Recall of Device Recall MyLab Ultrasound System

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by ESAOTE S.P.A..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    79547
  • Event Risk Class
    Class 2
  • Event Number
    Z-1137-2018
  • Event Initiated Date
    2018-01-03
  • Event Status
    Open, Classified
  • Event Country
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    System, imaging, pulsed doppler, ultrasonic - Product Code IYN
  • Reason
    The ultrasound system used in certain settings may result in overheating of the ultrasound probe head.
  • Action
    The January 3, 2018, notification letter requests the consignee complete and return a recall response form. The letter states that Esaote customer service will contact the consignee about the software update. In the meantime, it is recommended that the consignee does not use the PA023E probe in Pulsed Wave (PW) Doppler mode and that they refrain from using the system in Elaxto mode where the patient is unconscious or sedated. The consignee is asked to check the temperature of the probe before use.

Device

  • Model / Serial
    Software versions 4.0 - 13.60
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    US distribution in the states: AZ, FL, OK, PA, and TX.
  • Product Description
    MyLab Ultrasound System; Models 6100, 6150, 6200, 6250; PA023E Probe 7.5/10
  • Manufacturer

Manufacturer