Recall of Device Recall Siemens syngo Dynamics

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Siemens Medical Solutions USA, Inc.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    67559
  • Event Risk Class
    Class 2
  • Event Number
    Z-1216-2014
  • Event Initiated Date
    2014-01-27
  • Event Date Posted
    2014-03-18
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2015-03-16
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Picture Archiving and Communication System (PACS) - Product Code LLZ
  • Reason
    Potential patient data mixup. improper error handling could allow for two different patients to be displayed in syngo dynamics. images shown for one patient and reporting system shown for a different patient, leading to the potential for misdiagnosis.
  • Action
    A recall notification letter, dated January 27, 2014, was sent to end Users. Customer Safety Advisory Letter SY016/14/S. Siemens is preparing a Service Patch modification of syngo Dynamics v9.5 which will be available in March 2014.

Device

  • Model / Serial
    Model nos.10091804, 10091805, 10091806, 10091807, 10091673. syngo Dynamics Version 9.5 using Sensis integration.
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    Nationwide distribution to MN, AR, MA, TX, NJ, WV, GA, CA, PA, OH, MI, WI, TN, NY, KS, CT, NY, MO, IL, IN, OR, IA, and WA.
  • Product Description
    syngo Dynamics Picture Archiving and Communication System (PACS). Model numbers 10091804, 10091805, 10091806, 10091807, 10091673
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Siemens Medical Solutions USA, Inc, 51 Valley Stream Pkwy, Malvern PA 19355
  • Manufacturer Parent Company (2017)
  • Source