Recall of Device Recall ONQ PainBuster with OnDemand

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by I-Flow Corporation.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    44933
  • Event Risk Class
    Class 2
  • Event Number
    Z-0057-2008
  • Event Initiated Date
    2007-03-27
  • Event Date Posted
    2007-10-06
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2009-10-07
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Infusion pump - Product Code MED
  • Reason
    Misassembly: the on-q painbuster pmb01 package may contain a soaker catheter instead of a non-soaker catheter. as a result, the bolus button may remain latched in the down position which may cause the the device to deliver continuously at 7 ml/hr instead of 2 ml/hr.
  • Action
    An Important Recall Notification Letter was faxed to affected customers on March 27, 2007. Affected customers were notified of the recall, the reason for recall, and instructed to immediately quarantine any remaining inventory. A Customer Response (and Fax back) form was included with the Recall Notification. Affected customers were requested to return the FAX back form to I-Flow within 48 hours of receipt. Replacement product will be sent out as applicable.

Device

  • Model / Serial
    6A4171 (exp. MAR 2009), 694133 (exp. FEB 2009), 654920 (exp. OCT 2008), 5A4732 (exp. MAR 2008), 584699 (exp. JAN 2008)
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    Worldwide, USA and Australia
  • Product Description
    ON-Q PainBuster with OnDemand (270 ml, 2 ml/hr + 5 ml bolus /60 min refill), Model PMB01, I-Flow Corporation, Lake Forest, CA 92630 U.S.A.
  • Manufacturer

Manufacturer

  • Manufacturer Address
    I-Flow Corporation, 20202 Windrow Dr, Lake Forest CA 92630-8152
  • Source
    USFDA