Recall of Device Recall P.F.C.

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Depuy Orthopaedics, Inc..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    32192
  • Event Risk Class
    Class 2
  • Event Number
    Z-1275-05
  • Event Initiated Date
    2005-05-26
  • Event Date Posted
    2005-08-12
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2005-12-02
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Metal/Polymer, Mobile Bearing - Product Code NJL
  • Reason
    An incomplete seal on the inner pouch allows exposure to oxygen, which may result in oxidation and would affect long-term performance of the implant.

Device

  • Model / Serial
    Lots 84470A, 86123A, 86300A and 86585A.
  • Product Classification
  • Device Class
    3
  • Implanted device?
    Yes
  • Distribution
    Nationwide, Argentina, Brazil, Canada, Caguas, Colombia, Ireland, Malaysia, Mexico, Panama, Peru, Singapore, South Africa, Sweden, Uruguay and Venezuela.
  • Product Description
    Depuy P.F.C. E Knee System, stabilized all-polyethylene tibia, Sz 2.5, 12.5 mm; Ref 96-2632.
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Depuy Orthopaedics, Inc., 700 Orthopaedic Dr, Warsaw IN 46582-3994
  • Source
    USFDA