Recall of Device Recall INTELIS II pacemaker

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Guidant Corporation.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    32681
  • Event Risk Class
    Class 1
  • Event Number
    Z-1038-05
  • Event Initiated Date
    2005-07-18
  • Event Date Posted
    2005-07-28
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2007-01-05
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Implantable Pulse Generator, Pacemaker (Non-Crt) - Product Code LWP
  • Reason
    A hermetic sealing component utilized in the device may experience a gradual degradation, resulting in a higher than normal moisture content within the pacemeaker case late in the device's service life.
  • Action
    INITIAL RECALL Urgent medical Device Safety Information & Corrective Action letter dated 07/18/05 was issued to physicians listing clinical Implications and patient management recommendations. EXPANDED RECALL Press was issued January 23, 2006. Urgent Medical Device Safety Information & Corrective Action letter, dated 01/21/06, was issued to physicians listing a summary of previous communication to physicians, advisory population update, and information relative to a second population of devices manufactured between October 19, 1998 and December 5, 2000. This letter gives the observed rate of occurrence, clinical implications, projected rate of occurrence and recommendations.

Device

  • Model / Serial
    Affected devices may be contained in the following serial number ranges.  Model 1349: Serial numbers 100003 thru 100076. Model 1384: Serial numbers 300000 thru 300073. Model 1385: INITIAL RECALL serial numbers 400005 thru 400019. EXPANDED RECALL serial numbers 400003, 400004. Model 1483: Serial numbers 600005 thru 600082. Model 1484: Serial numbers 700005 thru 700064. Model 1485: INITIAL RECALL serial numbers 800005 thru 800053. EXPANDED RECALL serial numbers 800000, 800001, 800002, 800003, 800004. Model 1499: Serial numbers 200001 thru 200073.
  • Product Classification
  • Device Class
    3
  • Implanted device?
    Yes
  • Distribution
    worldwide. Including the United States and countries such as: Argentina, Australia, Austria, Belgium, Bolivia, Brazil, Canada, Chile, Colombia, Costa Rica, Cyprus, Czechoslovakia, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Greece, Guadeloupe, Guyana, Hong Kong, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Lebanon, Luxembourg, Malaysia, Mexico, Netherlands, New Zealand, Norway, Panama, Peru, Poland, Polynesia, Portugal, Reunion, Russia, Saudi Arabia, Singapore,Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan ROC, Turkey, United Kingdom, Uruguay and Venezuela. EXPANDED RECALL additional countries include: China, Croatia, Korea. Lithuania, Madagascar, Maldives, Malta, Monaco, Nepal and Thailand.
  • Product Description
    INTELIS II (model nos. 1349, 1384, 1385, 1483, 1484, 1485 and 1499) pacemaker. Sterilized with gaseous ethylene oxide. Guidant . INTELIS II pacemakers are multiprogrammable pacemakers from Guidant. The family consists of both dual-chamber and single-chamber models, offering both adaptive-rate therapy and conventional pacing therapy. These pacemakers provide various levels of therapeutic and diagnostic functionality. INTELIS II adaptive-rate models have an accelerometer, which is a motion sensor that responds to patient activity.
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Guidant Corporation, 4100 Hamline Ave N, Saint Paul MN 55112-5700
  • Source
    USFDA