Recall of Device Recall Emit(R) 2000 Phenobarbital Assay (OSR4D229) and Syva(R) Emit(R) Phenobarbital Assay (4D019UL)

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Siemens Healthcare Diagnostics, Inc..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    63892
  • Event Risk Class
    Class 2
  • Event Number
    Z-0664-2013
  • Event Initiated Date
    2012-12-10
  • Event Date Posted
    2013-01-10
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2014-08-27
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Enzyme immunoassay, phenobarbital - Product Code DLZ
  • Reason
    Firm has confirmed higher imprecision with the recalled lots on beckman coulter au(r) systems and the syva(r) emit(r) 2000 phenobarbital (4d019ul) lots e1, e2, and e3 as an application if the au(r) system is used, the issue may be experienced. firm has found that within-run imprecision of the recalled lots frequency of outliers that differ from the expected value by 20%. this is an expansion of.
  • Action
    Siemens sent an "URGENT MEDICAL DEVICE RECALL" notice letter dated November 2012 to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to immediately discontinue use and discard any remaining inventory of the affected product. Customers were asked to share the content of the letter with their laboratory director regarding the need to review previous Phenobarbital results, conduct patient follow up and/or repeat testing. For a no charge replacement of remaining inventory customers were instructed to contact Siemens Customer Service Center at 888-588-3916. For questions contact Siemens Technical Solutions Center at 800-227-8994.

Device

  • Model / Serial
    Emit(R) 2000 Phenobarbital Assay (OSR4D229) lots 1237 and 1264 and/or Syva(R) Emit(R) Phenobarbital Assay (4D019UL), lots E1, E2, and E3.
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    Distributed in the states of CA, FL, IL, KS, MA, MI, MN, MS, NM, NY, NC, OH, OR, PA, TN, TX, UT, and WA.
  • Product Description
    Emit(R) 2000 Phenobarbital Assay (OSR4D229) and Syva(R) Emit(R) Phenobarbital Assay (4D019UL). || Homogeneous enzyme immunoassay intended for use in the quantitative analysis of Phenobarbital in human serum or plasma.
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Siemens Healthcare Diagnostics, Inc., 500 GBC Drive, Mailstop 514, PO BOX 6101, Newark DE 19714-6101
  • Manufacturer Parent Company (2017)
  • Source
    USFDA