Recall of Device Recall Merge HEMO software

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Merge Healthcare, Inc..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    74577
  • Event Risk Class
    Class 2
  • Event Number
    Z-2341-2016
  • Event Initiated Date
    2012-07-02
  • Event Date Posted
    2016-08-04
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2017-06-26
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Computer, diagnostic, programmable - Product Code DQK
  • Reason
    The international normalized ratio (inr) value displayed in the study report does not match the value that is imported on the pre-procedure labs screen from hl7.
  • Action
    The firm issued an "Important Product Information" bulletin to customers dated 7/2/2012 via e-mail letting them know there was an upgrade available. On 1/30/2016, the recalling firm initiated phone calls to their customers letting them know there was an upgrade available and if they declined it, they would like a written response.

Device

  • Model / Serial
    Software versions 8.10, 9.10, 9.20.
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    The medical facilities who currently have the affected versions are located nationwide. There was government distribution but no military or foreign distribution.
  • Product Description
    Merge HEMO software.