Recall of Device Recall Outlook Pump IV Set w/Universal Spike, 3 Safeline Inj Sites, Spin Lock Conn & Pressure Limited Check

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by B. Braun Medical, Inc..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    53955
  • Event Risk Class
    Class 2
  • Event Number
    Z-0703-2010
  • Event Initiated Date
    2009-11-24
  • Event Date Posted
    2010-01-26
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2011-02-14
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Intravascular administration set - Product Code FPA
  • Reason
    System error 9 alarms. a system error 9 alarm indicates a potential pressure loss in the pump cassette delivery chamber (chamber closest to the patient). this alarm is generated during the pumps initiation sequence and will not allow the set to be used. this requires a new set to be obtained, thus creating a potential for delay in therapy.
  • Action
    The recalling firm issued a Product Removal Notice via US Postal Service Certified Mail with registered return receipt or FedEx to all customers and distributors in receipt of the affected product. The notices were sent 11/30/09 and 12/1/09. Direct customers also received letters addressed to Material Management that informed them of the recall and provided instruction for disposition of affected product in their inventory. Distributors received the direct customer recall letter along with a cover letter outlining their responsibilities to execute the recall and notify their customers. Questions are directed to the company's Customer Support Departments: US customers please call (800) 227-2862. Canadian customers, please call (800) 624-2920. International customers, please call (610) 691-5400.

Device

  • Model / Serial
    Product Code NF3140. Lot numbers: 61040801 exp 2/28/2012, 61040802 exp 2/28/2012, 61040803 exp 2/28/2012, 61040804 exp 3/30/2012, 61040805 exp 3/30/2012, 61040806 exp 3/30/2012, 61044303 exp 3/30/2012, 61044304 exp 4/30/2012, 61049787 exp 4/30/2012, 61049790 exp 4/30/2012, 61049793 exp 4/30/2012, 61049794 exp 4/30/2012, 61049795 exp 5/31/2012, 61049796 exp 5/31/2012, 61049797 exp 5/31/2012, 61049798 exp 5/31/2012, 61049799 exp 5/31/2012, 61049800 exp 5/31/2012, 61049801 exp 5/31/2012, 61057015 exp 6/30/2012, 61057016 exp 6/30/2012, 61057017 exp 6/30/2012, 61057018 exp 6/30/2012, 61057019 exp 6/30/2012, 61057200 exp 6/30/2012, 61057201 exp 7/30/2012, 61072172 exp 7/30/2012, 61072173 exp 7/30/2012, 61072174 exp 7/30/2012, and 61072175 exp 7/30/2012.
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    The products were shipped to distributors and medical facilities nationwide. The product was also shipped to Afghanistan, Canada and LU.
  • Product Description
    Outlook Pump IV Set w/Universal Spike, 3 Safeline Inj Sites, Spin Lock Conn & Pressure Limited Check Valve 134 in
  • Manufacturer

Manufacturer

  • Manufacturer Address
    B. Braun Medical, Inc., 901 Marcon Boulevard, Allentown PA 18109-9512
  • Manufacturer Parent Company (2017)
  • Source
    USFDA