850 is a multi-parameter high acuity patient monitor intended for use in multiple areas within a professional healthcare facility. The CARESCAPE Monitor B850 is intended for use on adult, pediatric, and neonatal patients and on one patient at a time. The CARESCAPE Monitor B850 system is indicated for monitoring of Hemodynamic (including ECG, ST Segment, Arrhythmia Detection, ECG Diagnostic Analysis and Measurement, Invasive Pressure, Noninvasive Blood Pressure, Pulse Oximetry, Cardiac Output, Temperature, Impedance Respiration and SvO2 (Mixed Venous Oxygen Saturation)), Airway Gases (Fi/Et C02, 02, N20 and Anesthetic Agent), Spirometry, Gas Exchange (02 Consumption (VO2), C02 production (VCO2), energy expenditure (EE), and respiratory quotient (RQ)) and neurophysiological (including electroencephalography (EEG), Entropy, Bispectral Index (BIS) and Neuromuscular Transmission (NMT) Monitoring) status. The CARESCAPE Monitor B850 provides alarms, trends, snapshots and events, and calculations and can be connected to displays, printers and recording devices. The CARESCAPE Monitor B850 can be a stand-alone monitor or interfaced to other devices. It can also be connected to other monitors for bed to bed viewing and to data management software devices via a network. The CARESCAPE Monitor B850 is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility. In addition to the healthcare practitioner, the CARESCAPE Monitor B850 is designed to provide configuration and troubleshooting capabilities to qualified service personnel. The CARESCAPE Monitor B850 is not intended for use during MRI.
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Ge Healthcare, Llc, 3000 N Grandview Blvd, Waukesha WI 53188-1615
  • Manufacturer Parent Company (2017)
  • Source
    USFDA