Recall of Device Recall Medtronic SIGMA 100/200/300 Series IPGs

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Medtronic Inc. Cardiac Rhythm Disease Managment.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    52076
  • Event Risk Class
    Class 1
  • Event Number
    Z-1518-2009
  • Event Initiated Date
    2009-05-18
  • Event Date Posted
    2009-06-11
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2012-03-28
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Implantable Pacemakere Pulse-Generator - Product Code DXY
  • Reason
    One or more bond wire pairs will lift or separate from the bonding terminals on the device hybrid. this may present clinically as loss of rate response, premature battery depletion, loss of telemetry, or no output.
  • Action
    Medtronic Sales representative hand delivered an Important Patient Safety Information letter, dated May 2009, to physicians beginning 05/18/09. The letter describes the issues, provides a predicted failure rate for the 3 populations of devices, and provides Patient Management Recommendations. The letter recommended that physicians should advise their patients to seek medical attention immediately if they experience symptoms (e.g., fainting or lightheadedness). It was also recommended that physicians should consider device replacement for patients who are both pacemaker dependent and who have been implanted with a device in the affected subsets. Medtronic will offer supplemental device warranty if the device is not already at elective replacement time. The last recommendation was that physicians should continue routine follow up in accordance with standard practice for those patients who are not pacemaker dependent. The letter also provides Physician and Patient Support. Letters will be mailed to patients with registered devices beginning 05/27/09. Questions should be directed to a local Medtronic Representative or Medtronic Technical Services at 800-505-4636.

Device

  • Model / Serial
    Model SVVI103, serial numbers: PJG159902H, PJG159903H, PJG159904H, PJG159906H, PJG159909H, PJG159910H, PJG159912H, PJG159913H, PJG159916H, PJG159918H, PJG159922H, PJG159926H, PJG159929H, PJG159932H, PJG159935H, PJG159936H, PJG159939H, PJG159941H, PJG159943H, PJG159944H, PJG159945H, PJG159948H, PJG159952H, PJG159954H, PJG159956H, PJG159957H, PJG159960H, PJG159961H, PJG159962H, PJG159964H, PJG159966H, PJG159968H, PJG159971H, PJG159972H, PJG159973H, PJG159974H, PJG159976H, PJG159977H, PJG159979H, PJG159981H, PJG159982H, PJG159983H, PJG159985H, PJG159988H, PJG159989H, PJG159990H, PJG159993H, PJG159996H, PJG159997H, PJG159999H, PJG160000H, PJG160003H, PJG160004H, PJG160005H, PJG160006H, PJG160007H, PJG160008H, PJG160009H, PJG160010H, PJG160011H, PJG160018H, PJG160020H, PJG160026H, PJG160027H, PJG160039H, PJG160040H, PJG160151H, PJG160152H, PJG160153H, PJG160155H, PJG160156H, PJG160158H, PJG160159H, PJG160160H, PJG160161H, PJG160163H, PJG160165H, PJG160167H, PJG160171H, PJG160722H, PJG160723H, PJG160724H, PJG160725H, PJG160726H, PJG160727H, PJG160728H, PJG160729H, PJG160730H, PJG160731H, PJG160732H, PJG160733H, PJG160734H, PJG160735H, PJG160736H, PJG160737H, PJG160738H, PJG160739H, PJG160740H, PJG160741H, PJG160742H, PJG160743H, PJG160744H, PJG160745H, PJG160746H, PJG160747H, PJG160748H, PJG160749H, PJG160751H, PJG160752H, PJG160753H, PJG160754H, PJG160755H, PJG160756H, PJG160757H, PJG160758H, PJG160759H, PJG160760H, PJG160761H, PJG160762H, PJG160763H, PJG160764H, PJG160765H, PJG160766H, PJG160767H, PJG160768H, PJG160769H, PJG160770H, PJG160771H, PJG160772H, PJG160773H, PJG160775H, PJG160776H, PJG160777H, PJG160778H, PJG160779H, PJG160780H, PJG160781H, PJG160782H, PJG160783H, PJG160784H, PJG160785H, PJG160786H, PJG160787H, PJG160788H, PJG160789H, PJG160790H, PJG160791H, PJG160792H, PJG160793H, PJG160794H, PJG160795H, PJG160796H, PJG160797H, PJG160798H, PJG160799H, PJG160800H, PJG160801H, PJG160802H, PJG160803H, PJG160804H, PJG160805H, PJG160806H, PJG160807H, PJG160808H, PJG160809H, PJG160810H, PJG160811H, PJG160812H, PJG160813H, PJG160814H, PJG160815H, PJG160816H, PJG160817H, PJG160818H, PJG160819H, PJG160820H, PJG160821H, PJG160822H, PJG160823H, PJG160824H, PJG160825H, PJG160826H, PJG160827H, PJG160828H, PJG160829H, PJG160831H, PJG160832H, PJG160833H, PJG160834H, PJG160835H, PJG160836H, PJG160837H, PJG160838H, PJG160839H, PJG160840H, PJG160841H, PJG160842H, PJG160843H, PJG160844H, PJG160845H, PJG160846H, PJG160847H, PJG160848H, PJG160849H, PJG160850H, PJG160851H, PJG160852H, PJG160853H, PJG160854H, PJG160855H, PJG160856H, PJG160857H, PJG160858H, PJG160859H, PJG160860H, PJG160861H, PJG160862H, PJG160863H, PJG160864H, PJG160865H, PJG160866H, PJG160867H, PJG160868H, PJG160869H, PJG160870H, PJG160871H, PJG162608H, PJG162609H, PJG162610H, PJG162611H, PJG162612H, PJG162613H, PJG162614H, PJG162615H, PJG162616H, PJG162617H, PJG162618H, PJG162619H, PJG162620H, PJG162621H, PJG162622H, PJG162623H, PJG162624H, PJG162625H, PJG162626H, PJG162627H, PJG162629H, PJG162630H, PJG162631H, PJG162632H, PJG162633H, PJG162634H, PJG162635H, PJG162636H, PJG162637H, PJG162639H, PJG162640H, PJG162641H, PJG162642H, PJG162643H, PJG162644H, PJG162645H, PJG162646H, PJG162647H, PJG162648H, PJG162649H, PJG162650H, PJG162651H, PJG162652H, PJG162653H, PJG162654H, PJG162655H, PJG162656H, PJG162657H, PJG162658H, PJG162659H, PJG162660H, PJG162661H, PJG162662H, PJG162663H, PJG162664H, PJG162665H, PJG162666H, PJG162667H, PJG162668H, PJG162670H, PJG162671H, PJG162672H, PJG162673H, PJG162674H, PJG162675H, PJG162676H, PJG162677H, PJG162678H, PJG162679H, PJG162680H, PJG162681H, PJG162682H, PJG162683H, PJG162684H, PJG162685H, PJG162686H, PJG162687H, PJG162688H, PJG162689H, PJG162690H, PJG162691H, PJG162692H, PJG162693H, PJG162694H, PJG162695H, PJG162696H, PJG162697H, PJG162698H, PJG162699H, PJG162700H, PJG162701H, PJG162702H, PJG162703H, PJG162704H, PJG162705H, PJG162706H, PJG162707H, PJG162708H, PJG162709H, PJG162710H, PJG162711H, PJG162712H, PJG162713H, PJG162714H, PJG162715H, PJG162716H, PJG162717H, PJG162718H, PJG162719H, PJG162720H, PJG162721H, PJG162722H, PJG162723H, PJG162724H, PJG162725H, PJG162726H, PJG162727H, PJG162728H, PJG162729H, PJG162730H, PJG162731H, PJG162732H, PJG162733H, PJG162734H, PJG162735H, PJG162736H, PJG162737H, PJG162738H, PJG162739H, PJG162740H, PJG162741H, PJG162742H, PJG162743H, PJG162744H, PJG162745H, PJG162746H, PJG162747H, PJG162748H, PJG162749H, PJG162750H, PJG162751H, PJG162752H, PJG162753H, PJG162754H, PJG162755H, PJG162756H, PJG162757H, PJG164006H, PJG164007H, PJG164008H, PJG164009H, PJG164010H, PJG164011H, PJG164012H, PJG164013H, PJG164014H, PJG164015H, PJG164016H, PJG164017H, PJG164018H, PJG164019H, PJG164020H, PJG164021H, PJG164022H, PJG164023H, PJG164024H, PJG164025H, PJG164026H, PJG164027H, PJG164028H, PJG164029H, PJG164030H, PJG164031H, PJG164032H, PJG164033H, PJG164034H, PJG164035H, PJG164036H, PJG164037H, PJG164038H, PJG164039H, PJG164040H, PJG164041H, PJG164042H, PJG164043H, PJG164044H, PJG164045H, PJG164046H, PJG164047H, PJG164048H, PJG164049H, PJG164050H, PJG164051H, PJG164052H, PJG164053H, PJG164054H, PJG164055H, PJG164056H, PJG164057H, PJG164058H, PJG164059H, PJG164060H, PJG164061H, PJG164062H, PJG164063H, PJG164064H, PJG164065H, PJG164066H, PJG164067H, PJG164068H, PJG164069H, PJG164070H, PJG164071H, PJG164072H, PJG164073H, PJG164074H, PJG164075H, PJG164076H, PJG164077H, PJG164078H, PJG164079H, PJG164080H, PJG164081H, PJG164082H, PJG164083H, PJG164085H, PJG164088H, PJG164089H, PJG164090H, PJG164093H, PJG164096H, PJG164098H, PJG164099H, PJG164101H, PJG165063H, PJG165064H, PJG165065H, PJG165066H, PJG165067H, PJG165069H, PJG165070H, PJG165071H, PJG165072H, PJG165073H, PJG165074H, PJG165075H, PJG165076H, PJG165077H, PJG165078H, PJG165079H, PJG165080H, PJG165081H, PJG165082H, PJG165083H, PJG165084H, PJG165085H, PJG165086H, PJG165087H, PJG165088H, PJG165089H, PJG165090H, PJG165091H, PJG165092H, PJG165093H, PJG165094H, PJG165095H, PJG165096H, PJG165097H, PJG165098H, PJG165099H, PJG165100H, PJG165101H, PJG165102H, PJG165103H, PJG165104H, PJG165105H, PJG165106H, PJG165107H, PJG165108H, PJG165109H, PJG165110H, PJG165111H, PJG165112H, PJG165113H, PJG165114H, PJG165115H, PJG165116H, PJG165117H, PJG165118H, PJG165119H, PJG165121H, PJG165122H, PJG165123H, PJG165124H, PJG165125H, PJG165126H, PJG165127H, PJG165128H, PJG165131H, PJG165132H, PJG165133H, PJG165135H, PJG165136H, PJG165137H, PJG165138H, PJG165139H, PJG165140H, PJG165141H, PJG165142H, PJG165143H, PJG165144H, PJG165145H, PJG165146H, PJG165147H, PJG165148H, PJG165149H, PJG165150H, PJG165151H, PJG165152H, PJG165153H, PJG165154H, PJG165155H, PJG165156H, PJG165157H, PJG165158H, PJG165159H, PJG165160H, PJG165161H, PJG165162H, PJG165163H, PJG165164H, PJG165165H, PJG165166H, PJG165167H, PJG165168H, PJG165169H, PJG165170H, PJG166292H, PJG166293H, PJG166294H, PJG166295H, PJG166296H, PJG166297H, PJG166298H, PJG166299H, PJG166300H, PJG166301H, PJG166302H, PJG166303H, PJG166304H, PJG166305H, PJG166306H, PJG166307H, PJG166308H, PJG166309H, PJG166310H, PJG166311H, PJG166312H, PJG166313H, PJG166314H, PJG166315H, PJG166316H, PJG166317H, PJG166318H, PJG166319H, PJG166320H, PJG166321H, PJG166322H, PJG166323H, PJG166324H, PJG166325H, PJG166326H, PJG166327H, PJG166328H, PJG166329H, PJG166330H, PJG166331H, PJG166332H, PJG166334H, PJG166335H, PJG166336H, PJG166337H, PJG166338H, PJG166339H, PJG166340H, PJG166341H, PJG166342H, PJG166343H, PJG166344H, PJG166345H, PJG166346H, PJG166347H, PJG166348H, PJG166349H, PJG166350H, PJG166351H, PJG166352H, PJG166353H, PJG166354H, PJG166355H, PJG166356H, PJG166357H, PJG166358H, PJG166359H, PJG166360H, PJG166361H, PJG166362H, PJG166363H, PJG166364H, PJG166365H, PJG166366H, PJG166367H, PJG166368H, PJG166369H, PJG166370H, PJG166371H, PJG166517H, PJG167066H, PJG167070H, PJG167072H, PJG167073H, PJG167074H, PJG167076H, PJG167077H, PJG167078H, PJG167080H, PJG167081H, PJG167082H, PJG167083H, PJG167084H, PJG167085H, PJG167086H, PJG167087H, PJG167088H, PJG167089H, PJG167090H, PJG167091H, PJG167092H.PJG167093H, PJG167094H, PJG167095H, PJG167097H, PJG167100H, PJG167101H, PJG167102H, PJG167103H, PJG167104H, PJG167105H, PJG167106H, PJG167107H, PJG167108H, PJG167109H, PJG167110H, PJG167111H, PJG167112H, PJG167113H, PJG167114H, PJG167115H, PJG167116H, PJG167117H, PJG167118H, PJG167119H, PJG167120H, PJG167121H, PJG167122H, PJG167123H, PJG167124H, PJG167125H, PJG167126H, PJG167127H, PJG167128H, PJG167129H, PJG167131H, PJG167132H, PJG167133H, PJG167134H, PJG167135H, PJG167136H, PJG167137H, PJG167138H, PJG167139H, PJG167140H, PJG170012H, PJG170015H, PJG170016H, PJG170020H, PJG170021H, PJG170025H, PJG170026H, PJG170027H, PJG170033H, PJG172475H, PJG172476H, PJG172493H, PJG172494H, PJG172495H, PJG172496H, PJG172497H, PJG172498H, PJG172499H, PJG172500H, PJG172501H, PJG172502H, PJG172504H, PJG172505H, PJG172507H, PJG172508H, PJG172509H, PJG172510H, PJG172511H, PJG172513H, PJG172514H, PJG172515H, PJG172516H, PJG172517H, PJG172518H, PJG172519H, PJG172520H, PJG172521H, PJG172522H, PJG172523H, PJG172524H, PJG172525H, PJG172526H, PJG172527H, PJG172528H, PJG172529H, PJG172531H, PJG172532H, PJG172533H, PJG172534H, PJG172535H, PJG172537H, PJG172538H, PJG172539H, PJG172540H, PJG172541H, PJG172542H, PJG172543H, PJG172544H, PJG172545H, PJG172546H, PJG172547H, PJG172548H, PJG172549H, PJG172550H, PJG172551H, PJG172552H, PJG172553H, PJG172554H, PJG172556H, PJG172557H, PJG172558H, PJG172559H, PJG172560H, PJG172561H, PJG172562H, PJG172563H, PJG172564H, PJG172565H, PJG172566H, PJG172567H, PJG172568H, PJG172569H, PJG172570H, PJG172571H, PJG172572H, PJG172573H, PJG172574H, PJG172575H, PJG172576H, PJG172577H, PJG172578H, PJG172579H, PJG172580H, PJG172581H, PJG172582H, PJG172584H, PJG172585H, PJG172587H, PJG172588H, PJG172589H, PJG172591H, PJG172592H, PJG172593H, PJG172594H, PJG172596H, PJG172597H, PJG172598H, PJG172599H, PJG172601H, PJG172603H, PJG172604H, PJG172609H, PJG172611H, PJG172613H, PJG172615H, PJG172617H, PJG172621H, PJG172622H, PJG172626H, PJG172630H, PJG172631H, PJG172646H, PJG172650H, PJG172652H, PJG172653H, PJG172659H, PJG172662H, PJG173568H, PJG173569H, PJG173570H, PJG173571H, PJG173572H, PJG173573H, PJG173574H, PJG173575H, PJG173576H, PJG173577H, PJG173578H, PJG173579H, PJG173580H, PJG173581H, PJG173582H, PJG173583H, PJG173584H, PJG173585H, PJG173586H, PJG173587H, PJG173588H, PJG173589H, PJG173590H, PJG173591H, PJG173592H, PJG173593H, PJG173594H, PJG173595H, PJG173596H, PJG173597H, PJG173598H, PJG173599H, PJG173600H, PJG173601H, PJG173602H, PJG173603H, PJG173604H, PJG173605H, PJG173606H, PJG173607H, PJG173608H, PJG173609H, PJG173610H, PJG173611H, PJG173612H, PJG173613H, PJG173614H, PJG173615H, PJG173616H, PJG173617H, PJG173618H, PJG173619H, PJG173620H, PJG173621H, PJG173622H, PJG173623H, PJG173624H, PJG173625H, PJG173626H, PJG173627H, PJG173628H, PJG173629H, PJG173630H, PJG173631H, PJG173632H, PJG173633H, PJG173634H, PJG173635H, PJG173636H, PJG173637H, PJG173638H, PJG173639H, PJG173640H, PJG173641H, PJG173642H, PJG173643H, PJG173644H, PJG173646H, PJG173647H, PJG174289H, PJG174290H, PJG174292H, PJG174293H, PJG174294H, PJG174295H, PJG174296H, PJG174297H, PJG174298H, PJG174300H, PJG174301H, PJG174302H, PJG174303H, PJG174304H, PJG174305H, PJG174306H, PJG174307H, PJG174308H, PJG174309H, PJG174310H, PJG174312H, PJG174313H, PJG174314H, PJG174315H, PJG174318H, PJG174319H, PJG174320H, PJG174321H, PJG174322H, PJG174323H, PJG174324H, PJG174326H, PJG174327H, PJG174328H, PJG174329H, PJG174330H, PJG174331H, PJG174332H, PJG174334H, PJG174335H, PJG174336H, PJG174337H, PJG174338H, PJG174339H, PJG174340H, PJG174341H, PJG174342H, PJG174343H, PJG174344H, PJG174345H, PJG174346H, PJG174347H, PJG174348H, PJG174349H, PJG174350H, PJG174351H, PJG174352H, PJG174353H, PJG174354H, PJG174355H, PJG174356H, PJG174357H, PJG174358H, PJG174359H, PJG174360H, PJG174361H, PJG174362H, PJG174363H, PJG174364H, PJG174365H, PJG174367H, PJG174368H, PJG174369H, PJG174370H, PJG174371H, PJG174372H, PJG174373H, PJG174374H, PJG174375H, PJG174376H, PJG174377H, PJG174378H, PJG174379H, PJG174381H, PJG174383H, PJG174384H, PJG174385H, PJG175001H, PJG175002H, PJG175003H, PJG175004H, PJG175005H, PJG175006H, PJG175007H, PJG175008H, PJG175009H, PJG175010H, PJG175011H, PJG175012H, PJG175013H, PJG175014H, PJG175015H, PJG175016H, PJG175017H, PJG175018H, PJG175019H, PJG175020H, PJG175021H, PJG175022H, PJG175023H, PJG175024H, PJG175025H, PJG175026H, PJG175027H, PJG175028H, PJG175029H, PJG175030H, PJG175031H, PJG175032H, PJG175033H, PJG175034H, PJG175035H, PJG175036H, PJG175037H, PJG175038H, PJG175039H, PJG175040H, PJG175041H, PJG175042H, PJG175043H, PJG175044H, PJG175045H, PJG175046H, PJG175047H, PJG175048H, PJG175049H, PJG175052H, PJG175053H, PJG175054H, PJG175055H, PJG175056H, PJG175057H, PJG175058H, PJG175060H, PJG175061H, PJG175062H, PJG175063H, PJG175064H, PJG175065H, PJG175066H, PJG175067H, PJG175068H, PJG175069H, PJG175070H, PJG175071H, PJG175072H, PJG175073H, PJG175074H, PJG175126H, PJG175127H, PJG175129H, PJG175130H, PJG175131H, PJG175132H, PJG175133H, PJG175134H, PJG175135H, PJG175136H, PJG175137H, PJG175138H, PJG175139H, PJG175140H, PJG175141H, PJG175143H, PJG175144H, PJG175145H, PJG175146H, PJG175147H, PJG175148H, PJG175149H, PJG175150H, PJG175151H, PJG175152H, PJG175153H, PJG175154H, PJG175155H, PJG175156H, PJG175157H, PJG175158H, PJG175159H, PJG175160H, PJG175161H, PJG175162H, PJG175163H, PJG175164H, PJG175165H, PJG175166H, PJG175167H, PJG175168H, PJG175220H, PJG175221H, PJG175222H, PJG175223H, PJG175224H, PJG175225H, PJG175226H, PJG175227H, PJG175228H, PJG175229H, PJG175230H, PJG175231H, PJG175232H, PJG175233H, PJG175234H, PJG175235H, PJG175236H, PJG175237H, PJG175238H, PJG175239H, PJG175240H, PJG175241H, PJG175242H, PJG175243H, PJG175244H, PJG175245H, PJG175246H, PJG175247H, PJG175248H, PJG175249H, PJG175250H, PJG175429H, PJG175430H, PJG175431H, PJG175432H, PJG175433H, PJG175434H, PJG175435H, PJG175436H, PJG175437H, PJG175438H, PJG175439H, PJG175440H, PJG175441H, PJG175442H, PJG175443H, PJG175444H, PJG175445H, PJG175446H, PJG175447H, PJG175448H, PJG175449H, PJG175450H, PJG175451H, PJG175452H, PJG175453H, PJG175454H, PJG175455H, PJG175456H, PJG175457H, PJG175458H, PJG175459H, PJG175460H, PJG175461H, PJG175462H, PJG175463H, PJG175464H, PJG175465H, PJG175466H, PJG175467H, PJG175468H, PJG175469H, PJG175470H, PJG175471H, PJG175473H, PJG175474H, PJG175475H, PJG175476H, PJG175477H, PJG175478H, PJG175479H, PJG175480H, PJG175481H, PJG175482H, PJG175483H, PJG175484H, PJG175485H, PJG175486H, PJG175487H, PJG175488H, PJG175489H, PJG175490H, PJG175491H, PJG175492H, PJG175493H, PJG175494H, PJG175495H, PJG175496H, PJG175497H, PJG175498H, PJG175499H, PJG175500H, PJG175501H, PJG175502H, PJG175503H, PJG175525H, PJG175526H, PJG175527H, PJG175528H, PJG175529H, PJG175530H, PJG175531H, PJG175532H, PJG175533H, PJG175534H, PJG175535H, PJG175537H, PJG175538H, PJG175539H, PJG175540H, PJG175541H, PJG175542H, PJG175543H, PJG175544H, PJG175545H, PJG175546H, PJG175547H, PJG175548H, PJG175549H, PJG175550H, PJG175551H, PJG175552H, PJG175553H, PJG175554H, PJG175555H, PJG175556H, PJG175557H, PJG175558H, PJG175559H, PJG175560H, PJG175561H, PJG175562H, PJG175563H, PJG175564H, PJG175565H, PJG175566H, PJG175567H, PJG175568H, PJG175569H, PJG175570H, PJG175571H, PJG175572H, PJG175573H, PJG175574H, PJG175575H, PJG175576H, PJG175577H, PJG175578H, PJG175579H, PJG175580H, PJG175581H, PJG175582H, PJG175583H, PJG175584H, PJG175585H, PJG175586H, PJG175587H, PJG175588H, PJG175589H, PJG175590H, PJG175591H, PJG175592H, PJG175593H, PJG175594H, PJG175595H, PJG175596H, PJG175597H, PJG175598H, PJG175599H, PJG175939H, PJG176143H, PJG176532H, PJG176533H, PJG176534H, PJG176535H, PJG176536H, PJG176537H, PJG176538H, PJG176539H, PJG176540H, PJG176541H, PJG176542H, PJG176543H, PJG176544H, PJG176545H, PJG176546H, PJG176547H, PJG176548H, PJG176549H, PJG176550H, PJG176551H, PJG176552H, PJG176553H, PJG176554H, PJG176555H, PJG176556H, PJG176557H, PJG176558H, PJG176559H, PJG176561H, PJG176562H, PJG176563H, PJG176564H, PJG176565H, PJG176566H, PJG176567H, PJG176568H, PJG176569H, PJG176570H, PJG176571H, PJG176572H, PJG176573H, PJG176574H, PJG176575H, PJG176576H, PJG176577H, PJG176578H, PJG176579H, PJG176580H, PJG176581H, PJG176582H, PJG176583H, PJG176584H, PJG176585H, PJG176586H, PJG176587H, PJG176588H, PJG176589H, PJG176590H, PJG176591H, PJG176592H, PJG176593H, PJG176594H, PJG176595H, PJG176596H, PJG176597H, PJG176598H, PJG176599H, PJG176600H, PJG176601H, PJG176602H, PJG176603H, PJG176604H, PJG176605H, PJG176606H, PJG176607H, PJG176608H, PJG176609H, PJG176610H, PJG176611H, PJG176612H, PJG176615H, PJG176616H, PJG176617H, PJG176618H, PJG176619H, PJG176620H, PJG176622H, PJG176623H, PJG176624H, PJG176625H, PJG176627H, PJG176628H, PJG176629H, PJG176630H, PJG176631H, PJG176632H, PJG176633H, PJG176634H, PJG176635H, PJG176636H, PJG176637H, PJG176638H, PJG176639H, PJG176640H, PJG176641H, PJG176642H, PJG176643H, PJG176644H, PJG176645H, PJG176646H, PJG176647H, PJG176648H, PJG176649H, PJG176650H, PJG176652H, PJG176653H, PJG176654H, PJG176655H, PJG176656H, PJG176657H, PJG176658H, PJG176659H, PJG176660H, PJG176661H, PJG176662H, PJG176663H, PJG176664H, PJG176665H, PJG176666H, PJG176668H, PJG176669H, PJG176670H, PJG176671H, PJG176672H, PJG176673H, PJG176675H, PJG176676H, PJG176677H, PJG176678H, PJG176679H, PJG176680H, PJG176681H, PJG176682H, PJG176683H, PJG176684H, PJG176685H, PJG176686H, PJG176687H, PJG176688H, PJG176689H, PJG176690H, PJG176691H, PJG176692H, PJG176693H, PJG176695H, PJG176696H, PJG176697H, PJG176698H, PJG176699H, PJG176700H, PJG176701H, PJG176702H, PJG176703H, PJG176704H, PJG176705H, PJG176706H, PJG176707H, PJG176709H, PJG176711H, PJG176712H, PJG176713H, PJG176714H, PJG176715H, PJG176716H, PJG176717H, PJG176886H, PJG176887H, PJG176888H, PJG176889H, PJG176890H, PJG176891H, PJG176892H, PJG176893H, PJG176894H, PJG176895H, PJG176896H, PJG176936H, PJG176937H, PJG177538H, PJG177544H, PJG177547H, PJG177549H, PJG177555H, PJG177562H, PJG177565H, PJG177569H, PJG177570H, PJG177571H, PJG177572H, PJG177575H, PJG177577H, PJG177582H, PJG177584H, PJG177587H, PJG177591H, PJG177593H, PJG177611H, PJG178450H, PJG178454H, PJG178496H, PJG178530H, PJG178531H, PJG178532H, PJG178553H, PJG178577H, PJG178585H, PJG178594H, PJG178596H, PJG178597H, PJG178603H, PJG178619H, PJG178620H, PJG178625H, PJG178630H, PJG178716H, PJG178759H, PJG179813H, PJG183552H, PJG183555H, PJG183556H, PJG183558H, PJG183559H, PJG183560H, PJG183562H, PJG183563H, PJG183566H, PJG183570H, PJG183571H, PJG183575H, PJG230582M, PJG230584M, PJG230587M, PJG230590M, PJG230592M, PJG230596M, PJG230598M, PJG230600M, PJG230601M, PJG230606M, PJG230607M, PJG230609M, PJG230610M, PJG230612M, PJG230613M, PJG230615M, PJG230617M, PJG230619M, PJG230622M, PJG230623M, PJG230624M, PJG230625M, PJG230626M, PJG230628M, PJG230629M, PJG230633M, PJG230634M, PJG230635M, PJG230638M, PJG230639M, PJG230640M, PJG230642M, PJG230643M, PJG230645M, PJG230743M, PJG230744M, PJG230745M, PJG230746M, PJG230747M, PJG230748M, PJG230750M, PJG230751M, PJG230752M, PJG230753M, PJG230754M, PJG230756M, PJG230757M, PJG230758M, PJG230759M, PJG230760M, PJG230761M, PJG230762M, PJG230764M, PJG230765M, PJG230766M, PJG230767M, PJG230769M, PJG230770M, PJG230771M, PJG230772M, PJG230773M, PJG230774M, PJG230775M, PJG230776M, PJG230777M, PJG230778M, PJG230779M, PJG230780M, PJG230781M, PJG230782M, PJG230783M, PJG230784M, PJG230785M, PJG230786M, PJG230787M, PJG230788M, PJG230789M, PJG230791M, PJG230792M, PJG230794M, PJG230795M, PJG230796M, PJG230797M, PJG230798M, PJG230799M, PJG230800M, PJG230802M, PJG230803M, PJG230804M, PJG230805M, PJG230806M, PJG230807M, PJG230808M, PJG230809M, PJG230810M, PJG230811M, PJG230812M, PJG230813M, PJG230814M, PJG230815M, PJG230816M, PJG230817M, PJG230818M, PJG230819M, PJG230820M, PJG230821M, PJG230822M, PJG230823M, PJG230824M, PJG230825M, PJG230826M, PJG230827M, PJG230828M, PJG230829M, PJG230830M, PJG230831M, PJG230832M, PJG230833M, PJG230835M, PJG230836M, PJG230837M, PJG230838M, PJG230839M, PJG230840M, PJG230841M, PJG230842M, PJG230843M, PJG230844M, PJG230845M, PJG230846M, PJG230847M.PJG230848M, PJG230850M, PJG230851M, PJG230852M, PJG230853M, PJG230854M, PJG230855M, PJG230856M, PJG230857M, PJG230858M, PJG230859M, PJG230860M, PJG230969M, PJG230970M, PJG230971M, PJG230972M, PJG230973M, PJG230974M, PJG230975M, PJG230976M, PJG230977M, PJG230978M, PJG230979M, PJG230981M, PJG230982M, PJG230983M, PJG230984M, PJG230985M, PJG230986M, PJG230987M, PJG230988M, PJG230991M, PJG230992M, PJG230993M, PJG230995M, PJG230996M, PJG230997M, PJG230998M, PJG230999M, PJG231000M, PJG231001M, PJG231002M, PJG231003M, PJG231005M, PJG231006M, PJG231007M, PJG231008M, PJG231009M, PJG231010M, PJG231011M, PJG231012M, PJG231013M, PJG231014M, PJG231015M, PJG231016M, PJG231017M, PJG231018M, PJG231019M, PJG231020M, PJG231021M, PJG231023M, PJG231024M, PJG231025M, PJG231026M, PJG231027M, PJG231028M, PJG231029M, PJG231031M, PJG231032M, PJG231033M, PJG231034M, PJG231035M, PJG231036M, PJG231037M, PJG231038M, PJG231039M, PJG231040M, PJG231041M, PJG231042M, PJG231043M, PJG231044M, PJG231045M, PJG231206M, PJG231207M, PJG231208M, PJG231210M, PJG231211M, PJG231212M, PJG231214M, PJG231216M, PJG231217M, PJG231218M, PJG231219M, PJG231220M, PJG231222M, PJG231223M, PJG231224M, PJG231225M, PJG231226M, PJG231227M, PJG231228M, PJG231229M, PJG231230M, PJG231232M, PJG231235M, PJG231236M, PJG231237M, PJG231238M, PJG231239M, PJG231240M, PJG231241M, PJG231242M, PJG231244M, PJG231245M, PJG231246M, PJG231247M, PJG231248M, PJG231249M, PJG231250M, PJG231251M, PJG231252M, PJG231253M, PJG231254M, PJG231255M, PJG231256M, PJG231258M, PJG231261M, PJG231262M, PJG231263M, PJG231265M, PJG231266M, PJG231267M, PJG231268M, PJG231269M, PJG231271M, PJG231272M, PJG231273M, PJG231274M, PJG231275M, PJG231276M, PJG231277M, PJG231278M, PJG231279M, PJG231280M, PJG231281M, PJG231285M, PJG231286M, PJG231287M, PJG231289M, PJG231290M, PJG231291M, PJG231292M, PJG231293M, PJG231294M, PJG231295M, PJG231296M, PJG231297M, PJG231298M, PJG231300M, PJG231301M, PJG231302M, PJG231303M, PJG231304M, PJG231306M, PJG231307M, PJG231308M, PJG231310M, PJG231311M, PJG231312M, PJG231313M, PJG231314M, PJG231315M, PJG231317M, PJG231318M, PJG231319M, PJG231320M, PJG231321M, PJG231322M, PJG231323M, PJG231324M, PJG231325M, PJG231327M, PJG231328M, PJG231329M, PJG231331M, PJG231332M, PJG231333M, PJG231334M, PJG231335M, PJG231336M, PJG231337M, PJG231338M, PJG231340M, PJG231341M, PJG231343M, PJG231344M, PJG231345M, PJG231348M, PJG231349M, PJG231351M, PJG231353M, PJG231354M, PJG231355M, PJG231357M, PJG231358M, PJG231359M, PJG231360M, PJG231361M, PJG231362M, PJG231363M, PJG231364M, PJG231366M, PJG231367M, PJG231368M, PJG231369M, PJG231370M, PJG231371M, PJG231372M, PJG231373M, PJG231374M, PJG231375M, PJG231376M, PJG231377M, PJG231378M, PJG231379M, PJG231380M, PJG231381M, PJG231383M, PJG231384M, PJG231388M, PJG231390M, PJG231391M, PJG231392M, PJG231393M, PJG231395M, PJG231396M, PJG231397M, PJG231398M, PJG231399M, PJG231401M, PJG231402M, PJG231403M, PJG231404M, PJG231406M, PJG231407M, PJG231408M, PJG231410M, PJG231411M, PJG231412M, PJG231413M, PJG231414M, PJG231415M, PJG231417M, PJG231419M, PJG231420M, PJG231421M, PJG231422M, PJG231423M, PJG231424M, PJG231425M, PJG231426M, PJG231427M, PJG231428M, PJG231429M, PJG231430M, PJG231431M, PJG231432M, PJG231433M, PJG231434M, PJG231435M, PJG231436M, PJG231437M, PJG231438M, PJG231439M, PJG231440M, PJG231441M, PJG231442M, PJG231443M, PJG231444M, PJG231445M, PJG231446M, PJG231447M, PJG231448M, PJG231449M, PJG231450M, PJG231451M, PJG231452M, PJG231453M, PJG231454M, PJG231455M, PJG231457M, PJG231458M, PJG231460M, PJG231461M, PJG231462M, PJG231463M, PJG231465M, PJG231466M, PJG231467M, PJG231468M, PJG231469M, PJG231470M, PJG231472M, PJG231474M, PJG231475M, PJG231476M, PJG231477M, PJG231478M, PJG231479M, PJG231480M, PJG231482M, PJG231483M, PJG231484M, PJG231486M, PJG231487M, PJG231488M, PJG231489M, PJG231490M, PJG231491M, PJG231711M, PJG231719M, PJG231735M, PJG231736M, PJG231737M, PJG231739M, PJG231740M, PJG231741M, PJG231742M, PJG231743M, PJG231744M, PJG231745M, PJG231747M, PJG231749M, PJG231750M, PJG231751M, PJG231752M, PJG231753M, PJG231754M, PJG231756M, PJG231757M, PJG231758M, PJG231759M, PJG231760M, PJG231761M, PJG231762M, PJG231763M, PJG231764M, PJG231765M, PJG231767M, PJG231768M, PJG231769M, PJG231770M, PJG231773M, PJG231774M, PJG231775M, PJG231776M, PJG231777M, PJG231778M, PJG231779M, PJG231781M, PJG231782M, PJG231783M, PJG231784M, PJG231785M, PJG231786M, PJG231788M, PJG231789M, PJG231790M, PJG231791M, PJG231792M, PJG231793M, PJG231794M, PJG231795M, PJG231796M, PJG231797M, PJG231798M, PJG231799M, PJG231800M, PJG231802M, PJG231803M, PJG231804M, PJG231805M, PJG231806M, PJG231807M, PJG231809M, PJG231810M, PJG231811M, PJG231812M, PJG231814M, PJG231815M, PJG231817M, PJG231818M, PJG231819M, PJG231821M, PJG231822M, PJG231823M, PJG231824M, PJG231825M, PJG231826M, PJG231827M, PJG231828M, PJG231829M, PJG231831M, PJG231832M, PJG231833M, PJG231834M, PJG231836M, PJG231837M, PJG231838M, PJG231839M, PJG231840M, PJG231841M, PJG231842M, PJG231843M, PJG231844M, PJG231845M, PJG231846M, PJG231847M, PJG231848M, PJG231849M, PJG231850M, PJG231851M, PJG231852M, PJG231853M, PJG231855M, PJG231856M, PJG231857M, PJG231858M, PJG231859M, PJG231860M, PJG231861M, PJG231862M, PJG231863M, PJG231864M, PJG231865M, PJG231866M, PJG231867M, PJG231868M, PJG231869M, PJG231871M, PJG231873M, PJG231874M, PJG231876M, PJG231877M, PJG231878M, PJG231879M, PJG231880M, PJG231882M, PJG231883M, PJG231884M, PJG231885M, PJG231886M, PJG231887M, PJG231888M, PJG231889M, PJG231890M, PJG231891M, PJG231893M, PJG231894M, PJG231896M, PJG231897M, PJG231899M, PJG231900M, PJG231901M, PJG231902M, PJG231903M, PJG231904M, PJG231905M, PJG231907M, PJG231908M, PJG231909M, PJG231910M, PJG231911M, PJG231912M, PJG231913M, PJG231914M, PJG231915M, PJG231916M, PJG231917M, PJG231918M, PJG231919M, PJG231920M, PJG231921M, PJG231922M, PJG231923M, PJG231925M, PJG231926M, PJG231927M, PJG231928M, PJG231930M, PJG231931M, PJG231932M, PJG231933M, PJG231934M, PJG231936M, PJG231937M, PJG231938M, PJG231939M, PJG231940M, PJG231941M, PJG231942M, PJG231943M, PJG231944M, PJG231945M, PJG231946M, PJG231947M, PJG231948M, PJG231949M, PJG231950M, PJG231951M, PJG231952M, PJG231953M, PJG231954M, PJG231955M, PJG231956M, PJG231957M, PJG231958M, PJG231959M, PJG231960M, PJG231961M, PJG231963M, PJG231964M, PJG231965M, PJG231966M, PJG231967M, PJG231968M, PJG231969M, PJG231970M, PJG231971M, PJG231972M, PJG231973M, PJG231975M, PJG231976M, PJG231978M, PJG231979M, PJG231980M, PJG231981M, PJG231982M, PJG231983M, PJG231985M, PJG231986M, PJG231987M, PJG231988M, PJG231989M, PJG231991M, PJG231992M, PJG231993M, PJG231994M, PJG231995M, PJG231996M, PJG231997M, PJG231998M, PJG231999M, PJG232001M, PJG232002M, PJG232003M, PJG232004M, PJG232005M, PJG232006M, PJG232007M, PJG232008M, PJG232009M, PJG232010M, PJG232011M, PJG232012M, PJG232013M, PJG232014M, PJG232015M, PJG232016M, PJG232017M, PJG232018M, PJG232019M, PJG232020M, PJG232021M, PJG232022M, PJG232023M, PJG232024M, PJG232025M, PJG232026M, PJG232028M, PJG232029M, PJG232030M, PJG232031M, PJG232032M, PJG232034M, PJG232035M, PJG232036M, PJG232037M, PJG232038M, PJG232039M, PJG232041M, PJG232042M, PJG232044M, PJG232045M, PJG232046M, PJG232047M, PJG232048M, PJG232050M, PJG232051M, PJG232052M, PJG232053M, PJG232054M, PJG232055M, PJG232056M, PJG232057M, PJG232058M, PJG232059M, PJG232060M, PJG232062M, PJG232063M, PJG232064M, PJG232065M, PJG232066M, PJG232067M, PJG232068M, PJG232069M, PJG232070M, PJG232072M, PJG232073M, PJG232074M, PJG232076M, PJG232077M, PJG232078M, PJG232079M, PJG232080M, PJG232081M, PJG232082M, PJG232083M, PJG232084M, PJG232085M, PJG232086M, PJG232087M, PJG232089M, PJG232090M, PJG232091M, PJG232092M, PJG232093M, PJG232094M, PJG232095M, PJG232096M, PJG232097M, PJG232098M, PJG232099M, PJG232100M, PJG232101M, PJG232102M, PJG232103M, PJG232106M, PJG232107M, PJG232108M, PJG232110M, PJG232111M, PJG232112M, PJG232113M, PJG232114M, PJG232115M, PJG232116M, PJG232117M, PJG232118M, PJG232119M, PJG232120M, PJG232121M, PJG232122M, PJG232124M, PJG232125M, PJG232126M, PJG232127M, PJG232128M, PJG232129M, PJG232131M, PJG232132M, PJG232133M, PJG232134M, PJG232135M, PJG232136M, PJG232137M, PJG232138M, PJG232139M, PJG232140M, PJG232141M, PJG232142M, PJG232143M, PJG232144M, PJG232146M, PJG232147M, PJG232148M, PJG232149M, PJG232151M, PJG232152M, PJG232153M, PJG232154M, PJG232155M, PJG232156M, PJG232157M, PJG232159M, PJG232160M, PJG232161M, PJG232162M, PJG232163M, PJG232164M, PJG232165M, PJG232166M, PJG232167M, PJG232168M, PJG232169M, PJG232170M, PJG232171M, PJG232172M, PJG232173M, PJG232174M, PJG232175M, PJG232176M, PJG232177M, PJG232178M, PJG232179M, PJG232180M, PJG232181M, PJG232182M, PJG232183M, PJG232184M, PJG232185M, PJG232186M, PJG232187M, PJG232188M, PJG232189M, PJG232190M, PJG232191M, PJG232192M, PJG232193M, PJG232194M, PJG232195M, PJG232196M, PJG232197M, PJG232198M, PJG232199M, PJG232200M, PJG232201M, PJG232202M, PJG232203M, PJG232204M, PJG232205M, PJG232206M, PJG232207M, PJG232208M, PJG232209M, PJG232211M, PJG232212M, PJG232213M, PJG232214M, PJG232215M, PJG232216M, PJG232217M, PJG232218M, PJG232219M, PJG232220M, PJG232221M, PJG232222M, PJG232223M, PJG232224M, PJG232225M, PJG232226M, PJG232227M, PJG232228M, PJG232229M, PJG232230M, PJG232231M, PJG232232M, PJG232233M, PJG232234M, PJG232235M, PJG232236M, PJG232237M, PJG232238M, PJG232239M, PJG232240M, PJG232241M, PJG232242M, PJG232243M, PJG232244M, PJG232245M, PJG232247M, PJG232248M, PJG232249M, PJG232250M, PJG232251M, PJG232252M, PJG232253M, PJG232255M, PJG232256M, PJG232257M, PJG232264M, PJG232265M, PJG232266M, PJG232269M, PJG232271M, PJG232272M, PJG232273M, PJG232274M, PJG232275M, PJG232276M, PJG232277M, PJG232278M, PJG232279M, PJG232280M, PJG232281M, PJG232282M, PJG232283M, PJG232284M, PJG232286M, PJG232287M, PJG232288M, PJG232289M, PJG232358M, PJG232359M, PJG232396M, PJG233014M, PJG233015M, PJG233016M, PJG233017M, PJG233018M, PJG233019M, PJG233020M, PJG233021M, PJG233022M, PJG233023M, PJG233024M, PJG233025M, PJG233026M, PJG233027M, PJG233028M, PJG233029M, PJG233030M, PJG233031M, PJG233032M, PJG233033M, PJG233034M, PJG233035M, PJG233036M, PJG233037M, PJG233039M, PJG233040M, PJG233041M, PJG233042M, PJG233043M, PJG233044M, PJG233045M, PJG233046M, PJG233047M, PJG233048M, PJG233050M, PJG233051M, PJG233052M, PJG233053M, PJG233054M, PJG233055M, PJG233056M, PJG233057M, PJG233058M, PJG233059M, PJG233060M, PJG233061M, PJG233062M, PJG233064M, PJG233065M, PJG233066M, PJG233067M, PJG233068M, PJG233069M, PJG233070M, PJG233071M, PJG233072M, PJG233073M, PJG233074M, PJG233075M, PJG233076M, PJG233077M, PJG233078M, PJG233079M, PJG233080M, PJG233081M, PJG233082M, PJG233083M, PJG233084M, PJG233085M, PJG233086M, PJG233087M, PJG233088M, PJG233089M, PJG233090M, PJG233091M, PJG233092M, PJG233093M, PJG233094M, PJG233095M, PJG233096M, PJG233097M, PJG233098M, PJG233099M, PJG233100M, PJG233101M, PJG233102M, PJG233103M, PJG233104M, PJG233105M, PJG233107M, PJG233108M, PJG233109M, PJG233110M, PJG233111M, PJG233112M, PJG233113M, PJG233114M, PJG233115M, PJG233116M, PJG233117M, PJG233118M, PJG233119M, PJG233120M, PJG233121M, PJG233122M, PJG233123M, PJG233124M, PJG233125M, PJG233126M, PJG233127M, PJG233128M, PJG233129M, PJG233130M, PJG233132M, PJG233133M, PJG233134M, PJG233136M, PJG233137M, PJG233138M, PJG233139M, PJG233141M, PJG233142M, PJG233143M, PJG233146M, PJG233147M, PJG233148M, PJG233149M, PJG233150M, PJG233151M, PJG233152M, PJG233153M, PJG233154M, PJG233155M, PJG233156M, PJG233157M, PJG233158M, PJG233159M, PJG233160M, PJG233161M, PJG233162M, PJG233163M, PJG233165M, PJG233166M, PJG233168M, PJG233169M, PJG233170M, PJG233171M, PJG233172M, PJG233173M, PJG233174M, PJG233176M, PJG233177M, PJG233178M, PJG233179M, PJG233180M, PJG233181M, PJG233182M, PJG233183M, PJG233184M, PJG233185M, PJG233186M, PJG233187M, PJG233188M, PJG233189M, PJG233191M, PJG233192M, PJG233193M, PJG233194M, PJG233195M, PJG233197M, PJG233198M, PJG233199M, PJG233200M, PJG233201M, PJG233202M, PJG233203M, PJG233204M, PJG233206M, PJG233207M, PJG233208M, PJG233209M, PJG233210M, PJG233211M, PJG233212M, PJG233213M.PJG233215M, PJG233216M, PJG233217M, PJG233218M, PJG233219M, PJG233220M, PJG233221M, PJG233222M, PJG233290M, PJG233291M, PJG233292M, PJG233293M, PJG233294M, PJG233295M, PJG233296M, PJG233297M, PJG233298M, PJG233300M, PJG233301M, PJG233303M, PJG233304M, PJG233305M, PJG233306M, PJG233307M, PJG233308M, PJG233309M, PJG233310M, PJG233311M, PJG233312M, PJG233313M, PJG233314M, PJG233315M, PJG233316M, PJG233317M, PJG233318M, PJG233319M, PJG233320M, PJG233321M, PJG233322M, PJG233323M, PJG233324M, PJG233325M, PJG233326M, PJG233328M, PJG233329M, PJG233331M, PJG233332M, PJG233333M, PJG233334M, PJG233335M, PJG233336M, PJG233337M, PJG233338M, PJG233339M, PJG233340M, PJG233341M, PJG233342M, PJG233343M, PJG233344M, PJG233347M, PJG233348M, PJG233349M, PJG233350M, PJG233351M, PJG233352M, PJG233353M, PJG233354M, PJG233356M, PJG233357M, PJG233358M, PJG233359M, PJG233360M, PJG233361M, PJG233362M, PJG233363M, PJG233364M, PJG233365M, PJG233366M, PJG233367M, PJG233368M, PJG233369M, PJG233370M, PJG233745M, PJG233747M, PJG233749M, PJG233750M, PJG233751M, PJG233752M, PJG233755M, PJG233757M, PJG233761M, PJG233763M, PJG233764M, PJG233768M, PJG233770M, PJG233771M, PJG233772M, PJG233773M, PJG233774M, PJG233775M, PJG233776M, PJG233777M, PJG233780M, PJG233781M, PJG233782M, PJG233785M, PJG233786M, PJG233788M, PJG233789M, PJG233790M, PJG233791M, PJG233792M, PJG233793M, PJG233795M, PJG233796M, PJG233800M, PJG233806M, PJG233807M, PJG233808M, PJG233809M, PJG233811M, PJG233815M, PJG233817M, PJG233818M, PJG233819M, PJG233821M, PJG233822M, PJG233823M, PJG233824M, PJG233830M, PJG233838M, PJG233846M, PJG233857M, PJG235260M, PJG432779H, PJG432780H, PJG432781H, PJG432782H, PJG432783H, PJG432784H, PJG432785H, PJG432786H, PJG432787H, PJG432788H, PJG432789H, PJG432791H, PJG432792H, PJG432793H, PJG432794H, PJG432795H, PJG432796H, PJG432797H, PJG432798H, PJG432799H, PJG432800H, PJG432801H, PJG432802H, PJG432803H, PJG432804H, PJG432805H, PJG432806H, PJG432807H, PJG432808H, PJG432857H, PJG432858H, PJG432859H, PJG432860H, PJG432861H, PJG432862H, PJG432865H, PJG432867H, PJG432870H, PJG432872H, PJG432875H, PJG432877H, PJG432878H, PJG432879H, PJG432880H, PJG432881H, PJG432882H, PJG432883H, PJG432884H, PJG432885H, PJG432886H, PJG432887H, PJG432888H, PJG432889H, PJG432890H, PJG432891H, PJG432892H, PJG432893H, PJG432895H, PJG677442S, PJG677443S, PJG677444S, PJG677445S, PJG677446S, PJG677447S, PJG677448S, PJG677449S, PJG677450S, PJG677451S, PJG677453S, PJG677454S, PJG677455S, PJG677456S, PJG677457S, PJG677458S, PJG677459S, PJG677461S, PJG677462S, PJG677463S, PJG677465S, PJG677466S, PJG677467S, PJG677468S, PJG677470S, PJG677471S, PJG677472S, PJG677473S, PJG677475S, PJG677476S, PJG677477S, PJG677478S, PJG677481S, PJG677482S, PJG677483S, PJG677484S, PJG677485S, PJG677486S, PJG677487S, PJG677488S, PJG677489S, PJG677490S, PJG677491S, PJG677492S, PJG677493S, PJG677494S, PJG677495S, PJG677496S, PJG677497S, PJG677498S, PJG677499S, PJG677500S, PJG677501S, PJG677502S, PJG677503S, PJG677504S, PJG677505S, PJG677506S, PJG677507S, PJG677508S, PJG677509S, PJG677510S, PJG677511S, PJG677512S, PJG677513S, PJG677514S, PJG677515S, PJG677516S, PJG677518S, PJG677519S, PJG677521S, PJG677522S, PJG677523S, PJG677524S, PJG677526S, PJG677528S, PJG677529S, PJG677531S, PJG677532S, PJG677533S, PJG677534S, PJG677535S, PJG677536S, PJG677537S, PJG677538S, PJG677540S, PJG677541S, PJG677542S, PJG677544S, PJG677545S, PJG677546S, PJG677547S, PJG677548S, PJG677549S, PJG677550S, PJG677551S, PJG677552S, PJG677553S, PJG677554S, PJG677555S, PJG677556S, PJG677559S, PJG677560S, PJG677561S, PJG677562S, PJG677563S, PJG677564S, PJG677565S, PJG677567S, PJG677568S, PJG677570S, PJG677571S, PJG677572S, PJG677573S, PJG677574S, PJG677575S, PJG677576S, PJG677578S, PJG677580S, PJG677581S, PJG678256S, PJG678257S, PJG678258S, PJG678260S, PJG678261S, PJG678262S, PJG678268S, PJG678269S, PJG678270S, PJG678271S, PJG678273S, PJG678275S, PJG678278S, PJG678279S, PJG678282S, PJG678283S, PJG678284S, PJG678285S, PJG678286S, PJG678287S, PJG678288S, PJG678289S, PJG678291S, PJG678308S, PJG678313S, PJG678320S, PJG678327S, PJG678328S, PJG678329S, PJG678330S, PJG678331S, PJG678332S, PJG678333S, PJG678334S, PJG678335S, PJG678356S, PJG678357S, PJG678358S, PJG678363S, PJG678368S, PJG678375S, PJG678392S, PJG678395S, PJG678397S, PJG678400S, PJG678401S, PJG678403S, PJG678405S, PJG678406S, PJG678586S, PJG678768S, PJG678771S, PJG678772S, PJG678774S, PJG678776S, PJG678778S, PJG678781S, PJG678782S, PJG678785S, PJG678786S, PJG678793S, PJG678797S, PJG678798S, PJG678800S, PJG678801S, PJG679448S, PJG679449S, PJG679451S, PJG679452S, PJG679453S, PJG679454S, PJG679455S, PJG679456S, PJG679457S, PJG679458S, PJG679459S, PJG679460S, PJG679462S, PJG679463S, PJG679465S, PJG679466S, PJG679467S, PJG679468S, PJG679470S, PJG679471S, PJG679472S, PJG679473S, PJG679478S, PJG679479S, PJG679484S, PJG679488S, PJG679489S, PJG679490S, PJG679491S, PJG679524S, PJG679525S, PJG679527S, PJG679528S, PJG679529S, PJG679532S, PJG679533S, PJG679943S, PJG679944S, PJG679947S, PJG679951S, PJG679952S, PJG679958S, PJG679959S, PJG679966S, PJG679976S, PJG679977S, PJG679980S, PJG679985S, PJG679993S, PJG679996S, PJG679997S, PJG679999S, PJG680003S, PJG680004S, PJG680006S, PJG680008S, PJG680011S, PJG680013S, PJG680017S, PJG680018S, PJG680019S, PJG680021S, PJG680023S, PJG680025S, PJG680026S, PJG680029S, PJG680037S, PJG680039S, PJG680042S, PJG680043S, PJG680044S, PJG680051S, PJG680054S, PJG680057S, PJG680061S, PJG680064S, PJG680068S, PJG680078S, PJG680081S, PJG680083S, PJG680085S, PJG680088S, PJG680095S, PJG680101S, PJG680107S, PJG680116S, PJG680117S, PJG680120S, PJG680125S, PJG680126S, PJG680132S, PJG681406S, PJG681407S, PJG681426S, PJG681427S, PJG681459S, PJG681460S, PJG681463S, PJG681473S, PJG681484S, PJG681503S, PJG681510S, PJG681525S, PJG681530S, PJG681542S, PJG681566S, PJG681572S, PJG681588S, PJG681591S, PJG681612S, PJG681892S, PJG681894S, PJG681896S, PJG681899S, PJG681900S, PJG681902S, PJG681908S, PJG681910S, PJG681911S, PJG681912S, PJG681919S, PJG681921S, PJG681935S, PJG681939S, PJG681943S, PJG682338S, PJG682346S, PJG682348S, PJG682361S, PJG682368S, PJG682369S, PJG682370S, PJG682374S, PJG682387S, PJG682391S, PJG682712S, PJG684092S, PJG690024S, PJG690052S, PJG690054S, PJG690087S, PJG696014S, PJG696433S, PJG696434S, PJG696435S, PJG696437S, PJG696438S, PJG696440S, PJG696442S, PJG696443S, PJG696445S, PJG696446S, PJG696448S, PJG696451S, PJG696452S, PJG696457S, PJG696460S, PJG696462S, PJG696464S, PJG696466S, PJG696467S, PJG696468S, PJG696469S, PJG696470S, PJG696471S, PJG696472S, PJG696475S, PJG696477S, PJG696478S, PJG696479S, PJG696481S, PJG696482S, PJG696483S, PJG696485S, PJG696487S, PJG696488S, PJG696489S, PJG696490S, PJG696496S, PJG696497S, PJG696498S, PJG696502S, PJG696503S, PJG696505S, PJG696508S, PJG696513S, PJG696514S, PJG696515S, PJG696517S, PJG696518S, PJG696519S, PJG696520S, PJG697007S, PJG697064S, PJG697084S, PJG697085S, PJG697086S, PJG697089S, PJG697090S, PJG697092S, PJG697093S, PJG697094S, PJG697095S, PJG697096S, PJG697097S, PJG697098S, PJG697099S, PJG697101S, PJG697102S, PJG697103S, PJG697104S, PJG697105S, PJG697107S, PJG697108S, PJG697109S, PJG697110S, PJG697111S, PJG697112S, PJG697113S, PJG697114S, PJG697115S, PJG697116S, PJG697117S, PJG697118S, PJG697119S, PJG697120S, PJG697121S, PJG697122S, PJG697123S, PJG697124S, PJG697125S, PJG697126S, PJG697127S, PJG697128S, PJG697129S, PJG697130S, PJG697131S, PJG697132S, PJG697133S, PJG697134S, PJG697573S, PJG697990S, PJG698640S, PJG698660S, PJG747654C, PJG747663C, PJG747666C, PJG747690C, PJG747692C, PJG747700C, PJG747717C, PJG747721C, PJG747726C, PJG747727C, PJG747731C, PJG747734C, PJG747741C, PJG747746C, PJG748310S, PJG748321S, PJG748322S, PJG748324S, PJG748337S, PJG748338S, PJG748340S, PJG748347S, PJG748352S, PJG748357S, PJG748363S, PJG748364S, PJG748366S, PJG748368S, PJG748369S, PJG748372S, PJG748375S, PJG748376S, PJG748379S, PJG748380S, PJG748384S, PJG748388S, PJG748390C, PJG748391C, PJG748392C, PJG748393C, PJG748396C, PJG748398C, PJG748399C, PJG748403C, PJG748405C, PJG748408C, PJG748409C, PJG748410C, PJG748411C, PJG748412C, PJG748413C, PJG748414C, PJG748416C, PJG748417C, PJG748418C, PJG748419C, PJG748422C, PJG748423C, PJG748424C, PJG748425C, PJG748426C, PJG748428C, PJG748430C, PJG748433C, PJG748436C, PJG748437C, PJG748438C, PJG748439C, PJG748440C, PJG748441C, PJG748442C, PJG748444C, PJG748445C, PJG748446C, PJG748468C, PJG748469C, PJG748470C, PJG748471C, PJG748472C, PJG748479C, PJG748480C, PJG748481C, PJG748482C, PJG748492C, PJG748497C, PJG748503C, PJG748511C, PJG748524C, PJG748527C, PJG748531C, PJG748536C, PJG748539C, PJG748545C, PJG748546C, PJG748558C, PJG748563C, PJG748565C, PJG748566C, PJG748572C, PJG748586C, PJG748587C, PJG748588C, PJG748591C, PJG748592C, PJG748593C, PJG748597C, PJG748598C, PJG748599C, PJG748600C, PJG748602C, PJG748610C, PJG748633C, PJG748639C, PJG748640C, PJG748641C, PJG748642C, PJG748644C, PJG748645C, PJG748647C, PJG748652C, PJG748653C, PJG748655C, PJG748657C, PJG748658C, PJG748666C, PJG748668C, PJG748714C, PJG748715C, PJG748716C, PJG748719C, PJG748723C, PJG748724C, PJG748842S, PJG748844S, PJG748846S, PJG748866S, PJG748871S, PJG748878S, PJG748879S, PJG748880S, PJG748896S, PJG748913S, PJG748920S, PJG748939S, PJG748940S, PJG748945S, PJG748949S, PJG748951S, PJG748971S, PJG748974S, PJG748983S, PJG748985S, PJG749020S, PJG749046S, PJG749050S, PJG749060S, PJG749089S, PJG749091S, PJG749103S, PJG749112S, PJG749162S, PJG749164S, PJG749166S, PJG749169S, PJG749173S, PJG749176S, PJG749184S, PJG749195S, PJG749211S, PJG749220S, PJG749229S, PJG749234S, PJG749235S, PJG749238S, PJG749240S, PJG749241S, PJG749243S, PJG749244S, PJG749248S, PJG749250S, PJG749252S, PJG749254S, PJG749255S, PJG749258S, PJG749259S, PJG749260S, PJG749261S, PJG749266S, PJG749273S, PJG749275S, PJG749277S, PJG749278S, PJG749280S, PJG749281S, PJG749282S, PJG749283S, PJG749286S, PJG749287S, PJG749288S, PJG749290S, PJG749293S, PJG749294S, PJG749296S, PJG749297S, PJG749299S, PJG749303S, PJG749304S, PJG749305S, PJG749308S, PJG749309S, PJG749311S, PJG749313S, PJG750351S, PJG750352S, PJG750355S, PJG750356S, PJG750358S, PJG750360S, PJG750362S, PJG750363S, PJG750365S, PJG750367S, PJG750368S, PJG750370S, PJG750371S, PJG750372S, PJG750373S, PJG750374S, PJG750375S, PJG750376S, PJG750377S, PJG750378S, PJG750380S, PJG750381S, PJG750382S, PJG750384S, PJG750385S, PJG750386S, PJG750387S, PJG750388S, PJG750389S, PJG750391S, PJG750392S, PJG750393S, PJG750394S, PJG750395S, PJG750396S, PJG750398S, PJG750399S, PJG750401S, PJG750426S, PJG750427S, PJG750428S, PJG750429S, PJG750430S, PJG750432S, PJG750433S, PJG750434S, PJG750436S, PJG750443S, PJG750890C, PJG750891C, PJG750893C, PJG750895C, PJG750896C, PJG750897C, PJG750898C, PJG750899C, PJG750900C, PJG750901C, PJG750902C, PJG750903C, PJG750904C, PJG750905C, PJG750907C, PJG750908C, PJG750909C, PJG750910C, PJG750911C, PJG750912C, PJG750913C, PJG750915C, PJG750916C, PJG750917C, PJG750918C, PJG750919C, PJG750920C, PJG750921C, PJG750922C, PJG750923C, PJG750924C, PJG750925C, PJG750927C, PJG750928C, PJG750929C, PJG750931C, PJG750932C, PJG750933C, PJG750934C, PJG750936C, PJG750940C, PJG750941C, PJG750942C, PJG750943C, PJG750944C, PJG750945C, PJG750946C, PJG750947C, PJG750948C, PJG750951C, PJG750952C, PJG750953C, PJG750954C, PJG750955C, PJG750956C, PJG750957C, PJG750958C, PJG750960C, PJG750962C, PJG750963C, PJG750964C, PJG750965C, PJG750967C, PJG750968C, PJG750969C, PJG750970C, PJG750971C, PJG750972C, PJG750973C, PJG750974C, PJG750976C, PJG750977C, PJG750978C, PJG750979C, PJG750980C, PJG750982C, PJG750983C, PJG750987C, PJG750988C, PJG750990C, PJG750991C, PJG750992C, PJG750994C, PJG750995C, PJG750996C, PJG750997C, PJG750999C, PJG751000C, PJG751002C, PJG751003C, PJG751004C, PJG751006C, PJG751008C, PJG751010C, PJG751012C, PJG751013C, PJG751160S, PJG751161S, PJG751162S, PJG751164S, PJG751165S, PJG751166S, PJG751167S, PJG751171S, PJG751178S, PJG751181S, PJG751186S, PJG751191S, PJG751192S, PJG751193S, PJG751196S, PJG751197S, PJG751198S, PJG751200S, PJG751202S, PJG751213S, PJG751218S, PJG751223S, PJG751232S.PJG751239S, PJG751243S, PJG751246S, PJG751252S, PJG751265S, PJG751268S, PJG751274S, PJG751275S, PJG751284S, PJG751307S, PJG751309S, PJG751325S, PJG751327S, PJG751333S, PJG751772C, PJG751776C, PJG751777C, PJG751786C, PJG751792C, PJG751794C, PJG751798C, PJG751800C, PJG751802C, PJG751808C, PJG751809C, PJG751834C, PJG751840C, PJG751863C, PJG751873C, PJG751904C, PJG751911C, PJG751919C, PJG753026S, PJG753037S, PJG753048S, PJG753050S, PJG753060S, PJG753063S, PJG753064S, PJG753081S, PJG753086S, PJG753096S, PJG753098S, PJG753100S, PJG753117S, PJG753120S, PJG753121S, PJG753125S, PJG753411S, PJG753416S, PJG753422S, PJG753424S, PJG753425S, PJG753426S, PJG753427S, PJG753431S, PJG753434S, PJG753435S, PJG753436S, PJG753437S, PJG753439S, PJG753441S, PJG753445S, PJG753447S, PJG753449S, PJG753450S, PJG753451S, PJG753458S, PJG753461S, PJG753463S, PJG753469S, PJG753470S, PJG753471S, PJG753473S, PJG753476S, PJG753478S, PJG753479S, PJG753480S, PJG753481S, PJG753485S, PJG753486S, PJG753487S, PJG753488S, PJG753489S, PJG753490S, PJG753496S, PJG753497S, PJG753502S, PJG753505S, PJG753506S, PJG753508S, PJG753509S, PJG753511S, PJG753513S, PJG753515S, PJG753517S, PJG753518S, PJG753519S, PJG753521S, PJG753522S, PJG753523S, PJG753525S, PJG753526S, PJG753587S, PJG753588S, PJG753589S, PJG753590S, PJG753591S, PJG753592S, PJG753594S, PJG753595S, PJG753596S, PJG753977S, PJG754027S, PJG754496S, PJG754501S, PJG754503S, PJG754504S, PJG754509S, PJG754511S, PJG754513S, PJG754517S, PJG754522S, PJG754525S, PJG754528S, PJG754534S, PJG754541S, PJG754544S, PJG754550S, PJG754551S, PJG754557S, PJG754558S, PJG754559S, PJG754564S, PJG754565S, PJG754567S, PJG754572S, PJG754575S, PJG754576S, PJG754581S, PJG754588S, PJG754589S, PJG754590S, PJG754591S, PJG754592S, PJG754593S, PJG754974C, PJG754999C, PJG755024C, PJG755028C, PJG755052C, PJG755079C, PJG755107C, PJG755118C, PJG755128C, PJG755130C, PJG755171C, and PJG755468S.
  • Product Classification
  • Device Class
    3
  • Implanted device?
    Yes
  • Distribution
    Worldwide Distribution -- USA, including Washington D.C., Guam, Puerto Rico, and Virgin Islands; Albania, Algeria, Argentina, Armenia, Aruba, Australia, Austria, Bahamas, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, Guadeloupe, Honduras, Hong Kong, Hungary, India, Iran, Ireland, Israel, Italy, Japan, Jordan, Kenya, Kuwait, Latin America, Latvia, Lebanon, Luxembourg, Martinique, Mauritius, Mexico, Morocco, Nepal, Netherlands, Netherlands Antilles, Norway, Oman, Pakistan, Palestinian Territory Occupied, Peru, Poland, Portugal, Qatar, Reunion, Romania, Saudi Arabia, Senegal, Slovakia (Slovak Republic), Slovenia, South Africa, Spain, Sri Lanka, Sudan, Suriname, Sweden, Switzerland, Trinidad and Tobago, Tunisia, Turkey, United Arab Emirates, United Kingdom, Uruguay, Vatican City State (Holy See), and Venezuela.
  • Product Description
    Medtronic SIGMA 100/200/300 Series IPGs, Single Chamber Pacemaker, model SVVI103, for ventricular use only.
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Medtronic Inc. Cardiac Rhythm Disease Managment, 8200 Coral Sea St. N.E., Saint Paul MN 55112
  • Manufacturer Parent Company (2017)
  • Manufacturer comment
    “If our surveillance systems identify a potential performance issue, our personnel promptly evaluate the problem, including, when appropriate, conducting root cause investigations and internal testing to assess whether the product continues to meet specifications and defined performance criteria,” Medtronic told ICIJ in a statement. “In some cases, based on this evaluation, Medtronic may determine that a recall is necessary.” The company said that it communicates with healthcare providers and/or patients and provide recommendations to address such issues. Medtronic noted that these communications can include letters, emails, calls, press releases, physician notifications and social media postings, as well as informing the FDA and other regulators of the actions.
  • Source
    USFDA