Recall of Device Recall Roche

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Roche Molecular Systems, Inc..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    73680
  • Event Risk Class
    Class 2
  • Event Number
    Z-1578-2016
  • Event Initiated Date
    2016-03-03
  • Event Date Posted
    2016-04-27
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2017-05-25
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Clinical sample concentrator - Product Code JJH
  • Reason
    The hand-held barcode scanner model it3800 used with the cobas ampliprep instrument mis-identified a sample barcode id.
  • Action
    Roche Molecular Systems sent an "Urgent Medical Device Correction (UMDC) notification letter dated March 3, 2016, to their affected customers. Customers were instructed to provide the UMDC to other affected parties who may need to be aware of this issue. Customers were asked to complete the attached fax form and fax it to 1-866-737-5396. Customers with questions should contact Roche Support Network Customer Support Center at 1-800-526-1247. For questions regarding this recall call 908-253-7200.

Device

Manufacturer

  • Manufacturer Address
    Roche Molecular Systems, Inc., 1080 US Highway 202 S, Branchburg NJ 08876-3733
  • Manufacturer Parent Company (2017)
  • Source
    USFDA