Recall of Device Recall MoblVac Nylon Transport Bag

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Ohio Medical Corporation.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    55638
  • Event Risk Class
    Class 2
  • Event Number
    Z-1778-2010
  • Event Initiated Date
    2010-05-10
  • Event Date Posted
    2010-06-07
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2011-08-19
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Pump, portable, aspiration (manual or powered) - Product Code BTA
  • Reason
    Ohio medical has received a report of an explosive event with a moblvac portable woundcare pump housing where parts were separated while in an enclosed nylon transport bag during use.
  • Action
    Ohio Medical sent Urgent Medical Device Recall letters dated 5/7/10 to the direct accounts on 5/10/10, with follow-up telephone calls on 5/11/10. The accounts were informed of the one report of an explosion of the pump, and were requested to take the pumps out of service and return the pumps to Ohio Medical. The accounts were requested to complete and fax back to Ohio Medical the enclosed acknowledgement form indicating the number of pumps being returned. Any questions were directed to Anita Christ at 847-855-6270. Follow-up letters dated 5/13/10 were sent to the accounts on 5/14/10, providing instructions for the return of the pumps and bags. Shipping boxes will be sent to the accounts based on the number of pumps reported on the acknowledgment form.

Device

  • Model / Serial
    model 769709, all units
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    Nationwide and internationally to Qatar and Oman
  • Product Description
    NPWT Small Mobile Bag No Logo; an adjunct to the MoblVac Portable Wound Care Pump; Ohio Medical Corporation, 1111 Lakeside Drive, Gurnee, IL 60031; Model 769709.
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Ohio Medical Corporation, 1111 Lakeside Dr, Gurnee IL 60031
  • Source
    USFDA