Recall of Device Recall Maquet Flowi

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Maquet Cardiovascular Us Sales, Llc.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    73137
  • Event Risk Class
    Class 1
  • Event Number
    Z-0804-2016
  • Event Initiated Date
    2015-12-10
  • Event Date Posted
    2016-02-22
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2016-07-06
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Gas-machine, anesthesia - Product Code BSZ
  • Reason
    Electrical issues in the control, monitoring or panel sub-systems, caused by the printed circuit boards (pcbs) that control the sub-systems. the problems manifested as multiple re-starts of the sub-systems and, in a few instances, disabling of the system has occurred. replacement pcbs are now available to correct the issues. only the model c20 is affected in the us.
  • Action
    Maquet sent "Urgent Medical Device Field Action Field Correction For Flow-i Anesthesia Systems" letters/Field Correction Response forms dated 12/10/2015 via UPS to affected customers. The notification informed customers of the issue, asked them to distribute the notification to appropriate staff; verify the serial number(s) and informed them they could continue to use the system if they were not experiencing electrical issues. Customers were informed that a Maquet service representative would be contacting them to arrange a time to replace their affected units. Customers were asked to complete and return the enclosed Field Correction Response Form to Maquet at FieldActions@maquet.com or by fax to 973.860.1675. If the customer had any related questions they were instructed to contact Maquet Technical Support at 1.888.627.8383 (option 3; option 1; option 1) Monday through Friday between 8:00 am and 5:00 pm EST.

Device

  • Model / Serial
    Model numbers: C20, C30 and C40 (Only Model # C20 for US)  Part numbers: 66 77 200 (US Only), 66 77 300 and 66 77 400  Serial numbers (2743, 2852, 2854, 2855, 2856, 2858
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    US Nationwide in the states of CA, NY, and OR
  • Product Description
    Flow-i Anesthesia System || Product Usage: || The indication for the Flow-i Anesthesia System is administering inhalation anesthesia while controlling the entire ventilation of patients with no ability to breath, as well as in supporting patients with a limited ability to breath. The system is intended for use on neonatal to adult patient populations. The system is intended for use in hospital environments, except MRI environment by healthcare professionals trained in inhalation anesthesia administration.
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Maquet Cardiovascular Us Sales, Llc, 45 Barbour Pond Drive, Wayne NJ 07470
  • Manufacturer Parent Company (2017)
  • Source
    USFDA