Recall of Device Recall Stryker Orthopaedics Restoration Gap Plate Screw

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Stryker Howmedica Osteonics Corp..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    52713
  • Event Risk Class
    Class 3
  • Event Number
    Z-2139-2009
  • Event Initiated Date
    2009-06-26
  • Event Date Posted
    2009-09-14
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2010-07-27
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Prosthesis, hip, semi-constrained, uncemented, metal/polymer, non-porous, calicum-phosphate - Product Code MEH
  • Reason
    Raw material supplied by a third party used to manufacture various medical devices does not comply with metallurgical requirements outlines in astm standard for titanium surgical implants.
  • Action
    Stryker Orthopaedics issued "Urgent Product Recall" letters via Federal Express priority overnight beginning June 26, 2009 to Branches, Hospital Risk Management and Chief of Orthopaedics and Surgeons. All users are requested to examine their inventory, retrieve all affected product and return for reconciliation to Stryker Orthopaedics. Users are also requested to complete the enclosed Product Accountability Form and fax with necessary information to the firm at 1-201-831-6069. For further information, contact Stryker Orthopaedics at 1-201-831-5970.

Device

  • Model / Serial
    Catalog number: 2080-0040; GAP PLATE SCREWS PTDMKE 27-Jun-2013 GAP PLATE SCREWS T0MMKE 01-Jul-2013 GAP PLATE SCREWS PT9MKE 02-Jul-2013 GAP PLATE SCREWS 34HMLE 09-Jul-2013 GAP PLATE SCREWS 34EMLE 14-Jul-2013 GAP PLATE SCREWS JKRMLE 23-Jul-2013 GAP PLATE SCREWS JKPMLE 23-Jul-2013 GAP PLATE SCREWS NL3MME 27-Aug-2013 GAP PLATE SCREWS NL6MME 27-Aug-2013 GAP PLATE SCREWS NL5MME 27-Aug-2013 GAP PLATE SCREWS NL4MME 28-Aug-2013 GAP PLATE SCREWS 7E4MNE 17-Sep-2013 GAP PLATE SCREWS 7E5MNE 17-Sep-2013 GAP PLATE SCREWS MERN50 05-Dec-2013 GAP PLATE SCREWS MERN4Y 05-Dec-2013 GAP PLATE SCREWS MET81P 18-Dec-2013 GAP PLATE SCREWS MET81T 18-Dec-2013 GAP PLATE SCREWS MET81V 18-Dec-2013 GAP PLATE SCREWS MET81R 22-Dec-2013 GAP PLATE SCREWS MET969 30-Dec-2013 GAP PLATE SCREWS MET967 30-Dec-2013 GAP PLATE SCREWS MET9H3 30-Dec-2013 GAP PLATE SCREWS MHA5LL 18-Jan-2014 GAP PLATE SCREWS MHA5LM 18-Jan-2014 GAP PLATE SCREWS MHA5LN 18-Jan-2014 GAP PLATE SCREWS MHA5LK 18-Jan-2014 GAP PLATE SCREWS MHA779 18-Jan-2014
  • Product Classification
  • Device Class
    2
  • Implanted device?
    Yes
  • Distribution
    Worldwide Distribution
  • Product Description
    Stryker Orthopaedics Restoration Gap Plate Screw; || Dia. 6.5 mm, Lnth 40 mm; || Made in USA, Sterile; || Howmedica Osteonics Corp. || 325 Corporate Drive || Mahwah, NJ 07430
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Stryker Howmedica Osteonics Corp., 325 Corporate Drive, Mahwah NJ 07430
  • Manufacturer Parent Company (2017)
  • Source
    USFDA