Recall of Device Recall MONOPLANE Xray Imaging Systems with Velara Generator

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Philips Medical Systems North America Co. Phillips.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    56093
  • Event Risk Class
    Class 2
  • Event Number
    Z-2448-2010
  • Event Initiated Date
    2010-02-04
  • Event Date Posted
    2010-09-22
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2011-09-23
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    System, X-Ray, Angiographic - Product Code IZI
  • Reason
    A lock-up might occur inside the velara generator due to intermittent shorts inside the system. the problem could only resolved by a warm or cold system restart. in exceptional cases the power converters of the kvma unit in the generator are damaged and the system can not be restarted.
  • Action
    On 02/04/2010, Philips Healthcare began sending out the URGENT-Device Correction Notices to their consignees. The firm issued two notifications, one for the Allura Integris and Xper FD10C, FD10F, FD10F/C and FD20 Monoplane X-Ray imaging systems with the Velara Generator. The second notification is for the Allura Integris Xper FD10 and FD20 Biplane X-ray imaging systems with the Velara Generator. This notice informs the consignees that a system locks up during a critical interventional case and it might impose a serious injury to patients. This lock up issue could only be resolved by a warm or cold system restart. A warm restart would take about 90 seconds or a cold restart would take about four minutes, and the device could not be used during this time. If the system does not work after the restart, the consignees should contact the Philips Healthcare Customer Service Representative at 1-800-772-9377 #5 (then follow prompts). A service representative will visit each consignee and conduct a software and hardware upgrade.

Device

  • Model / Serial
    Each unit is identified with a SITE number:  38358, 101952, 102560, 102564, 102730, 102923, 102948, 103128,  103265, 103275, 103289, 103331, 103380, 103525, 103561, 103562, 103564, 103696, 103729, 103751, 103780, 103781, 103927, 104060, 104075, 104106, 104139, 104149, 104464, 104501, 104512, 104514, 104515, 104536, 104642, 104668, 104670, 104890, 104925, 104927, 104952, 104954, 104955, 104956, 104957, 104958, 105073, 105081, 105082, 105100, 105136, 105137, 105139, 105140, 105153, 105156, 105157, 105186, 105191, 105304, 105343, 105386, 105390, 105402, 105409, 105583, 105598, 105615, 105631, 105655, 105656, 105670, 105674, 105684, 105686, 105742, 105779, 105784, 105866, 105908, 105912, 105917, 105940, 105946, 105947, 105950, 105951, 105952, 105953, 105961, 500004, 500028, 500029, 500038, 500041, 500042, 500052, 504344, 504493, 504636, 504638, 504646, 504659, 504665, 504666, 504730, 504747, 504750, 504769, 504770, 504776, 504789, 504792, 504795, 504797, 504798, 504806, 504807, 504871, 504872, 504903, 504904, 504908, 504913, 504915, 504916, 504920, 504940, 504977, 504988, 505008, 505017, 505064, 505106, 505107, 505111, 505135, 505250, 505252, 505254, 505270, 505271, 505281, 505363, 505421, 505424, 505486, 505520, 505538, 505546, 505572, 505587, 505650, 505651, 505668, 505679, 505680, 505706, 505719, 505759, 505803, 505804, 505805, 505807, 505809, 505860, 505861, 505886, 505925, 505968, 505989, 506009, 506047, 506048, 506049, 506060, 506121, 506148, 506169, 506170, 506178, 506222, 506225, 506230, 506232, 506322, 506369, 506370, 506397, 506403, 506408, 506462, 506469, 506476, 506538, 506539, 506540, 506541, 506564, 506616, 506621, 506637, 506654, 506657, 506658, 506693, 506778, 506785, 519115, 519179, 519195, 519198, 519199, 519211, 519212, 519213, 519214, 519533, 519534, 519535, 519536, 519537, 519541, 519548, 519576, 519594, 519673, 519676, 519677, 519680, 519733, 519738, 519767, 519771, 519812, 519874, 519894, 519899, 519901, 520004, 520147, 520189, 520236, 520251, 520309, 520319, 520430, 520431, 520481, 520491, 520517, 520523, 520524, 520525, 520526, 520598, 520630, 520631, 520632, 520660, 520680, 520681, 520755, 520761, 520779, 520783, 520809, 520814, 520840, 520882, 520948, 520956, 520977, 521001, 521131, 521154, 521222, 521223, 521310, 521311, 521312, 521313, 521314, 521315, 521318, 521320, 521463, 521512, 521537, 521545, 521548, 521562, 521589, 521650, 521659, 521661, 521675, 521711, 521763, 521769, 521770, 521802, 521820, 521853, 521854, 521890, 521892, 521918, 522038, 522065, 522118, 522119, 522135, 522138, 522162, 522177, 522191, 522224, 522243, 522245, 522247, 522250, 522293, 522397, 522418, 522440, 522446, 522458, 530361, 530368, 530370, 530390, 530391, 530392, 530393, 530421, 530459, 530460, 530463, 530464, 530466, 530485, 530584, 530655, 530656, 530658, 530662, 530663, 530664, 530666, 530667, 530668, 530669, 530715, 530783, 531651, 531728, 531752, 531753, 531760, 531762, 531805, 531806, 531808, 531813, 531817, 531834, 531837, 531931, 531942, 531943, 531944, 531945, 531946, 531948, 531949, 531951, 531957, 531961, 532074, 532077, 532080, 532148, 532150, 532172, 532178, 532207, 532215, 532225, 532226, 532227, 532234, 532235, 532239, 532244, 532245, 532246, 532249, 532262, 532263, 532264, 532292, 532294, 532445, 532446, 532532, 532533, 532534, 532535, 532611, 532612, 532613, 532615, 532616, 532617, 532649, 532652, 532799, 532800, 532986, 532990, 533084, 533085, 533236, 533278, 533282, 533283, 533284, 533285, 533286, 533287, 533289, 533325, 533343, 533394, 533395, 533398, 533401, 533418, 533476, 533541, 533543, 533551, 533567, 533619, 533738, 533762, 533763, 533772, 533796, 533857, 533861, 533897, 533922, 533955, 533956, 533966, 533967, 534108, 534150, 534152, 534155, 534158, 534159, 534160, 534215, 534217, 534218, 534221, 534232, 534233, 534237, 534246, 534250, 534251, 534252, 534255, 534268, 534270, 534374, 534380, 534418, 534420, 534427, 534431, 534440, 534452, 534476, 534477, 534478, 534479, 534545, 534565, 534569.534627, 534630, 534634, 534638, 534639, 534646, 534665, 534678, 534681, 534682, 534716, 534717, 534720, 534724, 534730, 534731, 534743, 534758, 534787, 534796, 534798, 534811, 534813, 534856, 534874, 534876, 534879, 534927, 534928, 534941, 535094, 535151, 535152, 535173, 535212, 535213, 535224, 535250, 535274, 535278, 535285, 535303, 535326, 535342, 535348, 535367, 535376, 535387, 535396, 535400, 535415, 535416, 535418, 535424, 535439, 535492, 535566, 535611, 535614, 535615, 535635, 535667, 535702, 535727, 535728, 535769, 535772, 535777, 535779, 535793, 535813, 535814, 535927, 535931, 535946, 535951, 535979, 535982, 536006, 536007, 536014, 536030, 536067, 536090, 536136, 536138, 536154, 536155, 536163, 536173, 536174, 536178, 536234, 536379, 536425, 536433, 536471, 536592, 536639, 536641, 536646, 536714, 536717, 536721, 536736, 536777, 536796, 536798, 536801, 536901, 536921, 536954, 536966, 536993, 537020, 537024, 537065, 537073, 537118, 537205, 537266, 537268, 537296, 537357, 537386, 537466, 537514, 537588, 537589, 537590, 537597, 537604, 537630, 537644, 537698, 537717, 537731, 537743, 537745, 537784, 537805, 537810, 537838, 537854, 537855, 537922, 537934, 537935, 537937, 537938, 537985, 538084, 538102, 538193, 538194, 538228, 538234, 538269, 538322, 538368, 538369, 538386, 538404, 538406, 538416, 538432, 538449, 538464, 538465, 538466, 538476, 538479, 538482, 538485, 538486, 538512, 538537, 538538, 538744, 538771, 538806, 538813, 538902, 538903, 538910, 538911, 539094, 539099, 539107, 539204, 539220, 539224, 539267, 539268, 539322, 539323, 539369, 539380, 539381, 539398, 539455, 539456, 539472, 539479, 539480, 539516, 539595, 539634, 539641, 539745, 540048, 540070, 540082, 540150, 540167, 540203, 540209, 540213, 540234, 540235, 540269, 540275, 540284, 540288, 540301, 540311, 540312, 540321, 540325, 540338, 540374, 540386, 540421, 540456, 540460, 540468, 540487, 540488, 540622, 540684, 540790, 540808, 540841, 540887, 540912, 540930, 540952, 540973, 541026, 541029, 541030, 541046, 541047, 541049, 541053, 541080, 541223, 541251, 541253, 541259, 541409, 541509, 541510, 541572, 541588, 541592, 541597, 541657, 541659, 541667, 541715, 541742, 541748, 541783, 541822, 541848, 541873, 541880, 542109, 542113, 542117, 542306, 542478, 542486, 542518, 542522, 542526, 542744, 543087, 543108, 543194, 543198, 543209, 543213, 543216, 543229, 543386, 543397, 543398, 543399, 543408, 543433, 543610, 543611, 543702, 543759, 543762, 543799, 543829, 543850, 543873, 543877, 543931, 543971, 543972, 543984, 543988, 543999, 544005, 544048, 544051, 544052, 544057, 544086, 544103, 544120, 544122, 544130, 544145, 544184, 544371, 544433, 544489, 544506, 544513, 544550, 544572, 544581, 544600, 544610, 544631, 544636, 544637, 544655, 544728, 544729, 544750, 544989, 544990, 544992, 545023, 545024, 545037, 545052, 545069, 545073, 545090, 545280, 545282, 545293, 545295, 545296, 545357, 545365, 545395, 545443, 545521, 545539, 545640, 545714, 545733, 545749, 545750, 545752, 545776, 545779, 545799, 545827, 545838, 545839, 545840, 545853, 545905, 545912, 545921, 545941, 545942, 545964, 545966, 545975, 545985, 545991, 546012, 546014, 546025, 546060, 546201, 546456, 546517, 546560, 546561, 546572, 546585, 546586, 546634, 546668, 546679, 546693, 546886, 546908, 546934, 547005, 547058, 547118, 547159, 547165, 547234, 547279, 547287, 547321, 547334, 547364, 547392, 547406, 547600, 547601, 547619, 547620, 547621, 547627, 547628, 547660, 547700, 547720, 547729, 547748, 547749, 547753, 547761, 547762, 547763, 547770, 547775, 547777, 547795, 547800, 547822, 547833, 547838, 547852, 547855, 547861, 547878, 547893, 547896, 547898, 547908, 547914, 547918, 547921, 547925, 547932, 547948, 547952, 547965, 547976, 547977, 548022, 548029, 548038, 548039, 548049, 548051, 548054, 548061, 548067, 548108, 548147, 548192, 548432, 548496, 548525, 548601, 548603, 548616, 548640, 548645, 548660, 548764, 548775, 548778, 548785, 548966, 549013, 549170, 549245, 549250, 549252, 549253, 549264.549306, 549378, 549379, 549385, 549400, 549451, 549738, 549750, 549792, 549902, 549925, 549967, 550031, 550053, 550122, 550124, 550133, 550172, 550173, 550244, 550253, 550255, 550304, 550360, 550371, 550399, 550404, 550451, 551050, 551051, 551146, 551178, 551340, 551379, 551607, 551608, 551633, 551725, 551763, 551851, 551853, 552178, 552192, 552202, 552279, 552381, 552382, 552425, 552440, 552469, 552470, 552546, 552559, 552585, 552586, 552590, 552736, 553097, 553137, 553234, 553288, 553330, 553351, 553352, 553397, 553432, 553445, 553466, 553486, 553487, 553603, 553697, 553723, 553798, 553803, 553846, 553867, 553868, 553869, 553871, 553873, 553881, 553925, 553964, 553981, 553984, 554018, 554019, 554044, 554049, 554054, 554136, 554370, 554427, 554431, 554442, 554444, 554447, 554499, 554518, 554550, 554556, 554631, 554680, 554768, 554775, 554776, 554784, 554791, 554804, 554811, 554896, 555206, 555300, 555321, 555343, 555349, 555350, 555420, 555447, 555470, 555605, 555742, 555781, 555927, 556017, 556053, 556095, 556156, 556311, 556323, 556390, 556470, 556601, 556605, 556609, 556701, 556708, 556709, 556760, 556773, 556783, 556810, 556811, 556814, 556820, 556855, 556857, 556893, 557385, 557512, 557653, 558238, 558448, 558499, 558582, 560060,  41416356, 41416358, 41443581, 41443671, 41443700, 41443707, 41443774, 41443855, 41443882, 41443928, 41444385, 41444398, 41444399, 41444419, 41444446, 41444454, 41444461, 41444463, 41444558, 41444597, 41444603, 41444609, 41444646, 41444662, 41444680, 41444728, 41444743, 41444747, 41444753, 41444758, 41444791, 41444797, 41445001, 41445010, 41445014, 41445026, 41445032, 41445070, 41445119, 41445125, 41445169, 41445173, 41445202, 41445207, 41445229, 41445233, 41445235, 41445246, 41445250, 41445253, 41445254, 41445255, 41445256, 41445261, 41445270, 41445285, 41445303, 41445309, 41445310, 41445315, 41445332, 41445336, 41445344, 41445376, 41445377, 41445385, 41445390, 41445427, 41445434, 41445438, 41445452, 41445486, 41445495, 41445502, 41445514, 41445515, 41445516, 41445578, 41445582, 41445616, 41445645, 41445676, 41445677, 41445692, 41445701, 41445744, 41445766, 41445785, 41445804, 41445809, 41445816, 41445817, 41445875, 41445916, 41445978, 41445989, 41446007, 41446024, 41446039, 41446044, 41446080, 41446117, 41446144, 41446150, 41446211, 41446212, 41446216, 41446217, 41446223, 41446277, 41446290, 41446292, 41446323, 41446363, 41446796, 41446807, 41446808, 41455880, 41455893, 41455894, 41455896, 41455900, 41455902, 41455906, 41455910, 41455924, 41455932, 41455955, 41586608, 41612034, 41622975, 41630856, 41691748, 41747146, 41755552, 41765492, 41779600, 41779610, 41787786, 41792527, 41792800, 41792812, 41793875, 41793888, 41795224, 41796153, 41851761, 41903418, 41903960, 41956849, 42028108, 42038023, 42046538, 42057714, 42057739, 42063953, 42066520, 42067846, 42095488, 42172062, 42197144, 42206171, 42225697, 42288802, 42304719, 42313004, 42323439, 42365781, 42366186, 42371287, 42393465, 42393532, 42394081, 42478916, 42508767, 42552595, 42635362, 42635423, 42635426, 42638228, 42638512, 42638525, 42639096, 42668010, 42799507, 42799589, 42799745, 42802803, 42811500, 42812038, 42816302, 42823758, 42826856, 42826863, 42878485, 42881036, 42881266, 42892018, 42892404, 42892430, 42892474, 42919054, 42919085, 42967816, 42985101, 42985218, 43067304, 43082276, 43084042, 43084220, 43084414, 43119111, 43120325, 43151051, 43165011, 43179280, 43180544, 43209312, 43209348, 43209356, 43209360, 43230259, 43259235, 43263118, 43270340, 43305020, 43320930, 43340698, 43392727, 43420965, 43421081, 43433587, 43438771, 43443629, 43445198, 43446897, 43592438, 43594036, 43594037, 43613819, 43618965, 43622121, 43689832, 43704085, 43708891, 43709318, 43718660, 43719245, 43720499, 43720519, 43720660, 43720677, 43720776, 43721160, 43721227, 43721256, 43721670, 43756720, 43848518, 43902667, 43914262, 43931289, 43971849, 43973609, 43973613, 44004308, 44004309, 44015656, 44026775, 44035662, 44038398, 44065220, 44203373, 44320671, 44339387, 44388661, 44450326, 44536044, 44536281, 44604746, 44620455, 44621056, 44642682, 44643966, 44644567, 44649231, 44650811, 44654610, 44654626, 44654678, 44654764, 44743567, 44743568, 44830153, 44866223, 44879032, 44881204, 44891499, 45072647, 45103403, 45109943, 45168574, 45217802, 45321194, 45405221, 45412976, 45414258, 45428128, 45428705, 45428708, 45485899, 45509016, 45632472, 45632486, 47158877, 47312619, 49599010, 49599013,  and 49599014.
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    Nationwide Distribution
  • Product Description
    MONOPLANE X-ray Imaging Systems include the Allura XPER FD10 F/C, Allura XPER FD20, Integris H5000/Allura 9 F/C, Integris CV, and Allura 12-15 Monoplane. || The Monoplane system has one C-arm and it could be mounted on a ceiling or on the floor stand. || The product's brochure is labeled in part: "Philips Healthcare Global Information Center...P.O. Box 1286...5602 BG Eindhoven...The Netherlands".
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Philips Medical Systems North America Co. Phillips, 22100 Bothell Everett Hwy, Bothell WA 98021-8431
  • Source
    USFDA