Recall of Device Recall Malibu Polyaxial Screwdriver

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Integra LifeSciences Corp..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    67350
  • Event Risk Class
    Class 2
  • Event Number
    Z-0938-2014
  • Event Initiated Date
    2014-01-22
  • Event Date Posted
    2014-02-07
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2014-12-17
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Orthosis, spinal pedicle fixation, for degenerative disc disease - Product Code NKB
  • Reason
    Integra lifesciences has identified through an investigation of complaints that there may be the potential for the shaft of the malibu polyaxial drivers to break during use. none of the complaints reported a patient injury. however, some complaints did indicate the surgery was slightly prolonged while the additional driver provided in the kit was retrieved and used.
  • Action
    Integra LifeSciences initiated this recall by sending recall notification letter to US consignees, delivered by traceable courier service. Non US consignees notified of the recall by either traceable e-mail or facsimile. The letter dated, January 23, 2014, titled "URGENT: VOLUNTARY MEDICAL DEVICE RECALL Integra¿ Malibu Polyaxial Drivers Catalogue Numbers: 91-2109 & 91- 2110", informed customers of the recall and provided product description with codes, picture of the product, advised customers to: Inspect their inventory of Malibu Polyaxial Drivers. - If you have product(s) with a lot number listed on the following page, STOP USING THEM IMMEDIATLY. - If you do not have Drivers with a lot number listed on the following page, you can continue to use them. Customers were also instructed that regardless if they have Drivers with the affected lots, to complete the attached form and return it to Integra as indicated. Customers were instructed to keep a copy of the form for their records., and contact information. Customers with questions were instructed to contact Customers Service at 1-866-942-8698. For questions regarding this recall call 609-275-0500.

Device

  • Model / Serial
    Model Number(s):  91-2109 (8.0 Plus)  91-2110 (Standard).  Lot AL549784F: 91-2109 Malibu Polyaxial Screwdriver 8.0 Plus. Lot AL552404F: 91-2110 Malibu Polyaxial Screwdriver Standard. Lot AL546768F: 91-2110 Malibu Polyaxial Screwdriver Standard.  The devices are not distributed as sterile and do not have a expiration date.
  • Product Classification
  • Device Class
    2
  • Implanted device?
    Yes
  • Distribution
    Worldwide Distribution - USA including IA, FL, LA, IL, CA, TX, AZ, NC, MD, NM, VA, and Internationally to South Africa.
  • Product Description
    Malibu Polyaxial Screwdriver, 8.0 Plus, Model Number 91-2109; || Malibu Polyaxial Screwdriver, Standard, Model Number 91-2110. || 510(k): || 91-2109 K061342. || 91-2110 K051942. || The Malibu" System is intended to be used as a temporary or permanent posterior, non-cervical implant to correct spinal disorders and provide stabilization of the spine to permit the biological process of spinal fusions to occur. The System is comprised of single use pedicle screws of varying reduction height, rods, rod connectors, etc. and an accompanying set of reusable surgical instruments. Some of the surgical instruments are specific to the system and others are general surgical instruments. The polyaxial screwdriver is specific to the polyaxial screws in the system.
  • Manufacturer

Manufacturer