Recall of Device Recall MagNa Pure

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Roche Diagnostics Corp..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    35790
  • Event Risk Class
    Class 2
  • Event Number
    Z-1372-06
  • Event Initiated Date
    2006-07-17
  • Event Date Posted
    2006-08-05
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2007-12-04
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Test, Factor V Leiden Mutations, Genomic Dna Pcr - Product Code NPQ
  • Reason
    If the user opens the instrument lid during decontamination, the cycle will stop, but the timer will continue, resulting in the user believing the instrument has been decontaminated when it has not .
  • Action
    Consignees were notified via recall letter dated 7/17/06 to restart the sterilization cycle if they open the instrument for any reason during the sterilization cycle.

Device

  • Model / Serial
    All units.
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    Nationwide.
  • Product Description
    Roche MagNA Pure LC Instrument; Catalog nos. 12236931444, 12236931692 ( Refurbished) and 12236931690 (Return). An automated, general purpose instrument for isolation and purification of nucleic acids and assembly of PCR reactions.