Recall of Device Recall PKG, TROCAR, 11MM, PYRAMIDAL TIP

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Stryker Endoscopy.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    69838
  • Event Risk Class
    Class 2
  • Event Number
    Z-0663-2015
  • Event Initiated Date
    2014-11-17
  • Event Date Posted
    2014-12-12
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2015-07-17
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Endoscope and/or accessories - Product Code KOG
  • Reason
    The parameters provided in the laparoscopic manual instruments instructions for use (ifu 1000-401-070 revision g or prior) do not support the unwrapped gravity cycle and the ethylene oxide cycle sterilization methods.
  • Action
    Stryker Endoscopy sent an Urgent Device Correction letter dated November 17, 2014, to affected customers. The letter informed the customers of the potential hazards and actions required. Customers were instructed to discard any old Instructions for Use for their Laparoscopic Manual Instruments and Accessories. Download the new Instuctions for Use (IFU1000-401-070 Revision J) from the URL link below: www.stryker.com/IFU1000401070 Customers were instructed to complete and sign the enclosed "Acknowledgement of Receipt" form and email them to endorecall@stryker.com or fax to (408) 855-6314. Customers with questions were instructed to contact Stryker Endoscopy at 408-855-6238 or via email at endorecall@stryker.com. For questions regarding this recall call 408-754-2000.

Device

  • Model / Serial
    197 38410 38539 39800 43905 44707 45337 45663 46563 47266 47477 47499 47500 47957 48256 48346 48694 49042 49141 50045 51396 51915 53499 54427 57166 57384 57979 58704 59130 59453 60572 61045 62137 63233 64030 64480 66874 67351 67645 67981 69931 73431 75611 75612 76147 76197 76198 79810 80433 83797 84016 84191 85143 85382
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.
  • Product Description
    PKG, TROCAR, 11MM, PYRAMIDAL TIP, P/N 0250080156. || Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Stryker Endoscopy, 5900 Optical Ct, San Jose CA 95138-1400
  • Manufacturer Parent Company (2017)
  • Source
    USFDA