Recall of Device Recall Medline Strider Midi 3 Scooter

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Medline Industries Inc.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    36171
  • Event Risk Class
    Class 3
  • Event Number
    Z-1522-06
  • Event Initiated Date
    2006-07-20
  • Event Date Posted
    2006-09-23
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2009-09-02
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    scooter - Product Code INI
  • Reason
    The p&g; solo 60a electronic control unit on the midi scooters may overheat.
  • Action
    Medline sent 'Voluntary Field Correction' letters dated 7/20/06 to the direct accounts on the same date, informing them that the Strider Midi 3 and Midi 4 scooters have an older version of the electronic control unit that may overheat, which Medline will replace with a newer version of the controller. Any questions were directed to 800-289-9793, press 1, then press 3 and ask to transfer to 4674 (8:00 AM to 5:00 PM CST).

Device

  • Model / Serial
    model MDS807500 - red and model MDS807500B - blue, scooter no., controller no.: EV10DB-*US00110*,13010613; EV10DB-*US00102*,13010615; EV10DB-*US00103*,53010802; EV10DB-*US00101*,53010803; EV10DB-*US00106*,53010806; EV10DB-*US00107*,53010807; EV10DB-*US00111*,53011072; EV10DB-*US00118*,53050871; EV10DB-*US00112*,53050872; EV10DB-*US00115*,53050873; EV10DB-*US00116*,53050874; EV10DB-*US00117*,53050875; EV10DB-*US00105*,73010706; EV10DB-*US00108*,73010710; EV10DB-*US00109*,73010711; EV10DB-*US00119*,73050684; EV10DB-*US00113*,73050685; EV10DB-*US00120*,73050686; EV10DB-*US00121*,73050688; EV10DB-*US00130*,73050805; EV10DB-*US00123*,73050806; EV10DB-*US00127*,73050807; EV10DB-*US00126*,73050808; EV10DB-*US00129*,73050809; EV10DB-*US00128*,73050810; EV10DB-*US00125*,73050812; EV10DB-*US00122*,73060000; EV10DB-*US00151*,13010484; EV10DB-*US00131*,23010651; EV10DB-*US00137*,33010608; EV10DB-*US00133*,53010863; EV10DB- *US00152*,53010928; EV10DB-*US00135*,53010929; EV10DB-*US00157*,53050630; EV10DB-*US00155*,53050631; EV10DB-*US00153*,53050632; EV10DB-*US00142*,53050865; EV10DB-*US00143*,53050866; EV10DB-*US00149*,53050867; EV10DB-*US00145*,53050868; EV10DB-*US00140*,53050869; EV10DB-*US00146*,53050870; EV10DB-*US00132*,73010716; EV10DB-*US00136*,73010765; EV10DB-*US00150*,73010824; EV10DB-*US00138*,73010825; EV10DB-*US00156*,73050441; EV10DB-*US00159*,73050443; EV10DB-*US00158*,73050444; EV10DB-*US00147*,73050680; EV10DB-*US00148*,73050681; EV10DB-*US00141*,73050682; EV10DB-*US00139*,73050683; EV10DB-*US00160*,73050442; EV10DB-*UK00150*,53010988; EV10DB-*UK00155*,53010676; EV10DB-*UK00158*,53010677; EV10DB-*UK00163*,73010884; EV10DB-*UK00173*,13010532; EV10DB-*UK00179*,53010668; EV10DB-*UK00181*,53010670; EV10DB-*UK00189*,13010530; EV10DB-*UK00192*,73010599; EV10DB-*UK00195*,53010791; EV10DB-*UK00200*,53010785; EV10DB-*UK00202*,53010789; EV10DB-*UK00203*,53010847; EV10DB-*UK00205*,73010750; EV10DB-*UK00206*,73010749; EV10DB-*UK00209*,53010787; EV10DB-*UK00220*,73010780; EV10DB-*UK00222*,53010776; EV10DB-*UK00227*,33010692; EV10DB-*UK00228*,73010691; EV10DB-*UK00230*,53010779; EV10DB-*UK00231*,73010689; EV10DB-*UK00232*,43010700; EV10DB-*UK00233*,53010788; EV10DB-*UK00236*,33010695; EV10DB-*UK00241*,13010525; EV10DB-*UK00248*,53010908; EV10DB-*UK00251*,33010620; EV10DB-*UK00252*,73010802; EV10DB-*UK00253*,53010907; EV10DB-*UK00256*,33010625; EV10DB-*UK00261*,23010532; EV10DB-*UK00265*,43010632; EV10DB-*UK00268*,73010628; EV10DB-*UK00270*,53010698; EV10DB-*UK00272*,53010688; EV10DB-*UK00276*,53010687; EV10DB-*UK00278*,23010545; EV10DB-*UK00279*,13010528; EV10DB-*UK00282*,23010533; EV10DB-*UK00286*,23010531; EV10DB-*UK00289*,73010616; EV10DB-*UK00292*,43010832; EV10DB-*UK00297*,33010838; EV10DB-*UK00298*,33010839; EV10DB-*UK00299*,43010831; EV10DB-*UK00303*,53010962; EV10DB-*UK00307*,73010859; EV10DB-*UK00309*,43010808; EV10DB-*UK00310*,43010807; EV10DB-*UK00311*,33010810; EV10DB-*UK00313*,43010811; EV10DB-*UK00315*,13010609; EV10DB-*UK00319*,43010809; EV10DB-*UK00326*,33010808; EV10DB-*UK00330*,73010858; EV10DB-*UK00341*,43010238; EV10DB-*UK00349*,53010825; EV10DB-*UK00359*,73010665; EV10DB-*UK00360*,53010738; EV10DB-*UK00361*,73010663; EV10DB-*UK00366*,73010871; EV10DB-*UK00367*,53010734; EV10DB-*UK00370*,73010776; EV10DB-*UK00376*,53010876; EV10DB-*UK00383*,43010232; EV10DB-*UK00385*,73010651; EV10DB-*UK00390*,43010602; EV10DB-*UK00398*,33010591; EV10DB-*UK00400*,33010592; EV10DB-*UK00403*,53010978; EV10DB-*UK00409*,53010722; EV10DB-*UK00411*,53010909; EV10DB-*UK00421*,23010568; EV10DB-*UK00425*,13010562; EV10DB-*UK00426*,13010563; EV10DB-*UK00429*,13010503; EV10DB-*UK00430*,73010588; EV10DB-*UK00431*,13010561; EV10DB-*UK00432*,53010914; EV10DB-*UK00433*,73010810; EV10DB-*UK00435*,53010910; EV10DB-*UK00446*,43010624; EV10DB-*UK00452*,73010754; EV10DB-*UK00461*,43010695; EV10DB-*UK00475*,53050530; EV10DB-*UK00477*,73050341; EV10DB-*UK00479*,73050344; EV10DB-*UK00487*,73050676; EV10DB-*UK00490*,43050817; EV10DB-*UK00492*,53050898; EV10DB-*UK00500*,73050732; EV10DB-*UK00501*,73050729; EV10DB-*UK00502*,73050733; EV10DB-*UK00505*,73050678; EV10DB-*UK00507*,53050915; EV10DB-*UK00510*,43050821; EV10DB-*UK00512*,53050914; EV10DB-*UK00521*,73050460; EV10DB-*UK00529*,53050677; EV10DB-*UK00560*,73050719; EV10DB-*UK00581*,53050826; EV10DB-*UK00582*,43050567; EV10DB-*UK00583*,43050806; EV10DB-*UK00595*,53050881; EV10DB-*UK00598*,73050693; EV10DB-*UK00600*,73050311; EV10DB-*UK00606*,73050696; EV10DB-*UK00613*,43050804; EV10DB-*UK00619*,73050387; EV10DB-*UK00621*,53050577; EV10DB-*UK00623*,53050576; EV10DB-*UK00627*,73050495; EV10DB-*UK00628*,53050687; EV10DB-*UK00629*,73050500; EV10DB-*UK00630*,73050497; EV10DB-*UK00636*,53050686; EV10DB-*UK00637*,73050462; EV10DB-*UK00639*,53050655; EV10DB-*UK00640*,53050618; EV10DB-*UK00643*,53050617; EV10DB-*UK00645*,73050508; EV10DB-*UK00648*,53050699; EV10DB-*UK00653*,53050654; EV10DB-*UK00655*,53050697; EV10DB-*UK00657*,53050696; EV10DB-*UK00659*,43050572; EV10DB-*UK00660*,53050653; EV10DB-*UK00661*,73050467; EV10DB-*UK00666*,73050468; EV10DB-*UK00668*,43050809; EV10DB-*UK00670*,53050906; EV10DB-*UK00671*,43050812; EV10DB-*UK00673*,73050456; EV10DB-*UK00675*,53050644; EV10DB-*UK00676*,73050455; EV10DB-*UK00678*,33050776; EV10DB-*UK00679*,53050890; EV10DB-*UK00680*,43050811; EV10DB-*UK00681*,43050813; EV10DB-*UK00683*,73050720; EV10DB-*UK00685*,53050908; EV10DB-*UK00687*,53050929; EV10DB-*UK00689*,53050579; EV10DB-*UK00691*,73050394; EV10DB-*UK00693*,53050580; EV10DB-*UK00695*,73050324; EV10DB-*UK00696*,73050392; EV10DB-*UK00698*,43050457; EV10DB-*UK00699*,53050924; EV10DB-*UK00700*,73050737; EV10DB-*UK00701*,53050926; EV10DB-*UK00702*,73050741; EV10DB-*UK00703*,53050925; EV10DB-*UK00706*,73050742; EV10DB-*UK00709*,53050646; EV10DB-*UK00712*,73050740; EV10DB-*UK00713*,53050907; EV10DB-*UK00717*,73050722; EV10DB-*UK00721*,73050381; EV10DB-*UK00723*,53050571; EV10DB-*UK00725*,43050837; EV10DB-*UK00726*,33050793; EV10DB-*UK00727*,53050922; EV10DB-*UK00730*,33050794; EV10DB-*UK00731*,33050431; EV10DB-*UK00733*,73050383; EV10DB-*UK00736*,53050611; EV10DB-*UK00737*,33050792; EV10DB-*UK00739*,73050382; EV10DB-*UK00740*,73050420; EV10DB-*UK00741*,73050323; EV10DB-*UK00742*,53050509; EV10DB-*UK00743*,73050422; EV10DB-*UK00746*,73050325; EV10DB-*UK00747*,53050508; EV10DB-*UK00748*,73050659; EV10DB-*UK00750*,53050511; EV10DB-*UK00753*,73050321; EV10DB-*UK00755*,53050612; EV10DB-*UK00756*,73050421; EV10DB-*UK00758*,73050419; EV10DB-*UK00760*,73050347; EV10DB-*UK00761*,73050346; EV10DB-*UK00762*,53050534; EV10DB-*UK00763*,73050658; EV10DB-*UK00766*,43050857; EV10DB-*UK00767*,43050432; EV10DB-*UK00768*,53050935; EV10DB-*UK00769*,43050434; EV10DB-*UK00770*,43050430; EV10DB-*UK00771*,73050652; EV10DB-*UK00773*,73050744; EV10DB-*UK00777*,53050844; EV10DB-*UK00778*,53050934; EV10DB-*UK00781*,73050384; EV10DB-*UK00782*,53050568; EV10DB-*UK00785*,53050569; EV10DB-*UK00786*,73050747; EV10DB-*UK00788*,53050536; EV10DB-*UK00789*,53050613; EV10DB-*UK00791*,73050348; EV10DB-*UK00792*,43050854; EV10DB-*UK00795*,53050837; EV10DB-*UK00797*,53050610; EV10DB-*UK00800*,73050423; EV10DB-*UK00805*,73050385; EV10DB-*UK00808*,43050433; EV10DB-*UK00809*,33050430; EV10DB-*UK00811*,53050931; EV10DB-*UK00812*,53050535; EV10DB-*UK00816*,43050856; EV10DB-*UK00817*,73050738; EV10DB-*UK00818*,53050538; EV10DB-*UK00819*,33050429;
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    Nationwide
  • Product Description
    Medline Strider Midi 3 Scooter; a three wheeled battery operated scooter; manufactured by KYMCO, Kaohsiung, Taiwan for Medline Industries, Inc., One Medline Place, Mudelein, IL 60060 USA; model MDS807500 - red and MDS807500B - blue