Recall of Device Recall OEC 6800 Miniview

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by GE OEC Medical Systems, Inc.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    62382
  • Event Risk Class
    Class 1
  • Event Number
    Z-2040-2012
  • Event Initiated Date
    2006-11-20
  • Event Date Posted
    2012-07-24
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2013-05-13
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Image-intensified fluoroscopic x-ray system, mobile - Product Code OXO
  • Reason
    Please be aware that this is not a new recall. the firm has taken action; but, due to administrative issues this recall is now being classified by the agency. the oec 6800 miniview was recalled in november 2006 due to several intermittent potential safety issues including navigation tracking inaccuracy, system lock-up, incorrect image display, incorrect dose area calculated air kerma in "film mo.
  • Action
    Please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. On 11/20/2006, two written notifications were sent to 1513 identified consignees, addressed to the Facility Administrator and the Radiology Administrator for a combined total of 3026 letters. The Urgent Recall Notice, dated 11/8/2006, advised customers that GE Healthcare had identified several intermittent potential safety issues that may occur with the identified affected product. The notification also provided interim or permanent solutions for each of the problems described. For the issues that did not have permanent solutions at the time, the firm was to contact customers and remedy the issues free of charge. For the issues that did have a permanent fix at the time, the GEHC OEC service team was to contact customers to install the correction. If customers had questions or concerns, they were to contact the service team for further information at 800-874-7378 option 8. On 04/23, 2007, a second written notification was sent via certified mail to the same 1513 consignees providing further clarification on the issue of "incorrect Image Display" and a temporary remedy to the issue.

Device

  • Model / Serial
    86-1834 , 86-1175 , 86-0238 , 86-0585 , 86-1230 , 86-1992 , 86-0184 , 86-1033 , 86-1480 , 86-1369 , 86-0579 , 86-1678 , 86-1059 , 86-0639 , 86-0389 & 86-0390 , 86-0775 , 86-0850 & 86-1703 , 86-0151 & 86-1311 , 86-1883 , 86-0048 & 86-2015 , 86-0915, 86-1626 & 86-1631 , 86-2118 , 86-0949 , 86-0937 , 86-0071 & 86-1015 , 86-0352 , 86-0527 , 86-0140 , 86-1697 , 86-0697 , 86-1162 , 86-1424 , 86-0795 , 86-0260 , 86-0077 , 86-2002 , 86-0811 , 86-2148 , 86-1582 , 86-1633 , 86-0454 , 86-0185, 86-1119 & 86-1885 , 86-0668 , 86-1026 , 86-0529 , 86-1754 , 86-0901 , 86-0228 , 86-1555 , 86-1212 , 86-1095 , 86-0997 , 86-1475 , 86-2137 , 86-0829 , 86-1737 , 86-1259 , 86-0013 , 86-1139 , 86-0058 , 86-0327 , 86-1346 , 86-0419 & 86-2047 , 86-1892 , 86-0359 & 86-1070 , 86-0331 , 86-1255 , 86-0785 , 86-1043 , 86-1854 , 86-0148 & 86-1036 , 86-0427 & 86-2068 , 86-1988 , 86-0092 , 86-1283 , 86-1955 , 86-0711 , 86-2035 , 86-2017 , 86-0770 & 86-2044 , 86-0716 , 86-0692 , 86-1610 , 86-0154 & 86-1822 , 86-1727 , 86-1209 & 86-1210 , 86-0046 , 86-1303 , 86-0468 , 86-1274 , 86-0113 , 86-0035 , 86-1481 , 86-1808 , 86-1332 , 86-0618 , 86-1860 , 86-0425 , 86-0912 , 86-0925 , 86-0306 & 86-0664 , 86-0987 , 86-2170 , 86-0308 , 86-1302 , 86-0399 , 86-1205 , 86-1615 , 86-2128 , 86-0162 , 86-0166 , 86-0105 , 86-1687 & 86-1844 , 86-0865 , 86-1045 , 86-1360 , 86-0903 , 86-1456 , 86-1213 , 86-1326 , 86-0745 , 86-0517 , 86-2140 , 86-0625 & 86-1399 , 86-0923 , 86-1319 & 86-2133 , 86-1499 , 86-1823 , 86-0401 , 86-0157 , 86-0685 , 86-0570 , 86-1446 , 86-1606 , 86-2030 , 86-0086 , 86-0075 & 86-0076 , 86-0435 , 86-1780 , 86-0247-RC , 86-1478 , 86-1894 , 86-0896 , 86-1249 , 86-0601, 86-0914 & 86-1248 , 86-1250 , 86-1836 , 86-0805, 86-0806 & 86-0972 , 86-1619 , 86-1576 , 86-0544 , 86-1886 , 86-1782 , 86-0475 , 86-0321 , 86-0511 , 86-1949 , 86-1150 , 86-1065 , 86-1614 , 86-1907 , 86-1625 , 86-0757 & 86-0786 , 86-2092 , 86-0501 , 86-2117 , 86-1879 , 86-1714 , 86-0416 , 86-1506 , 86-1732 , 86-0977 , 86-0964 , 86-1531 , 86-0994 & 86-1909 , 86-0091 , 86-0115 , 86-1154 , 86-2171 , 86-0739 , 86-1621 & 86-1721 , 86-1647 , 86-0589 , 86-1490 , 86-0485 , 86-0995 , 86-0657 , 86-0700 , 86-1620 , 86-2032 , 86-0487 & 86-0494 , 86-1343 , 86-0102 , 86-2005 , 86-0504 , 86-1178 , 86-0329 , 86-0777 & 86-1273 , 86-1187 , 86-0519 , 86-1536 , 86-1448 , 86-0572 , 86-1223 , 86-2199 , 86-1880 , 86-0827 , 86-0068, 86-0155 & 86-0239 , 86-0492 , 86-0558 & 86-1966 , 86-0347 & 86-0411 , 86-1324 , 86-0191 , 86-1166 , 86-0701 , 86-2008 , 86-1219 , 86-0737 , 86-0974 , 86-2095 , 86-1460 , 86-0381 , 86-1234 , 86-0890 , 86-0099 , 86-1420 , 86-0750 , 86-0259 , 86-1226 , 86-0204 , 86-0482 , 86-0378 , 86-2194 , 86-1293 , 86-0966 , 86-1111 , 86-1775 & 86-1965 , 86-0069 , 86-0137, 86-0428 & 86-0432 , 86-1813 , 86-0978 , 86-0609 , 86-0047 , 86-0070 , 86-1646 & 86-1838 , 86-0906 , 86-1432 , 86-1391 , 86-1431 , 86-1266 & 86-2238 , 86-2239 , 86-1375 , 86-0560 , 86-1686 , 86-1218 , 86-1308 , 86-0277 , 86-0755 & 86-1959 , 86-1179 , 86-0245 & 86-0253 , 86-2056 , 86-1151 , 86-2043 , 86-1192 & 86-1632 , 86-0840 , 86-1728 , 86-0025 , 86-1022 , 86-0871 , 86-0163 , 86-0271 , 86-1962 , 86-0153 & 86-1267 , 86-1523 & 86-1560 , 86-1611 , 86-0732 , 86-1540 , 86-1094 , 86-0573 & 86-1211 , 86-1494 , 86-0198 , 86-0388 , 86-0200 & 86-1381 , 86-2087 , 86-0981 & 86-1713 , 86-0120 , 86-1914 , 86-1972 , 86-1918 , 86-0106 , 86-1858 & 86-1889 , 86-1869 , 86-1583 , 86-1085 , 86-0020 , 86-0913 & 86-2126 , 86-0674-RC , 86-1652 , 86-0520 , 86-0131 , 86-1542 , 86-0036 , 86-0577 , 86-1544 , 86-0206 , 86-0731 , 86-2115 , 86-1056 , 86-0237 , 86-0358 , 86-0361 , 86-0870 , 86-0541 & 86-0980 , 86-1123 , 86-2193 , 86-2026 , 86-0298 , 86-0214 , 86-2011 , 86-0042 , 86-0918 , 86-0441 , 86-2147 , 86-1509 , 86-1428 & 86-2057 , 86-2077 , 86-1607 , 86-0299 , 86-0973 , 86-0583 & 86-0584 , 86-0575 , 86-0346, 86-0556 & 86-2113 , 86-2018 , 86-0552 , 86-2110 , 86-0624 , 86-0403 , 86-1443 , 86-0736 , 86-1579 & 86-1580 , 86-1029 , 86-1142 , 86-1009 , 86-2075 , 86-0804 , 86-1784 , 86-1513 , 86-0567 , 86-1807 , 86-0032 , 86-0255 , 86-1398 , 86-1067 , 86-1258 , 86-1960 , 86-2121 , 86-0832 & 86-2210 , 86-1317 , 86-0364 & 86-1565 , 86-1690 , 86-0064 , 86-1884 , 86-2165 , 86-1891 , 86-0594 , 86-1134 , 86-1068 & 86-1963 , 86-0452 , 86-0703 , 86-0790 , 86-2227 , 86-0838 , 86-1115 , 86-1099 , 86-0112 & 86-0715 , 86-1521 , 86-1736 , 86-2006 , 86-0996 , 86-1541 & 86-1812 , 86-0460 , 86-1133 , 86-0291 , 86-0720 , 86-0828 , 86-1489 , 86-0845 , 86-1849 , 86-0549, 86-0673 & 86-1575 , 86-1393 , 86-1388 , 86-2039 , 86-0695 , 86-0138 , 86-2157 , 86-0641 , 86-0443 , 86-0365 , 86-0430 , 86-0722 , 86-1113 , 86-0897 , 86-0109 & 86-1778 , 86-0173 , 86-0798 , 86-0262 , 86-1505 , 86-0917 , 86-0680 , 86-1841 , 86-1404 , 86-1217 , 86-0951 , 86-1831 , 86-0458 , 86-2119 , 86-0215, 86-1720 & 86-1920 , 86-0275 , 86-1698 , 86-1818 , 86-2187 , 86-1278 , 86-1657 , 86-0791 , 86-0478 , 86-1373 , 86-0328 , 86-0891, 86-0893 & 86-1338 , 86-0895 , 86-0894 , 86-1791 , 86-0554 , 86-1495 , 86-2101 , 86-0776 , 86-0107, 86-0167, 86-0766 & 86-1176 , 86-0121 , 86-0767 , 86-1550 , 86-2226 , 86-0424 , 86-0164 , 86-0844 , 86-0753 , 86-1125 , 86-1018 , 86-1126, 86-1268 & 86-1269 , 86-1156 , 86-0413 , 86-0282 , 86-1000 , 86-1037 , 86-0886 , 86-0406 , 86-0159 , 86-0332 , 86-0873 , 86-0586 , 86-1430 , 86-2205 , 86-0521 , 86-0348 , 86-0778 , 86-0402 , 86-0836 , 86-0887 , 86-0792 , 86-1551 , 86-2139 , 86-0085 & 86-2097 , 86-1999 , 86-0339 , 86-0864 , 86-0822 , 86-0506 , 86-0270-RC , 86-2003 , 86-1746 & 86-1749 , 86-1057 , 86-0802 , 86-0212 , 86-2122 , 86-0783 , 86-2135 , 86-0338, 86-0553, 86-0862 & 86-1900 , 86-1157 , 86-2107 , 86-1316 , 86-1627 , 86-0279 , 86-1587 , 86-0702 , 86-0935-RC , 86-1915 , 86-1368 , 86-1590 , 86-0831 , 86-0746 , 86-0257 , 86-2111 , 86-0311 , 86-0136 , 86-1702 , 86-0334 & 86-1733 , 86-0596, 86-0613, 86-0614, 86-0868 & 86-1724 , 86-0309 , 86-0921 , 86-1855 , 86-1527 , 86-1216 , 86-1385 , 86-1695 , 86-0826 , 86-1529 , 86-1515 , 86-0375 , 86-1761 & 86-1764 , 86-0616 , 86-0859 , 86-0843 , 86-1819 , 86-0263 , 86-1982 , 86-0649 , 86-0227 , 86-0742 & 86-2004 , 86-0663 , 86-1638 , 86-0170 , 86-0097 & 86-0100 , 86-1581 , 86-2168 & 86-2198 , 86-0223 , 86-0683 , 86-0201 , 86-0203 & 86-1688 , 86-0769 , 86-0156 , 86-2232 & 86-2234 , 86-0057 , 86-1201 , 86-1287 , 86-1856 , 86-1027 , 86-1772 , 86-1260 , 86-1310 , 86-0817 , 86-0984 , 86-1670 , 86-0957 , 86-0119 , 86-2203 , 86-0333 , 86-1926 , 86-1704 , 86-0067 , 86-1870 , 86-0610 , 86-0597 , 86-0171, 86-0763 & 86-0883 , 86-0030 , 86-1138 , 86-2136 , 86-0061, 86-0453, 86-1097 & 86-1357 , 86-1300 , 86-0985 , 86-1342 , 86-1304 , 86-1204 , 86-2176 & 86-2177 , 86-0867 , 86-1659 & 86-0547 , 86-1184 & 86-1487 , 86-1055 , 86-1106 , 86-0623 & 86-1693 , 86-1003 , 86-0125 , 86-2197 , 86-1636 , 86-1532 , 86-0360 , 86-1158 , 86-0869 , 86-0533 , 86-1671 , 86-0765 , 86-0310 & 86-1630 , 86-1305 , 86-1307 , 86-0369 , 86-0652 , 86-1314 , 86-1829 , 86-1155 , 86-0190 , 86-0307 , 86-1572 , 86-0513 , 86-0250 , 86-1107 , 86-1588 , 86-0486 , 86-1207 , 86-0124 , 86-0098 , 86-0193 , 86-0049 & 86-0648 , 86-1906 , 86-1277 , 86-1121 , 86-0095 , 86-1124 , 86-1117 , 86-2055 , 86-1421 & 86-1445 , 86-0861 , 86-0660 , 86-1923 , 86-1140 , 86-2222 , 86-0525 , 86-0599 , 86-1722 , 86-1779 , 86-0962 & 86-1851 , 86-1423 , 86-1806 & 86-1957 , 86-0706 , 86-1696 , 86-1866 , 86-1425 , 86-2099 , 86-1500 & 86-2207 , 86-1345 , 86-0412 , 86-0932 , 86-1561, 86-1876 & 86-1878 , 86-0684 , 86-2164 , 86-1719 , 86-0383 & 86-1589 , 86-0114 , 86-1102 , 86-0322 , 86-0983 , 86-1998 , 86-1832 , 86-0396 , 86-0342 , 86-1815 , 86-1110, 86-1933 & 86-1986 , 86-0283 , 86-1637, 86-2038 & 86-2060 , 86-1491 , 86-1108 , 86-1927 , 86-2191 & 86-2192 , 86-0080 , 86-2130 , 86-0699 & 86-1241 , 86-0374, 86-1603, 86-1604 & 86-1605 , 86-1752 , 86-1514 , 86-0074 & 86-1147 , 86-1821 , 86-1781 , 86-0251 , 86-0229 , 86-0857 , 86-0920 , 86-1974 , 86-0559 & 86-0719 , 86-0366 , 86-1001 , 86-1845 , 86-1661 , 86-1881 , 86-0122 , 86-1076 , 86-0916 & 86-1435 , 86-1052 , 86-0384, 86-0385, 86-0636 & 86-1066 , 86-0281 , 86-1379 , 86-1377 , 86-1867 , 86-0631 , 86-0417 & 86-1503 , 86-1221 , 86-1383 , 86-2209 , 86-2208 , 86-0244 , 86-0408 & 86-1562 , 86-0490 , 86-1916 & 86-1917 , 86-0910-RC , 86-0797 , 86-1047, 86-1786 & 86-1940 , 86-0545 & 86-0735 , 86-1535 , 86-1852 , 86-1622 , 86-2007 , 86-2014 , 86-0199 , 86-1941 , 86-0650 , 86-0789, 86-1612 & 86-1613 , 86-1476 , 86-1742 , 86-2146 , 86-0540 , 86-1014 , 86-0038 , 86-0420, 86-0421 & 86-1566 , 86-1069 , 86-1079, 86-1082 & 86-1098 , 86-1143 , 86-1188 , 86-1510 , 86-1112 , 86-0470 , 86-1465 , 86-1231 , 86-1041 , 86-1297 , 86-0813, 86-0880 & 86-1251 , 86-0236 , 86-0976 , 86-0855 , 86-0084 & 86-0698 , 86-1429 , 86-0093 , 86-1654 , 86-0226, 86-0979 & 86-1382 , 86-1729 , 86-0550 , 86-0264 , 86-1888 , 86-1785 , 86-0444 , 86-1252 , 86-1595 & 86-1596 , 86-2237 , 86-0542 , 86-0999 , 86-1279 , 86-1694 , 86-0415, 86-0833 & 86-0900 , 86-0841 , 86-0543 , 86-0174 , 86-1214 , 86-0219 , 86-1053 , 86-0508 , 86-0904 , 86-1093 , 86-0394 , 86-0371 , 86-0576 , 86-1196 , 86-1669 , 86-0286 , 86-2138 , 86-1634 , 86-0644 , 86-0738 , 86-0688 & 86-1790 , 86-1903 , 86-0712 , 86-1136 , 86-0955 , 86-0687 & 86-2034 , 86-1182 , 86-1286 , 86-0571 , 86-1254 , 86-0507 , 86-0045 , 86-1402 & 86-1537 , 86-0192 , 86-0370 & 86-0620 , 86-1276 , 86-0284 , 86-1450 , 86-1451 & 86-1452 , 86-1816 , 86-0637 , 86-0989 , 86-1597 , 86-0592 , 86-1890 , 86-0600 & 86-1272 , 86-1953 , 86-1993 , 86-0608 , 86-2215, 86-2216, 86-1810, 86-2000, 86-2188 & 86-2189 , 86-1864 & 86-1939 , 86-0866 , 86-1058 & 86-2083 , 86-0642 , 86-1242 , 86-2025 , 86-0988 , 86-1422 , 86-0393 , 86-0662 , 86-1987 , 86-0149 , 86-1483 , 86-0014 & 86-0017 , 86-0581, 86-0678 & 86-0679 , 86-0222 & 86-1725 , 86-1739 , 86-0040 , 86-0852 , 86-0343 , 86-0029 , 86-0691 , 86-0033 , 86-0764 , 86-1570 & 86-1574 , 86-1013 , 86-0548 , 86-0349 , 86-0462 , 86-1600 , 86-0555 , 86-0677 , 86-0605 , 86-1616 & 86-1618 , 86-0304 & 86-1677 , 86-1061 , 86-1449 , 86-0382 , 86-2229 , 86-0909 , 86-0782 & 86-2112 , 86-1454 , 86-0846 , 86-1682 , 86-1734 , 86-0380 , 86-1392 , 86-2149 , 86-2163 , 86-1240 , 86-0952 , 86-0211 & 86-0690 , 86-0647 & 86-1712 , 86-0875 , 86-1715 , 86-1372 , 86-1034 , 86-1945 , 86-0147 , 86-0761 & 86-2167 , 86-0325 , 86-2127 , 86-1904 , 86-1771 , 86-1760 & 86-1847 , 86-0655 & 86-1135 , 86-0276 , 86-0477 , 86-1011 , 86-1183 , 86-1320 , 86-0233 , 86-0320 , 86-1060 , 86-0863 , 86-0718 , 86-0231 & 86-0457 , 86-0422 , 86-0748 , 86-0078 & 86-1330 , 86-0693 , 86-0874 , 86-1050 , 86-1051 , 86-0788 , 86-0858 , 86-1679 , 86-1556 , 86-0016 & 86-1144 , 86-0150 , 86-1042 & 86-1783 , 86-1441 , 86-1077 , 86-0741 , 86-1709 , 86-1173 & 86-2103 , 86-0269 , 86-0465 , 86-0146 & 86-1376 , 86-1224 & 86-1271 , 86-0431 & 86-1768 , 86-1374 , 86-1658 , 86-1922 , 86-0986 , 86-0034 , 86-0230 , 86-1899 , 86-2028 , 86-2081 , 86-1169 , 86-0168 , 86-0704 , 86-1651 , 86-2041 , 86-0261 , 86-1924 , 86-1171 & 86-1501 , 86-0337 , 86-0953 , 86-1968 , 86-0450 , 86-1868 , 86-2230 , 86-0161 , 86-0357 , 86-0301, 86-0303 & 86-1186 , 86-0177 , 86-1983 , 86-0787 , 86-2131 , 86-1502 , 86-2067 , 86-2145 , 86-1208 , 86-1895 , 86-0669 & 86-1664 , 86-0653 , 86-0796 , 86-1012 , 86-0646 , 86-0423 & 86-2072 , 86-0397 , 86-0039 , 86-0128 , 86-2049 , 86-0178 , 86-0265 , 86-1763 , 86-2225 , 86-0931 , 86-0258 , 86-1457 , 86-1608 , 86-1081 , 86-0754 , 86-1358 , 86-0670 , 86-1396 , 86-2090 , 86-1774 , 86-1996 & 86-1997 , 86-1617 , 86-1181 , 86-1875 , 86-1717 , 86-0196 & 86-1681 , 86-0661 , 86-0825 , 86-0217 , 86-1528 , 86-0111 & 86-1517 , 86-1623 , 86-0537 , 86-1710 , 86-1261 , 86-0123 & 86-0433 , 86-0530 , 86-0116-RC , 86-0617 , 86-1141 , 86-1990 , 86-0235 , 86-1675 , 86-2061 , 86-0368 , 86-1447 , 86-2223 , 86-0793 , 86-0024 , 86-1835 , 86-1573 , 86-1039 , 86-1757 , 86-1554 , 86-0290 , 86-1306 , 86-0856 , 86-1913 , 86-0050 , 86-0946 , 86-0392, 86-1473 & 86-2042 , 86-1750 , 86-1723 , 86-0141 , 86-2098 , 86-0090 , 86-2109 , 86-1378 , 86-0445 , 86-1127 & 86-1361 , 86-1485 , 86-1078 , 86-1753 , 86-1100 , 86-0934 , 86-1969 , 86-1281 , 86-2123 , 86-2089 , 86-0336 , 86-2162 , 86-2202 , 86-0216 , 86-1486 , 86-0240 , 86-0466 , 86-1031 , 86-1648 , 86-1801 , 86-0942 , 86-1685 , 86-0129 , 86-0410 , 86-1200 , 86-1017 , 86-0768 , 86-1534 , 86-2086 , 86-0241 , 86-1301 , 86-1642 , 86-1025 , 86-0967 & 86-2022 , 86-1718 , 86-0363 & 86-1970 , 86-0289 & 86-1526 , 86-1165 , 86-1882 , 86-1132 , 86-0407 , 86-1893 , 86-0595-RC , 86-0254 & 86-1518 , 86-2040 , 86-0574 , 86-0028 , 86-1943 , 86-0096 , 86-1545 , 86-1253 , 86-0179 & 86-1461 , 86-2108 , 86-1512 , 86-1020 , 86-0249 , 86-0205 , 86-0502 , 86-1777 , 86-1474 , 86-1956 , 86-0800 , 86-1359 , 86-0635 , 86-0094 , 86-2106 , 86-1751 , 86-0654 & 86-1105 , 86-1233 , 86-0801 , 86-0892 , 86-0377 & 86-0484 , 86-1599 , 86-2220 , 86-0835 , 86-1340 , 86-1384 , 86-1951 , 86-0021 , 86-0528 & 86-0773 , 86-0051 , 86-1436 , 86-1814 , 86-1564 , 86-0062 , 86-1770 , 86-0588-RC , 86-0820 , 86-1160 , 86-1086 , 86-1048 , 86-0326 , 86-1859 , 86-1689 , 86-2033 , 86-0376 , 86-1901 , 86-1530 , 86-0524 , 86-0651 , 86-0312 , 86-0442 , 86-1023 , 86-0760 , 86-0593 , 86-1109 , 86-2012 , 86-0056 , 86-0627 , 86-1089 , 86-0340 , 86-0993 & 86-0998 , 86-0645 , 86-1504 , 86-1643 , 86-1309 , 86-1994 , 86-0902-RC , 86-1400 , 86-0752 , 86-2062 , 86-0975 , 86-1837 , 86-1180 , 86-0089 , 86-0451 , 86-1655 , 86-0073 , 86-1440 & 86-1442 , 86-0256 , 86-0330 , 86-0354 , 86-1295 , 86-1769 , 86-0954 , 86-0437 , 86-1850 & 86-2154 , 86-0418 , 86-1159 , 86-2114 , 86-0355 , 86-1862 , 86-2214 , 86-1129 , 86-0103 & 86-1394 , 86-0313 & 86-0400 , 86-0602 , 86-1232 , 86-0041 , 86-0335 , 86-1558 & 86-1563 , 86-0837 , 86-0088 , 86-2058 , 86-0707 , 86-1185 , 86-2069 , 86-0638 , 86-1280 , 86-0936 , 86-0449 & 86-0479 , 86-0234 & 86-1104 , 86-2213 , 86-0278 & 86-2212 , 86-1363 , 86-1789 , 86-0749 , 86-0152 , 86-0438 , 86-0172 & 86-1824 , 86-0072, 86-0087, 86-0434, 86-0938 & 86-1672 , 86-1758 , 86-0395 , 86-1674 , 86-0818 & 86-0819 , 86-0580 & 86-1971 , 86-2173 , 86-0022 , 86-1936 , 86-0296 , 86-0027 & 86-0812 , 86-0656 , 86-0207 , 86-1087 , 86-0183 , 86-0367 , 86-0604 , 86-0710 , 86-1049 , 86-1128 , 86-1479 , 86-1549 , 86-1755 , 86-0037 & 86-1458 , 86-0500 , 86-1275 , 86-0842 , 86-0175 , 86-0672 & 86-1327 , 86-0834 , 86-1364 , 86-0740 & 86-0965 , 86-0884 & 86-0885 , 86-1569 , 86-0830 , 86-0023 , 86-1692 , 86-0391 , 86-0118 , 86-0681 , 86-1028 , 86-0911 , 86-0982 , 86-0947 & 86-2174 , 86-0948 , 86-0682 , 86-0341 , 86-1193 , 86-1571 , 86-1322 , 86-0344 , 86-1063 , 86-0387 , 86-0619 & 86-0459 , 86-1389 , 86-1522 , 86-0079 & 86-1591 , 86-0130 , 86-0733 , 86-0958 , 86-0569 , 86-0724 , 86-0532, 86-0603 & 86-1177 , 86-1857 , 86-1592, 86-1601 & 86-1602 , 86-0436 , 86-0590 , 86-1477 , 86-1663 , 86-0694 & 86-1559 , 86-0615 , 86-1929 , 86-1825 , 86-0043 & 86-1439 , 86-1170 , 86-1773 , 86-0483 , 86-0345 , 86-0879 , 86-1318 & 86-1547 , 86-1593 , 86-0496 , 86-0509 , 86-1662 , 86-0971 , 86-1270 , 86-0243 , 86-0285 & 86-2104 , 86-1932 , 86-2102 , 86-0940 , 86-1103 , 86-1735 , 86-1942 , 86-0031 & 86-1116 , 86-0568 & 86-1337 , 86-1245 , 86-1743 , 86-2169 , 86-0467 , 86-0405 , 86-0562 , 86-1683 , 86-0629 , 86-1937 , 86-1676 , 86-1977 , 86-2124 , 86-0063 , 86-0246-RC , 86-1064 , 86-1328 , 86-2190 , 86-2027 , 86-0429 , 86-0882 , 86-1244 & 86-2175 , 86-0824 , 86-0696 & 86-2233 , 86-1482 , 86-0725 , 86-0065 , 86-1496 , 86-0232 & 86-1229 , 86-0780 , 86-1080 , 86-1054 , 86-0379 , 86-1707 & 86-1708 , 86-1594 , 86-1705 , 86-1578 , 86-1265 , 86-1024 , 86-2059 , 86-0751 , 86-1030 , 86-0248 , 86-1700 , 86-2228 , 86-2120 , 86-0762 , 86-1800 , 86-1455 , 86-0512, 86-0860 & 86-1684 , 86-1842 , 86-1313 , 86-2036 , 86-1902 , 86-0714 , 86-1291 , 86-0474 , 86-0705 & 86-0794 , 86-1538 , 86-1131 , 86-0643 , 86-0561 , 86-2009 , 86-1344 , 86-0881 , 86-0872 , 86-0019 , 86-1691 , 86-2218 , 86-1152 , 86-1492 , 86-0839 & 86-1557 , 86-1325 , 86-0591 , 86-1756 , 86-0242 , 86-1741 & 86-2088 , 86-1323 , 86-2204 , 86-0404 , 86-1497 , 86-0823 & 86-1995 , 86-0110 , 86-1865 , 86-1641 , 86-0213 , 86-1146 , 86-0426 , 86-1032 , 86-2079 , 86-0101 , 86-1964 , 86-1908 , 86-1706 , 86-2132 , 86-0456 , 86-0469, 86-0611 & 86-1887 , 86-0104 & 86-1227 , 86-0816 & 86-1228 , 86-2023 & 86-2024 , 86-0158 , 86-0582 , 86-1148 , 86-2206 , 86-2074 , 86-0734 & 86-1928 , 86-0758 & 86-0759 , 86-1362 , 86-1438 , 86-1730 , 86-0922 , 86-1312 , 86-1493 , 86-1021, 86-1161 & 86-1976 , 86-1975 , 86-0939 , 86-0060 , 86-0026, 86-0489, 86-1130 & 86-1329 , 86-1699 , 86-0165 , 86-1114 & 86-1437 , 86-1246, 86-1339 & 86-1826 , 86-1586 , 86-1759 , 86-0372 , 86-1804 , 86-0188 , 86-2096 , 86-0743 , 86-1628 , 86-2217 & 86-2221 , 86-1820 , 86-1673 , 86-1958 , 86-1084 , 86-0956 , 86-0252 & 86-1805 , 86-0526 , 86-0493 , 86-1366 , 86-1827 , 86-1853 , 86-1905 , 86-1910 , 86-1944 , 86-0963 , 86-1921 , 86-2219 , 86-1453 , 86-0686 , 86-0044 , 86-0626-RC , 86-0854 , 86-0578 , 86-0409 , 86-1367, 86-1433 & 86-1981 , 86-0495, 86-0498 & 86-1985 , 86-2076 , 86-2029 , 86-0139 , 86-1660 , 86-1846 , 86-1243 , 86-1434 , 86-0557 , 86-1650 , 86-1863 & 86-2001 , 86-2082 , 86-0774 & 86-1016 , 86-1802 , 86-1292 , 86-2020, 86-2021 & 86-2093 , 86-0108 & 86-0536 , 86-0186 , 86-1498 , 86-1083 , 86-0015-R , 86-1282 , 86-0160 , 86-0566 & 86-1843 , 86-0300 , 86-1989 , 86-1649 , 86-2129 , 86-0055 , 86-1711 , 86-1222 , 86-0781 , 86-1738 , 86-2016 , 86-0747 , 86-0676 , 86-1002 , 86-1315 , 86-1680 , 86-1787 , 86-1321 , 86-0905 & 86-1731 , 86-1896 , 86-1149 , 86-0675 , 86-2125 , 86-1744 , 86-0294, 86-2053 & 86-2054 , 86-2019 , 86-1386 & 86-1387 , 86-0455 & 86-0689 , 86-1062, 86-1198, 86-1347, 86-1484, 86-2166 & 86-2172 , 86-1256 , 86-0851 , 86-0218 & 86-0853 , 86-1459 , 86-2116 , 86-1163 & 86-1164 , 86-1035 , 86-0464, 86-1194 & 86-1195 , 86-0814 , 86-0353 , 86-1046 , 86-1828 , 86-1044 , 86-0176 , 86-2236 , 86-0505 , 86-1225 , 86-0709 , 86-1740 & 86-1950 , 86-0531 , 86-0356 , 86-0744 , 86-1911 & 86-1912 , 86-0132 , 86-0809 & 86-1553 , 86-0499 & 86-1809 , 86-0926 , 86-0471 , 86-1947 , 86-0472-RC 86-0018 , 86-0052 , 86-0053 , 86-0054 , 86-0059 , 86-0066 , 86-0081 , 86-0083 , 86-0126 , 86-0127 , 86-0133 , 86-0134 , 86-0135 , 86-0142 , 86-0143 , 86-0144 , 86-0145 , 86-0169 , 86-0180 , 86-0181 , 86-0187 , 86-0195 , 86-0197 , 86-0202 , 86-0208 , 86-0209 , 86-0210 , 86-0220 , 86-0221 , 86-0225 , 86-0266 , 86-0267 , 86-0268 , 86-0272 , 86-0273 , 86-0274 , 86-0280 , 86-0287 , 86-0288 , 86-0292 , 86-0293 , 86-0295 , 86-0297 , 86-0302 , 86-0305 , 86-0314 , 86-0315 , 86-0316 , 86-0317 , 86-0318 , 86-0319 , 86-0323 , 86-0324 , 86-0350 , 86-0351 , 86-0362 , 86-0373 , 86-0386 , 86-0398 , 86-0414 , 86-0439 , 86-0440 , 86-0446 , 86-0447 , 86-0448 , 86-0461 , 86-0463 , 86-0473 , 86-0476 , 86-0480 , 86-0481 , 86-0488 , 86-0491 , 86-0497 , 86-0503 , 86-0510 , 86-0514 , 86-0516 , 86-0518 , 86-0522 , 86-0523 , 86-0534 , 86-0535 , 86-0538 , 86-0539 , 86-0563 , 86-0564 , 86-0565 , 86-0587 , 86-0598 , 86-0606 , 86-0607 , 86-0612 , 86-0621 , 86-0622 , 86-0630 , 86-0632 , 86-0633 , 86-0634 , 86-0659 , 86-0665 , 86-0666 , 86-0667 , 86-0671 , 86-0708 , 86-0713 , 86-0717 , 86-0721 , 86-0723 , 86-0726 , 86-0727 , 86-0728 , 86-0729 , 86-0730 , 86-0756 , 86-0771 , 86-0772 , 86-0779 , 86-0784 , 86-0803 , 86-0807 , 86-0808 , 86-0810 , 86-0815 , 86-0821 , 86-0847 , 86-0848 , 86-0849 , 86-0876 , 86-0877 , 86-0878 , 86-0888 , 86-0889 , 86-0898 , 86-0899 , 86-0907 , 86-0908 , 86-0919 , 86-0924 , 86-0927 , 86-0928 , 86-0929 , 86-0930 , 86-0933 , 86-0941 , 86-0943 , 86-0944 , 86-0945 , 86-0950 , 86-0959 , 86-0960 , 86-0961 , 86-0968 , 86-0969 , 86-0970 , 86-0990 , 86-0991 , 86-0992 , 86-1004 , 86-1005 , 86-1006 , 86-1007 , 86-1008 , 86-1010 , 86-1019 , 86-1038 , 86-1040 , 86-1071 , 86-1072 , 86-1073 , 86-1074 , 86-1075 , 86-1088 , 86-1090 , 86-1091 , 86-1092 , 86-1096 , 86-1101 , 86-1118 , 86-1120 , 86-1122 , 86-1137 , 86-1145 , 86-1153 , 86-1168 , 86-1172 , 86-1174 , 86-1189 , 86-1190 , 86-1191 , 86-1197 , 86-1199 , 86-1202 , 86-1203 , 86-1206 , 86-1215 , 86-1220 , 86-1235 , 86-1236 , 86-1237 , 86-1238 , 86-1239 , 86-1247 , 86-1257 , 86-1262 , 86-1263 , 86-1264 , 86-1284 , 86-1285 , 86-1288 , 86-1289 , 86-1290 , 86-1294 , 86-1296 , 86-1298 , 86-1299 , 86-1333 , 86-1334 , 86-1335 , 86-1336 , 86-1341 , 86-1348 , 86-1349 , 86-1350 , 86-1351 , 86-1352 , 86-1353 , 86-1354 , 86-1355 , 86-1356 , 86-1365 , 86-1370 , 86-1371 , 86-1380 , 86-1390 , 86-1395 , 86-1397 , 86-1401 , 86-1403 , 86-1405 , 86-1406 , 86-1407 , 86-1408 , 86-1409 , 86-1410 , 86-1411 , 86-1412 , 86-1413 , 86-1414 , 86-1415 , 86-1416 , 86-1417 , 86-1418 , 86-1419 , 86-1426 , 86-1427 , 86-1444 , 86-1462 , 86-1463 , 86-1464 , 86-1466 , 86-1467 , 86-1468 , 86-1469 , 86-1470 , 86-1471 , 86-1472 , 86-1507 , 86-1508 , 86-1511 , 86-1516 , 86-1519 , 86-1520 , 86-1524 , 86-1525 , 86-1533 , 86-1543 , 86-1546 , 86-1552 , 86-1567 , 86-1568 , 86-1577 , 86-1584 , 86-1585 , 86-1598 , 86-1624 , 86-1629 , 86-1639 , 86-1644 , 86-1645 , 86-1653 , 86-1656 , 86-1665 , 86-1666 , 86-1667 , 86-1668 , 86-1716 , 86-1726 , 86-1745 , 86-1747 , 86-1748 , 86-1762 , 86-1765 , 86-1766 , 86-1767 , 86-1776 , 86-1788 , 86-1792 , 86-1793 , 86-1803 , 86-1811 , 86-1817 , 86-1833 , 86-1848 , 86-1861 , 86-1871 , 86-1872 , 86-1873 , 86-1877 , 86-1897 , 86-1898 , 86-1919 , 86-1930 , 86-1931 , 86-1934 , 86-1935 , 86-1948 , 86-1952 , 86-1961 , 86-1967 , 86-1973 , 86-1978 , 86-1979 , 86-1980 , 86-1991 , 86-2010 , 86-2013 , 86-2037 , 86-2045 , 86-2046 , 86-2048 , 86-2050 , 86-2051 , 86-2052 , 86-2063 , 86-2064 , 86-2066 , 86-2091 , 86-2094 , 86-2105 , 86-2134 , 86-2141 , 86-2142 , 86-2143 , 86-2144 , 86-2150 , 86-2151 , 86-2152 , 86-2153 , 86-2155 , 86-2156 , 86-2158 , 86-2159 , 86-2160 , 86-2161 , 86-2195 , 86-2196 , 86-2200 , 86-2201 , 86-2231 , and 86-2235.
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    Worldwide Distribution
  • Product Description
    OEC 6800 Miniview, Image-intensified fluoroscopic x-ray system used during diagnostic, surgical and interventional procedures. || Designed to provide the physician with general fluoroscopic visualization of the patient including but not limited to surgical orthopedic and extremity imagine. The device is not intended for whole-body pediatric imaging.
  • Manufacturer

Manufacturer

  • Manufacturer Address
    GE OEC Medical Systems, Inc, 384 Wright Brothers Dr, Salt Lake City UT 84116-2862
  • Source
    USFDA