Recall of Device Recall CIVCO Rectangular Extension

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Med Tec Inc.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    55999
  • Event Risk Class
    Class 2
  • Event Number
    Z-0394-2011
  • Event Initiated Date
    2010-06-21
  • Event Date Posted
    2010-11-16
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2012-04-17
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Table, radiographic, stationary top - Product Code IXQ
  • Reason
    Under certain conditions, the extensions can become detached from the main body of the couch.
  • Action
    The recalling firm, CIVCO, issued an 'Urgent Field Safety Notice" dated June 15, 2010, via overnight mail on June 21, 2010, regarding the first issue for the three extension models. The letter described the product, problem, and actions to be taken by the customers. The customers were instructed to load the patients on the couch from the bottom of the couch and not from the superior end with the extension and do not let patients crawl to the top of the couch with their full upper body weight at the end of the couch; notify others within their organization that may have the extension, and complete and return the enclosed response form via fax: Attention: CIVCO MTL6600 Rectangular Extension at 877-271-2310. The letter informs the customer that they will be receiving a kit to replace the locking tab so that the extension will be less likely to come loose. Once the kit is received, the customers are ask to return the replaced part and sign a confirmation form for their records. CIVCO sent a second "Urgent Field Safety Notice" dated July 20, 2010 via overnight mail on July 21, 2010, and by phone on July 22, 2010, regarding the second issue which covered only MTIL6600 regarding the weight being dropped onto the corner of the rectangular extension. The letter described the product, problem and action to be taken by the customers. The customers were instructed to accept a one-piece Universal Couch Top replacement; sign the attached form stating the they will stop using the rectangular extension and return it to CIVCO or sign the form stating they have been advised of the recall and they will exercise the necessary precautions to prevent the event from occurring by always using the thumb screw lock and never allowing the patient to crawl on the couch top or extension, and to notify others within their organization that have the extension. The customer was informed that their current system would be replaced and sent back to the recalling firm. Please contac

Device

  • Model / Serial
    Lots M179980, M163620, M163110, M161470, M161460, M161450, M161440, M135840, and M131280.
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    Worldwide distribution: USA including states of: CA, FL, IA OH, and WI; and countires of: United Kingdom, Australia, and Switzerland.
  • Product Description
    CIVCO Rectangular Extension, REF MTIL6600, Reusable non-sterile extension for use with Universal Couchtop. The firm name on the label is CIVCO, Orange City, IA. || For use to support and aid in positioning a patient during radiologic and other medical procedures.
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Med Tec Inc, 1401 8th St, PO Box 320, Orange City IA 51041-0320
  • Manufacturer Parent Company (2017)
  • Source
    USFDA