Recall of Device Recall MODEL: Aegis Torque Handle CATALOG NO. : 287110800

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by DePuy Spine, Inc..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    72654
  • Event Risk Class
    Class 2
  • Event Number
    Z-0454-2016
  • Event Initiated Date
    2015-11-20
  • Event Date Posted
    2015-12-21
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2017-07-03
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Appliance, fixation, spinal intervertebral body - Product Code KWQ
  • Reason
    Manufacturing error within the torque handle may prohibit some handles from being able to fully assemble with the tightener shaft. the design of the inspection gage does not account for the additional shift length.
  • Action
    DePuy sent an Urgent Voluntary Product Recall Notification dated November 17, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. They were asked to do the following: 1. Review inventory located within your facility and quarantine any products listed in this recall notice. Please contact your DePuy Synthes Spine* Sales Consultant to return any inventory located within your facility. 2. Review, complete, sign and return the attached business reply form to in accordance with the directions on the business reply form within 5 business days of receipt of this notification. 3. Forward this notice to anyone in your facility that needs to be informed. 4. If any product listed below has been forwarded to another facility, contact that facility and provide them with this letter. 5. Keep a copy of this notice. For further questions please call (508) 828-3774

Device

  • Model / Serial
    LOT NUMBER: GB0906 GB0707 GB0907 GB58815 GB59244 GB61134
  • Product Classification
  • Device Class
    2
  • Implanted device?
    Yes
  • Distribution
    US Distribution to the states of : NC, MI, CO, MD, FL, ID, TX, CA and NY.
  • Product Description
    Aegis Torque Handle CATALOG NO. : 2871-10-800. A spinal surgical instrument used on the AEGIS Anterior Lumbar Plate System.
  • Manufacturer

Manufacturer

  • Manufacturer Address
    DePuy Spine, Inc., 325 Paramount Dr, Raynham MA 02767-5199
  • Manufacturer Parent Company (2017)
  • Source
    USFDA