Recall of Device Recall QuikClot Trauma Pad

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Z-Medica, LLC.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    79407
  • Event Risk Class
    Class 2
  • Event Number
    Z-0925-2018
  • Event Initiated Date
    2017-11-17
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2018-06-04
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Dressing, wound, drug - Product Code FRO
  • Reason
    A customer complained that one of the pouches in a box of 10 had "quikclot roll" pouch instead of "quikclot traumapad" pouch. the carton as well as the remaining 9 pouches were properly identified as quikclot trauma pad. the customer alleged that when the pouch was opened, it contained a quikclot traumapad. all 10 pouches in the box had the correct part number, lot number, and expiration date for the prod uct (quikclot traumapad).
  • Action
    On November 22, 2017 an URGENT ADVISORY NOTICE was issued to customers stating the following: The QuikClot TraumaPad is packaged with 10 pouches in a carton. This lot may contain some pouches that incorrectly identify the product as QuikClot Roll. All pouches (identified as TraumaPad or Roll) are identified with the correct Part N umber (REF 460), Lot Number, and Expiration Date for QuikClot TraumaPad, and contain the QuikClot TraumaPad product. Please note that the labeling on the cartons is also correct and we believe that this only impacts a small num ber of pouches. In addition, QuikClot TraumaPad and QuikCot Roll have the identical ingredients, chemical composition, indications for use, packaging material and surface area. The only difference is that the QuikClot TraumaPad is a 12 inch x 12 inch, 3-ply pad and the QuikClot Roll is a 3 inch x 4 yard roll. Both products are equally effective in being used to stop bleeding. This was the result of an isolated operator error. Please note that actions have been taken to prevent this issue in the future and further corrective actions are under review. Ifyou have any inventory of this lot number, you may choose to use the prod uct as the correct prod uct is packaged into each pouch (only pouch labeling is incorrect). Ifyou chose to destroy the prod uct or have any questions regardi ng this matter, please contact us at the email and telephone number below.

Device

  • Model / Serial
    Part# 460, Lot# 7397
  • Product Classification
  • Device Class
    Unclassified
  • Implanted device?
    No
  • Distribution
    Nationally
  • Product Description
    QuikClot TraumaPad, Part# 460 a topical dressing for local management of bleeding wounds such as cuts, lacerations, and abrasions. It may also be used for temporary treatment of severely bleeding wounds such as surgical wounds (operative, postoperative, dermatological, etc.) and traumatic injuries
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Z-Medica, LLC, 4 Fairfield Blvd, Wallingford CT 06492-1857
  • Manufacturer Parent Company (2017)
  • Source
    USFDA