Recall of Device Recall CELLDYN 3000 Hematology Analyzer

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Abbott Laboratories.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    38322
  • Event Risk Class
    Class 2
  • Event Number
    Z-1198-2007
  • Event Initiated Date
    2007-05-11
  • Event Date Posted
    2007-08-29
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2008-06-19
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Hematology Analyzer - Product Code GKZ
  • Reason
    The product failed electrostatic discharge testing, which could result in product failure.
  • Action
    Mandatory Technical Service Bulletins (TSB) are being issued to Field Service to upgrade all instruments in the field. All customers will be contacted by Field Service to schedule this mandatory upgrade. The upgrades are expected to take 12 months.

Device

  • Model / Serial
    Model number CELL-DYN 3000 CS: List Numbers 91323-01 and 91323-03, all serial numbers distributed on or before May 3, 2007. Model number CELL-DYN 3000 SL: List Numbers 91325-01 and 91325-03; all serial numbers distributed on or before May 3, 2007.
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    Worldwide, including USA, Puerto Rico, Canada, Mexico, Argentina, Australia, Bahamas, Brazil, Chile, China, Colombia, Costa Rica, Dominican Republic, El Salvador, Ecuador, Germany, Guatemala, Hong Kong, Jamaica, Japan, New Zealand, Panama, Paraguay, Peru, Singapore, South Korea, Taiwan, Thailand, Uruguay, and Venezuela.
  • Product Description
    Abbott CELL-DYN 3000 Hematology Analyzer