Recall of Device Recall STERIS 4085, 5085 & 5085SRT

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Steris Corporation.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    55850
  • Event Risk Class
    Class 2
  • Event Number
    Z-1941-2010
  • Event Initiated Date
    2010-05-14
  • Event Date Posted
    2010-06-30
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2012-02-24
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Surgical Tables - Product Code FQO
  • Reason
    Steris has learned that a limited quantity of manifold assemblies were manufactured by our supplier with incorrect washers and springs in the valves of the manifold. this supplier error has the potential to cause: a) an inability to articulate/move a table section when commanded by the operator; b) two different articulations occurring when a single articulation is commanded, or c) that the table.
  • Action
    On May 14, 2010 the firm sent FCL letters to their customers notifying them of the problem. They informed their customers that a service representative will contact them to arrange correction of the unit.

Device

  • Model / Serial
    Model #: 4085, 5085 & 5085SRT Serial Numbers: 0400710041, 0400710042, 0400710043, 0400710044, 0401510077, 0401510080, 0401510081, 0401510082, 0401510086, 0401510087, 0401510088, 0401510089, 0401510090, 0401510092, 0401910133, 0402010001, 0402010002, 0402010003, 0403210095, 0403210096, 0403210096, 0403210097, 0403210098, 0403610046, 0403610047, 0403610048, 0403610049, 0403610050, 0403610053, 0403610054, 0403610055, 0403610057, 0403610061, 0403610062, 0403610063, 0404610061, 0404610062, 0404610063, 0404610063, 0404610064, 0404610064, 0404610065, 0404610065, 0404610070, 0404610070, 0404610071, 0404610071, 0404610072, 0404610072, 0405410037, 0405410038, 0405410039, 0432709078, 0432709079, 0432709079, 0432709080, 0432709081, 0432709082, 0432709083, 0432709083, 0432709084, 0432709084, 0433609001, 0433609001, 0433609002, 0433609002, 0433609005, 0433609005, 0433609062, 0434309100, 0434309101, 0434309103, 0434809001, 0434809003, 0434809004, 0434809005, 0434809006, 0434809007, 0434809007, 0434809008, 0434809008, 0434809012, 0434809013, 0434809014, 0434809014, 0434809015, 0434809015, 0434809016, 0434809016, 0434809017, 0434809017, 0434809018, 0434809018, 0434809019, 0434809019, 0436309084, 0436309085, 0436509122, 0436509123, 0436509124 & 0436509124.
  • Product Classification
  • Device Class
    1
  • Implanted device?
    No
  • Distribution
    Worldwide distribution: USA, Australia, India, Korea & New Zealand
  • Product Description
    STERIS 4085, 5085 & 5085SRT Surgical Tables
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Steris Corporation, 5960 Heisley Rd, Mentor OH 44060
  • Manufacturer Parent Company (2017)
  • Source
    USFDA