Recall of Device Recall Cyberknife Robotic Radiosurgery System

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Accuray Inc.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    50207
  • Event Risk Class
    Class 2
  • Event Number
    Z-0629-2009
  • Event Initiated Date
    2008-11-14
  • Event Date Posted
    2009-01-17
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2010-12-02
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Accelerator, linear, medical - Product Code IYE
  • Reason
    System may use random incorrect data to calculate dose. resulting dose calculation error can exceed 100% of correct dose which may lead to serious patient injury.
  • Action
    An Urgent Advisory Notification letter dated November 14, 2008 was issued to all sites via Fed Ex. The letter identifies the issue and required actions. The user was instructed not to modify current extrapolation depth settings and to contact Accuray Incorporated Customer Support prior to making any changes. Please direct any questions or concerns regarding this issue to Accuray Incorporated Customer Support at 1-877-668-8667 (USA) or 1-408-716-4700 (outside of USA) or customersupport@accuray.com.

Device

  • Model / Serial
    All units that have MP software version 3.0 and the Iris Collimator in clinical use at the same time are subject to this recall.
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    Nationwide Distribution. Product was distributed to 4 medical facilities in the US.
  • Product Description
    Cyberknife Robotic Radiosurgery System. A radiation therapy device, MultiPlan (MP) Treatment Planning Software and Iris Variable Aperture Collimator, Software version 3.0. || Product is indicated for treatment planning and image guided stereotactic radiosurgery and precision radiotherapy of lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.