Recall of Device Recall FP1000 Cell Preparation System

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Beckman Coulter Inc.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    38369
  • Event Risk Class
    Class 2
  • Event Number
    Z-0002-2008
  • Event Initiated Date
    2007-02-22
  • Event Date Posted
    2007-10-16
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2012-04-26
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    flow cytometer cell preparation - Product Code LXG
  • Reason
    Pressure (from recapped tubes) or vacuum (from under-filled short-draw tubes) will cause inaccurate results if not properly vented.
  • Action
    A Product Corrective Action (PCA) letter was mailed on Feb 22, 2007, to all FP1000 customers to inform them of an issue associated with the FP1000 Cell Preparation System. Pressure (from recapped tubes) or vacuum (from under-filled short-draw tubes) will cause inaccurate results if not properly vented. Under these conditions the FP1000 will under-dispense specimen volume leading to lower than expected absolute counts. Specimen tubes that are recapped or underfilled must therefore be vented prior to running on the FP1000. If the initial state of the specimen tube is unknown, then it must be vented to ensure accurate results prior to running on the FP1000. Additional instructions are provided. A Fax Back Response Form was also provided.

Device

  • Model / Serial
    Part Number 624922. Software versions 1.1 & 1.2
  • Product Classification
  • Device Class
    1
  • Implanted device?
    No
  • Distribution
    Nationwide (MA, IN, IL, UT, RI, NJ), and Canada.
  • Product Description
    FP1000 Cell Preparation System Software versions 1.1 & 1.2, Part Number 624922, for in vitro diagnostic use.