Recall of Device Recall Seno Advantage 1.x, Seno Advantage 2.0, Seno Advantage 2.1

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by GE Healthcare, LLC.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    57077
  • Event Risk Class
    Class 2
  • Event Number
    Z-0476-2011
  • Event Initiated Date
    2009-08-28
  • Event Date Posted
    2010-11-26
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2012-06-05
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    System, image processing, radiological - Product Code LLZ
  • Reason
    The measurement values provided by the ge workstation may be incorrect when applied to magnification images acquired on non-ge digital mammography systems. if not noticed by the caregiver, this could lead to an overestimate of the size of the breast lesion that may impact patient treatment and safety. the measurement values for ge images are not affected.
  • Action
    GE Healthcare issued "Urgent Medical Device Correction" letters dated August 28, 2009 or August 31, 2009 to either Seno Advantage Customers (Hospital Administrators/Risk Mangers, Radiology Department Managers), and Image Diagnostic Customers (Hospital Administrators/Risk Managers, Radiology Department Managers). The letters described the Safety Issue, Affected Product Details, safety Instructions, Product correction and Contact Information. GE will linstall new software with the release of Field Modification Instruction 12129. Customers can contact GE Healthcare about this issue at 800-437-1171.

Device

  • Model / Serial
    00000073186VE9 00000073190VE1 00000073191VE9 00000073194VE3 00000073220VE6 00000073246VE1 00000073252VE9 00000073260VE2 00000073263VE6 00000073264VE9 00000073267VE7 00000075112VE3 00000075136VE2 00000075146VE1 00000092523VE0 00000092528VE9 00000092531VE3 00000092534VE7 00000092537VE0 00000092544VE6 00000092547VE9 00000092551VE1 00000092558VE6 00000092563VE6 00000092564VE4 00000092565VE1 00000092567VE7 00000092571VE9 00000092574VE3 00000092584VE2 00000092587VE5 00000092591VE7 00000092593VE3 00000092596VE6 00000092599VE0 00000092601VE4 00000092603VE0 00000092604VE8 00000092613VE9 00000092614VE7 00000092615VE4 00000094937VE0 00000094939VE6 00000094940VE4 00000094944VE6 00000094950VE3 00000094952VE9 00000094954VES 00000094955VE2 00000094970VE1 00000102693VE9 00000102702VE8 00000102703VE6 00000102710VE1 00000102720VE0 00000102721VE8 00000102722VE6 00000102727VE5 00000102729VE1 00000102750VE7 00000102755VE6 00000102757VE2 00000102781VE2 00000103336VE4 00000103366VE1 00000425154BU2 00000425548BU5 00000425549BU3 00000449452BU2 00000487114BU1 00000487121BU6 00000487144BU8 00000487147BU1 0000FRB4050CPR 0000FRB4240CMQ 0000FRB43705LG 11 00000092529VE7 00000092543VE8 00000092545VE3 00000092557VE8 00000092559VE4 00000092560VE2 00000092562VE8 00000092569VE3 00000092570VE1 00000092576VE8 00000092582VE6 00000092583VE4 00000092588VE3 00000092589VE1 00000092592VE5 00000092594VE1 00000092597VE4 00000092612VE1 00000092617VE0 00000094935VE4 00000094938VE8 00000094946VE1 00000094954VE5 00000094963VE6 00000094973VE5 00000102686VE3 00000102715VE0 00000102716VE8 00000102718VE4 00000102725VE9 00000102732VE5 00000102743VE2 00000103333VE1 00000103359VE6 00000103360VE4 00000103365VE3 00000103368VE7 00000103371VE1 00000103375VE2 00000103376VE0 00000103378VE6 00000103399VE2 00000103400VE8 00000103406VE5 00000103408VE1 00000092539VE6 00000102746VE5 00000102766VE3 00000102772VE1 00000102774VE7 00000428047BU5 00000102683VE0 00000102691VE3 00000102726VE7 00000102777VE0 00000102778VE8 00000103340VE6 00000103363VE8 00000487118BU2 00000092549VE5 00000092521VE4 00000092581VE8 00000094941VE2 00000102764VE8 00000103350VE5 00000102717VE6 00000425546BU9 00000075121VE4 00000087617VE7 00000094943VE8 00000102739VE0 00000103370VE3 00000073189VE3 00000073208VE1 00000073221VE4 00000073224VE8 00000073237VE0 00000073251VE1 00000073253VE7 00000073261VE0 00000073262VE8 00000073271VE9 00000073281VE8 00000075110VE7 00000075118VE0 00000075147VE9 00000092542VE0 00000092550VE3 00000092555VE2 00000092595VE8 00000092605VE5 00000092609VE7 00000094958VE6 00000094971VE9 00000102706VE9 00000102714VE3 00000102719VE2 00000102736VE6 00000102738VE2 00000102742VE4 00000102769VE7 00000102773VE9 00000102780VE4 00000103341VE4 00000103342VE2 00000103349VE7 00000103354VE7 00000103356VE2 00000103401VE6 00000103403VE2 00000103404VE0 00000103405VE7 00000103407VE3 00000103414VE9 00000103417VE2 00000425547BU7 00000102697VE0 00000092540VE4 00000092610VE5 00000094949VE5 00000094965VE1 00000102735VE8 00000102779VE6 00000103334VE9 00000103352VE1 00000103409VE9 00000103411VE5 00000102771VE3 00000103374VE5 00000092535VE4 00000092616VE2 00000094933VE9 00000094934VE7 00000103348VE9 00000103373VE7 00000075115VE6 00000075150VE3 00000092554VE5 00000094964VE4 00000094977VE6 00000094979VE2 00000102687VE1 00000102711VE9 00000102734VE1 00000102765VE5 00000102767VE1 00000102775VE4 00000102782VE0 00000103335VE6 00000103339VE8 00000103347VE1 00000103351VE3 00000103357VE0 00000425156BU7 00000487109BU1 00000487178BU6 00000102685VE5 00000102763VE0 00000094948VE7 00000102761VE4 00000103412VE3 00000487110BU9 00000487132BU3 00000092572VE7 00000073248VE7 00000073250VE3 00000092598VE2 00000092611VE3 00000102737VE4 00000102741VE6 00000102744VE0 00000103416VE4 00000073203VE2 00000073240VE4 00000075113VE1 00000087606VE0 00000087610VE2 00000092546VE1 00000094936VE2 00000094947VE9 00000102740VE8 00000102749VE9 00000102758VE0 00000103364VE6 00000073280VE0 00000073198VE4 00000102733VE3 00000102707VE7 00000487161BU2     0000CZC7065H15 000000U8162054 000000B8150003 000000U8162005 00002UA5421D6R 0000CZC93668ZC 0000CZC9280BKQ 0000CZC9280BKF 0000CZC9280BKP 0000CZC93668ZG 0000CZC931BJM1 0000CZC8236LSF 0000CZC9064P9M 0000CZC9064PB0 0000CZC9064PB1 0000CZC9064PB4 0000CZC9064P9N 0000CZC9064P9Q 0000CZC9064P9P 0000CZC9064PCP 0000CZC9064PCM 00000080151HP4 00000079777HP9 0000CZC81370NR 000000MI000375 00000088820HP6 0000CZC931BJLX 0000CZC9280BKL 0000CZC9280BKL 0000CZC9316RLM 00000080122HP5 00000079479HP2 00000089210HP9 0000CZC92457XD 0000CZC6502YB3 0000CZC8295TP7 0000CZC8326DR2 0000FRB4280DKC 0000CZC8207NVN 0000CZC8303V6S 0000CZC8295TP5 0000CZC931BJLT 0000CZC93668ZJ 0000CZC80602B1 000000B8150015 000000D8145020 0000CZC9316RLJ 0000CZC8207NVT 0000CZC8207NVY 0000CZC8191GG4 0000CZC81370NT 0000CZC81370NP 0000CZC81370NS 0000CZC81370NV 0000CZC81370P4 00002UA5402D6H 0000CZC813012P 00000088943HP6 0000CZC813012X 0000CZC8130131 0000CZC8120YCW 0000CZC7414GSF 0000CZC6303F8S 000000MI000382 000000L7332001 00000089195HP2 00000089214HP1 00000509654BU0 00002UA70305Q3 00000079744HP9 00000089173HP9 00000079949HP4 0000CZC80602B8 0000CZC7455QF6 0000CZC7521JWF 00000089307HP3 00000089339HP6 0000CZC6323QJY 00000089060HP8 0000CZC733IWV5 0000FRB3430V5V 00000089288HP5 000000MI000343 0000CZC7321Z52 00002UA608K4QH 00000089052HP5 0000CZC7280B2J 0000CZC7330YXC 0000CZC7395WWY 00000088934HP5 00000088851HP1 0000CZC7011GB3 00000088719HP0 0000CZC7333J7W 00000088828HP9 0000CZC7300YXR 00000089220HP8 0000CZC74003D9 00000534598BU8 00000089027HP7 00000088290HP2 00000089146HP5 00000089134HP1 0000CZC7280F7V 00000088854HP5 00000089292HP7 00000088682HP0 00000089189HP5 0000CZC7280F80 0000CZC72449PM 0000CZC7333J7L 0000CZC7330YXX 0000CZC7330YXK 0000CZC7330YXW 0000CZC7321Z5G 0000CZC8066YCX 0000CZC72544NQ 0000CZC72544NW 0000CZC72544NY 0000CZC7253SH9 0000CZC7253SHD 0000CZC7253SHJ 0000CZC7253SH5 0000CZC7280F7M 0000CZC7280F7P 0000CZC7114TZY 0000C7C6512VKG 0000CZC70416Y2 0000CZC7113CLC 0000FRB4460R03 0000CZC6342R34 0000CZC7262J7D 0000CZC72449PB 000000M1001468 0000CZC72449P9 00000080144HP9 0000CZC6510K9M 0000CZC7110RG5 0000CZC7182HNG 0000CZC7182HNW 0000CZC7182LW0 0000CZC7182LW1 000000M1000320 000000MI000377 00000080194HP4 0000CZC71108XV 0000CZC7113CL7 00000080149HP8 00000080009HP4 0000CZC7110RG3 00000509668BUO 0000FRB4050CRB 000000ANT22001 0000CZC7090XLR 00000534666BU3 0000CZC7072JXY 0000CZC7051FVF 0000CZC703221Q 0000CZC7012HHG 00000534595BU4 0000CZC7023M5M 00000534533BU5 0000CZC6512VKM 0000CZC7012HHJ 00000534616BU8 00000534661BU4 0000CZC6482BJ9 00000534597BU0 00000534590BU5 00000534621BU8 00000534637BU4 00000534591BU3 00000534601BU0 00000534631BU7 0000CZC9316RLK 0000CZC641057S 0000CZC6342R3B 0000CZC6300VHF 00000509587BU2 0000CZC6300NQQ 00000487144BU8 00000487120BU8 00000509579BU9 00002UA63102FC 00000035279AW9 0000CZC6350M9T 00000080266HPO 0000CZC6142043 0000CZC6084446 0000C2C55207GY 0000CZC55207GQ 00000020924AW7 00002UA54410CD 00002UA54412X6 00002UA5440KTY 00002UA5421D3R 00002UA5421D6Z 0000FRB437042V 00002UA535111Z 00002UA5420VPJ 00002UA52918Y8 0000FRB411000Q 00002UA5351123 00000103366VE1 0000FRB4460QZM 0000FRB4280DKM 0000FRB44107Q8 00000073252VE9 0000FRB44107PZ 00000142239HP3 0000FRB4370432 0000USU4440136 0000FRB43705LW 0000FRB42409MJ 00000092523VE0 00000092584VE2 0000FRB4380DKL 00000092567VE7 00000087618VE5 0000FRB4240CLR 0000FRB4240CN0 00000092575VE0 0000FRB42409M8 0000FRB4050CPF 0000CZC93668ZF 0000FRB41604C9 0000FRB41604CI 0000FRB4050D0D 00000075136VE2 0000FRB40304S0 0000FRB3390KNZ 0000FRB3390KNN 000000Q9099006 00002UA9080757 0000CZC7491JR1 000000Q9145004 0000CZC8464GFW 0000CZC9064PCR 000000D8145029 000000A8177032 0000CZC9064PCJ 0000CZC9064PBN 0000CZC83331FP 0000CZC9064P9W 0000CZC9064P9Z 0000CZC9064P9R 0000CZC8462DCC 0000CZC8464GFK 0000CZC8464GFH 0000CZC8464GFQ 0000CZC7393C2B 0000CZC84643GZ 0000CZC8464GFR 0000CZC83331FS 0000CZC9064PD4 0000CZC9064PC1 0000CZC9064PCY 0000CZC9064PC4 0000CZC9064PC9 000000J8351002 0000CZC9020WYG 0000CZC8393VM2 0000CZC9064PD1 0000CZC9064PCQ 0000CZC8464GFY 000000B7166006 0000CZC7182HNQ 0000CZC60445DV 000000C7247005 000000C7240008 000000U7347003 000000V8057014 000000Y8309002 0000CZC82388GF 00002UA642094Z 0000CZC8393VLX 0000CZC8326DRJ 000000C8159015 000000U8162059 000000U8162035 0000CZC7262J79 000000U7262015 000000ANT22775 0000CZC7182LWC 0000CZC82388GN 0000CZC8393VM3 000000J8351005 0000CZC8236LS4 0000CZC82388GP 0000CZC8187QPR 0000CZC8187QPH 0000CZC8393VM9 0000CZC83331FL 0000CZC8191GFP 0000CZC8295TNY 00000023712AW3 0000CZC8326DRD 0000CZC8303V6J 000000Y8323014 0000CZC8236LS5 000000B8329006 000000Y8323006 0000CZC825612Y 0000CZC83331FG 0000CZC83331FY 0000CZC8303V6G 0000CZC8303V67 000000U8344004 000000A7208005 000000H7267009 0000CZC80602B5 0000CZC81370P0 0000CZC641057Y 0000CZC8295TP8 0000CZC8303V69 0000CZC8236LSQ 0000CZC8236LSH 0000CZC8303V6M 000000E7297005 000000C8159006 000000B8150002 0000CZC7150M5K 000000B7241013 0000CZC7333J7H 000000B7339005 000000D8145023 000000A8070008 000000D8145018 000000C7342009 000000B8150005 0000CZC73706DH 000000A8045011 000000B7241001 000000B8150036 000000B8077002 000000C8159018 0000CZC8295TPH 0000CZC8295TP4 0000CZC8295TP9 0000CZC825613J 0000CZC8464GFX 0000CZC8464GFJ 000000U8162027 0000CZC8223P56 0000CZC8326DRK 0000CZC8191GFT 0000CZC8326DQZ 000000X8294007 0000CZC81370NQ 0000CZC8236LS8 0000CZC5432MVN 0000CZC8326DR9 0000CZC8236LSC 0000CZC825613C 000000A8302007 0000CZC841ONL5 00002UA54410CH 000000B9153012 0000CZC8326DR6 0000CZC8223P5N 0000CZC8295TP3 00000ZC7521LZZ 0000CZC73706D2 0000CZC8223P5R 0000CZC7521LZX 0000CZC8295TPB 000000U8162029 0000CZC8024SDS 0000CZC8191GFR 0000CZC8207NVR 0000CZC7521RVZ 0000CZC8256131 0000CZC8191GFL 000000U8162001 0000CZC8191GFZ 000000U8162002 000000U8162036 000000U8162018 000000U8162003 0000CZC7383456 0000CZC81370P5 0000CZC8191GFX 0000CZC8303V66 0000CZC8191GFN 0000CZC8191GG0 0000CZC8191GG3 0000CZC81370P2 0000CZC8191GFS 0000CZC8191GFM 0000CZC8191GG2 0000CZC7521LZS 000000B8150020 000000D8145021 0000CZC8223P5Q 0000CZC8130134 000000U8162007 000000U8162021 0000CZC825613G 000000073706D4 000000U8162008 0000CZC8303V6R 000000U8162025 000000U8162053 000000U8162012 0000CZC7333J7S 0000CZC8187QPX 000000C8159012 0000CZC8256132 0000CZC813012V 0000CZC82202QK 00000CZC81301Y 000000B8150022 000000B8150019 000000D8145005 000000D8145006 0000CZC7253SHL 000000B8150032 000000B8150031 000000H7314011 000000U8162028 000000B8150027 0000CZC73834S2 0000CZC8207NW2 0000CZC7414MHJ 0000CZC7393C1N 000000U8162019 000CZC7023-M63 0000CZC7521JWG 0000CZC8130132 0000CZC8187QPB 0000CZC8223P5C 0000CZC7414GSC 000000B8150033 0000CZC825612X 0000CZC8187QPK 0000CZC8120YCV 0000CZC8223P59 0000CZC8223P57 0000CZC8223P58 0000CZC8207NVL 000000D8145008 0000CZC813012L 000000V8057010 0000CZC7521RW0 0000CZC73706D5 0000CZC74003D0 0000CZC7414GSL 0000CZC806029Z 0000CZC7521RVP 0000CZC806029V 0000CZC7471QX5 0000CZC7321Z4X 000000B8077017 0000CZC73834RY 0000CZC82202QJ 0000CZC82202QL 0000CZC8207NVK 0000CZC82202QN 0000CZC82202Q5 0000CZC813012N 000CZC7521RWS4 0000CZC8191GFY 0000CZC7395WWZ 0000CZC8207NVQ 0000CZC8187QPL 0000CZC8187QPW 0000CZC8187QPC 0000CZC8187QPP 0000CZC8187QPG 0000CZC8187QPV 0000CZC73747Z7 0000CZC7452MNW 000000H8109016 0000CZC73834SC 0000CZC7090XLS 0000CZC7471QX2 0000CZC7182LWB 00000042927AW4 0000CZC80602BC 0000CZC806029S 0000CZC806029Y 000000A8045014 0000CZC7253SH0 0000CZC6391D1B 0000CZC7393C1Z 0000CZC73706DC 0000CZC7150M5M 0000CZC73834S9 0000CZC7491JQX 0000CZC7333J7T 0000CZC74003D7 0000CZC73834SB 000000C7282002 0000000A801901 0000CZC7061B11 0000CZC7064SCP 000000B9153009 0000CZC73834S5 0000CZC6071433 0000CZC7380HMZ 0000CZC73747ZB 000000A8019003 0000CZC7330YXS 0000CZC7280B2W 0000CZC7023M5R 0000CZC7110RG2 0000CZC7333J7X 0000CZC7253SH4 000000ANT26045 0000CZC74003CZ 000000H7349001 000CZC73933C22 000000T7331001 000CZC7521JWSW 000000A8019001 000000C7342007 0000CZC7330YXV 000000D7303001 0000CZC7521RW5 0000CZC614204B 0000CZC7380HMY 0000CZC7452MP3 0000CZC8024SDW 0000CZC74003D5 0000CZC62701R7 0000CZC73747Z1 000000U7262017 0000CZC74003DB 000000H7267012 0000CZC62217PB 0000CZC7321Z4W 0000CZC6321RT4 0000CZC6391D17 0000CZC7182HNT 0000CZC64025XJ 000000ANT21247 0000CZC74003D8 0000CZC75109B3 0000CZC74003CX 000000D7303010 0000CZC74003D4 0000CZC80602B7 0000CZC8024SDX 0000CZC8024SDZ 0000CZC7182HNV 0000CZC7452MPG 0000CZC64025X9 000000U7262002 0000CZC7105R6Q 0000CZC60619J3 0000CZC7452MPC 0000CZC7253SGZ 0000CZC7280B2G 0000CZC7395WWV 0000CZC7395WWX 000000B7205018 0000CZC7432L2C 0000CZC7331WV4 0000CZC7452MPH 0000CZC7432L27 000000D7303009 0000CZC7072JXH 000000K7311003 0000CZC7262J6Y 000000ANT23711 0000CZC7280B2P 0000CZC7393C1Q 0000CZC7432L1W 0000CZC549318Y 0000CZC72544NT 0000CZC7333J7G 0000CZC7280B2D 0000CZC7380HMM 00000CZC2701RG 000000K7170010 0000CZC7414GS6 000000J7226012 000000B7205001 0000CZC5500626 0000CZC7380HMT 0000CZC7280B2Y 0000CZC73747Z9 000000C7247011 0000CZC7331WV6 0000CZC7330YXN 0000CZC7280F7W 0000CZC72449PC 000000U7262020 00000CZC7333J7 000000ZA718401 0000CZC7321Z5C 0000CZC7105R6V 0000CZC7262J76 000000A7208011 000000F7229015 000000B7205011 000000J7226011 000000B7241008 0000CZC7072JXQ 0000CZC549318K 00002VA54412XJ 0000CZC7032PHF 0000CZC7321Z5B 0000CZC7333J7Y 0000CZC7280B2R 0000CZC7090XLX 0000CZC7330YXM 0000CZC7333J7P 000000H7267002 0000CZC73706D9 00000483835BU5 00002UA7340YTR 000000A7208015 0000CZC72535H2 0000CZC7333J7K 0000CZC7330YXQ 0000CZC7150M5D 0000CZC72449PD 00000016736AW1 0000CZC7330YXG 0000CZC7193XTY 0000CZC7331WV9 0000CZC7193XV4 000000C7257019 00000020950AW2 00000020925AW4 0000CZC73706D3 0000CZC7380HMR 00000030956AW7 00000040332AW9 0000CZC7280F7R 0000CZC7262J70 0000CZC7280F7S 000000F7229001 0000CZC549318W 00000020963AW5 0000CZC7182HNH 0000CZC6512VKG 000000ANT23699 0000CZC7280B2S 0000CZC7280F82 0000CZC7193XTV 000000C7247009 000000J7226003 0000CZC7280F7Y 00000CZC7280FN 000000B7205014 00002UA52918YG 0000CZC7280B2L 0000CZC7365LVW 0000CZC6512VKH 000000J7226002 00000057063AW0 0000CZC7331WV7 0000CZC7331WVC 0000CZC7262J78 0000CZC7262J75 0000CZC71521VB 0000CZC7182LWM 0000CZC7262J6Z 0000CZC7262J73 0000CZC7262J77 000000A7208014 0000CZC7150M5B 0000CZC7150M5J 0000CZC7262J7H 00000056434AW4 00000030163GE0 0000CZC72449PJ 0000CZC72449PS 0000CZC71521V9 0000CZC7182HP0 0000CZC6280K4M 00000042964AW7 000000ANT23702 0000CZC7280B2B 0000CZC7280F7T 000000ANT20560 0000CZC7262J7C 0000CZC7193XTT 0000CZC7090XM0 0000CZC7262J6W 000000ANT23722 0000CZC7280F7X 0000CZC7193XVM 0000CZC7193XV9 0000CZC7182LW5 0000CZC71108XQ 0000CZC71108XL 0000CZC71108XD 0000CZC711ATZT 0000CZC7114TZM 000000ANT22759 000000ANT23708 0000CZC7083GZW 000000ANT22761 000000ANT20545 0000CZC641057T 0000CZC7182LWG 000000ANT23709 00000056367AW6 000000ANT20767 000000ANT21596 0000CZC7105R6M 000000T7250003 00000056441AW9 00000057090AW3 000000ANT20168 0000CZC6391D16 00002UA5390FQQ 000000ANT20556 00000023720AW6 00000042921AW7 00000042947AW2 00000042990AW2 00000025828GE5 0000CZC7113CL8 000000ANT21237 0000CZC6391D14 00000057095AW2 0000CZC7105R71 0000CZC60407F4 0000CZC7193XV8 0000CZC70414HR 00002386410DHF 0000CZC7083H00 00000056383AW3 0000CZC7072JXW 0000CZC5450Q8C 0000CZC7023M5K 0000CZC7072JXM 0000CZC7090XLQ 00000057076AW2 0000CZC6391D1C 0000CZC7083GZY 0000CZC7072JXJ 0000CZC7072JXV 00000057088AW7 0000CZC7083GZZ 0000CZC7031XXS 00000543701BU7 0000CZC7012HG0 00000CZC694GLT 0000CZC6481ZJX 0000CZC7051FVK 0000CZC60445DR 0000CZC703221N 0000CZC7090XLZ 0000CZC7065H1C 00000054370BU1 00000056428AW6 00000056422AW9 0000CZC70414HS 00000534573BU1 00000042977AW9 00000057057AW2 0000CZC7031XXR 0000CZC7023M5L 0000CZC6502Y9Y 00000044349AW9 00000056384AW1 0000CZC62806L4 0000CZC7061B12 00000534538BU4 0000CZC6502YB2 0000CZC6502YB7 0000CZC6435BFP 00000016742AW9 0000CZC6512VKJ 00000056394AWO 0000CZC62806LB 0000CZC6482BJ8 0000CZC6494GLN 00000030170GE5 0000CZC6435BFQ 0000CZC5432MVK 00000025825GE1 0000CZC6502YB1 00000026956GE3 00000025829GE3 00000026954GE8 00000025273GE4 0000004432OAWO 00000056403AW9 0000CZC6391D19 00000025196GE7 0000CZC6502YBB 00000025267GE6 0000CZC6502YB5 0000CZC6071434 0000CZC71108XH 00000030162GE2 0000CZC6421C4F 0000CZC6481ZDK 0000CZC55207GV 0000CZC6482BJ5 0000CZC64025XC 0000CZC6512VKD 0000CZC6502YBD 00000043000AW9 00000016705AW6 0000CZC6481ZJW 0000CZC6040X2Z 0000CZC60407F0 0000CZC608443Q 0000CZC62806L5 00000023739AW6 00000025830GE1 00000025824GE4 0000CZC6321RT7 0000CZC6321RT8 0000CZC6391D13 0000CZC643501J 00002UA5421D72 00000044355AW6 0000CZC6391D15 0000CZC6482BJ3 0000CZC6142047 0000CZC6482BJ6 0000CZC6200CZR 0000CZC641057W 0000CZC64148ZJ 0000CZC641057X 00000042999AW3 0000CZC62806LJ 00000043005AW8 0000CZC62217PC 00000035325AW0 0000CZC62806L8 0000CZC62806L7 0000CZC62806LH 0000CZC6040X31 0000CZC60445DW 0000CZC60619J4 00000020967AW6 00000031009AW4 0000CZC64025XD 0000CZC6280K4H 0000CZC6103V2D 0000CZC626329R 00000035297AW1 00000042976AW1 0000CZC626329Q 0000CZC62806L3 0000CZC64025XK 0000CZC64025XF 00002UA5421D76 00000042925AW8 0000CZC64025XG 0000CZC6181NLD 0000CZC6303F8Q 00000042950AW6 0000CZC62806L9 00000023709AW9 00000042933AW2 00000043620AW4 0000CZC62806LM AW HP XW8200 S 00000020912AW2 0000CZC6300NQV 00000042941AWS 0000CZC6280K4K 00000040342AW8 0000CZC62806LF 0000CZC6130XXK 00000040336AW0 0000CZC6083SBR 0000CZC614203Z 0000CZC626329T 0000CZC6300VHJ 0000CZC6181NL9 00000040362AW6 0000CZC6130XX0 0000CZC626329P 0000CZC6300NQT 0000CZC6132TJW 0000CZC6122LCB 0000CZC6122LC0 00000483885BU0 0000CZC607140Y 00000483890BU0 0000CZC6192FTS 0000CZC62701R8 00000CZC613TJT 0000CZC62701R6 0000CZC5432MVQ 0000CZC5450Q8D 0000CZC6130XXD 0000CZC6130XXN 0000CZC6103V2M 00000030370AW1 000CZC61222LC6 0000CZC6071432 0000CZC6071437 0000CZC60445DQ 0000CZC6130XXP 00000035259AW1 0000CZC6122LC3 0000CZC6130XXG 000000F7229009 0000CZC6071436 0000CZC6084442 0000CZC6132TJN 0000CZC6142041 00002UA5421D6L 00000029429GE8 0000CZC608443Z 0000CZC6130XWZ 0000CZC60445DS 00002UA54412XH 0000CZC6071410 0000CZC60407DW 0000CZC60445DZ 0000CZC608443M 0000CZC60445F0 0000CZC6083SBZ 0000CZC6040X38 0000CZC6083SBV 00000023754AW5 0000CZC608443Y 0000002091SAW5 00000023727AW1 0000CZC6084444 0000CZC608443S 0000CZC60445DT 0000CZC608443W 0000CZC60619J0 0000CZC607140Z 0000CZC60619J6 0000CZC6084441 0000CZC60619J7 0000CZC6071438 0000CZC5493191 0000CZC55207GS 0000CZC60619HY 00000016749AW4 00000023744AW6 00000020916AW3 0000CZC549318T 00000020935AW3 00000020982AW5 00002UA5440KV1 00000016743AW7 00000020922AW1 0000CZC549318M 00000016761AW9 00000023730AW5 0000CZC60407F2 00002UA5420VPC 0000CZC5480XPD 0000CZC55207GY 00000016746AWO 00000016747AWB 00000483827BU2 00000030165GE5 00002UA5421D6V 00002UA5440KTZ 00000483820BU7 00000023725AW5 00000483808BU2 00000028460GE4 00000483807BU4 00000483815BU7 00000483810BU8 00002UA5421D41 00000029434GE8 00000023726AW3 00000483844BU7 00000023721AW4 00000016734AW6 00000483822BU3 00002UA5421D3W 00000026951GE4 0000CZC5382370 00000483804BU1 00002UA5421D3Z 00000029430GE6 00000030176GE2 00000030177GE0 00000483797BU7 00000029438GE9 00000025271GE8 00002UA5421D6Y 00002UA5421D6M 00000029977GE6 00002UA54410C9 00000025840GE0 0000CZC5400KQK 00000029439GE7 00000029441GE3 00000025832GE7 00000030160GE6 00000030159GE8 00000028458GE8 00000026955GE5 00000029979GE2 00000026953GE0 00000028459GE6 00000026947GE2 00000002648GE0 00000025827GE7 00000025826GE9 00000025841GE8 00000025274GE2 00000042983AW7 0000CZC9064P9Y 0000CZC9064PBX 000000A8177070 000000A8177058 0000CZC707342T 00000167474II6 000000Y8323011 0000CZC8295TPD 000000D8145022 0000CZC7521LZZ 000000C8159025 000000D8145007 0000CZC81370NN 0000CZC8120YCP 00000CZC7400CR 0000CZC7521JWP 000000A7274025 0000CZC7452MP6 000000B7239002 000000H7267004 000000C7247012 0000CZC7333J7R 0000CZC7333J7N 0000CZC72544NR 000000MI000383 000000MI001470 000000ANT20962 0000CZC7114TZX 0000CZC7105R6K 0000CZC6481ZK2 0000CZC7182LW3 00000020635AW9 000000ANT22430 0000CZC7105R6Z 00000480900BU0 0000CZC7090XLY 00000080148HP0 00000080166HP2 000000M1000302 0000CZC7092FHS 00000056395AW7 0000CZC6482BJ7 00000534610BU1 0000CZC641057V 0000CZC62806L6 00000040370AW9 0000CZC615450R 0000CZC6142042 0000CZC6084445 0000CZC6132TJQ 0000CZC608443J 00000020995AW7 0000CZC6083SC3 0000CZC60619HW 0000CZC6071430 0000CZC60407F1 00000020937AW9 0000CZC60407F3 0000CZC55207GX 00002UA54412XJ 00000023715AW6 00000483842BU1 0000CZC5400KQQ 00000029440GE5 00000025270GE0 00000028462GE0 00000025842GE6 00002UA5390FQT 00000024541GE5 00000023809GE7 0000FRB4510BG6 00000092528VE9 00000092532VE1 00000092615VE4 00000075143VE8 0000FRB3630V7K 0000CZC8223P5G 0000CZC73747Z3 0000CZC6103V2J 0000CZC7521JWH 0000CZC7521JW5 0000CZC80602BB 0000CZC80602B2 0000CZC8393VMB 0000CZC83331FZ 0000CZC82388GK 0000CZC8024SDT 0000CZC8236LSK 0000CZC8223P5K 0000CZC8223P5P 000000A8177015 0000CZC8207NVW 0000CZC8207NVM 0000CZC8207NVV 0000CZC8120YCQ 0000CZC7491JQW 0000CZC7471QXG 0000CZC75109B5 0000CZC7491JR2 0000CZC71816TF 0000CZC7393C1P 0000CZC7432L29 000000C7257012 0000CZC73747Z5 0000CZC7321Z58 0000CZC7092FHV 0000CZC7193XVF 0000CZC7182HNN 0000CZC7114TZR 00000CZC7108XF 0000CZC7051FVG 000000ANT20537 0000CZC7065H18 0000CZC7065H1B 0000CZC70623H1 0000CZC7042DB7 0000CZC7100BB6 0000CZC6152BYL 0000CZC7031XXT 000000ANT20538 000000ANT21248 000000ANT20470 0000CZC64607S2 0000CZC7023M5S 0000CZC6482BJB 00000043018AW1 0000CZC64148ZL 00000043621AW2 00000044354AW9 0000CZC6152BYM 0000CZC6323QJV 0000CZC6323QJX 00000042937AW3 0000CZC6311JZY 0000CZC6231QQW 0000CZC60407DT 0000CZC60407DZ 0000CZC6481ZJV 0000CZC61007Z7 0000CZC60619J5 0000CZC5500627 00000483812BU4 0000FRB4110002 00000016712AW2 00000021002AW1 0000CZC9280BKG 0000CZC8024SDL 0000CZC6481ZK9 0000CZC6130XXJ 0000CZC615450M 0000CZC615450L 0000CZC71521V8 0000CZC82388GT 000000M8281002 0000CZC8120YCJ 0000CZC9280BKW 0000CZC7193XVL 00000483831BU4 0000CZC6174218 0000CZC9280BKT 0000CZC9064PCD 0000CZC84643GH 000000K8336003 0000CZC8303V65 0000CZC82388GJ 0000CZC8236LSD 0000CZC825613B 0000CZC82388GZ 0000CZC8223P55 000000C8159001 0000CZC8187QPD 0000CZC8207NVJ 000000B8150038 0000CZC8012RPV 0000CZC8120YCG 0000CZC7414MHM 0000CZC7471QXK 0000CZC7393C1R 0000CZC7380HMN 0000CZC71145V4 0000CZC7114TZJ 0000CZC7114V00 0000CZC6481ZK1 0000CZC6280K4J 0000CZC5321V38 0000CZC61007Z6 0000CZC5360X9D 0000CZC549318N 0000CZC5321V3F 0000CZC5321V36 0000CZC5321V3C 0000CZC9064PBJ 00000089234HP9 00000089473HP3 00000008956HP0 00000089505HP2 00000023749AW5 0000CZC8256135 000000N8197006 0000CZC75109BB 0000CZC8120YCK 0000CZC7414GSP 0000CZC7253SHC 0000CZC72449PQ 0000CZC71320NW 0000CZC7114TZN 0000CZC71320NV 0000CZC64607S1 0000CZC703221K 0000CZC6481ZJS 0000CZC6132TJX 00002UA5420VPG 0000CZC7373G22 0000CZC70414HT 0000CZC7042DK2 0000CZC6435BFN 0000CZC626329X 0000CZC62701RB 0000CZC62701RF 000000U8162038 0000CZC82202Q3 0000CZC9316RLX 0000CZC9280BKD 0000CZC9280BKV 0000FRB42409MC 0000CZC7182HNF 0000CZC7521JWN 0000CZC9064PBT 0000CZC8466Y95 0000CZC8466Y92 0000CZC8466Y90 0000CZC8464GG2 0000CZC8462DC7 0000CZC8410NLK 0000CZC8236LSM 0000CZC82388GL 0000CZC8236LSR 0000CZC8223P5M 0000CZC7432L28 0000CZC7193XVH 0000CZC7113CLB 000000ANT21393 0000CZC7280F7Z 0000CZC72449PK 000000ANT20574 0000CZC7065H16 00000044352AW3 0000CZC643086R 0000CZC6192FTP 00000035333AW4 0000CZC6173PJ8 0000CZC6083SBX 0000CZC5480XP8 0000CZC5432MVM 0000CZC5351KH7 0000CZC8462DC1 000000U8344011 0000CZC8393VMC 0000CZC80602B0 0000CZC7521RVR 0000CZC7380HMX 0000CZC7330YXH 0000CZC7330YXJ 0000CZC7330YXL 0000CZC7432L2B 000000T7331007 0000CZC7330YXP 0000CZC7042DK7 0000CZC7121CXR 0000CZC7090XM2 000000ANT21654 0000CZC6512VKK 0000CZC6303F8W 0000CZC6323QJW 0000CZC645066D 00000040366AW7 0000CZC5480XP4 00000103370VE3 00000102739VE0 00000087617VE7 00000075121VE4 000000Q7179012 0000CZC6181NLC 0000CZC71108XJ 000000Q7278005 0000CZC74003CT 0000CZC6321RT5 0000CZC63204M1 00000043017AW3 0000CZC8207NW3 0000CZC82388GM 0000CZC9280BKM 0000CZC8256138 000000A8177050 0000CZC7090XM3 0000CZC6481ZKB 00000483877BU7 000000Q9099001 0000CZC9064PBZ 0000CZC7432L20 0000CZC9064PC5 0000CZC8256137 0000CZC8410NLC 000000A8302002 0000CZC7432L25 0000CZC8187QPF 0000CZC7393C1X 00000020999AW9 0000CZC6452TY5 00000044347AW3 0000CZC6103V2K 0000CZC6103V2G 0000CZC7376B5B 000000D8145026 0000CZC6181NLB 0000CZC549318S 0000CZC7193XV0 0000CZC5350K5M 0000CZC8466Y8S 0000CZC8236LSL 00000543719BU9 000000B8150040 0000CZC8462DC3 0000CZC8393VLT 0000CZC73747ZC 0000CZC82388H0 0000CZC82202QF 0000CZC75109B7 0000CZC7414GSJ 0000CZC7262J74 0000CZC7100BB5 0000CZC5360X9B 0000CZC8256136 0000CZC6421C4J 0000CZC6421C4C 0000CZC9064PB9 0000CZC7373G1T 0000CZC7373G1X 0000CZC7373G1S 0000CZC7373G1V 0000CZC6512VKF 0000CZC7380HMQ 0000CZC8120YCB 0000CZC81370P1 0000CZC75109B1 0000CZC7491JQM 0000CZC7380HMP 0000CZC7193XV6 0000CZC7193XVG 000000ANT21380 0000CZC7051FVM 00000016696AW7 0000CZC7042DB8 00000057087AW9 0000CZC6450XYL 0000CZC6300VHG 0000CZC6083SC0 0000CZC7321Z56 0000CZC7193XVJ 0000CZC6321RT3 0000CZC6321RT6 0000CZC9064P9S 000000Q9099009 0000CZC8236LS9 0000CZC82388GG 0000CZC751099Z 0000CZC84643H2 0000CZC73747ZL 00000040365AW9 00000042985AW2 00000042938AW1 0000CZC8303V6D 0000CZC82202QC 000000Q9145007 0000CZC8256134 0000CZC8120YCH 0000CZC7452MP5 0000CZC8120YCC 0000CZC8120YCL 0000CZC73706D7 0000CZC6122LBZ 00000029432GE2 000000B7205019 0000CZC6200XWK 00002UA5421D78 000000B8150007 0000CZC5500628 0000CZC5500625 00000029435GE5 00000029433GE0 0000CZC7380HN0 0000CZC7373G21 000000A8045012 0000CZC8024SDJ 0000CZC75109B4 000000R8051002 000000A8045020 000000A8045023 000000V8057009 0000CZC75109B9 0000CZC8024SDH 0000CZC8024SDY 0000CZC7521JWQ 0000CZC7521JWC 0000CZC75109B2 0000CZC7110RG6 0000CZC7414MHN 0000CZC8462DCH 0000CZC608443N 0000CZC82202Q7 0000CZC7491JQY 0000CZC7471QX7 0000CZC72449PL 0000CZC7521RVS 0000CZC7114TZZ 0000CZC9280BKS 000000C9168001 0000CZC92457X9 0000CZC9064PBS 0000CZC9064PBB 0000CZC9064PBL 0000CZC9064PBY 0000CZC8464GFP 0000CZC8462DBZ 0000CZC8466Y94 0000CZC8462DC0 0000CZC8462DC5 0000CZC8462DCK 0000CZC8466Y8M 0000CZC8462DCB 0000CZC84643GP 000000Q8354002 0000CZC8326DR8 0000CZC7471QX8 0000CZC83331FF 0000CZC8410NLB 0000CZC8120YCY 0000CZC6083SC1 0000CZC8303V6K 0000CZC8295TNW 0000CZC8295TNX 000000H8109006 000000H8109020 0000CZC7012HHH 0000CZC81370P3 0000CZC81370P7 0000CZC8120YCX 0000CZC81370NY 0000CZC8012RPR 0000CZC8012RPW 0000CZC71816TJ 0000CZC8012RPS 0000CZC7491JQJ 0000CZC751099V 0000CZC8024SDF 0000CZC9280BKJ 0000CZC7521JW9 0000CZC7491JQS 0000CZC608443X 0000CZC7491JQQ 0000CZC7414GSH 0000CZC7414GS2 0000CZC7414GS9 0000CZC7414GSM 0000CZC7471QXJ 0000CZC7471QXL 0000CZC541226K 0000CZC7393C2F 0000CZC74003D6 0000CZC7414GSD 000000D7303006 00000089716HP5 0000CZC73834RZ 0000CZC7321Z5F 0000CZC7193XTS 0000CZC7280B2T 000000ANT20768 0000CZC71108XX 0000CZC7193XVD 0000CZC71108XF 0000CZC7193XTW 00000509717BU5 0000CZC71320NX 0000CZC7114V01 0000CZC7100BB7 0000CZC7011J4C 0000CZC70623H2 000000ANT20671 0000CZC7042DB9 0000CZC70414HW 0000CZC7090XLV 0000CZC7090XLW 0000CZC7042DK6 0000CZC7071NWC 0000CZC703221L 0000CZC7023M5N 0000CZC6512VKL 0000CZC7065H17 0000CZC6502YB9 0000CZC6481ZK3 0000CZC7011J4D 0000CZC70414HQ 0000CZC70414HV 00000043002AW5 0000CZC7031XXQ 0000CZC6494GLQ 0000CZC7042DK3 0000CZC7042DK5 0000CZC70416YN 0000CZC7042DK8 0000CZC7023M5Z 0000CZC6200XWD 0000CZC6421C4H 0000CZC6481ZK6 00000056363AW5 0000CZC6482BJ4 00000020932AW0 0000CZC6452TY8 0000CZC63204LV 0000CZC643086Q 0000CZC6481ZK5 0000CZC62806LG 0000CZC6481ZK4 0000CZC6174215 0000CZC6421C45 0000CZC64148ZP 00000043013AW2 0000CZC63204LX 0000CZC6311JZW 0000CZC6311JZZ 0000CZC61007Z5 0000CZC6200CZT 0000CZC6300VHL 00000023731AW3 0000CZC63204LW 00000031021AW9 0000CZC615450N 0000CZC615450Q 0000CZC6280K4N 0000CZC6173PJC 0000CZC6173PJD 0000CZC6181NLF 0000CZC61504VV 0000CZC6200XWH 0000CZC6173PJ7 0000CZC6174216 00000030999AW7 0000CZC61504VW 00000483871BU0 0000CZC6173PJ9 0000CZC61007Z4 0000CZC61007YY 0000CZC61007Z0 00000030369AW3 0000CZC6142040 0000CZC61007Z3 0000CZC6130XXH 0000CZC61007Z8 00000031020AW1 00000031010AW2 0000CZC608443V 0000CZC60619J2 0000CZC55207GT 0000CZC55207GW 0000CZC608443L 0000CZC549318Z 0000CZC5480XPB 0000CZC60619J1 0000CZC6040X39 0000CZC5480XP9 0000CZC5480XP6 0000CZC5480XPC 00000020921AW3 0000CZC55207GZ 00000483845BU4 0000CZC5412WNB 0000CZC541226H 00000023718AW0 0000CZC5420T81 0000CZC5432MVP 00000483823BU1 00002UA5420VPH 00002UA5421D3V 0000CZC5292LNK 00000023728AW9 0000CZC5420T80 0000CZC5400KQL 0000CZC5400KQP 00000483793BU6 00000483801BU7 0000CZC541226G 0000CZC541226J 0000CZC5350K5N 0000CZC5321V37 00000016681AW9 0000CZC5351KH8 0000CZC5350K5P 0000CZC5321V39 00000016695AW9 0000CZC5292LNL 0000CZC5292LNP 00000016672AW8 0000CZC5292LNR 00000016666AW0 0000CZC5321KVK 0000CZC7121CXT 0000CZC6421C4G 0000CZC6130XXF 0000CZC6040X36 000000ANT21995 0000CZC84643GX 0000CZC7521RVL 0000CZC7521RW2 0000CZC73706D8 0000CZC9064PBH 000000J9026002 0000CZC7061B13 0000CZC643086P 0000CZC63204M0 0000CZC7262J71 0000CZC7521RVX 0000CZC9064PCG 0000CZC9064PC0 0000CZC9064PBW 0000CZC8393VMF 0000CZC8462DC6 0000CZC8410NL3 000000A8177009 0000CZC8295TPG 0000CZC8223P5S 0000CZC7491JQN 0000CZC813012T 0000CZC8223P5L 000000A8177023 000000A8177033 0000CZC8236LSG 0000CZC82388H1 0000CZC8120YCR 0000CZC82202QB 0000CZC82202Q6 0000CZC7491JR0 0000CZC81370P6 000000H8109005 0000CZC751099T 0000CZC7521RW3 0000CZC8120YCN 0000CZC8024SF0 0000CZC8024SDD 0000CZC7432L2G 0000CZC7414GSN 0000CZC8024SDV 0000CZC7393C27 0000CZC7521JWL 0000CZC7471QX4 0000CZC7121CXS 000000A7274004 0000CZC7373G1Y 0000CZC7432L23 0000CZC7432L1V 0000CZC7393C1W 0000CZC7380HMW 0000CZC7452MNY 0000CZC7365LVZ 0000CZC7432L21 0000CZC813012W 0000CZC7331WV8 0000CZC7280F7L 0000CZC7193XV1 0000CZC7280F84 0000CZC72544NS 000000B7205015 0000CZC7121CXP 0000CZC64148ZQ 0000CZC6421C4D 00000044345AW7 0000CZC64148ZR 0000CZC643086L 00000044326AW7 00000044367AW1 00000043016AW5 0000CZC63204LY 0000CZC6181NL8 0000CZC6303F8V 00000035324AW3 0000CZC6200XWF 0000CZC55207GR 00000035335AW9 0000CZC6132TJV 0000CZC6152BYH 00000020941AW1 0000CZC541226F 0000CZC549318Q 0000CZC5400KQN 0000CZC541226L 0000CZC5420T82 0000CZC5412WNC 0000CZC8303V6N 0000CZC8464GFN 0000CZC7491JQG 0000CZC81370NZ 00000089502HP9 0000CZC6132TJM 0000CZC9064PBK 0000CZC9064PBP 0000CZC9064PC7 0000CZC9064PCC 0000CZC84643GJ 0000CZC7521RW1 0000CZC7182LW9 0000CZC7182HNY 00000483832BU2 0000CZC6173PJB 00000035284AW9 0000CZC6311JZX 00002UA54410CC 0000CZC7393C21 0000CZC7333J82 0000CZC72449PN 000000C7242004 0000CZC5351KH9 0000CZC61007Z9 0000CZC8466Y8X 0000CZC84643GM 0000CZC83331FM 0000CZC7182HNL 0000CZC61908DH 0000CZC5321V3B 0000CZC5321V3D 0000CZC73706DB 0000CZC7193XVP 0000CZC73706D6 0000CZC9280BKH 0000CZC9064PB3 0000CZC9064PBQ 0000CZC83331FK 0000CZC8223P5B 000000D7303012 0000CZC9064PB2 0000CZC9064PBF 0000CZC92457XC 0000CZC9064PCX 0000CZC8466Y8P 0000CZC8466Y93 000000B9153008 0000CZC8462DC4 0000CZC83331FV 0000CZC83331FH 0000CZC83331FW 0000CZC8236LS7 0000CZC8207NVS 000000L7289002 0000CZC7491JQL 0000CZC8024SDK 0000CZC7491JQT 0000CZC7521JW7 0000CZC7414GS5 0000CZC7471QX9 0000CZC74003CS 0000CZC7432L1Y 00000089494HP9 0000CZC74003D2 0000CZC6452TY4 0000CZC7105R6J 0000CZC7023M5Y 0000CZC64607S6 00000016703AW1 0000CZC9280BKX 0000CZC8393VM8 00000509591BU4 0000CZC6174219 00000089517HP7 0000CZC62701RC 0000CZC8303V6B 0000CZC7393C20 0000CZC7373G1W 0000CZC7373G24 0000CZC7491JQP 0000CZC8236LS6 000000A8070006 0000CZC7105R6T 00000483840BU5 0000CZC8024SDN 0000CZC7491JQR 0000CZC7280F81 0000CZC7280F7Q 0000CZC7321Z55 000000C7242001 0000CZC7253SH6 0000CZC72544NX 0000CZC72449PR 0000CZC7182LW8 0000CZC549318P 0000CZC7090XM5 00000040345AW1 0000CZC6300NQW 0000CZC62701R9 0000CZC6200WXJ 0000CZC6200CZP 0000CZC61504VX 0000CZC61504VY 0000CZC6192FTN 0000CZC6200CZS 0000CZC61007YZ 0000CZC7373G1R 0000CZC8466Y82 0000FRB4230PRD 0000FRB4510BFN 0000FRB44107QH 000000C8159008 0000CZC8462DCG 0000CZC8236LSJ 0000CZC7051FVP 0000CZC7042DK0 00000089602HP7 00000089480HP8 0000FRB3430VC1 0000CZC7121CXN 00000089580HP5 00000543703BU3 0000FRB4330BBZ 0000FRB44107PQ 00000102734VE1 00000102775VE4 0000FRB4490005 0000FRB4370CY2 0000CZC9064PCF 0000CZC9064PBM 0000CZC7114TZQ 00000044366AW3 0000CZC7452MPF 0000CZC7380HMV 0000CZC9064PCH 0000CZC8466Y8Q 000000M8357001 0000CZC92457XH 0000CZC8466Y8L 000000M8357007 0000CZC8462DCD 0000CZC84643GQ 0000CZC8466Y8V 0000CZC84643GW 0000CZC8410NL9 0000CZC82388GW 0000CZC83331FC 0000CZC83331FN 0000CZC82388GR 000000A8177053 0000CZC8326DR1 0000CZC82388GH 0000CZC8303V6C 0000CZC8256130 0000CZC8236LSB 0000CZC82388GY 0000CZC8223P5J 0000CZC82202Q9 0000CZC8191GFW 0000CZC8120YCS 000000B8077007 000000A8086003 000000A8086004 000000ANT23158 0000CZC8012RPQ 0000CZC80602B6 0000CZC806029W 000000U7347006 0000CZC751099W 0000CZC7414GS8 000000A8070011 0000CZC7491JR4 0000CZC7253SHG 0000CZC7253SH3 0000CZC751099R 0000CZC7471QXH 0000CZC74003CY 0000CZC7321Z5D 000000C7282004 0000CZC7393C2C 0000CZC7432L1Z 0000CZC73834S8 0000CZC5291LNQ 0000CZC6142048 0000CZC7330YXT 0000CZC7373G1Z 0000CZC7150M59 0000CZC7253SH8 0000CZC7061B16 0000CZC7253SH7 0000CZC7193XV2 0000CZC7253SHH 0000CZC7150M5G 0000CZC7150M5H 000000Q7179001 0000CZC7193XV7 0000CZC71108XY 0000CZC7150M5L 0000CZC7114TZK 0000CZC7114TZW 0000CZC7121CXQ 0000CZC7105R73 0000CZC7100BB8 0000CZC7100BB9 0000CZC7113CL3 0000CZC7011J4F 00000056426AW0 0000CZC7092FHW 0000CZC7012HJY 0000CZC7042DK1 000000ANT20171 000000ANT20862 0000CZC5493190 00000056446AW8 0000CZC7064SCM 0000CZC7064SCN 0000CZC7092FHQ 0000CZC7061B18 0000CZC703221J 0000CZC6152BYJ 0000CZC7051FVN 0000CZC7072JXL 0000CZC7072JXX 0000CZC7065H13 000000ANT20422 0000CZC6200CZQ 000000ANT20174 00000543713BU2 0000CZC70416YQ 0000CZC7012HG1 00000044327AW5 00000044340AW8 00000023708AW1 00000042986AW0 0000CZC6481ZKC 0000CZC6481ZJT 0000CZC64607S3 0000CZC64148ZK 0000CZC6452TY3 0000CZC645066F 00000CZC6040X3 0000CZC62701RD 00000040369AW1 00000035323AW5 00000035304AW5 0000CZC6132TJP 00000483875BU1 00000020979AW1 0000CZC60445DX 00002UA5440Z1K 00002UA5421D6J 00000044342AW4 00000044343AW2 00000044358AW0 00000044364AW8 00000056369AW2 0000CZC7072JXS 0000CZC6481ZK0 0000CZC6040X33 0000CZC6040X30 00000016738AW7 0000CZC73747YZ 0000CZC73747ZG 0000CZC73747Z2 0000CZC6481ZJY 0000CZC8326DRH 0000CZC8326DRB 0000CZC9064PBR 0000CZC82202Q8 0000CZC8326DR7 0000CZC8326DRF 000000A8273001 0000CZC8303V68 0000CZC8326DRG 0000CZC82202QH 0000CZC6132TJY 0000CZC7471QXN 0000CZC7471QX6 000000D8145009 000000M8281006 0000CZC84643GG 0000CZC7491JQV 0000CZC73747ZH 0000CZC73747Z6 0000CZC7193XV5 000000ANT20519 0000CZC61908DD 0000CZC6391D18 0000CZC6481ZK8 0000CZC7072JXR 0000CZC82202QG 0000CZC7471QXM 00000016739AW5 0000CZC6452TY6 00002UA5421D6G 0000CZC9064P9T 0000CZC8464GG1 000000A8177003 0000CZC7182LW2 0000CZC9064PCW 0000CZC9064PD3 0000CZC9064PD0 000000B9153005 0000CZC9064PCN 0000CZC9064PB8 000000B9153014 000000B9153002 0000CZC9064PBG 0000CZC8393VM6 0000CZC9280BKC 0000CZC9064PCB 0000CZC9064PCZ 0000CZC8466Y8Y 0000CZC8464GFS 000000Q8354019 0000CZC9280BKN 0000CZC84643GK 0000CZC8393VLW 0000CZC8410NL2 000000A8177071 0000CZC8393VLZ 0000CZC83331FT 000000A8177060 0000CZC8393VM1 0000CZC8393VM7 000000A8177030 0000CZC8410NLL 0000CZC83331FQ 0000CZC8256139 000000A8177067 0000CZC82388GQ 0000CZC8236LSP 000000A8177076 0000CZC71108XP 0000CZC82202QM 0000CZC825613H 000000A8177049 0000CZC82202Q2 000000U8162040 0000CZC7521RVV 0000CZC7471QXB 0000CZC7414MHP 0000CZC7414MHK 000000B8077012 000000Y8024004 0000CZC71108XC 0000CZC7491JQK 0000CZC7110RG4 0000CZC7471QX3 000000C7342005 0000CZC73747ZK 0000CZC7023M5X 000000ANT21600 0000CZC73747Z4 0000CZC73834S4 0000CZC73747Z8 0000CZC74003CV 000000H7314008 0000CZC92457X6 0000CZC7114TZP 0000CZC7023M5W 00000089702HP5 0000CZC71108XG 0000CZC6494GLP 0000CZC7065H19 000000ANT22435 0000CZC7105R72 000000ANT23160 0000CZC7114TZS 000000ANT21879 0000CZC6481ZKF 0000CZC71145V3 000000ANT23161 0000CZC7114TZG 0000CZC61504VT 00000035281AW5 0000CZC6494GLS 00000056396AW5 0000CZC7072JXP 000000ANT20418 0000CZC6421C4B 0000CZC6481ZK7 00000056370AW0 00000044321AW8 0000CZC6481ZJZ 00000056387AW4 0000CZC6323QJS 0000CZC5330471 0000CZC6311K01 00000035338AW3 0000CZC6200CZN 0000CZC62701RJ 00000040375AW8 0000CZC6231QQX 00000023716AW4 0000CZC6084443 0000CZC9280BKK 0000CZC608443T 0000CZC6083SBS 00000483841BU3 0000CZC6103V2C 0000CZC608443H 00000483837BU1 00000023732AW1 0000CZC60713RB 0000CZC6081FV3 0000CZC6081FV5 0000CZC5493192 0000CZC5292LNN 00000023717AW2 00000016674AW4 00000035309AW4 0000CZC7393C1T 0000CZC8466Y8T 0000CZC9064PB6 0000CZC8466Y8K 0000CZC8466Y8N 0000CZC8012RPT 0000CZC7321Z53 0000CZC6192FTM 0000CZC8191GG1 0000CZC8410NLN 0000CZC8393VLY 0000CZC8410NLM 0000CZC8410NLJ 0000CZC8410NLH 0000CZC8410NL6 0000CZC71108XW 0000CZC8393VMD 0000CZC8462DC2 0000CZC84643GV 0000CZC8464GFL 0000CZC8393VM4 0000CZC82388GS 0000CZC8393VM0 0000CZC8295TPF 0000CZC7333J81 0000CZC8410NLD 0000CZC813012R 000CZC77414GS3 0000CZC8130135 0000CZC8207NVP 0000CZC8187QPJ 0000CZC7452MPB 0000CZC7521JWM 0000CZC813012Q 0000CZC813012Z 0000CZC73706DG 0000CZC74003CQ 0000CZC73706DD 0000CZC74003D3 0000CZC7414GSG 0000CZC7393C25 0000CZC7321Z54 0000CZC7393C23 0000CZC7393C1V 0000CZC7321Z4V 0000CZC7331WV3 0000CZC72449PG 0000CZC7321Z59 0000CZC7321Z51 00000042989AW4 00000042996AW9 00000043008AW2 00000483828BU0 0000CZC6071411 0000CZC6122LC7 0000CZC6152BYK 00000033855AW8 00000021001AW3 0000CZC6103V2F 0000CZC6122LC8 00000020943AW7 00000483800BU9 00000016690AW0 00000016676AW9 0000CZC8207NVH 0000CZC6040X37 0000CZC61908DF 0000CZC9064P9L 0000CZC8303V6Q 000000U9126001 0000CZC9064PCV 000000U8162014 0000CZC7414MHL 0000CZC8024SDG 0000CZC6152BYG 0000CZC64148ZM 0000CZC8295TNZ 0000CZC8295TPC 0000CZC80602B4 0000CZC9064PC8 0000CZC9064PBC 0000CZC8464GFM 0000CZC8464GFV 0000CZC8464GG3 000000A8177057 0000CZC84643GT 0000CZC8410NL4 0000CZC8295TP0 0000CZC83331FR 0000CZC8295TP1 0000CZC82388GV 0000CZC75109BC 0000CZC75109B6 0000CZC806029T 0000CZC751099S 0000CZC751099X 000000A7201008 0000CZC643501H 0000CZC7262J7G 0000CZC7193XVN 0000CZC7193XVN 0000CZC72449PP 0000CZC7182HNM 0000CZC7023M5P 00000044368AW9 00000044328AW3 0000CZC626329Y 0000CZC549318R 0000CZC5292LNJ 0000CZC84643H1 0000CZC617421B 0000CZC7182HNS 000000ANT23704 000000B7166009 0000CZC7521RVW 0000CZC75109B8 0000CZC7521JWK 0000CZC7432L1X 0000CZC7280B2M 0000CZC7182LWH 0000CZC70416YL 00000035352AW4 0000CZC6311K00 0000CZC5480XP5 0000CZC82202Q4 0000CZC7521RVQ 0000CZC80602B9 0000CZC80602B3 0000CZC5292LNM 0000CZC7182HNJ 0000CZC71145V5 00000030991AW4 00000359667CN3 0000CZC6083SBT
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    Worldwide Distribution: USA in the states of AL, AK, AZ, AR, CA, CO, CT, FL, GA, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI,MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WV, WI, and WY, and to the countries of UNITED KINGDOM, TURKEY, TAIWAN, SWITZERLAND, SPAIN, SOUTH, KOREA, SAUDI ARABIA , PORTUGAL, NEW ZEALAND, NETHERLANDS, KUWAIT, JAPAN, ITALY, GERMANY, FRANCE , CHINA, CANADA, BRAZIL, BELGIUM, AUSTRIA, AUSTRALIA, VENEZUELA, UNITED ARAB EMIRATES, THAILAND, SWEEDEN, SOUTH AFRICA, SLOVENIA, SINGAPORE, RUSSIA, PHILLIPNES, NORWAY, POLAND, MOROCCO, MEXICO, MARTINIQUE, MALAYSIA, LUXEMBOURG, LEBANON, JORDAN, JAMAICA, ISRAEL, IRELAND, INDIA, ICELAND, HUNGARY, HONG KONG, HONDURAS, GUADOPUO, GREECE, GEORGIA, FINLAND, EGYPT, DOMINICAN REPUBLIC, DENMARK, CZECH REPUBLIC, CROATIA, COLOMBIA, BULGARIA, BOTSWANA, and ALGERIA.
  • Product Description
    GE Seno Advantage 1.x, Seno Advantage 2.0, Seno Advantage 2.1
  • Manufacturer

Manufacturer

  • Manufacturer Address
    GE Healthcare, LLC, 3000 N Grandview Blvd, Waukesha WI 53188-1615
  • Manufacturer Parent Company (2017)
  • Source
    USFDA