Recall of Device Recall Medtronic SynchroMed II Programmable Pump

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Medtronic Neuromodulation.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    49603
  • Event Risk Class
    Class 2
  • Event Number
    Z-0592-2009
  • Event Initiated Date
    2008-08-25
  • Event Date Posted
    2009-01-26
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2012-05-17
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Pump, infusion, implanted, programmable - Product Code LKK
  • Reason
    Medtronic issued a letter to healthcare professionals providing important safety information regarding mri (magnetic resonance imaging) effects on synchromed el and synchromed ii implantable infusion pumps. as stated in product labeling, the magnetic field of an mri will temporarily stop the rotor of the pump motor and suspend drug infusion for the duration of mri exposure for all synchromed pump.
  • Action
    An Urgent Medical Device Correction, Important Information on Potential MRI Effects letter, dated August 2008, was sent to health care professionals beginning August 25, 2008. The letter described the issue, provided patient management recommendations and included enclosures: Roller Study Procedure to confirm pump function, Post MRI Pump Interrogation Guidelines and New Information Regarding Potential MRI Effects.

Device

  • Model / Serial
    All SynchroMed EL and SynchroMed II pumps.
  • Product Classification
  • Device Class
    3
  • Implanted device?
    Yes
  • Distribution
    Worldwide Distribution --- USA including Washington D.C., and Puerto Rico, and countries of Algeria, Aruba, Austria, Belarus, Belgium, Croatia, Cyprus Czech Republic, Denmark, Egypt, Finland, France, French Polynesia, Germany, Greece, Hungary, Iceland, India, Iran, Iraq, Ireland, Israel, Italy , Jordan Kuwait, Lebanon, Luxembourg, Malta, Netherlands, Netherlands Antilles, Norway, Pakistan, Poland, Portugal, Reunion, Romania, Russian Federation, San Marino, Saudi Arabia, Serbia and Montenegro, Slovakia, Slovenia, Sough Africa, Spain, Sweden, Switzerland, Tunisia, Turkey, United Arab Emirates, United Kingdom, Vatican City State, New Caledonia, Brazil, Dominican Republic, Mexico, Hong Kong, Singapore, Canada, Australia and Japan.
  • Product Description
    Medtronic SynchroMed II Programmable Pump, Model 8637-40. 40 mL Reservoir. The contents [STERILE] of the inner package have been sterilized by ethylene oxide gas. The implantable Medtronic SynchroMed II programmable Pump is part of the SynchroMed II Infusion System designed to contain and administer prescribed drugs to a specific site. The implantable components of the SynchroMed II Infusion System include the pump, catheter, and catheter accessories. Medtronic, Inc., 710 Medtronic Parkway, Minneapolis, MN 55432-5604, USA. || The SynchroMed II Programmable Pump is indicated when patient therapy requires the chronic infusion of the follosing drugs or fluids: -chronic intraspinal (epidural and intrathecal) infusion of preservative-free morphine sulfate sterile solution in the treatment of chronic intractable pain. Preservative-free sterile saline, USP, can be used to achieve the physician-prescribed concentration of preservative-free morphine sulfate sterile solution. - chronic intrathecal infusion of preservative-free ziconotide sterile solution for the management of serve chronic pain. Preservative-free sterile saline, USP, can be used to achieve the physician-prescribed concentration of preservative-free ziconotide sterile solution after the initial fill of the pump with this drug. -chronic intrathecal infusion of Lioresal Intrathecal (baclofen injection) in the management of severe spasticity. Preservative-free sterile saline, USP, can be used to achieve the physician-prescribed concentration of Lioresal Intrathecal (baclofen Injection). -Chronic intravascular infusion of floxuridine, (FUDR) or methotrexate for the treatment of primary or metastatic cancer. Bacteriostatic water or preservative-free sterile saline, USP, can be used to achieve the physician-prescribed concentration of chemotherapy drugs or to flush the pump reservoir. Saline or heparinized saline (unless contraindicated) may be used during an interruption in chemotherapy to maintain pump and catheter patency.
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Medtronic Neuromodulation, 800 53rd Ave NE, PO Box 1250, Minneapolis MN 55440-1250
  • Source
    USFDA