Recall of Device Recall ContraAngle Torque Driver Kit

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Biomet 3i, LLC.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    67183
  • Event Risk Class
    Class 2
  • Event Number
    Z-1171-2014
  • Event Initiated Date
    2013-07-22
  • Event Date Posted
    2014-03-12
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2015-04-21
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    General surgery tray (kit) - Product Code LRO
  • Reason
    During recent revised sterilization validation testing, the included surgical trays did not meet the sterility assurance level required.
  • Action
    Biomet 3i sent an Urgent Medical Device Field Correction Notice dated July 2013 to all affected customer via email, facsimile, or postal mail. The letter identifed the affected product, problem and actions to be taken. Customers were instructed to complete the attached response form and fax to 561-514-6316. For questions contact Biomet 3i customer service at 1-800-342-5454.

Device

  • Model / Serial
    NPSDK0 aka PSKDO (PSKDT1)(
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    Worldwide Distribution - USA Nationwide in the states of AL, AZ, CA, CO, CT, DC, DE, FL, GA, IL, KS, KY, LA, MA, MD, MI, MN, MS, MT, NC, NE, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC, TN, TX, VA, WA, WI, WV, including Hawaii and Puerto Rico and the countries of : Argentina, Austria, Canada, China, Costa Rica, Germany, Dominican Republic, Spain, France, Great Britain, Hong Kong, Hungary, Israel, Japan, Korea, Mexico, Poland, Portugal, Paraguay, Sweden, Singapore, Taiwan, and Uruguay.
  • Product Description
    Contra-Angle Torque Driver Kit For Certain and External Connection || Product Usage: Molded plastic tray used in holding various dental instruments for autoclave processing.
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Biomet 3i, LLC, 4555 Riverside Dr, Palm Beach Gardens FL 33410-4200
  • Manufacturer Parent Company (2017)
  • Manufacturer comment
    “The safety of patients has always been, and continues to be, Zimmer Biomet’s top priority and it is our honor to be a leader in this industry for the past 90 years,” Zimmer Biomet told ICIJ in a statement. “We adhere to strict regulatory standards, and work closely with the FDA and all applicable regulatory agencies in each of our regions as part of our commitment to operating a first-rate quality management system across our global manufacturing network. The company added that it is focused on staying at the forefront of innovation and doing right by the millions of patients who rely on the company’s products.
  • Source
    USFDA