Recall of Device Recall Symbiq One Channel Infuser

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Hospira Inc..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    63994
  • Event Risk Class
    Class 2
  • Event Number
    Z-0305-2017
  • Event Initiated Date
    2013-02-04
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2016-12-15
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
    ICU Medical, Inc. acquired from Pfizer in early 2017, Hospira Infusion Systems, the portion of Hospira dedicated to develop infusion pumps.
  • Extra notes in the data
    Pump, infusion - Product Code FRN
  • Reason
    Symbiq infusers have the potential to experience a white screen during titration of a keep vein open (kvo) delivery. this can allow the clinician to select the "next" button prior to completion of the programming. if this error occurs, a delay in therapy could occur and potentially result in significant injury.
  • Action
    Hospira sent an Urgent Device Field Correction notice dated February 4, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to follow the steps provided in the letter while Hospira is developing a software change. Customers were asked to complete the attached reply form and return it via fax to the number on the form. Customers with questions were instructed to contact Stericycle at 1-888-201-9332. For questions regarding this recall call 224-212-4892.

Device

  • Model / Serial
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    Worldwide Distribution - US (nationwide) and Canada
  • Product Description
    Symbiq One Channel Infuser; an Rx medical device infusion pump used to administer I.V. fluids. List number 16026. Note that Hospira retired this pump June 30, 2015. || The Hospira Symbiq Infusion System is intended for the delivery of fluids, solutions, drugs, agents, nutritionals, electrolytes, blood, and blood products via parenteral, enteral, intravenous, intra-arterial, subcutaneous, epidural, or irrigation routes of administration.
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Hospira Inc., 275 N Field Dr, Lake Forest IL 60045-2579
  • Manufacturer Parent Company (2017)
  • Source
    USFDA