Recall of Device Recall ShapeMatch Cutting Guides

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Stryker Howmedica Osteonics Corp..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    64308
  • Event Risk Class
    Class 1
  • Event Number
    Z-1052-2013
  • Event Initiated Date
    2012-11-08
  • Event Date Posted
    2013-04-12
  • Event Status
    Open, Classified
  • Event Country
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Prosthesis, knee, patello/femorotibial, semi-constrained, uncemented, porous, coated, polymer/metal/ - Product Code MBH
  • Reason
    With regards to shapematch cutting guides using mri imaging, stryker orthopaedics has become aware that there are potential issues associated with internal processes for planning cases and producing shapematch cutting guides.
  • Action
    Stryker sent Product Correction Bulletins on 11/8/2012 dated 11/9/2012 and 11/19/2012 to all affected customers. Stryker sent an updated Urgent Medical Device Product Notification letters and Acknowledgement letters dated 1/24/2013 through 1/28/2013 to all affected customers via Fed Ex with return receipt: The letter identified the affect product, problem, potential hazards, risk mitigating factors and actions to be taken. For questions call 201-831-6878. Stryker issued press release on 4/10/2013.

Device

  • Model / Serial
    Catalog No: TR7100-L and TR7100-R Lot Code: All
  • Product Classification
  • Device Class
    2
  • Implanted device?
    Yes
  • Distribution
    USA Nationwide Distribution
  • Product Description
    ShapeMatch Cutting Guides || Product Usage: The ShapeMatch Cutting Guides are intended to be used as patient-specific surgical instrumentation to assist in the positioning of total knee arthroplasty components intraoperatively and in guiding the marking of bone before cutting.
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Stryker Howmedica Osteonics Corp., 325 Corporate Drive, Mahwah NJ 07430
  • Manufacturer Parent Company (2017)
  • Source
    USFDA