Recall of Device Recall syngo.plaza

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Siemens Medical Solutions USA, Inc.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    62679
  • Event Risk Class
    Class 2
  • Event Number
    Z-2275-2012
  • Event Initiated Date
    2012-07-18
  • Event Date Posted
    2012-08-27
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2014-01-24
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    System, image processing, radiological - Product Code LLZ
  • Reason
    There is a potential malfunction in which a miscalculation occurs for images with pixel depth values greater that 12 bit in certain functions.
  • Action
    The firm initiated their recall/correction of this device on July 18, 2012 by sending a "Customer Safety Advisory Notice" letter, dated July 12, 2012, to their consignees. The letter identified the affected product along with the malfunction. It also discussed the steps the user can take to avoid the issue. Users were asked to not use the mentioned functions for images from those modalities until the issue was resolved and the system was updated with the software version containing the solution. Users were to contact their local Siemens Service in order to plan for the updated software version that will resolve the issue. Additionally, users should place the safety advisory notice in the system's instructions for use, and personnel should be instructed accordingly and maintain awareness. If the device was sold and is no longer in the users' possession, the notice should be forwarded to the new owner and the firm should be notified of the new owner.

Device

  • Model / Serial
    Serial Numbers: 100263, 100401, 100177, 100138, 100174, 100219, 100406, 100249, 200040, 100201, 100198, 100181, 100301, & 100146.
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    Nationwide Distribution -- CA, FL, GA, MA, MN, MO, NJ, NY, OH, WA, and WI.
  • Product Description
    syngo.plaza, Model number 10592457. || Intended use: radiological image processing system.
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Siemens Medical Solutions USA, Inc, 51 Valley Stream Pkwy, Malvern PA 19355
  • Manufacturer Parent Company (2017)
  • Source
    USFDA