Recall of Device Recall Phiips Medical Systems

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Philips Healthcare Inc..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    57846
  • Event Risk Class
    Class 2
  • Event Number
    Z-1609-2011
  • Event Initiated Date
    2011-02-07
  • Event Date Posted
    2011-03-08
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2017-03-16
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Automated External Defibrillators (non-wearable) - Product Code MKJ
  • Reason
    Software: etco2 and ventilation rate values may be labeled incorrectly.
  • Action
    Philips Healthcare representatives issued an Urgent Medical Device letter on 2/7/11. Customers are asked to follow the Action to be taken by Customer/User section of the Urgent Medical Device Correction Notification. In this section, customers are given the following instructions: Philips software (SW) version F.01.03 (for M3535A and M3536A) and R.01.00 (for M3536A, English only) corrects this problem. If you have affected product a Philips Service Engineer will contact you to install software version F.01.03 or R.01.00. Until your software is updated your HeartStart MRx monitor/defibrillator can remain in service. If you are using the device in one of the use modes listed under problem description, make sure that operators are aware that even though the EtCO2 and ventilation rate values can display interchangeably (unexpectedly or randomly switch) under either the ventilation rate label or the EtCO2 label, the values themselves can be identified by their color (for example if the ventilation rate label is white, the ventilation value is displayed as white). Customers are to contact their local Philips representative at 1-800-722-9377, if they need further information or support regarding this issue.

Device

  • Model / Serial
    Software versions: F.01.00 or F.01.01
  • Product Classification
  • Device Class
    3
  • Implanted device?
    No
  • Distribution
    Worldwide -- USA, Canada, AUSTRALIA, AUSTRIA, BAHRAIN, BELGIUM, BRAZIL, CHINA, COLOMBIA, CZECH REPUBLIC, DENMARK, EGYPT, ESTONIA, FRANCE, GERMANY, HONG KONG, INDIA, IRELAND, ISRAEL, ITALY, KENYA, LATVIA, MALAYSIA, MEXICO, NEW ZEALAND, NORWAY, PHILIPPINES, REUNION, SAUDI ARABIA, SINGAPORE, SOUTH AFRICA, SPAIN, SWITZERLAND, THAILAND, UNITED ARAB EMIRATES, and UNITED KINGDOM.
  • Product Description
    Philips HeartStart MRx Defibrillator Monitor (M3535A/M3536A) with software (SW) versions F.01.00 or F.01.01 and with all of the following hardware options EtCO2, IP and Temp. || Used for the termination of ventricular tachycardia and ventricular fibrillation.
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Philips Healthcare Inc., 3000 Minuteman Road, Andover MA 01810
  • Manufacturer Parent Company (2017)
  • Source
    USFDA