Recall of Device Recall cobas Liat Analyzer

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Roche Molecular Systems, Inc..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    74094
  • Event Risk Class
    Class 2
  • Event Number
    Z-2216-2016
  • Event Initiated Date
    2016-05-09
  • Event Status
    Completed
  • Event Country
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Respiratory virus panel nucleic acid assay system - Product Code OCC
  • Reason
    Device was released with default pcal/gain settings for photometers instead of the pcal/gain settings that are required for release. the default settings can potentially lead to the generation of erroneous or invalid results.
  • Action
    Customers were notified of the recall via phone and email sent on 5/9/2016. They were instructed to take the following actions: Immediately discontinue use of the affected instruments until they have been reworked by a Roche Field Engineering Specialist. A Roche representative will contact you to schedule this service. Complete the attached fax form and fax it to 1-317-521-4815. File this Urgent Medical Device Correction (UMDC) for future reference. Please contact the Roche Support Network Customer Support Center 1-800-800-5973 if you have questions.