Recall of Device Recall Stryker inTouch Critical Care

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Stryker Medical Div. of Stryker Corporation.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    45445
  • Event Risk Class
    Class 2
  • Event Number
    Z-0356-2008
  • Event Initiated Date
    2007-11-02
  • Event Date Posted
    2007-12-08
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2010-10-27
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Hospital Bed - Product Code FNL
  • Reason
    Multiple defects: 1) beds exhibit reverse motions when the fowler, gatch or foot section is activated 2) a battery charging defect which may result in a depletion of battery power 3) the touch screen on the footboard freezing, resulting in controls to powered, pressure-relief mattresses being difficult to access 4) beds equipped with zoom drive systems may have an overdrive problem which could re.
  • Action
    Consignees were notified via an Urgent Medical Device Correction letter dated 11/2/07, and instructed to remove the beds from service if they display any unintended motions and informed that one of the firm's field service representatives will visit the facility to make corrections to all beds under recall.

Device

Manufacturer

  • Manufacturer Address
    Stryker Medical Div. of Stryker Corporation, 3800 E. Centre Ave., Portage MI 49002
  • Source
    USFDA