Recall of Device Recall Mantis Tap Sleeve

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Stryker Spine.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    57665
  • Event Risk Class
    Class 2
  • Event Number
    Z-1247-2011
  • Event Initiated Date
    2011-01-07
  • Event Date Posted
    2011-02-15
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2015-03-08
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Orthopedic manual surgical instrument - Product Code LXH
  • Reason
    Stryker spine has received reports that after sterilization but prior to surgery, cracks were noticed at the tantalum marker / pei tube interface. this can likely be attributed to differences in the rate of thermal expansion between the markers and the sleeve during the sterilization cycle.
  • Action
    Stryker Spine sent an Urgent Product Recall letter dated January 07, 2011, to Stryker Spine sales branches and agencies where affected product was distributed. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to examine their inventory and hospital locations to identify the product. Reconcile any Mantis Tap Sleeves with the catalog and lot numbers referenced in the letter. Complete the Customer Response Form with distribution sheet provided to assist with product reconcilation. Customers were asked to fax a copy of the Customer Response Form to Regulatory Compliance at 201-760-8370. Customers were to retrieve and return the affected product using the enclosed pre-paid mailing label to Regulatory Compliance, Stryker Spine 6 Pearl Court, Allendale, New Jersey 07401. For questions call 201-760-8180.

Device

  • Model / Serial
    All lots - Specialty Mantis Modular Tap Set - Catalog #IS2162MAN; Custom Tap Sleeve - Catalog #IS830MAN; Custom Tap Sleeve, Mantis - Catalog #IS2271MAN; and Custom Mantis Modular Tap Set - Catalog #IS2080MAN. All lots
  • Product Classification
  • Device Class
    1
  • Implanted device?
    No
  • Distribution
    Nationwide Distribution including MA, MN, and PA.
  • Product Description
    Mantis Tap Sleeve,Mfg by: Stryker Spine, 2 Pearl Court, Allendale, NJ 07401-1677. packaged in the following products: || Specialty Mantis Modular Tap Set - Catalog #IS2162MAN; Custom Tap Sleeve - Catalog #IS830MAN; Custom Tap Sleeve, Mantis - Catalog #IS2271MAN; and Custom Mantis Modular Tap Set - Catalog #IS2080MAN || The Mantis Tap Sleeves with Markers was designed to be used in a manner identical to the predicate Standard Mantis Tap Sleeve 48281315. It is to be used in conjunction with Standard Mantis Cannulated Modular Taps: 1) to protect soft tissue from the cutting edge of the tap 2) to indicate the depth at which the Tap has been inserted.
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Stryker Spine, 2 Pearl Ct, Allendale NJ 07401-1611
  • Manufacturer Parent Company (2017)
  • Source
    USFDA