Recall of Device Recall MagNA Pure

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Roche Diagnostics Corp..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    30483
  • Event Risk Class
    Class 2
  • Event Number
    Z-0297-05
  • Event Initiated Date
    2004-11-05
  • Event Date Posted
    2004-12-08
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2005-08-29
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Test, Factor V Leiden Mutations, Genomic Dna Pcr - Product Code NPQ
  • Reason
    Test results may be adversely affected by software shutting down the cooling blocks in systems with software version 3.09.
  • Action
    Consignees were notified via letter dated 11/5/04, which provides the user with a workaround.

Device

  • Model / Serial
    All systems using software version 3.09.
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    Nationwide.
  • Product Description
    MagNA Pure LC Instrument; catalog numbers 2236931 and 03670325001.
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Roche Diagnostics Corp., 9115 Hague Rd, Indianapolis IN 46256-1025
  • Source
    USFDA