Recall of Device Recall Gemstar Infusion System

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Hospira Inc..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    64832
  • Event Risk Class
    Class 1
  • Event Number
    Z-0349-2014
  • Event Initiated Date
    2013-03-15
  • Event Date Posted
    2013-11-21
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2017-03-07
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
    ICU Medical, Inc. acquired from Pfizer in early 2017, Hospira Infusion Systems, the portion of Hospira dedicated to develop infusion pumps.
  • Extra notes in the data
    Pump, infusion - Product Code FRN
  • Reason
    The proximal and distal pressure sensor calibration can drift resulting in the pump failing the proximal or distal occlusion operational test, as described in the gemstar technical service manual, or reporting one of the following errors during device setup or infusion: 1) cassette check - d; 2) cassette check - p; 3) proximal occlusion; 4) distal occlusion; 5) pressure calibration error; 6) bad pressure sensor event; 7) bad pressure state; 8) distal pressure is out of range; 9) proximal sensor is out of range. a pump with this issue may, instead of reporting an error, not detected occlusions or issue false occlusion alarms, which will stop the infusion and invoke visual and audible warnings to the user.
  • Action
    Hospira sent an Urgent Device Field Correction letter dated March 15, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. The letter asks that customers immediately perform proximal and distal occlusion tests as defined i nthe GemStar Technical Service Manual. If the device fails either of the tests, remove it from clinical service and contact Hospira at 1-800-441-4100 (Monday - Friday, 8:00 AM - 5:00 PM) to report the issue and arrange for the return of the device for recalibration. The letter further reminds customers that the proximal and distal occlusion tests should be performed on an annual basis as a part of the yearly GemStar maintenance schedule. Customers were also requested to complete and return the attached reply form and return it to Hospira by fax to 1-888-714-5077 or by e-mail to Hospira7906@stericycle.com. Customers were directed to notify their consignees if they further distributed the product and confirm they have done so by return of the reply form to Stericycle. For questions regarding this recall call 800-441-4100.

Device

  • Model / Serial
    *** 1) List Number 13000, Serial Numbers:  98121336 ,  98121173 ,  98120835 ,  98120730 ,  98120768 ,  98120782 ,  98120788 ,  98120822 ,  98120824 ,  98120839 ,  98120856 ,  98120880 ,  98120885 ,  98120895 ,  98120915 ,  98120749 ,  98120769 ,  98120774 ,  98120784 ,  98120859 ,  98120884 ,  98121019 ,  98121066 ,  98121079 ,  98121083 ,  98121135 ,  98121153 ,  98121163 ,  98121197 ,  98120522 ,  98120653 ,  13431222 ,  98120635 ,  98120405 ,  98120508 ,  98120634 ,  98121147 ,  98120543 ,  12470380 ,  98120512 ,  98120815 ,  98120681 ,  97971490 ,  97970893 ,  97970240 ,  98070264 ,  98070295 ,  99170068 ,  99170032 ,  98070345 ,  98070476 ,  98070905 ,  98070995 ,  97972107 ,  97972609 ,  97971538 ,  98021653 ,  98021821 ,  99400401 ,  99400435 ,  97970456 ,  13431303 ,  98021816 ,  97972825 ,  98021121 ,  97971767 ,  99107521 ,  98020395 ,  14180727 ,  98120459 ,  98120495 ,  98120777 ,  98020964 ,  98070286 ,  98121157 ,  97972571 ,  97970541 ,  97971254 ,  98120535 ,  98120640 ,  98120840 ,  97970194 ,  98021636 ,  98020575 ,  97972639 ,  99222573 ,  98021550 ,  98021603 ,  98021682 ,  98370030 ,  98020523 ,  98020446 ,  98020441 ,  99107526 ,  97971754 ,  98020929 ,  98020932 ,  98022223 ,  13434785 ,  98020072 ,  98020317 ,  98021476 ,  98021478 ,  98021732 ,  98021891 ,  98021912 ,  98021252 ,  99170080 ,  98020623 ,  98020018 ,  98020293 ,  98020574 ,  98021031 ,  98021423 ,  98021449 ,  98020013 ,  97972018 ,  98021177 ,  98020264 ,  99221974 ,  98020817 ,  98021687 ,  98070125 ,  98021304 ,  98021546 ,  98022039 ,  98020281 ,  97972802 ,  97972086 ,  97971122 ,  12470492 ,  98121130 ,  97972659 ,  97972489 ,  98120869 ,  13433080 ,  97972047 ,  97972824 ,  98020549 ,  98021584 ,  97970313 ,  97970073 ,  98021364 ,  98020312 ,  98021961 ,  98020127 ,  98020606 ,  98021194 ,  99222658 ,  98020490 ,  98021141 ,  98021276 ,  98021171 ,  98020460 ,  98021969 ,  99107535 ,  99221815 ,  98021977 ,  98021120 ,  98021412 ,  99222620 ,  14180122 ,  99400456 ,  98071220 ,  98070607 ,  98070371 ,  98070781 ,  98070815 ,  98070696 ,  98070812 ,  97972981 ,  97970766 ,  98020074 ,  98020902 ,  14180174 ,  99219028 ,  98070190 ,  13433667 ,  97970613 ,  98021101 ,  98021099 ,  97970666 ,  97970308 ,  97970071 ,  98021367 ,  98021508 ,  98021308 ,  98021309 ,  98021208 ,  97972842 ,  97972841 ,  99222510 ,  98020727 ,  98020235 ,  98021264 ,  98021708 ,  98021794 ,  97971750 ,  99170302 ,  98070212 ,  98070006 ,  97972808 ,  98021554 ,  98021900 ,  99400482 ,  98020054 ,  98021195 ,  98021496 ,  98120534 ,  98020866 ,  98021179 ,  98070850 ,  98070448 ,  98070794 ,  97972040 ,  97971807 ,  98350110 ,  97971723 ,  98350105 ,  98350089 ,  98021587 ,  98021999 ,  98021207 ,  98021968 ,  98020855 ,  98021551 ,  99400235 ,  98070699 ,  97972950 ,  99222016 ,  98070351 ,  97971659 ,  98020901 ,  98020949 ,  98020474 ,  98021592 ,  98021932 ,  97971790 ,  97971766 ,  97972867 ,  97972283 ,  13432084 ,  99170140 ,  98070868 ,  98070028 ,  98070717 ,  98070296 ,  98070688 ,  99220393 ,  98121080 ,  12474733 ,  99222601 ,  97971952 ,  99222578 ,  98121160 ,  97970769 ,  97970839 ,  99222563 ,  99222322 ,  97970124 ,  97972801 ,  13430060 ,  14180300 ,  99221044 ,  97972664 ,  97970952 ,  97970873 ,  13430773 ,  97972475 ,  97972824 ,  13431359 ,  12471354 ,  99221849 ,  99221798 ,  98120493 ,  98120726 ,  98120696 ,  99222537 ,  99220861 ,  13432293 ,  13434708 ,  97970948 ,  14180791 ,  13433680 ,  13431566 ,  13431924 ,  99222636 ,  99222656 ,  12474581 ,  13433821 ,  12474354 ,  12471791 ,  13433340 ,  13432931 ,  13431284 ,  13440709 ,  13433382 ,  99221243 ,  13433983 ,  97971063 ,  97971874 ,  99222356 ,  99221852 ,  14180397 ,  13430576 ,  12474729 ,  13432281 ,  12473137 ,  13432245 ,  12470955 ,  13431610 ,  13436245 ,  99222612 ,  13434287 ,  12470279 ,  13433445 ,  13431619 ,  12473303 ,  12470403 ,  13431307 ,  98120382 ,  99222532 ,  13435380 ,  98020906 ,  12474665 ,  13430152 ,  99221958 ,  14180515 ,  99222578 ,  12473180 ,  13434080 ,  13439991 ,  97970460 ,  97970303 ,  13430198 ,  13430743 ,  13435220 ,  13438795 ,  13430039 ,  13432326 ,  13431725 ,  13431485 ,  13431224 ,  13433164 ,  99222614 ,  99222331 ,  13431989 ,  13432293 ,  13434726 ,  13431827 ,  13430627 ,  13431950 ,  13430182 ,  13434556 ,  12473423 ,  13436180 ,  13432449 ,  99222375 ,  13431069 ,  13430881 ,  97972801 ,  99222384 ,  13431420 ,  13432862 ,  13432588 ,  12474249 ,  13439388 ,  98021812 ,  99222471 ,  99222379 ,  13437386 ,  99222608 ,  99222423 ,  99221856 ,  13434243 ,  13433174 ,  13430135 ,  13433314 ,  12474465 ,  12473145 ,  13433684 ,  12471384 ,  14180622 ,  13440159 ,  99222418 ,  13431720 ,  12474332 ,  12471844 ,  13431992 ,  13439898 ,  13433072 ,  99222523 ,  13434042 ,  13432775 ,  99222578 ,  99220435 ,  13430344 ,  14180036 ,  13439887 ,  97970496 ,  97970510 ,  13431002 ,  13432432 ,  14180638 ,  13438445 ,  12471365 ,  98120641 ,  13432828 ,  99222539 ,  99220428 ,  12470727 ,  13431121 ,  13434646 ,  12472866 ,  13434339 ,  13433351 ,  13433021 ,  13433746 ,  13433597 ,  14180301 ,  13431965 ,  13433449 ,  99222633 ,  12470294 ,  99222348 ,  98021598 ,  12472916 ,  12472926 ,  12472854 ,  13431168 ,  98020252 ,  13432809 ,  12474227 ,  97970117 ,  13430007 ,  99222505 ,  13440463 ,  12471872 ,  13432893 ,  13430543 ,  14180874 ,  99222047 ,  14181058 ,  12470501 ,  12471513 ,  12474272 ,  13431727 ,  12474057 ,  99220921 ,  12471545 ,  98020372 ,  12470387 ,  13433411 ,  13433763 ,  12472231 ,  13434621 ,  13434369 ,  12471484 ,  13431471 ,  97972256 ,  12471430 ,  13431425 ,  14180319 ,  13431269 ,  13430443 ,  13430263 ,  13433150 ,  13430109 ,  99401327 ,  97971904 ,  13439881 ,  97970664 ,  97971885 ,  97970638 ,  13431396 ,  13430382 ,  12471101 ,  13430160 ,  13435527 ,  13436516 ,  97972950 ,  97972981 ,  97970743 ,  97970687 ,  97970844 ,  97970447 ,  97970796 ,  97972983 ,  99220427 ,  97970914 ,  13437108 ,  12472089 ,  13431393 ,  13431918 ,  99221429 ,  13440255 ,  14180725 ,  99220597 ,  99220574 ,  97970397 ,  14180479 ,  13440421 ,  99220925 ,  98022207 ,  13431186 ,  13431809 ,  13432210 ,  97971939 ,  97971936 ,  12474716 ,  13430489 ,  99222353 ,  99222678 ,  98120538 ,  13439829 ,  13431923 ,  12473802 ,  14180289 ,  13430070 ,  97970291 ,  99222501 ,  99222273 ,  12474114 ,  13432728 ,  13440049 ,  13431059 ,  13431830 ,  13432534 ,  13432513 ,  13433799 ,  13430665 ,  13432915 ,  13431361 ,  13431739 ,  98120478 ,  98120395 ,  13438672 ,  98022167 ,  13438727 ,  13432289 ,  14180895 ,  13432091 ,  13431252 ,  13432255 ,  13433270 ,  13430279 ,  13430538 ,  13430844 ,  13431479 ,  14180264 ,  14180898 ,  13434337 ,  13440017 ,  12474538 ,  97971882 ,  13431358 ,  13433089 ,  13431063 ,  99221063 ,  99222367 ,  12470447 ,  14180278 ,  12471371 ,  12470672 ,  13430539 ,  99221119 ,  13431568 ,  13432824 ,  13431741 ,  13431617 ,  13432250 ,  14181041 ,  13437404 ,  13437428 ,  12473076 ,  12473089 ,  13430016 ,  13430020 ,  13431344 ,  13430258 ,  12471929 ,  13431184 ,  13431443 ,  13431774 ,  13433246 ,  13431631 ,  13431697 ,  13431116 ,  13431119 ,  13430192 ,  13430604 ,  13431519 ,  13433865 ,  12471630 ,  13440771 ,  13436018 ,  13439275 ,  13439683 ,  13435913 ,  13431418 ,  13430093 ,  13437435 ,  13434268 ,  13432887 ,  13437573 ,  99222578 ,  13431642 ,  97971605 ,  13436546 ,  12471697 ,  12471390 ,  13431701 ,  13434523 ,  13432514 ,  13431728 ,  13432122 ,  13433114 ,  13430480 ,  13432090 ,  13430955 ,  14180208 ,  13439760 ,  99222337 ,  13430273 ,  97971904 ,  12473444 ,  14180765 ,  13430148 ,  12472880 ,  13434655 ,  97972588 ,  13432324 ,  13440380 ,  13433045 ,  13431994 ,  13438524 ,  99220352 ,  99220854 ,  14180184 ,  13431384 ,  13431127 ,  13438450 ,  99220882 ,  13431162 ,  13433002 ,  13432723 ,  13431908 ,  99222759 ,  13432874 ,  13435219 ,  13430262 ,  12472143 ,  98120435 ,  97972490 ,  97970390 ,  13438345 ,  97970470 ,  13434431 ,  13434442 ,  99220353 ,  13433454 ,  13431991 ,  12471555 ,  14180342 ,  97970621 ,  13431543 ,  12471517 ,  13438440 ,  12472157 ,  12473611 ,  13432276 ,  13440636 ,  99221054 ,  99222361 ,  99221310 ,  13430256 ,  13431360 ,  13433041 ,  12471186 ,  13430195 ,  13430815 ,  13436356 ,  13430201 ,  12471405 ,  12474235 ,  97970854 ,  99222615 ,  13430562 ,  99222479 ,  12473981 ,  12470581 ,  13432710 ,  12471922 ,  13438017 ,  97970362 ,  13431921 ,  97970326 ,  97971803 ,  13433335 ,  99222377 ,  13435064 ,  13435427 ,  99401385 ,  99401617 ,  99222126 ,  13432525 ,  13436949 ,  13430681 ,  13438671 ,  98020979 ,  12473779 ,  12471513 ,  13438805 ,  13438767 ,  12471749 ,  13433432 ,  99220613 ,  99222507 ,  14180029 ,  12471778 ,  13437662 ,  13431798 ,  13430736 ,  12471804 ,  12471355 ,  13430130 ,  13433176 ,  13433596 ,  13439582 ,  97970578 ,  13440172 ,  13430868 ,  13430816 ,  13432788 ,  13431480 ,  12472460 ,  12474554 ,  13434676 ,  13435905 ,  99221060 ,  13433299 ,  13434817 ,  99222089 ,  97970974 ,  99170292 ,  13440406 ,  13434759 ,  12471030 ,  14181047 ,  12471742 ,  13432918 ,  12474110 ,  13433825 ,  13432794 ,  12470871 ,  97970116 ,  13439657 ,  98021726 ,  13440224 ,  12474186 ,  99221786 ,  14180621 ,  13431541 ,  13435455 ,  13430207 ,  14180440 ,  13430792 ,  13431829 ,  14180178 ,  12470619 ,  12474746 ,  13439433 ,  13434210 ,  13436451 ,  99222280 ,  97970389 ,  12470644 ,  13431360 ,  12474613 ,  13430599 ,  12473175 ,  12473377 ,  13434603 ,  13434020 ,  98020101 ,  98020417 ,  98022207 ,  98021803 ,  12473807 ,  14180351 ,  13434095 ,  12473182 ,  12473203 ,  12474695 ,  99222333 ,  13431397 ,  13430298 ,  97970559 ,  13431474 ,  12474341 ,  12470615 ,  13431951 ,  13431755 ,  13430853 ,  12474717 ,  13438602 ,  99220388 ,  14180450 ,  13437416 ,  99401584 ,  99401641 ,  12470520 ,  13430030 ,  12470858 ,  13432939 ,  13435085 ,  12470465 ,  13430945 ,  98120683 ,  98120690 ,  98120692 ,  98120851 ,  98121198 ,  98121177 ,  98120765 ,  98121212 ,  98121228 ,  13432762 ,  12471473 ,  13431281 ,  13432411 ,  13438839 ,  99220844 ,  13438699 ,  13438677 ,  12471452 ,  13439841 ,  99222720 ,  99401657 ,  13440154 ,  99222498 ,  13431663 ,  99222476 ,  13434655 ,  13430325 ,  14180328 ,  13433568 ,  13431193 ,  13433097 ,  12471445 ,  14180997 ,  12473546 ,  13438206 ,  97970487 ,  97970573 ,  97970579 ,  13438631 ,  13439025 ,  99222765 ,  13432838 ,  14180860 ,  13430778 ,  12473176 ,  12470399 ,  13430675 ,  14180159 ,  97972797 ,  12470573 ,  13431775 ,  12472960 ,  14180734 ,  13431744 ,  13430373 ,  13439497 ,  13434520 ,  13432237 ,  14180310 ,  13440659 ,  13430131 ,  14180053 ,  13435327 ,  13436405 ,  13436423 ,  13432332 ,  13437020 ,  13438629 ,  14180977 ,  14180651 ,  13432397 ,  14180131 ,  13433506 ,  13439733 ,  13435275 ,  13436547 ,  12470736 ,  99222578 ,  99220589 ,  97970477 ,  13439822 ,  13433190 ,  13436022 ,  13431029 ,  12471666 ,  99400011 ,  13433257 ,  13430319 ,  13438556 ,  13435539 ,  13432632 ,  13432221 ,  13432481 ,  13433169 ,  14180354 ,  13430994 ,  13433330 ,  13434671 ,  99222622 ,  99222389 ,  99222501 ,  13433194 ,  13430395 ,  13431083 ,  12470257 ,  12473173 ,  13435152 ,  14180094 ,  13431537 ,  13434473 ,  13433438 ,  14180658 ,  13432406 ,  99222730 ,  99222401 ,  97970395 ,  97970690 ,  97970755 ,  97970788 ,  13439062 ,  13431573 ,  13432930 ,  12472023 ,  13435316 ,  12474702 ,  14180269 ,  12473627 ,  13434245 ,  13434301 ,  99221489 ,  99221071 ,  99222568 ,  99222739 ,  99222639 ,  99220646 ,  13433565 ,  13432050 ,  13437722 ,  99220930 ,  13436046 ,  13440126 ,  13440212 ,  13430876 ,  12472500 ,  12471057 ,  14180749 ,  13432776 ,  14180311 ,  13435358 ,  13436443 ,  99222590 ,  99220908 ,  12471146 ,  12472109 ,  13432212 ,  13438476 ,  13431588 ,  97972950 ,  99220584 ,  12473120 ,  12470872 ,  13432662 ,  97970040 ,  97971938 ,  12473967 ,  12471672 ,  13434500 ,  13432552 ,  97970992 ,  12470317 ,  13435270 ,  13435585 ,  13436034 ,  13436181 ,  99220622 ,  13432755 ,  13430654 ,  13439030 ,  12471281 ,  12471354 ,  99220462 ,  13432014 ,  13440040 ,  13437132 ,  13431449 ,  99221429 ,  13435974 ,  13435238 ,  13431552 ,  13431819 ,  99222713 ,  99222439 ,  99221239 ,  13433235 ,  13436226 ,  13438711 ,  99222540 ,  13431822 ,  12470375 ,  13434249 ,  99744476 ,  13433143 ,  13436216 ,  97971597 ,  13430996 ,  13433258 ,  13435109 ,  14181090 ,  13432693 ,  99222738 ,  13430471 ,  13431404 ,  13432842 ,  14180065 ,  13437483 ,  13430399 ,  13433839 ,  13431782 ,  13431517 ,  99222399 ,  13431635 ,  13431639 ,  13436277 ,  12470506 ,  99220439 ,  99222692 ,  13440109 ,  99222406 ,  13434550 ,  99220848 ,  13438812 ,  99222642 ,  13430125 ,  12471135 ,  13435415 ,  13432376 ,  13437425 ,  13433283 ,  13437385 ,  13439089 ,  99222695 ,  13433444 ,  13433451 ,  13433558 ,  13440615 ,  13433410 ,  99220900 ,  13432969 ,  13440202 ,  12470973 ,  13432575 ,  14180410 ,  13430928 ,  13434649 ,  13439075 ,  98020117 ,  13439057 ,  13430497 ,  12471859 ,  13433175 ,  13439900 ,  13439851 ,  97972983 ,  99222046 ,  99220866 ,  12471600 ,  12470351 ,  13431196 ,  14181069 ,  12474734 ,  12474429 ,  12470550 ,  14180220 ,  12474317 ,  13433296 ,  13433289 ,  97970902 ,  97971667 ,  13431194 ,  12472889 ,  13432345 ,  13432654 ,  13431201 ,  14180766 ,  13436096 ,  13431619 ,  99222654 ,  14180258 ,  12474714 ,  98022264 ,  12473896 ,  97970590 ,  97970663 ,  13434686 ,  13430555 ,  13430460 ,  14180887 ,  13432957 ,  13432260 ,  13430831 ,  13431309 ,  12474435 ,  99222365 ,  13433437 ,  12471337 ,  13435946 ,  97970443 ,  12473471 ,  99220412 ,  12473751 ,  99222014 ,  99221257 ,  13432925 ,  13430845 ,  13435079 ,  12473009 ,  99220638 ,  13433287 ,  12474266 ,  13432501 ,  99222395 ,  99220488 ,  12473267 ,  13434524 ,  13432256 ,  13436262 ,  13438909 ,  99744321 ,  99222718 ,  13432028 ,  99220436 ,  12472468 ,  12470350 ,  12470676 ,  12474785 ,  99220928 ,  99221172 ,  99222757 ,  13435998 ,  13435961 ,  99222262 ,  13436223 ,  13435375 ,  13435440 ,  13439698 ,  99222751 ,  13437487 ,  13433123 ,  13438356 ,  13438527 ,  99220923 ,  99220406 ,  12471409 ,  13434656 ,  13433149 ,  12472760 ,  98022207 ,  99222181 ,  99222491 ,  13435130 ,  13435408 ,  13435049 ,  13430041 ,  97970415 ,  12471644 ,  99220602 ,  13434024 ,  99220498 ,  14180998 ,  97970286 ,  98020428 ,  13440139 ,  99222171 ,  13433179 ,  12473483 ,  13439694 ,  13430363 ,  13440601 ,  13437365 ,  99222346 ,  14180214 ,  14180132 ,  13432462 ,  13434530 ,  13436209 ,  13439922 ,  13430772 ,  13439496 ,  13433627 ,  13436025 ,  14181077 ,  14180605 ,  14180423 ,  12471050 ,  12472876 ,  13439896 ,  13436529 ,  13434661 ,  12474025 ,  99221643 ,  13434475 ,  13434240 ,  13434638 ,  97970339 ,  13436365 ,  99220868 ,  14181024 ,  14180907 ,  14181050 ,  99221170 ,  97970125 ,  97970364 ,  13432391 ,  97972791 ,  99222623 ,  12473460 ,  13433816 ,  99220489 ,  99222371 ,  99221840 ,  97970614 ,  13432178 ,  13431099 ,  13431650 ,  13437180 ,  13435997 ,  99222710 ,  98120671 ,  99220621 ,  13433559 ,  13431102 ,  12474000 ,  13434155 ,  12474633 ,  13436030 ,  99401667 ,  13436255 ,  13436306 ,  97970335 ,  99401227 ,  97970623 ,  99222582 ,  13438472 ,  97970526 ,  13430284 ,  13432770 ,  13434041 ,  13434689 ,  13440290 ,  13432409 ,  13438509 ,  12470911 ,  13430498 ,  13436714 ,  13430145 ,  99220444 ,  12473866 ,  12471904 ,  98022262 ,  12473023 ,  97971868 ,  13432890 ,  12472018 ,  99222470 ,  13431368 ,  13433054 ,  13432576 ,  13435197 ,  13431958 ,  99220555 ,  97970003 ,  13433191 ,  12473171 ,  12472888 ,  12473859 ,  13435437 ,  97972952 ,  13437794 ,  13430474 ,  13433091 ,  13431619 ,  13432765 ,  13433093 ,  12471724 ,  13438598 ,  99222749 ,  13438601 ,  13431170 ,  13436161 ,  13434769 ,  98020399 ,  13430236 ,  13433251 ,  12472098 ,  12473115 ,  13433171 ,  12474053 ,  13432296 ,  99222319 ,  99222660 ,  99222275 ,  97970480 ,  97970444 ,  99222307 ,  13436019 ,  99220866 ,  13436535 ,  13436542 ,  14180747 ,  12471437 ,  12472997 ,  12472978 ,  13434696 ,  13433290 ,  13435860 ,  13432751 ,  14180707 ,  13435351 ,  13432563 ,  13435183 ,  13431674 ,  13431431 ,  98121018 ,  98120863 ,  98120830 ,  13437496 ,  12470351 ,  98121125 ,  13439669 ,  99220667 ,  99222711 ,  99222252 ,  13436971 ,  13438783 ,  13438454 ,  99220159 ,  99222341 ,  12473001 ,  13438826 ,  13431904 ,  14180780 ,  13438821 ,  12471756 ,  99221431 ,  13434637 ,  13432871 ,  13437737 ,  13433422 ,  13434571 ,  99220445 ,  97970461 ,  97970358 ,  99222715 ,  99400830 ,  13440208 ,  99220559 ,  13436649 ,  14180129 ,  99220897 ,  13430303 ,  13438619 ,  13431633 ,  13437663 ,  99222416 ,  13436620 ,  99221695 ,  13431993 ,  13439772 ,  13434577 ,  99222165 ,  99222352 ,  99222267 ,  13439863 ,  99222709 ,  14180223 ,  13437685 ,  13431704 ,  13430036 ,  12472998 ,  97971965 ,  99744235 ,  99221938 ,  13433440 ,  99221639 ,  13435893 ,  13436013 ,  99221928 ,  13432814 ,  13432943 ,  13432761 ,  13435337 ,  13435579 ,  99222754 ,  99222328 ,  97971652 ,  98022207 ,  97970987 ,  97971591 ,  97970545 ,  13435319 ,  13435317 ,  13434208 ,  13434241 ,  98022275 ,  99222680 ,  13438441 ,  13437148 ,  13437270 ,  99401593 ,  13433172 ,  98411600 ,  13439470 ,  12472287 ,  13432988 ,  99400030 ,  99222764 ,  13440086 ,  97972646 ,  97972688 ,  97970616 ,  99222116 ,  13432678 ,  13431388 ,  13439887 ,  13440184 ,  13440186 ,  13433100 ,  99220502 ,  14180073 ,  13440007 ,  13440068 ,  13440550 ,  12470578 ,  13432652 ,  14180076 ,  13438645 ,  99221585 ,  99222360 ,  13435359 ,  12474618 ,  99222489 ,  98022164 ,  97971911 ,  13438585 ,  13439727 ,  14180645 ,  14180742 ,  14180884 ,  14180938 ,  14180771 ,  14180833 ,  13431332 ,  13432895 ,  13431988 ,  13430606 ,  13431064 ,  12472993 ,  99222681 ,  99220464 ,  13431262 ,  13431078 ,  13431583 ,  13435485 ,  13434175 ,  13431579 ,  13432934 ,  13432483 ,  97970079 ,  13432559 ,  13432802 ,  13433647 ,  13432670 ,  13434344 ,  12474015 ,  99221291 ,  99222648 ,  99222537 ,  13433870 ,  99220379 ,  12472781 ,  99222562 ,  13432136 ,  99221734 ,  99220795 ,  14180808 ,  13439402 ,  13434118 ,  13430514 ,  13432972 ,  13437263 ,  99221265 ,  99220458 ,  99222598 ,  99222292 ,  13436385 ,  97970454 ,  13439684 ,  13432322 ,  13432615 ,  12473710 ,  97970585 ,  97971109 ,  97972479 ,  99220243 ,  13438475 ,  99222565 ,  13432098 ,  13434368 ,  97971897 ,  13436275 ,  12474321 ,  13433707 ,  13433815 ,  13436588 ,  13433033 ,  12474302 ,  13434615 ,  14180209 ,  13436544 ,  14180232 ,  99222354 ,  97970412 ,  13437320 ,  99222646 ,  99222357 ,  99222627 ,  99221289 ,  14181052 ,  13435311 ,  13439371 ,  13438573 ,  12474163 ,  13433241 ,  99222541 ,  99220595 ,  99220577 ,  99222512 ,  13439379 ,  13433507 ,  14180909 ,  12470576 ,  13431100 ,  13439534 ,  13433834 ,  99220581 ,  12474042 ,  13431129 ,  13439658 ,  12472777 ,  13432029 ,  13431654 ,  13438885 ,  14180794 ,  97971804 ,  97972720 ,  99220929 ,  13431772 ,  99221184 ,  99222657 ,  99222266 ,  97970398 ,  13434059 ,  13432215 ,  13432156 ,  12472802 ,  99221006 ,  12473544 ,  13439724 ,  13438589 ,  13433157 ,  99221801 ,  13433464 ,  97970459 ,  99221824 ,  99221240 ,  99222342 ,  12471546 ,  13436090 ,  13431895 ,  13437847 ,  13436218 ,  13435477 ,  13432614 ,  12472074 ,  12473457 ,  13432517 ,  13436240 ,  13436887 ,  13430022 ,  13430058 ,  14180617 ,  14180554 ,  13435382 ,  13438710 ,  99222301 ,  13438790 ,  13430979 ,  13438478 ,  13438538 ,  97971061 ,  13433964 ,  13437972 ,  13432526 ,  99222298 ,  97970546 ,  13434450 ,  13437599 ,  13431066 ,  99221483 ,  99221201 ,  12472298 ,  13439711 ,  13433375 ,  99400862 ,  13430738 ,  13438604 ,  13431459 ,  99221500 ,  12473226 ,  13438519 ,  13434984 ,  13435122 ,  13434990 ,  99401748 ,  13437366 ,  12473372 ,  13431738 ,  13432758 ,  13430972 ,  99401587 ,  99221123 ,  13439103 ,  13437029 ,  13431212 ,  13438888 ,  13435841 ,  13436295 ,  99221221 ,  13432240 ,  12471354 ,  99221143 ,  99221160 ,  13434044 ,  13437067 ,  13434562 ,  14180300 ,  12474742 ,  14180421 ,  99221411 ,  12474636 ,  97970760 ,  13439676 ,  14180322 ,  13432465 ,  99222428 ,  13438667 ,  13438646 ,  13432285 ,  12471264 ,  13432699 ,  13439672 ,  13431395 ,  13436196 ,  13435949 ,  97970445 ,  97971948 ,  97970871 ,  13433840 ,  13430738 ,  99222278 ,  99222442 ,  13430436 ,  13436106 ,  99220619 ,  13430890 ,  99220624 ,  99222683 ,  12473794 ,  14181051 ,  99401596 ,  99222649 ,  99222499 ,  13434672 ,  99221275 ,  98020128 ,  98021824 ,  99222637 ,  13430976 ,  13437150 ,  13437811 ,  13437882 ,  99221728 ,  99221130 ,  99220424 ,  13433477 ,  12472808 ,  13432226 ,  13435357 ,  99222706 ,  13432914 ,  99221211 ,  13430750 ,  13439117 ,  13437348 ,  13437277 ,  13437903 ,  13436130 ,  99222686 ,  13438501 ,  13433343 ,  13430762 ,  12474499 ,  13430487 ,  13439624 ,  13439202 ,  13439784 ,  13430744 ,  99222574 ,  97971891 ,  99221118 ,  99222150 ,  13431721 ,  97970120 ,  99221379 ,  14180931 ,  13436237 ,  13434402 ,  13434555 ,  13433452 ,  13436063 ,  13435158 ,  13436078 ,  12471569 ,  13433730 ,  13439021 ,  13439135 ,  13438726 ,  98121196 ,  13436684 ,  13432272 ,  13436584 ,  13434687 ,  13434688 ,  13435587 ,  13438304 ,  12471790 ,  13436765 ,  13436353 ,  13435539 ,  13438613 ,  13432488 ,  13432521 ,  13436194 ,  12471539 ,  99220514 ,  97970794 ,  13440178 ,  13440204 ,  13440173 ,  13440125 ,  13440213 ,  13440080 ,  13440196 ,  13440209 ,  99222391 ,  13439889 ,  13436344 ,  13437624 ,  99221196 ,  99221188 ,  12473801 ,  99221259 ,  13437120 ,  13430265 ,  13432821 ,  13432801 ,  12471071 ,  99220453 ,  13437692 ,  13432746 ,  13433210 ,  99221648 ,  13438230 ,  13430790 ,  13435172 ,  99220585 ,  99222182 ,  12470621 ,  12473400 ,  13434306 ,  13434258 ,  13434718 ,  13435263 ,  13434122 ,  13438455 ,  13438412 ,  13438470 ,  13438822 ,  12472013 ,  13432000 ,  99220918 ,  99222679 ,  99222486 ,  99221503 ,  99221524 ,  13436205 ,  12474607 ,  13433976 ,  13434037 ,  12472153 ,  12473899 ,  12473858 ,  12473863 ,  12471618 ,  99220913 ,  13432356 ,  12471179 ,  13431455 ,  12470618 ,  12474658 ,  13439028 ,  13434828 ,  13438857 ,  13439041 ,  13439019 ,  13438856 ,  13439032 ,  13438851 ,  13437501 ,  13431475 ,  12474773 ,  13431481 ,  12470457 ,  13431558 ,  99220433 ,  99744517 ,  13439775 ,  99401691 ,  13437137 ,  13436307 ,  99220526 ,  99222466 ,  99221387 ,  99222702 ,  13438840 ,  14180679 ,  13434608 ,  13434536 ,  99221567 ,  99221786 ,  99222131 ,  99222334 ,  99222454 ,  13439919 ,  99220809 ,  13439717 ,  13439504 ,  13439483 ,  99222653 ,  13440538 ,  13430076 ,  13430107 ,  13439828 ,  13439744 ,  13439686 ,  13439808 ,  13439832 ,  13439825 ,  13439834 ,  13439691 ,  98022314 ,  99221323 ,  13430584 ,  13434035 ,  13434429 ,  13438868 ,  13439043 ,  12472955 ,  99220459 ,  13438784 ,  99220365 ,  13438451 ,  12471924 ,  13430888 ,  13432650 ,  13431762 ,  13440148 ,  13437172 ,  12473635 ,  13433345 ,  14180909 ,  14180051 ,  14181053 ,  13431081 ,  13430206 ,  13437540 ,  98022235 ,  99222469 ,  13440242 ,  12473164 ,  13436386 ,  12470739 ,  13436491 ,  13436387 ,  14180567 ,  13437211 ,  13435323 ,  99744506 ,  13432390 ,  99222520 ,  99221288 ,  13439650 ,  13438354 ,  13437863 ,  13437928 ,  13430285 ,  13440214 ,  12471440 ,  13437291 ,  13439078 ,  13439826 ,  13435245 ,  13434117 ,  13438411 ,  98020970 ,  99222276 ,  13432384 ,  12473080 ,  13439191 ,  13437334 ,  13432398 ,  13437505 ,  13437387 ,  12474619 ,  99220588 ,  99220568 ,  99220570 ,  99401625 ,  99220670 ,  99744581 ,  99222269 ,  99221482 ,  99222388 ,  13438833 ,  12471507 ,  99401670 ,  99221261 ,  14180381 ,  13436118 ,  12470589 ,  13439196 ,  13440203 ,  13434981 ,  99222742 ,  99221612 ,  99401763 ,  99222620 ,  99221208 ,  99222310 ,  13439845 ,  12472521 ,  99220815 ,  99222728 ,  99221189 ,  13430496 ,  13439171 ,  99222715 ,  99220662 ,  13440211 ,  13433505 ,  13436458 ,  13433262 ,  13437169 ,  13431535 ,  13430455 ,  13440119 ,  12473642 ,  99222393 ,  13440221 ,  13431241 ,  99221232 ,  99222513 ,  99220461 ,  13439633 ,  13439715 ,  13439729 ,  99222264 ,  99401673 ,  99401623 ,  99221205 ,  13438517 ,  99222543 ,  99401581 ,  13430737 ,  14180891 ,  14180944 ,  13438877 ,  99222075 ,  13436734 ,  97970044 ,  98022169 ,  13434273 ,  12473216 ,  13435333 ,  13440518 ,  13434292 ,  12474218 ,  99222516 ,  13438779 ,  14180701 ,  13432657 ,  13432304 ,  13439666 ,  13439756 ,  13431629 ,  13432660 ,  13433308 ,  99222410 ,  13437135 ,  13434234 ,  13439789 ,  99221074 ,  13432900 ,  13437208 ,  13431277 ,  14180514 ,  12473595 ,  13436204 ,  99220784 ,  99221117 ,  13435491 ,  13435357 ,  13440265 ,  13431039 ,  98022294 ,  13430433 ,  13431068 ,  13438537 ,  13435119 ,  13435426 ,  14180667 ,  13440379 ,  13434640 ,  13434763 ,  13437423 ,  13430450 ,  13438707 ,  13438794 ,  13432006 ,  14180630 ,  99220565 ,  99220737 ,  13439099 ,  99220914 ,  13437843 ,  97970599 ,  13439061 ,  13438875 ,  12470841 ,  14180325 ,  99220676 ,  99221255 ,  13430006 ,  12470672 ,  13432355 ,  99222763 ,  14180380 ,  13440639 ,  12472259 ,  13439763 ,  13430706 ,  99222311 ,  99222315 ,  13433934 ,  99222125 ,  12471218 ,  13434395 ,  13433472 ,  12471626 ,  99221985 ,  12473898 ,  13438336 ,  99221066 ,  13440146 ,  13431103 ,  99221412 ,  14180829 ,  13438892 ,  99221456 ,  12472049 ,  13436638 ,  13430198 ,  13439723 ,  99222553 ,  13436429 ,  13438546 ,  13434690 ,  12474382 ,  12470653 ,  99222005 ,  13430018 ,  99222602 ,  13437443 ,  12473854 ,  99222556 ,  12473382 ,  12473152 ,  12473155 ,  13433502 ,  99220829 ,  13433480 ,  13433509 ,  99220364 ,  13437313 ,  13438550 ,  12470517 ,  99221062 ,  99220580 ,  13430374 ,  13440270 ,  13433339 ,  13430953 ,  99221083 ,  99222154 ,  99222445 ,  99221003 ,  99220859 ,  99222265 ,  13435543 ,  13435954 ,  14180835 ,  13439259 ,  13437481 ,  14180855 ,  13437378 ,  13436836 ,  13437420 ,  14180712 ,  14180714 ,  99221936 ,  99222012 ,  99221836 ,  99222450 ,  13439778 ,  13434006 ,  13439739 ,  13433434 ,  14180843 ,  14180496 ,  12470354 ,  13437238 ,  13439620 ,  99221043 ,  13438447 ,  99222685 ,  13430071 ,  99221819 ,  99221318 ,  13436732 ,  99222126 ,  99222659 ,  13437413 ,  13440268 ,  13432898 ,  99220837 ,  13440604 ,  13438343 ,  13438355 ,  99221092 ,  13433247 ,  13431114 ,  99222336 ,  99221530 ,  12473066 ,  13438879 ,  13438881 ,  99220416 ,  13438008 ,  13432612 ,  99222180 ,  13438724 ,  13438639 ,  13436690 ,  13439722 ,  99220658 ,  99220954 ,  13432476 ,  99222533 ,  14180752 ,  14181025 ,  13433399 ,  14180957 ,  99221237 ,  99221304 ,  99222650 ,  99401664 ,  13433923 ,  13432196 ,  99220671 ,  13440298 ,  13440458 ,  99221335 ,  99401591 ,  99222285 ,  13439887 ,  99221719 ,  12473109 ,  13432556 ,  13433616 ,  13432457 ,  13433346 ,  99221182 ,  99222415 ,  99222444 ,  13433439 ,  14180910 ,  13437455 ,  99222152 ,  99401671 ,  13438653 ,  14180909 ,  13439811 ,  13439795 ,  99220867 ,  13434049 ,  13434000 ,  99220823 ,  14180753 ,  13436318 ,  99222365 ,  13437268 ,  13432208 ,  13440245 ,  13440371 ,  13437215 ,  13433938 ,  13438514 ,  13434833 ,  13437236 ,  99222506 ,  13434453 ,  99222736 ,  99221535 ,  99222261 ,  13433548 ,  12471077 ,  12473736 ,  13438437 ,  99221198 ,  13432798 ,  13430124 ,  13440014 ,  12470607 ,  13431900 ,  12471138 ,  99221401 ,  99220813 ,  13440323 ,  13438208 ,  13432711 ,  12472008 ,  13430278 ,  12474512 ,  13430407 ,  13430417 ,  12474529 ,  13433112 ,  13438690 ,  99221274 ,  12474220 ,  13439450 ,  13439105 ,  98121329 ,  98121022 ,  98121126 ,  98121300 ,  98121131 ,  98120719 ,  98121346 ,  98121028 ,  98121068 ,  98120863 ,  13439738 ,  13439367 ,  13440307 ,  12470616 ,  99222609 ,  99222563 ,  12472824 ,  12473824 ,  99221227 ,  13434588 ,  13433917 ,  13432179 ,  13432597 ,  98121219 ,  98070965 ,  13438900 ,  13433546 ,  13431153 ,  13435179 ,  99401616 ,  99401717 ,  99222449 ,  99221687 ,  99220856 ,  13438004 ,  13439538 ,  99220788 ,  13432837 ,  99222629 ,  13430593 ,  99221578 ,  99220614 ,  12473950 ,  13430422 ,  14181071 ,  99222564 ,  13436723 ,  13432933 ,  99220691 ,  99222594 ,  99221053 ,  99222529 ,  99221229 ,  13431074 ,  13430269 ,  13439204 ,  99221667 ,  99222586 ,  13433204 ,  13439092 ,  99221099 ,  99221216 ,  99221283 ,  99220576 ,  13431900 ,  99221608 ,  13437578 ,  13440483 ;   *** 2) List Number 13100, Serial Numbers:   13530228 ,  13530338 ,  13530203 ,  13530205 ,  13530328 ,  13530312 ,  13530387 ,  13530209 ,  13530312 ,  12771348 ,  13530386 ,  13530385 ,  13530331 ,  13530373 ,  13530370 ;   *** 3) List Number 13150, Serial Numbers:   98431537 ,  98430945 ,  98431018 ,  98431041 ,  13637227 ,  99271058 ,  98431008 ,  99270417 ,  98930032 ,  98930073 ,  98930074 ,  98930083 ,  13631592 ,  99749369 ,  98430113 ,  98430192 ,  99450473 ,  98410749 ,  13630146 ,  98431142 ,  98431205 ,  13630031 ,  98930058 ,  98431409 ,  98410005 ,  98410015 ,  98411582 ,  98410379 ,  98411404 ,  99450476 ,  98410027 ,  98410353 ,  98410363 ,  98411239 ,  98431167 ,  98410231 ,  98410048 ,  98410580 ,  99270394 ,  13631348 ,  98431086 ,  99749251 ,  98410752 ,  99450050 ,  98410387 ,  99749532 ,  13631527 ,  98410741 ,  98410836 ,  98930402 ,  99749112 ,  98430321 ,  98411440 ,  12720251 ,  99450479 ,  98431440 ,  13636963 ,  98431000 ,  99270409 ,  98431149 ,  98411588 ,  98410556 ,  98410563 ,  98410910 ,  13630816 ,  98431200 ,  98410569 ,  98411468 ,  99450565 ,  98432002 ,  12720318 ,  13630186 ,  13630384 ,  98410078 ,  99450463 ,  98410722 ,  99450478 ,  98410579 ,  98410596 ,  98411312 ,  98411581 ,  98930497 ,  98410972 ,  98411080 ,  98411582 ,  98411622 ,  98410622 ,  99270406 ,  99270571 ,  98430968 ,  99270689 ,  98432021 ,  98430724 ,  99749554 ,  13630112 ,  12721156 ,  99749431 ,  98431174 ,  13631400 ,  13630492 ,  98432187 ,  99749552 ,  98432043 ,  12721163 ,  98431184 ,  13630909 ,  13631823 ,  12721091 ,  13635636 ,  13630480 ,  13631107 ,  13631673 ,  13630317 ,  13630426 ,  99270387 ,  13630880 ,  13631711 ,  13636201 ,  12720577 ,  13630338 ,  13630015 ,  12721082 ,  13633967 ,  12720583 ,  12720873 ,  12720160 ,  13632682 ,  13633145 ,  12720055 ,  13630583 ,  12721276 ,  13636416 ,  13636362 ,  13631294 ,  13637031 ,  12720614 ,  12720554 ,  13630739 ,  13631251 ,  12721250 ,  99270878 ,  98431435 ,  13631811 ,  13634026 ,  99749538 ,  13630419 ,  13631754 ,  12720275 ,  13631063 ,  12720631 ,  13633172 ,  12720765 ,  13631177 ,  13631192 ,  13631010 ,  12721182 ,  13631863 ,  12720082 ,  12721184 ,  13636378 ,  13631227 ,  99270505 ,  13633174 ,  13633143 ,  13630948 ,  12721167 ,  12720621 ,  13630976 ,  12720410 ,  13631671 ,  13630109 ,  13630489 ,  12721118 ,  13635670 ,  13635603 ,  99749580 ,  13632189 ,  12720382 ,  13637420 ,  13637301 ,  13631136 ,  13630868 ,  13630858 ,  13630640 ,  99270058 ,  13630536 ,  13632824 ,  13631882 ,  99749575 ,  13633173 ,  13630924 ,  13635652 ,  13631744 ,  13630900 ,  13630633 ,  13630740 ,  13631345 ,  98431170 ,  13631809 ,  13631652 ,  98430984 ,  12720822 ,  98431158 ,  13630401 ,  13635834 ,  13635794 ,  13631210 ,  13631243 ,  12720388 ,  98410485 ,  12721108 ,  13630585 ,  12720917 ,  12720850 ,  12720480 ,  13631691 ,  99749571 ,  98430910 ,  13630062 ,  13630823 ,  98432023 ,  13631551 ,  13631311 ,  13631605 ,  13630911 ,  13633941 ,  12720210 ,  12721018 ,  13631850 ,  13631923 ,  99270067 ,  98431140 ,  98431121 ,  13632028 ,  13631916 ,  13631548 ,  12720032 ,  13631451 ,  13631559 ,  13630537 ,  13635336 ,  13630596 ,  13635403 ,  13631515 ,  12720219 ,  13634316 ,  13630159 ,  12720125 ,  13632002 ,  12720791 ,  13630165 ,  13631300 ,  13631069 ,  13636142 ,  13631398 ,  13632160 ,  13630505 ,  12721209 ,  13631993 ,  99270698 ,  99270829 ,  13630726 ,  98432275 ,  99270470 ,  13636580 ,  13631655 ,  13631622 ,  12720720 ,  13632104 ,  99749540 ,  13637859 ,  99270747 ,  99270327 ,  99749514 ,  13630158 ,  98432111 ,  13636539 ,  13630422 ,  13630337 ,  13633129 ,  12720013 ,  13631009 ,  13631918 ,  13631898 ,  13630397 ,  13631763 ,  12721245 ,  12721227 ,  13630695 ,  13631800 ,  13637326 ,  99749506 ,  13632582 ,  13631545 ,  13636822 ,  13631344 ,  12720162 ,  12721014 ,  12720537 ,  13630837 ,  13632818 ,  13631241 ,  13631714 ,  13630567 ,  13630071 ,  13632035 ,  13634335 ,  12721168 ,  12721093 ,  13636349 ,  13636285 ,  13636483 ,  12720373 ,  12721044 ,  98432194 ,  98431010 ,  98431082 ,  98430964 ,  13634368 ,  13630322 ,  13636363 ,  12720228 ,  13631838 ,  13631604 ,  13634238 ,  13631055 ,  13631350 ,  13631258 ,  13630918 ,  13630838 ,  13631821 ,  12720630 ,  12721210 ,  13635545 ,  13637151 ,  13633127 ,  13630334 ,  13636171 ,  13631661 ,  13637779 ,  12720544 ,  13633944 ,  13630939 ,  12720681 ,  12720802 ,  13635936 ,  13631026 ,  13630460 ,  13631862 ,  13635249 ,  13635292 ,  12720650 ,  13634514 ,  12720202 ,  13631193 ,  13634240 ,  13635438 ,  13635655 ,  13635597 ,  13631404 ,  13631143 ,  13635799 ,  12720153 ,  13633170 ,  13630581 ,  12721226 ,  13631128 ,  13632683 ,  13636106 ,  98431518 ,  13630221 ,  13636197 ,  98430970 ,  13636222 ,  13636259 ,  13636272 ,  13630716 ,  12720920 ,  12720494 ,  13635896 ,  13636408 ,  13633607 ,  13631117 ,  13630499 ,  13634126 ,  13631339 ,  13630842 ,  12720137 ,  13630810 ,  13630682 ,  13631734 ,  12720392 ,  13630016 ,  12720668 ,  13635419 ,  13630815 ,  13636220 ,  98430971 ,  12720425 ,  13632203 ,  99270141 ,  13631563 ,  12721160 ,  13630915 ,  12721129 ,  13632186 ,  13630182 ,  13630709 ,  13630701 ,  12721101 ,  13631431 ,  13631500 ,  12720572 ,  12721046 ,  99270125 ,  99749527 ,  13631608 ,  13635993 ,  12720302 ,  13631040 ,  13632250 ,  98432177 ,  98410238 ,  98410671 ,  99749541 ,  12720114 ,  99749509 ,  13631002 ,  13631315 ,  99749523 ,  99749539 ,  13630378 ,  13635607 ,  13630267 ,  12721140 ,  13631245 ,  12720586 ,  13636487 ,  13630790 ,  13630827 ,  13630605 ,  12720448 ,  12720787 ,  13630775 ,  12721157 ,  13635590 ,  13631615 ,  13631226 ,  13631038 ,  12721047 ,  12721217 ,  13630434 ,  12720111 ,  99271144 ,  13634349 ,  98432020 ,  13630932 ,  12720229 ,  13630246 ,  13631237 ,  13636506 ,  98431301 ,  99270338 ,  13631878 ,  13631722 ,  12720664 ,  13633194 ,  13634767 ,  13630981 ,  13632075 ,  13632827 ,  12721251 ,  13631414 ,  12720360 ,  13636204 ,  13631913 ,  13631649 ,  13636117 ,  13630754 ,  13632010 ,  13632082 ,  13630654 ,  13630105 ,  13635797 ,  13633511 ,  13632152 ,  12720258 ,  13635816 ,  13631394 ,  13631768 ,  13632102 ,  12720190 ,  13631223 ,  12720447 ,  13631514 ,  13631264 ,  12720077 ,  98430891 ,  13636532 ,  13632206 ,  13630984 ,  13630836 ,  13630982 ,  13634411 ,  12720184 ,  13631558 ,  13632834 ,  13630149 ,  12721223 ,  13635488 ,  12720722 ,  12720054 ,  12720993 ,  13634208 ,  99749360 ,  99749252 ,  12721054 ,  98432072 ,  98432224 ,  13630211 ,  13631169 ,  13631999 ,  13631927 ,  13631997 ,  13631846 ,  13636102 ,  98432261 ,  13631447 ,  13631990 ,  13632025 ,  13631865 ,  13631879 ,  13630504 ,  13630530 ,  12720603 ,  12720078 ,  13630394 ,  13632708 ,  13630923 ,  12721040 ,  13632034 ,  13634552 ,  13633626 ,  13630336 ,  13633601 ,  13631609 ,  13632051 ,  13631472 ,  13635381 ,  13636809 ,  13633594 ,  13634073 ,  12721141 ,  12720171 ,  13634261 ,  13630069 ,  13636466 ,  13630687 ,  13633552 ,  13635752 ,  13632076 ,  13633677 ,  13634154 ,  13631956 ,  13630256 ,  13631832 ,  12720602 ,  13635544 ,  13631302 ,  13630617 ,  13635445 ,  99270871 ,  13630913 ,  13631682 ,  12720169 ,  13633951 ,  13630784 ,  13635442 ,  98432167 ,  13631481 ,  13630623 ,  13636329 ,  12720512 ,  13630104 ,  13631282 ,  12721197 ,  13631352 ,  13630467 ,  12721084 ,  13631740 ,  13630141 ,  13631015 ,  12720433 ,  13632089 ,  13631820 ,  13632125 ,  13630027 ,  13631713 ,  13631573 ,  13630719 ,  13634370 ,  13632588 ,  13635534 ,  98432097 ,  98431056 ,  13631502 ,  12720322 ,  13631626 ,  13631586 ,  13632346 ,  13633570 ,  13635869 ,  98431297 ,  13631417 ,  13632221 ,  13632191 ,  13631743 ,  99271127 ,  13636802 ,  13632240 ,  99271084 ,  99271080 ,  99271091 ,  13633633 ,  13631519 ,  12720983 ,  13631462 ,  12721222 ,  13637206 ,  13631584 ,  13631599 ,  13637326 ,  13635800 ,  12720737 ,  13632138 ,  12721081 ,  13630248 ,  13634917 ,  98431260 ,  13635706 ,  13635632 ,  13635758 ,  13635625 ,  98431180 ,  13636299 ,  13630372 ,  13631426 ,  13631046 ,  13630525 ,  13632172 ,  12721203 ,  13630352 ,  13630783 ,  98430673 ,  98432184 ,  99270486 ,  13637032 ,  13633958 ,  13631541 ,  13630954 ,  13630861 ,  13630910 ,  13635285 ,  12720140 ,  12720095 ,  13630125 ,  13631824 ,  13635764 ,  13637272 ,  13630710 ,  13631041 ,  13635374 ,  13631232 ,  12720238 ,  13635849 ,  13635497 ,  13631001 ,  13634214 ,  13635298 ,  98430636 ,  98432199 ,  13636202 ,  13630951 ,  13633647 ,  13636059 ,  13631164 ,  13631316 ,  13634405 ,  13635449 ,  13630662 ,  12720282 ,  98432098 ,  12720191 ,  13632081 ,  13632241 ,  13632033 ,  13631697 ,  13630207 ,  13631577 ,  12720729 ,  13630933 ,  13631413 ,  13631695 ,  13633951 ,  13631636 ,  13631639 ,  13631200 ,  99270518 ,  13636502 ,  12721143 ,  13635681 ,  13631668 ,  13633939 ,  13635410 ,  13635310 ,  13631202 ,  13631091 ,  13631664 ,  13634133 ,  13633945 ,  13631209 ,  13631843 ,  13636310 ,  12720684 ,  13633649 ,  13630316 ,  13633554 ,  13631785 ,  13633506 ,  13631939 ,  13632442 ,  13637155 ,  13637071 ,  13630059 ,  12721268 ,  13631837 ,  12720429 ,  12720089 ,  13637161 ,  13631978 ,  12720498 ,  13630442 ,  13630758 ,  13634834 ,  13637159 ,  12720732 ,  12720686 ,  12720361 ,  98432237 ,  13635855 ,  99270662 ,  13631709 ,  99270378 ,  13637804 ,  13630886 ,  12720811 ,  13634573 ,  13631627 ,  13635538 ,  13631342 ,  13631277 ,  12720147 ,  13637920 ,  13630622 ,  12720035 ,  12720792 ,  13632048 ,  13630658 ,  13634350 ,  13634383 ,  13630970 ,  12720957 ,  13633943 ,  13630445 ,  13631566 ,  13630300 ,  13631775 ,  13630568 ,  98432001 ,  13631653 ,  13634517 ,  13631029 ,  13631581 ,  13631547 ,  13631645 ,  98432230 ,  99270457 ,  12720495 ,  13633705 ,  13631925 ,  13632162 ,  98432060 ,  13635437 ,  13631753 ,  13632471 ,  13631930 ,  13630665 ,  13632020 ,  13636016 ,  13632756 ,  13636055 ,  13637283 ,  98432022 ,  13631090 ,  13635714 ,  13634665 ,  13633603 ,  98431225 ,  13631411 ,  12721165 ,  13631147 ,  13635599 ,  13635628 ,  13630862 ,  13631371 ,  13634399 ,  13631561 ,  13630752 ,  98431152 ,  13635244 ,  13630215 ,  13630680 ,  13631826 ,  13634599 ,  13630813 ,  13630660 ,  99270977 ,  99749419 ,  98432181 ,  98430870 ,  98430641 ,  99749061 ,  98432253 ,  12721004 ,  13636526 ,  12720545 ,  98432176 ,  98431800 ,  98431071 ,  13630940 ,  13636291 ,  13636294 ,  13635440 ,  13635392 ,  13635425 ,  13631284 ,  13631522 ,  12720131 ,  13630405 ,  13630140 ,  13630193 ,  13631706 ,  13631098 ,  13630853 ,  13634081 ,  13630400 ,  13635380 ,  12720240 ,  12720004 ,  13633122 ,  13632063 ,  13631944 ,  13635441 ,  13631953 ,  13635930 ,  13635899 ,  12720311 ,  13631252 ,  99270442 ,  12721155 ,  13630991 ,  13631225 ,  13631025 ,  13631336 ,  13635314 ,  13631286 ,  12720756 ,  13635760 ,  13635740 ,  13636449 ,  13630553 ,  12720234 ,  13630285 ,  98432277 ,  13630908 ,  98431602 ,  99749546 ,  12720286 ,  13630028 ,  13630788 ,  98431214 ,  13635317 ,  99270001 ,  98411617 ,  12720214 ,  13630889 ,  12721248 ,  13634166 ,  13630054 ,  13633133 ,  13630878 ,  13630603 ,  13634331 ,  13636986 ,  13632005 ,  13631534 ,  13631555 ,  13636642 ,  13630953 ,  13630635 ,  13635929 ,  98431264 ,  13634506 ,  13630233 ,  13636050 ,  98432188 ,  12721070 ,  12721017 ,  98432209 ,  13631508 ,  13631635 ,  12721009 ,  13631479 ,  12720872 ,  13631528 ,  13631681 ,  13631287 ,  13631488 ,  13631456 ,  12720188 ,  13630638 ,  12721144 ,  13635498 ,  13635645 ,  12721193 ,  13636194 ,  98431290 ,  98432018 ,  13633982 ,  13631248 ,  98431217 ,  98431285 ,  13633496 ,  13636153 ,  13631084 ,  13630111 ,  12720566 ,  13631390 ,  13635766 ,  13637137 ,  12721001 ,  12721126 ,  13637245 ,  98432265 ,  13635687 ,  13635982 ,  13630251 ,  13630406 ,  13630766 ,  13633638 ,  12721264 ,  13631313 ,  13634528 ,  13631194 ,  13630129 ,  12720567 ,  12720525 ,  13631303 ,  13630943 ,  13632047 ,  12720192 ,  13633151 ,  13631629 ,  13637687 ,  12720247 ,  13630561 ,  13633713 ,  13637590 ,  99271055 ,  13636081 ,  13637481 ,  13635840 ,  13631816 ,  12720723 ,  13635145 ,  13635796 ,  13632113 ,  13636750 ,  13636845 ,  13633562 ,  13630203 ,  13631594 ,  13630367 ,  13635399 ,  13631520 ,  13632038 ,  13630877 ,  12720869 ,  13633589 ,  12720301 ,  13630522 . 13637685 ,  99270325 ,  13633640 ,  13637106 ,  13634010 ,  13633998 ,  13630914 ,  13632060 ,  13636083 ,  13635793 ,  13631780 ,  13636290 ,  99270179 ,  13631557 ,  13637209 ,  13633650 ,  12720018 ,  13633590 ,  12720023 ,  13631752 ,  13633659 ,  13633701 ,  13633673 ,  13633616 ,  13630628 ,  13635829 ,  98431097 ,  98432210 ,  13633543 ,  12721247 ,  13630074 ,  13630044 ,  13630363 ,  13632123 ,  12720404 ,  13630358 ,  13631568 ,  13631517 ,  13634132 ,  13634289 ,  12720705 ,  13634890 ,  13631623 ,  13632214 ,  13637870 ,  12720295 ,  13633711 ,  13631688 ,  13632217 ,  13635359 ,  13635306 ,  13633586 ,  13633669 ,  98432251 ,  13633195 ,  13630254 ,  13637133 ,  13631080 ,  13637010 ,  13633995 ,  13630685 ,  12720164 ,  98432228 ,  99271057 ,  12720591 ,  13630068 ,  13637234 ,  13637099 ,  13631889 ,  12720176 ,  13631694 ,  13636008 ,  13632054 ,  13633718 ,  13636818 ,  13635267 ,  13634343 ,  13634813 ,  13635333 ,  13635332 ,  13635344 ,  13635237 ,  13634024 ,  12720003 ,  12720303 ,  12720408 ,  13630803 ,  12721096 ,  13630619 ,  13635662 ,  13632216 ,  13630435 ,  13631773 ,  13630216 ,  13635698 ,  13636330 ,  13631660 ,  13630184 ,  13631921 ,  13631957 ,  12720547 ,  12720719 ,  13637039 ,  13635405 ,  13631060 ,  13630308 ,  13631942 ,  13631869 ,  13631949 ,  13632170 ,  13637635 ,  13630436 ,  13630725 ,  13635388 ,  13635245 ,  13635266 ,  13635205 ,  13635877 ,  13632198 ,  13631772 ,  13636331 ,  13634365 ,  13630271 ,  12720029 ,  13633628 ,  13633732 ,  13630209 ,  13630494 ,  12720581 ,  13635120 ,  13631081 ,  13636038 ,  13631552 ,  13636017 ,  13635127 ,  13633148 ,  13631905 ,  13633734 ,  98431530 ,  13632019 ,  13634369 ,  12720779 ,  13634315 ,  13631443 ,  13631650 ,  12720771 ,  99270915 ,  13631027 ,  12720138 ,  13630830 ,  13633981 ,  13634221 ,  13636691 ,  13630449 ,  13635253 ,  13632052 ,  13631621 ,  12721033 ,  98431313 ,  13631218 ,  13636275 ,  13637185 ,  13633188 ,  99270646 ,  99749420 ,  98431714 ,  13632006 ,  13636341 ,  13636316 ,  13635416 ,  13637448 ,  13634060 ,  13630713 ,  13634263 ,  13634389 ,  98431512 ,  13630866 ,  12720890 ,  13630967 ,  13634689 ,  13630228 ,  13630850 ,  13630050 ,  13634265 ,  13630883 ,  13636048 ,  12720891 ,  13635667 ,  12720278 ,  13632735 ,  13631337 ,  13634789 ,  13630226 ,  13632594 ,  12721166 ,  99271109 ,  12721006 ,  98431749 ,  12721131 ,  13635148 ,  13632110 ,  13636112 ,  13637732 ,  12720986 ,  13630870 ,  13632757 ,  13636295 ,  13635483 ,  13630032 ,  13635732 ,  13635772 ,  98431050 ,  12720414 ,  13630904 ,  99749534 ,  13634327 ,  13634181 ,  13630696 ,  13630274 ,  13631174 ,  13631618 ,  13631445 ,  13635162 ,  13631322 ,  13633545 ,  13631764 ,  13634376 ,  13630828 ,  13630702 ,  13630440 ,  13637639 ,  13632091 ,  13635436 ,  13635358 ,  13634183 ,  13634210 ,  13634334 ,  13635742 ,  13635761 ,  13633513 ,  13636719 ,  13637129 ,  13637204 ,  13631895 ,  13632184 ,  13635415 ,  13631979 ,  12720345 ,  13636753 ,  13630297 ,  13635478 ,  13636688 ,  13635624 ,  13631859 ,  13635593 ,  12720083 ,  13635407 ,  12721094 ,  99749327 ,  99749558 ,  99749563 ,  13635547 ,  13636189 ,  13636485 ,  13634283 ,  13630358 ,  13635791 ,  13631415 ,  13631485 ,  13631428 ,  13634684 ,  12720587 ,  12721275 ,  12720700 ,  13634381 ,  13630785 ,  13637140 ,  13632136 ,  13631158 ,  99270929 ,  13637036 ,  13635499 ,  13635734 ,  13634021 ,  13637050 ,  13632055 ,  12720077 ,  13630094 ,  13636148 ,  13631389 ,  13634929 ,  13633694 ,  13637152 ,  13636057 ,  13633707 ,  13630941 ,  13635832 ,  12720687 ,  13631593 ,  13631518 ,  13635475 ,  13635458 ,  13635397 ,  13635493 ,  12720254 ,  13630067 ,  12720246 ,  13630832 ,  13630857 ,  12720552 ,  99271105 ,  13636806 ,  13635519 ,  99749234 ,  99749317 ,  13633639 ,  13632200 ,  13630439 ,  13635456 ,  13634169 ,  13633726 ,  13636296 ,  13631585 ,  13631442 ,  13637509 ,  13637491 ,  13632105 ,  13632145 ,  12720682 ,  13632753 ,  13633631 ,  13637527 ,  13631915 ,  99271018 ,  13634558 ,  13634486 ,  13637892 ,  13634330 ,  13637115 ,  13636103 ,  13632249 ,  13632023 ,  13631459 ,  13637917 ,  13632785 ,  13635844 ,  13635831 ,  13634585 ,  13637006 ,  13630113 ,  13632134 ,  13635466 ,  13636833 ,  13636646 ,  13635455 ,  13637795 ,  13635513 ,  13631000 ,  13637158 ,  13636282 ,  12720271 ,  13631733 ,  13631590 ,  13635418 ,  13634733 ,  13637052 ,  13635859 ,  13636786 ,  13633679 ,  13632100 ,  13632150 ,  12720619 ,  13635559 ,  13636167 ,  13637312 ,  13634042 ,  12720306 ,  13630521 ,  12721170 ,  13632159 ,  13636093 ,  13631173 ,  13637128 ,  13634271 ,  13633975 ,  13635972 ,  13637278 ,  13633190 ,  13636582 ,  13630029 ,  13634630 ,  13636255 ,  12720755 ,  13634850 ,  13636080 ,  12720502 ,  13636839 ,  13631146 ,  13631589 ,  13636859 ,  13636205 ,  13634001 ,  12720548 ,  12720775 ,  13635426 ,  13635512 ,  13635506 ,  13632158 ,  13631022 ,  12720593 ,  13631992 ,  12720344 ,  13631467 ,  13631600 ,  13636229 ,  13635421 ,  13635432 ,  13635549 ,  13634641 ,  13632111 ,  12720557 ,  13635412 ,  13636381 ,  13632114 ,  13636663 ,  13636163 ,  13637758 ,  13630779 ,  13635481 ,  13632124 ,  13633671 ,  13634025 ,  13637402 ,  13630952 ,  13637067 ,  13632824 ,  13634520 ,  13631928 ,  13635642 ,  13633985 ,  13636710 ,  13634409 ,  13636619 ,  13636601 ,  13637120 ,  13633964 ,  13637351 ,  13631297 ,  13634148 ,  13630122 ,  13631233 ,  13630930 ,  12720159 ,  13633668 ,  13637840 ,  13636452 ,  13634815 ,  99270960 ,  13631220 ,  13631206 ,  13637648 ,  13637626 ,  13636268 ,  13634360 ,  13630162 ,  13630214 ,  13636828 ,  13636669 ,  13637012 ,  13637005 ,  12721204 ,  13631506 ,  13630812 ,  13631933 ,  13632202 ,  13631405 ,  13635790 ,  12720734 ,  13637486 ,  13637432 ,  13637560 ,  13636467 ,  13636280 ,  13637405 ,  13631103 ,  13632210 ,  13630674 ,  13637910 ,  13637727 ,  13634097 ,  13633128 ,  13635833 ,  13635406 ,  13631872 ,  12720577 ,  13631016 ,  12721130 ,  13630591 ,  12721028 ,  13636643 ,  99749572 ,  13637403 ,  12720487 ,  13631132 ,  13630500 ,  13635339 ,  13634362 ,  12720439 ,  13636593 ,  13635434 ,  12720908 ,  13633629 ,  13631611 ,  13635783 ,  13635716 ,  13632544 ,  13634515 ,  13630048 ,  13631061 ,  13637077 ,  13631802 ,  13633519 ,  13633523 ,  13633662 ,  13631030 ,  13631746 ,  12720966 ,  12720721 ,  13633961 ,  13634091 ,  13633942 ,  13631932 ,  13630741 ,  13637943 ,  13634119 ,  13636793 ,  13636607 ,  13633731 ,  12720136 ,  13632690 ,  13634912 ,  12721146 ,  13633714 ,  13630303 ,  12721138 ,  13635598 ,  13630060 ,  13634581 ,  13631940 ,  13630700 ,  12720346 ,  12720456 ,  13636455 ,  13633598 ,  13631188 ,  13636996 ,  12720428 ,  13637737 ,  13631203 ,  13634361 ,  99749567 ,  13632230 ,  13636286 ,  13636247 ,  13636248 ,  13634300 ,  13634028 ,  12720980 ,  13636795 ,  13637136 ,  13633704 ,  13630399 ,  13630544 ,  13630341 ,  13632795 ,  13636686 ,  13636762 ,  13637589 ,  13637015 ,  13637146 ,  12720767 ,  12721117 ,  12720129 ,  13633182 ,  13635940 ,  12720417 ,  13637708 ,  13637802 ,  12720390 ,  13634363 ,  13631149 ,  13636221 ,  13636645 ,  13633952 ,  13634079 ,  13630295 ,  13635922 ,  13632149 ,  13636907 ,  13631727 ,  13635617 ,  12720531 ,  13637821 ,  13636546 ,  13637702 ,  13632011 ,  13634182 ,  13631065 ,  13636061 ,  13635735 ,  13636567 ,  13630572 ,  13635684 ,  12721111 ,  13634382 ,  13636063 ,  13636342 ,  13636141 ,  13636332 ,  13636121 ,  99270916 ,  13637616 ,  13631935 ,  12720657 ,  13636924 ,  13633547 ,  13637258 ,  13632802 ,  13637273 ,  12720688 ,  13637915 ,  13634820 ,  13633646 ,  13637546 ,  13630487 ,  13636219 ,  13631703 ,  98410659 ,  13631747
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    Worldwide Distribution - USA including AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY. Internationally to Australia, New Zealand, China, Hong Kong, Taiwan, Japan, Korea, Philippines, Thailand, Malaysia, Canada, Austria, Bahrain, Belgium, Croatia, Denmark, Egypt, Finland, France, Germany, Gibraltar, Greece, Hungary, Iceland, Ireland, Israel, Italy, Jordan, Kuwait, Lebanon, Libya, Luxembourg, Malta, Netherlands, Norway, Oman, Portugal, Qatar, Romania, Saudi Arabia, Slovenia, South Africa, Spain, Sweden, Switzerland, Turkey, UK, United Arab Emirates, Brazil, Chile and Colombia.
  • Product Description
    The GemStar Pump is a small and lightweight, single channel infusion device designed for use in the home, in the hospital, or anywhere electronic infusion is required. The GemStar Pump can be powered by AC mains adaptor, rechargeable battery pack, docking station, or two disposable AA alkaline batteries. When powered by batteries, The GemStar Pump is ideal for ambulatory patients.
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Hospira Inc., 275 N Field Dr, Lake Forest IL 60045-2579
  • Manufacturer Parent Company (2017)
  • Source
    USFDA