Recall of Device Recall Stago IVD, Liquid AntiXa

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Diagnostica Stago, Inc..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    69070
  • Event Risk Class
    Class 3
  • Event Number
    Z-2650-2014
  • Event Initiated Date
    2014-07-18
  • Event Date Posted
    2014-09-16
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2015-01-16
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Assay, heparin - Product Code KFF
  • Reason
    Diagnostica stago inc. received some customer complaints regarding a misreading of reagent 1 (substrate) barcode on sta analyzers.
  • Action
    Diagnostica Stago sent a "Quality Information" letter/Customer Acknowledgement form to its customers. The letter identified the product, problem, and actions to be taken by the customers. For additional information and further support contact the Diagnostica Stago Hotline at 800-725-0607.

Device

  • Model / Serial
    510(k) #K111822, Lot #112366, Part # REF 00311US
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    Distributed USA (nationwide) and the country of Canada.
  • Product Description
    Stago IVD, Liquid Anti-Xa, NDC 3607450003116. || The STA - Liquid Anti-Xa kits are intended for use with STA-R, STA Compact and STA Satellite Analyzers for the quantitative determination of the plasma levels of unfractionated (UFH) and low molecular weight (LMWH) heparins by measuring their anti-Xa activity on antithrombin in a competitive assay using a synthetic chromogenic substrate.
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Diagnostica Stago, Inc., 5 Century Dr, Parsippany NJ 07054-4607
  • Source
    USFDA