Recall of Device Recall S/5 Anesthesia Delivery Unit

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by GE Healthcare.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    34631
  • Event Risk Class
    Class 2
  • Event Number
    Z-1196-06
  • Event Initiated Date
    2006-03-22
  • Event Date Posted
    2006-07-12
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2007-05-29
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Anesthesia Delivery Unit - Product Code BSZ
  • Reason
    There exists the possibility that the adu fresh gas control unit (a-fgc1) can shut off during normal handling of the anesthesia machine and result in no anesthetic agent output.
  • Action
    An Urgent Medical Device Correction letter dated March 21, 2006, was sent to customers. It warns of the potential for no anesthetic agent output. The letter contains short term recommendations for use of the device until a GE Healthcare representative contacts the facility to schedule a vist and implelment corrective action.

Device

  • Model / Serial
    S/5 Anesthesia Delivery Unit serial numbers: 6028818, 6048768, 6048777, 6048778, 6048795, 6048797, 6048799, 6048808, 6048811, 6048812, 6048815, 6048827, 6067704, 6067706, 6085914, 6085926, 6085927, 6085929, 6085935, 6085937, 6086576, 6086577, 6086583, 6086585, 6086588, 6086589, 6086592, 6086593, 6086595, 6086596, 6086600, 6086601, 6086602, 6086603, 6086608, 6086610, 6086611, 6086612, 6086613, 6097355, 6097356, 6097357, 6097364.  A-FGC1 Units serial numbers: 6020250, 6044947, 6044950, 6044951, 6045113, 6045120, 6045122, 6045127, 6045138, 6045151, 40022777, 40100065, 40106020, 40114446, 40115565, 40119019, 40121119, 40149506, 40173259.
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    World wide-(including the States of IA, CA, MI, FL, MI, TX, IL, WI, OH, and countries of Finland, Singapore, Austria, South Africa, Sweden, Israel, Norway, Taiwan, Germany, Turkey, Portugal, Italy, Switzerland, Great Britain, and France
  • Product Description
    Fresh Gas Control (A-FGC1) Unit, a component of the S/5 Anesthesia Delivery Unit. The Anesthesia Delivery Unit is designed to mix and dose respiratory gases and to ventilate the patient. The delivery unit is the basis for a flexible anesthesia system with full monitoring and data management capabilities.
  • Manufacturer

Manufacturer

  • Manufacturer Address
    GE Healthcare, 3030 Ohmeda Drive, PO Box 7550, zip 53707-7550, Madison WI 53718
  • Source
    USFDA