Recall of Device Recall NexGen Complete Knee Solution MIS Total Knee Procedure Stemmed Tibial Component, Precoat

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Zimmer, Inc..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    68457
  • Event Risk Class
    Class 2
  • Event Number
    Z-1937-2014
  • Event Initiated Date
    2014-05-19
  • Event Date Posted
    2014-06-27
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2015-10-29
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer/metal/polymer - Product Code JWH
  • Reason
    Zimmer is initiating a voluntary recall of specific identified lots due to the potential that the threads may be out of specification.
  • Action
    Zimmer sent an Urgent Medical Device Recall letter June 2014 to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to review the notification and ensure affected personnel are aware of the contents. Locate and quarantine any affected product and notify their Zimmer sales representative. Their Zimmer sales representative will remove the recalled product from their facility. For patients that previously had this product implanted, it is recommended that they continue their normal operative follow up routine. Customers with questions were instructed to call 1-877-946-2761.

Device

  • Model / Serial
    Size 6, includes the following Lot Numbers: 370319 37213647 37218135 37218727 37219556 37220950 61414605 61414606 61442322 61451224 61451225 61451226 61479488 61479489 61479490 61518197 61518198 61518199 61538053 61543978 61592093 61592094 61627261 61643938 61672569 61682157 61685470 61721675 61721676 61774189 61774190 61774191 61774195 61863158 61863159 61863160 61863161 61863162 61979034 61979035 62016163 62022560 62037896 62041306 62041308 62069549 62077856 62141261 62143609 62168186 62177113 62227421 62227422 62251569 62251570 62261977 62261978 62326316 62330687 62341281 62369796 62369797 62377652 62473295 62473296 62602804 62602805 62602806 62602807 62602808
  • Product Classification
  • Device Class
    2
  • Implanted device?
    Yes
  • Distribution
    Worldwide Distribution - USA (nationwide) and Internationally to Australia, China, Hong Kong, India, Japan, Korea, Malaysia, Nicaragua, Singapore, and Taiwan.
  • Product Description
    NexGen Complete Knee Solution, MIS Total Knee Procedure, Stemmed Tibial Component, Precoat, Fixed Bearing, Size 6. Part of the NexGen systems of semi-constrained, non-linked condylar knee prosthesis.
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Zimmer, Inc., 1800 W Center St, Warsaw IN 46580-2304
  • Manufacturer Parent Company (2017)
  • Manufacturer comment
    “The safety of patients has always been, and continues to be, Zimmer Biomet’s top priority and it is our honor to be a leader in this industry for the past 90 years,” Zimmer Biomet told ICIJ in a statement. “We adhere to strict regulatory standards, and work closely with the FDA and all applicable regulatory agencies in each of our regions as part of our commitment to operating a first-rate quality management system across our global manufacturing network. The company added that it is focused on staying at the forefront of innovation and doing right by the millions of patients who rely on the company’s products.
  • Source
    USFDA