Recall of Device Recall Integra Small Rickham Style Reservoir

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Integra LifeSciences Corp..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    67025
  • Event Risk Class
    Class 2
  • Event Number
    Z-0621-2014
  • Event Initiated Date
    2013-11-13
  • Event Date Posted
    2014-01-03
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2015-02-13
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Shunt, central nervous system and components - Product Code JXG
  • Reason
    Integra became aware through an adverse complaint trend that rickham style reservoirs with bent or broken connectors that were shipped in sleeve-type package were found prior to use but after the sterile package was open.
  • Action
    Integra sent an Urgent Voluntary Medical Device Recall letter dated November 13, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to inspect their inventory for the affected product and if located stop using immediately. Customers were asked to complete the attached form and return it to Integra as indicated and keep a copy of the form for their records. Once the form was received by Integra they will send an RMA number, directions to return the product, and input an order to replace the quantity they indicated on the form. Customers with questions were instructed to call 1-855-532-1723. For questions regarding this recall call 855-532-1723.

Device

  • Model / Serial
    1091271, 1094671, 1104165, 1104631, 1104813, 1110404, 1120638, 1120816, 1122316, 1123940, 11311659, 1132989, 1133427
  • Product Classification
  • Device Class
    2
  • Implanted device?
    Yes
  • Distribution
    Worldwide Distribution - USA including FL, NE, WI, WA, CT, ND, AL, NY, NH, OR, OK, MN, and TX. Internationally to Belgium, Czech Republic, Germany, France, Great Britian, Italy, Poland, and Portugal.
  • Product Description
    Integra Small Rickham -Style Reservoir || Rx Only Sterile for Single Use Only || Integra NeuroSciences 311 Enterprise Drive Plainsboro, NJ 08536 USA Made in the USA || NL 850-1121 || Provides access to the lateral cerebral ventricles via hypodermic puncture. It is useful in obtaining CSF (cerebral spinal fluid_ samples for cytological and chemical studies, for monitoring ventricular fluid pressure and for ventricular drainage. The reservoir provides easy access to the lateral ventricles and to cystic tumors for the injections of chemotherapeutic agents and/or radio-isotopes. The rickham reservoir may be utilized in hydrocephalic patients as a component in systems designed to shunt CSF from the lateral ventricles into either the right atrium of the heart of the peritoneum.
  • Manufacturer

Manufacturer