Recall of Device Recall GE OEC 9800

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by GE OEC Medical Systems, Inc.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    62691
  • Event Risk Class
    Class 2
  • Event Number
    Z-2240-2012
  • Event Initiated Date
    2009-11-06
  • Event Date Posted
    2012-08-22
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2013-01-28
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    System, x-ray, fluoroscopic, image-intensified - Product Code JAA
  • Reason
    Ge oec is recalling certain 9800 image intensified flouroscopic x-ray systems due to customer complaint analysis showing that a replacement hitachi 160 gb hard drive, when installed on certain models is susceptible of causing unanticipated system shut downs, no boots, data loss or data mix, or unexpected hard drive corruption.
  • Action
    GE OEC sent a Urgent Recall Notice dated November 6, 2009, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The notification advised them that GE Healthcare had identified a potential safety issue associated with a "Significant increase on Data Loss, Data Mix, and System file corruption of the Hard Drives" on the 9800 Systems. The field correction consisted of installing a "Smart Power Switch" between the power control PCB and the system power switch on 9800 Systems with a Celeron single board computer and software version 29. This will remedy the potential for data loss, data mix, and system file corruption of the hard drives. For further questions please call 800-874-7378 Option 8.

Device

  • Model / Serial
    Model Number HDS721616PLAT80; GE Healthcare part number 5304649.
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    Worldwide Distribution -- USA (nationwide)
  • Product Description
    892.1650 Image-intensified fluoroscopic x-ray system; the device is designed to provide fluoroscopic and spot film imaging of the patient during diagnostic surgical and interventional procedures. The systems include features specifically designed for use in diagnostic and interventional cardiac imaging procedures, and are also intended for cholangiography, endoscopic, urologic, orthopedic, neurologic, vascular, critical care and emergency room procedures. The systems may be used for other imaging applications at the physician's discretion.
  • Manufacturer

Manufacturer

  • Manufacturer Address
    GE OEC Medical Systems, Inc, 384 Wright Brothers Dr, Salt Lake City UT 84116-2862
  • Source
    USFDA