Recall of Device Recall Roche COBAS 6000

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Roche Diagnostics Corp..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    45984
  • Event Risk Class
    Class 2
  • Event Number
    Z-0576-2008
  • Event Initiated Date
    2007-12-04
  • Event Date Posted
    2008-03-04
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2008-08-04
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Clinical Chemistry Analyzer - Product Code JJY
  • Reason
    Incorrect results: crp assays may be calculated using incorrect calibration parameters which could result in falsely high or falsely low patient results being reported.
  • Action
    Roche Diagnostics contacted the consignee by phone on 12/4/07 and by follow-up letter dated 12/4/07. The user was instructed to discontinue use of two CRP assays on this analyzer. The letter described a workaround and that a new revision of software, due out first quarter 2008, will fix this.

Device

  • Model / Serial
    All cce and cc system configurations with core unit software version 02-03.
  • Product Classification
  • Device Class
    1
  • Implanted device?
    No
  • Distribution
    USA Distribution in the state of Nebraska.
  • Product Description
    Roche COBAS 6000 Clinical Chemistry analyzer; Catalog Number: 04745914001, Roche Diagnostics Corporation, Indianapolis, IN 46250.