Recall of Device Recall MiniCAT CT Scanner

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Xoran Technologies, Inc..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    66344
  • Event Risk Class
    Class 2
  • Event Number
    Z-0041-2014
  • Event Initiated Date
    2009-12-14
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2014-09-11
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    System, x-ray, tomography, computed - Product Code JAK
  • Reason
    Failure of the gantry to rotate. the complaint alleged that the scanner would emit radiation but would not rotate.
  • Action
    The service report in response to this complaint stated that the Head select relay board, chip and fan were replaced. The machine (ENTOOOO I) was subsequently tested. All tests were successful.

Device

  • Model / Serial
    minicat
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    Worldwide Distribution - US nationwide
  • Product Description
    Xoran MiniCAT CT Scanner ID ENT00001
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Xoran Technologies, Inc., 5210 S State Rd, Ann Arbor MI 48108-7936