Recall of Device Recall Integra

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Integra LifeSciences Corp..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    68519
  • Event Risk Class
    Class 2
  • Event Number
    Z-1976-2014
  • Event Initiated Date
    2014-06-11
  • Event Date Posted
    2014-07-02
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2014-12-19
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Headlamp, operating, battery-operated - Product Code HPP
  • Reason
    Led battery chargers may prematurely fail and will not charge the led battery as intended.
  • Action
    Integra initiated a voluntary recall on June 11, 2014 to consignees that have been shipped affected product lots of Integra¿ LED Battery Chargers, Catalogue No. 90523. The written notifications were delivered to US consignees by traceable courier service and non-US consignees by traceable email or facsimile. The consignee notification: Advises them of the nature of the issue and the potential for an adverse patient consequence. Advises them to review their inventory and how to determine if they have affected battery chargers. Advises them to immediately stop the use or distribution of any affected product lot numbers they may have. Requests they identify and report to Integra if they do or do not have any unexpired affected product lots. Advises them replacements battery chargers will be shipped for any affected products they affirm are in their possession and how to return their affected product. Questions regarding these instructions, please contact Integra Customer Service at 1-800-431-1123 or your Integra sales

Device

  • Model / Serial
    Lot Numbers:  IE123610, IE123710, IE123810, IE123910, IE124010, IE124110, IE124210, IE124310, IE124410, IE124510, IE124610, IE124710, IE124810, IE124910, IE125010, IE125110, IE125210, IE130110, IE130210, IE130310, IE130410, IE130510, IE130610, IE130710, IE130810, IE130910, IE131010
  • Product Classification
  • Device Class
    1
  • Implanted device?
    No
  • Distribution
    Worldwide Distribution - US Nationwide and the countries of Canada, Australia, Arab Emirates, Denmark, Germany, Great Britain, India, Ireland, Israel, Japan, Kuwait, Lebanon, Malta., Netherlands, Panama, Poland, Saudi Arabia, Singapore, South Africa, Spain, Sweden, and Tunisia.
  • Product Description
    Integra¿ LED Battery Charger, Single Bay. || Product Number: 90523 || The LED Headlight System Battery Charger (90523) is an accessory to the LED Headlight System: || 90520US - LED Headlight w/Battery & AC/DC Power Supply - US || 90520EU - LED Headlight w/Battery & AC/DC Power Supply - EU || 90520UK - LED Headlight w/Battery & AC/DC Power Supply - UK || 90520AU - LED Headlight w/Battery & AC/DC Power Supply - AU || Product Usage: The Integra LED Headlight System Is designed to provide illumination to aid visualization during minor surgical, diagnostic, or therapeutic procedures