Recall of Device Recall Cardiovive 92532 automated external defibrillator

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Cardiac Science Corporation.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    53827
  • Event Risk Class
    Class 1
  • Event Number
    Z-0773-2010
  • Event Initiated Date
    2009-11-13
  • Event Date Posted
    2010-03-04
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2014-10-24
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    Automated external defibrillators (non-wearable) - Product Code MKJ
  • Reason
    Potential for devices not to deliver therapy.
  • Action
    Cardiac Science issued an "Urgent - Voluntary Medical Device Correction AED Resistor Issue" letter dated November 13, 2009. Consignees were informed of the affected product and instructed on preventive measures. For further information, contact Cardiac Science at 1-877-901-1788 in the US and +44-161-926-0011 for outside the US.

Device

  • Model / Serial
    SERIAL NUMBERS: 4022185, 92532-0001124, 4022096, 4022095, 4022094, 4022093, 4022092, 4022091, 4022089, 4022088, 4022087, 4022086, 4022085, 4022084, 4022233, 4022234, 4022235, 4022236, 4022090, 4022112, 4022126, 4022125, 4022124, 4022123, 4022122, 4022121, 4022120, 4022119, 4022118, 4022117, 4022116, 4022115, 4022099, 4022113, 4022098, 4022111, 4022110, 4022109, 4022108, 4022107, 4022106, 4022105, 4022104, 4022103, 4022102, 4022101, 4022250, 4022100, 4022114, 4022268, 4022248, 4022278, 4022277, 4022276, 4022275, 4022274, 4022273, 4022272, 4022271, 4022280, 4022269, 4022258, 4022267, 4022266, 4022265, 4022264, 4022263, 4022262, 4022261, 4022260, 4022127, 4022203, 4022270, 4022251, 4022238, 4022239, 4022240, 4022241, 4022242, 4022243, 4022244, 4022245, 4022246, 4022279, 4022249, 4022237, 4022252, 4022253, 4022254, 4022255, 4022256, 4022259, 4022283, 4022282, 4022257, 4022281, 4022247, 4022190, 4022206, 4022232, 4022205, 4022204, 4022200, 4022198, 4022197, 4022196, 4022195, 4022128, 4022192, 4022209, 4022189, 4022188, 4022187, 4022184, 4022183, 4022182, 4022181, 4022199, 4022178, 4022194, 4022219, 4022231, 4022230, 4022229, 4022228, 4022227, 4022225, 4022224, 4022223, 4022222, 4022207, 4022220, 4022208, 4022218, 4022217, 4022216, 4022215, 4022214, 4022213, 4022212, 4022211, 4022210, 4022177, 4022221, 4022139, 4022179, 4022151, 4022150, 4022149, 4022147, 4022146, 4022145, 4022144, 4022143, 4022142, 4022152, 4022140, 4022153, 4022138, 4022137, 4022136, 4022135, 4022134, 4022133, 4022132, 4022131, 4022130, 4022129, 4022141, 4022165, 4022175, 4022174, 4022173, 4022172, 4022171, 4022170, 4022169, 4022168, 4022167, 4022166, 4022176, 4022163, 4022162, 4022161, 4022160, 4022159, 4022158, 4022157, 4022156, 4022155, 4022154, 4035467, 4035457, 4035458, 4035459, 4035460, 4035461, 4035462, 4035463, 4035464, 4037389, 4035466, 4035454, 4035468, 4035469, 4035470, 4035471, 4035472, 4035473, 4035474, 4035465, 4035445, 4035435, 4035436, 4035437, 4035438, 4035439, 4035440, 4035441, 4035442, 4035456, 4035444, 4035455, 4035446, 4035447, 4035448, 4035449, 4035450, 4035451, 4035452, 4035453, 4035477, 4035443, 4037400, 4035475, 4037414, 4037413, 4037412, 4037409, 4037407, 4037406, 4037405, 4037416, 4037402, 4037417, 4037399, 4037397, 4037396, 4037395, 4037394, 4037393, 4037392, 4037391, 4035338, 4037403, 4037439, 4035431, 4035478, 4035479, 4035481, 4035482, 4035483, 4037446, 4037445, 4037415, 4037441, 4035476, 4037435, 4037432, 4037426, 4037425, 4037423, 4037422, 4037421, 4037420, 4037419, 4037444, 4035374, 4035361, 4035363, 4035364, 4035365, 4035367, 4035368, 4035370, 4035371, 4035433, 4035373, 4035358, 4035375, 4035376, 4035377, 4035378, 4035379, 4035380, 4035381, 4035382, 4035372, 4035349, 4035339, 4035340, 4035341, 4035342, 4035343, 4035344, 4035345, 4035346, 4035360, 4035348, 4035359, 4035350, 4035351, 4035352, 4035353, 4035354, 4035355, 4035356, 4035357, 4035385, 4035347, 4035422, 4035383, 4035412, 4035413, 4035414, 4035416, 4035417, 4035418, 4035419, 4035410, 4035421, 4035409, 4035423, 4035424, 4035425, 4035426, 4035427, 4035428, 4035429, 4035430, 4037388, 4035420, 4035397, 4035432, 4035386, 4035387, 4035389, 4035391, 4035392, 4035393, 4035394, 4035411, 4035396, 4035384, 4035398, 4035399, 4035400, 4035401, 4035402, 4035403, 4035405, 4035406, 4035408, 4035395, 4037316, 4033025, 4033023, 4033022, 4033021, 4033020, 4033019, 4033018, 4033017, 4033048, 4037317, 4033028, 4037315, 4037314, 4037312, 4037311, 4037308, 4037306, 4037305, 4037304, 4037318, 4033037, 4037390, 4033046, 4033045, 4033044, 4033043, 4033042, 4033041, 4033040, 4033026, 4033038, 4033027, 4033036, 4033035, 4033034, 4033033, 4033032, 4033031, 4033030, 4033029, 4037300, 4033039, 4037260, 4037302, 4037270, 4037269, 4037267, 4037266, 4037265, 4037264, 4037263, 4037272, 4037261, 4037273, 4037258, 4037257, 4037256, 4037255, 4037253, 4037252, 4037251, 4037250, 4037248, 4037262, 4037287, 4033049, 4037298, 4037296, 4037295, 4037294, 4037293, 4037291, 4037290, 4037271, 4037288, 4037301, 4037286, 4037285, 4037284, 4037283, 4037281, 4037280, 4037279, 4037276, 4037275, 4037289, 4037339, 4037354, 4037353, 4037352, 4037351, 4037348, 4037347, 4037343, 4037342, 4033047, 4037340, 4037361, 4037338, 4037337, 4037335, 4037334, 4037332, 4037331, 4037330, 4037327, 4037341, 4037373, 4037387, 4037386, 4037385, 4037384, 4037383, 4037382, 4037380, 4037377, 4037357, 4037374, 4037360, 4037372, 4037370, 4037369, 4037368, 4037367, 4037365, 4037364, 4037362, 4037322, 4037375, 4033059, 4037326, 4033068, 4033067, 4033066, 4033065, 4033064, 4033063, 4033062, 4033070, 4033060, 4033071, 4033058, 4033057, 4033056, 4033055, 4033054, 4033053, 4033052, 4033051, 4033050, 4033061, 4033081, 4035388, 4037321, 4037320, 4037319, 4033087, 4033086, 4033085, 4033084, 4033069, 4033082, 4037324, 4033080, 4033079, 4033078, 4033077, 4033076, 4033075, 4033074, 4033073, 4033072, 4033083, 4035390, 4010619, 4010613, 4010614, 4010615, 4010616, 4010620, 4010618, 4010622, 4010605, 4010621, 4010617, 4010612, 4010610, 4010609, 4010601, 4010606, 4010604, 4010603, 4010602, 4010623, 4010637, 4010600, 4010599, 4010598, 4010607, 4010638, 4010647, 4010648, 4010560, 4010597, 4010646, 4010645, 4010644, 4010643, 4010642, 4010641, 4010634, 4010639, 4010624, 4010636, 4010633, 4010632, 4010631, 4010630, 4010629, 4010628, 4010627, 4010626, 4010625, 4010640, 4018005, 4010564, 4010562, 4010559, 4010558, 4010557, 4018012, 4018011, 4018010, 4018009, 4018008, 4010563, 4018006, 4010567, 4018004, 4018003, 4018002, 4018001, 4018000, 4017999, 4017998, 4017997, 4018072, 4010649, 4018007, 4010580, 4010595, 4010594, 4010593, 4010592, 4010590, 4010589, 4010588, 4010586, 4010584, 4010583, 4010565, 4010581, 4010566, 4010579, 4010578, 4010577, 4010576, 4010575, 4010574, 4010573, 4010571, 4010570, 4010569, 4010596, 4010582, 4018084, 4018101, 4018100, 4018098, 4018097, 4018096, 4018095, 4018094, 4018093, 4018091, 4018090, 4018089, 4018088, 4018087, 4018066, 4018077, 4018068, 4018069, 4018071, 4018073, 4018074, 4018086, 4018076, 4018085, 4018078, 4018079, 4018081, 4018082, 4018083, 4018105, 4018075, 4018131, 4018102, 4018125, 4018126, 4018127, 4018128, 4018122, 4018130, 4018121, 4018132, 4018134, 4018135, 4018136, 4018070, 4010650, 4018129, 4018114, 4018104, 4018107, 4018108, 4018109, 4018110, 4018123, 4018113, 4018103, 4018115, 4018116, 4018117, 4018118, 4018119, 4018120, 4018111, 4018033, 4018040, 4018026, 4018027, 4018028, 4018029, 4018030, 4018023, 4018032, 4018022, 4018034, 4018035, 4018036, 4018037, 4018038, 4018039, 4018031, 4018015, 4010651, 4010652, 4018106, 4010654, 4018065, 4010656, 4018025, 4018014, 4010655, 4018016, 4018017, 4018018, 4018019, 4018020, 4018021, 4018013, 4018051, 4018060, 4018058, 4018057, 4018061, 4018055, 4018054, 4018053, 4018052, 4018059, 4018064, 4018063, 4018050, 4018045, 4018048, 4018042, 4018047, 4018041, 4018046, 4018049, 4010653, 4018043, 4018044, 4001366, 4006659, 4001391, 4006660, 4006661, 4001363, 4006658, 4001365, 4001364, 4000752, 4001367, 4000753, 4001372, 4000754, 4000755, 4000756, 4000757, 4000758, 4000759, 4000776, 4000791, 4006662, 4000751, 4006670, 4006669, 4006668, 4006667, 4006666, 4006665, 4001379, 4001370, 4006663, 4001368, 4001378, 4001377, 4001376, 4001375, 4001374, 4001373, 4000760, 4001371, 4001369, 4006664, 4001430, 4001421, 4001439, 4001438, 4001437, 4001436, 4001435, 4001434, 4001441, 4001431, 4001442, 4001429, 4001428, 4001427, 4001426, 4001425, 4001424, 4001423, 4001389, 4001432, 4001451, 4001459, 4001458, 4001457, 4001456, 4001455, 4001454, 4001453, 4001440, 4000793, 4001420, 4001450, 4001449, 4001448, 4001447, 4001446, 4001445, 4001444, 4001443, 4001452, 4006672, 4001422, 4001397, 4001396, 4001395, 4001394, 4001393, 4001392, 4001400, 4001390, 4001401, 4001388, 4001387, 4001386, 4001385, 4001384, 4001383, 4001382, 4001381, 4006697, 4001410, 4001419, 4001418, 4001417, 4001416, 4001415, 4001414, 4001413, 4001398, 4001411, 4001380, 4001409, 4001408, 4001407, 4001406, 4001405, 4001404, 4001403, 4001402, 4001412, 4002156, 4002240, 4002146, 4002147, 4002148, 4002149, 4002150, 4002151, 4002152, 4002153, 4002144, 4002155, 4002143, 4002157, 4002159, 4002160, 4002161, 4002162, 4002163, 4002164, 4002165, 4002166, 4002167, 4002154, 4002228, 4006694, 4002238, 4002237, 4002235, 4000790, 4002234, 4002233, 4002232, 4002231, 4002145, 4002229, 4002170, 4002227, 4002226, 4002225, 4002224, 4002223, 4002222, 4002221, 4002220, 4002219, 4002218, 4002230, 4002206, 4002195, 4002196, 4002197, 4002198, 4002199, 4002200, 4002201, 4002202, 4002203, 4002168, 4002205, 4002192, 4002207, 4002208, 4002209, 4002210, 4002211, 4002212, 4002213, 4002214, 4002215, 4002216, 4002204, 4002182, 4002241, 4002171, 4002172, 4002173, 4002174, 4002175, 4002176, 4002177, 4002178, 4002179, 4002194, 4002181, 4002193, 4002183, 4002184, 4002185, 4002186, 4002187, 4002188, 4002189, 4002190, 4002191, 4002169, 4002180, 4006706, 4002239, 4006696, 4001484, 4006698, 4006699, 4006700, 4006701, 4006702, 4006703, 4006693, 4006705, 4006692, 4006707, 4006708, 4006709, 4006710, 4006711, 4006712, 4006713, 4006714, 4006715, 4006716, 4006704, 4006681, 4000775, 4000774, 4000773, 4000772, 4006674, 4006675, 4006676, 4006677, 4006678, 4001460, 4006680, 4006718, 4006682, 4006683, 4006684, 4006685, 4006686, 4006687, 4006688, 4000771, 4006689, 4006690, 4006679, 4006754, 4006743, 4006744, 4006745, 4006746, 4006747, 4006748, 4006749, 4006750, 4006751, 4000770, 4006753, 4006740, 4006755, 4006756, 4006757, 4000767, 4000766, 4000765, 4000761, 4000764, 4000763, 4002242, 4006752, 4000768, 4006673, 4006719, 4000769, 4006720, 4006723, 4006724, 4006725, 4006726, 4006727, 4006728, 4006742, 4006730, 4006741, 4006731, 4006732, 4006733, 4006734, 4006735, 4006736, 4006737, 4006738, 4006739, 4006717, 4006729, 4006333, 4006323, 4006345, 4006344, 4006343, 4000783, 4006342, 4006341, 4000782, 4006340, 4006339, 4006338, 4006337, 4006336, 4006347, 4006328, 4000786, 4000792, 4006325, 4001482, 4006326, 4006335, 4006327, 4006334, 4006329, 4006330, 4000788, 4006331, 4006332, 4006348, 4001461, 4000749, 4000828, 4000829, 4000830, 4000831, 4000785, 4000832, 4000833, 4000834, 4000835, 4000836, 4000838, 4006313, 4000840, 4006346, 4006356, 4006349, 4006350, 4006351, 4006352, 4006353, 4000841, 4006355, 4000842, 4006357, 4001344, 4000784, 4001343, 4000750, 4006322, 4006354, 4006299, 4006324, 4006286, 4006287, 4006288, 4006289, 4006290, 4000789, 4006291, 4006292, 4006293, 4006295, 4006296, 4006297, 4006284, 4006305, 4006312, 4006311, 4006310, 4006309, 4006308, 4006298, 4006306, 4000780, 4006304, 4006303, 4006302, 4006301, 4006300, 4006283, 4006307, 4001357, 4006321, 4000778, 4006320, 4006319, 4006318, 4006317, 4006316, 4006315, 4006314, 4000781, 4001362, 4001361, 4001360, 4006285, 4001351, 4001345, 4001346, 4001347, 4000779, 4001348, 4001359, 4001350, 4001358, 4001352, 4001353, 4001354, 4001355, 4001356, 4000837, 4001349, 4001504, 4001530, 4001511, 4001510, 4001509, 4001508, 4001507, 4001513, 4001505, 4001514, 4001503, 4001502, 4001501, 4001500, 4001499, 4001497, 4001506, 4001521, 4001529, 4001528, 4000827, 4001526, 4000839, 4001524, 4001512, 4001522, 4001494, 4001520, 4001519, 4001518, 4001517, 4001516, 4001515, 4001523, 4001468, 4001476, 4001475, 4001474, 4001473, 4001472, 4001471, 4001496, 4001469, 4001479, 4001467, 4001466, 4001465, 4001464, 4001463, 4001462, 4001470, 4001486, 4001527, 4001493, 4001492, 4001491, 4001490, 4001489, 4001477, 4001487, 4001478, 4001485, 4000794, 4001483, 4002217, 4001481, 4001480, 4001495, 4001488, 4000818, 4000804, 4000805, 4000806, 4000807, 4000808, 4000809, 4000811, 4001525, 4000812, 4000813, 4000743, 4000814, 4000815, 4000803, 4000748, 4000826, 4000825, 4000824, 4000823, 4000822, 4000816, 4000820, 4000817, 4000747, 4000746, 4000745, 4000744, 4000819, 4000810, 4000821, 4001539, 4001537, 4001536, 4001535, 4000802, 4001534, 4001533, 4001538, 4001532, 4001531, 4001540, 4001541, 4000798, 4000777, 4000795, 4000801, 4000800, 4000796, 4000799, 4000797, 4001542, 92532-0001014, 92532-0001010, 92532-0001004, 92532-0001011, 92532-0001074, 92532-0001015, 92532-0001008, 92532-0001007, 92532-0001016, 92532-0001012, 92532-0001006, 92532-0001009, 92532-0000988, 92532-0000974, 92532-0000975, 92532-0000976, 92532-0000978, 92532-0000973, 92532-0000979, 92532-0000980, 92532-0000981, 92532-0000982, 92532-0000983, 92532-0000984, 92532-0000985, 92532-0000991, 92532-0000987, 92532-0001003, 92532-0000989, 92532-0000990, 92532-0000992, 92532-0000993, 92532-0000994, 92532-0000995, 92532-0000996, 92532-0000997, 92532-0000998, 92532-0001000, 92532-0001001, 92532-0001002, 92532-0000986, 92532-0001062, 92532-0001050, 92532-0001048, 92532-0001049, 92532-0001051, 92532-0001052, 92532-0001053, 92532-0001054, 92532-0001055, 92532-0001056, 92532-0001057, 92532-0001058, 92532-0001044, 92532-0001061, 92532-0001043, 92532-0001063, 92532-0001064, 92532-0001065, 92532-0001066, 92532-0001067, 92532-0001068, 92532-0001069, 92532-0000972, 92532-0000772, 92532-0001073, 92532-0001072, 92532-0001060, 92532-0001031, 92532-0001018, 92532-0001019, 92532-0001020, 92532-0001021, 92532-0001022, 92532-0001023, 92532-0001024, 92532-0001025, 92532-0001026, 92532-0001027, 92532-0001028, 92532-0001046, 92532-0001030, 92532-0001017, 92532-0001032, 92532-0001033, 92532-0001034, 92532-0001035, 92532-0001036, 92532-0001037, 92532-0001038, 92532-0001039, 92532-0001040, 92532-0001041, 92532-0001042, 92532-0001029, 92532-0000786, 92532-0000769, 92532-0000770, 92532-0000771, 92532-0000773, 92532-0000776, 92532-0000777, 92532-0000778, 92532-0000779, 92532-0000780, 92532-0000781, 92532-0000782, 92532-0000783, 92532-0000774, 92532-0000785, 92532-0000766, 92532-0000787, 92532-0000788, 92532-0000789, 92532-0000790, 92532-0000791, 92532-0000792, 92532-0000793, 92532-0000794, 92532-0000795, 92532-0000796, 92532-0000797, 92532-0000798, 92532-0000800, 92532-0000784, 92532-0000753, 92532-0000738, 92532-0001298, 92532-0001076, 92532-0000739, 92532-0000740, 92532-0000741, 92532-0000742, 92532-0000743, 92532-0000744, 92532-0000746, 92532-0000747, 92532-0000748, 92532-0000749, 92532-0000768, 92532-0000759, 92532-0000903, 92532-0000765, 92532-0000764, 92532-0000763, 92532-0000762, 92532-0000750, 92532-0000760, 92532-0000752, 92532-0000758, 92532-0000757, 92532-0000756, 92532-0000755, 92532-0000754, 92532-0000767, 92532-0000761, 92532-0000957, 92532-0000901, 92532-0000937, 92532-0000938, 92532-0000939, 92532-0000941, 92532-0000942, 92532-0000943, 92532-0000944, 92532-0000946, 92532-0000947, 92532-0000948, 92532-0000951, 92532-0000952, 92532-0000935, 92532-0000963, 92532-0000970, 92532-0000969, 92532-0000968, 92532-0000967, 92532-0000966, 92532-0000953, 92532-0000964, 92532-0000955, 92532-0000962, 92532-0000961, 92532-0000960, 92532-0000959, 92532-0000958, 92532-0000934, 92532-0000965, 92532-0000920, 92532-0000971, 92532-0000904, 92532-0000905, 92532-0000906, 92532-0000907, 92532-0000908, 92532-0000909, 92532-0000910, 92532-0000912, 92532-0000913, 92532-0000914, 92532-0000915, 92532-0000917, 92532-0000936, 92532-0000926, 92532-0000933, 92532-0000932, 92532-0000931, 92532-0000930, 92532-0000929, 92532-0000918, 92532-0000927, 92532-0000919, 92532-0000925, 92532-0000924, 92532-0000923, 92532-0000922, 92532-0000921, 92532-0000902, 92532-0000928, 92532-0001243, 92532-0001198, 92532-0001229, 92532-0001230, 92532-0001232, 92532-0001233, 92532-0001234, 92532-0001235, 92532-0001236, 92532-0001237, 92532-0001238, 92532-0001239, 92532-0001227, 92532-0001242, 92532-0001226, 92532-0001244, 92532-0001245, 92532-0001246, 92532-0001247, 92532-0001248, 92532-0001249, 92532-0001250, 92532-0001251, 92532-0001252, 92532-0001253, 92532-0001254, 92532-0001255, 92532-0001240, 92532-0001213, 92532-0001296, 92532-0001200, 92532-0001201, 92532-0001202, 92532-0001203, 92532-0001204, 92532-0001205, 92532-0001206, 92532-0001208, 92532-0001209, 92532-0001210, 92532-0001228, 92532-0001212, 92532-0001258, 92532-0001214, 92532-0001215, 92532-0001216, 92532-0001217, 92532-0001218, 92532-0001219, 92532-0001220, 92532-0001221, 92532-0001222, 92532-0001223, 92532-0001224, 92532-0001225, 92532-0001211, 92532-0001300, 92532-0001287, 92532-0001288, 92532-0001289, 92532-0001290, 92532-0001291, 92532-0001292, 92532-0001293, 92532-0001294, 92532-0001295, 92532-0000737, 92532-0001297, 92532-0001256, 92532-0001299, 92532-0001284, 92532-0001301, 92532-0001302, 92532-0001303, 92532-0001304, 92532-0001305, 92532-0001306, 92532-0001307, 92532-0001308, 92532-0001309, 92532-0001310, 92533-0000000, 0000775, 92532-0001160, 92532-0001272, 92532-0001197, 92532-0001259, 92532-0001260, 92532-0001261, 92532-0001262, 92532-0001263, 92532-0001264, 92532-0001265, 92532-0001266, 92532-0001267, 92532-0001268, 92532-0001269, 92532-0001286, 92532-0001271, 92532-0001285, 92532-0001273, 92532-0001274, 92532-0001275, 92532-0001276, 92532-0001277, 92532-0001278, 92532-0001279, 92532-0001280, 92532-0001281, 92532-0001282, 92532-0001283, 92532-0001257, 92532-0001270, 92532-0001122, 92532-0001199, 92532-0001110, 92532-0001111, 92532-0001112, 92532-0001113, 92532-0001114, 92532-0001115, 92532-0001116, 92532-0001117, 92532-0001118, 92532-0001119, 92532-0001108, 92532-0001121, 92532-0001107, 92532-0001123, 92532-0001125, 92532-0001126, 92532-0001127, 92532-0001128, 92532-0001129, 92532-0001130, 92532-0001131, 92532-0001132, 92532-0001133, 92532-0001134, 92532-0001135, 92532-0001120, 92532-0001092, 92532-0001078, 92532-0001079, 92532-0001080, 92532-0001081, 92532-0001082, 92532-0001083, 92532-0001084, 92532-0001086, 92532-0001087, 92532-0001088, 92532-0001089, 92532-0001109, 92532-0001091, 92532-0001138, 92532-0001093, 92532-0001094, 92532-0001096, 92532-0001097, 92532-0001098, 92532-0001099, 92532-0001100, 92532-0001101, 92532-0001102, 92532-0001103, 92532-0001105, 92532-0001106, 92532-0001090, 92532-0001183, 92532-0001170, 92532-0001171, 92532-0001172, 92532-0001173, 92532-0001174, 92532-0001175, 92532-0001176, 92532-0001177, 92532-0001178, 92532-0001179, 92532-0001180, 92532-0001136, 92532-0001182, 92532-0001167, 92532-0001184, 92532-0001185, 92532-0001186, 92532-0001187, 92532-0001188, 92532-0001189, 92532-0001190, 92532-0001191, 92532-0001192, 92532-0001194, 92532-0001195, 92532-0001196, 92532-0001181, 92532-0001153, 92532-0001077, 92532-0001139, 92532-0001140, 92532-0001141, 92532-0001142, 92532-0001143, 92532-0001144, 92532-0001145, 92532-0001146, 92532-0001147, 92532-0001148, 92532-0001149, 92532-0001169, 92532-0001151, 92532-0001168, 92532-0001154, 92532-0001155, 92532-0001156, 92532-0001157, 92532-0001158, 92532-0001159, 92532-0001161, 92532-0001163, 92532-0001164, 92532-0001165, 92532-0001166, 92532-0001137, 92532-0001150, 92532-0000665, 92532-0000669, 92532-0000507, 92532-0000551, 92532-0000662, 92532-0000605, 92532-0000606, 92532-0000588, 92532-0000664, 92532-0000631, 92532-0000628, 92532-0000666, 92532-0000586, 92532-0000667, 92532-0000668, 92532-0000544, 92532-0000663, 92532-0000559, 92532-0000699, 92532-0000545, 92532-0000546, 92532-0000547, 92532-0000593, 92532-0000549, 92532-0000630, 92532-0000592, 92532-0000670, 92532-0000590, 92532-0000510, 92532-0000633, 92532-0000632, 92532-0000509, 92532-0000589, 92532-0000550, 92532-0000692, 92532-0000627, 92532-0000682, 92532-0000683, 92532-0000685, 92532-0000686, 92532-0000688, 92532-0000679, 92532-0000691, 92532-0000677, 92532-0000693, 92532-0000694, 92532-0000695, 92532-0000696, 92532-0000697, 92532-0000698, 92532-0000690, 92532-0000608, 92532-0000585, 92532-0000626, 92532-0000671, 92532-0000672, 92532-0000504, 92532-0000607, 92532-0000681, 92532-0000625, 92532-0000594, 92532-0000674, 92532-0000655, 92532-0000675, 92532-0000603, 92532-0000624, 92532-0000676, 92532-0000673, 92532-0000503, 92532-0000521, 92532-0000642, 92532-0000511, 92532-0000513, 92532-0000502, 92532-0000560, 92532-0000644, 92532-0000515, 92532-0000645, 92532-0000516, 92532-0001162, 92532-0000641, 92532-0000517, 92532-0000518, 92532-0000634, 92532-0000514, 92532-0000661, 92532-0000654, 92532-0000653, 92532-0000652, 92532-0000656, 92532-0000657, 92532-0000658, 92532-0000643, 92532-0000660, 92532-0000523.92532-0000651, 92532-0000650, 92532-0000649, 92532-0000648, 92532-0000647, 92532-0000646, 92532-0000659, 92532-0000542, 92532-0000519, 92532-0000637, 92532-0000537, 92532-0000636, 92532-0000538, 92532-0000601, 92532-0000602, 92532-0000541, 92532-0000535, 92532-0000635, 92532-0000599, 92532-0000598, 92532-0000597, 92532-0000596, 92532-0000543, 92532-0000540, 92532-0000528, 92532-0000640, 92532-0000524, 92532-0000525, 92532-0000561, 92532-0000527, 92532-0000639, 92532-0000536, 92532-0000638, 92532-0000584, 92532-0000529, 92532-0000530, 92532-0000531, 92532-0000532, 92532-0000533, 92532-0000534, 92532-0000604, 92532-0000579, 92532-0000712, 92532-0000614, 92532-0000715, 92532-0000505, 92532-0000621, 92532-0000570, 92532-0000557, 92532-0000571, 92532-0000714, 92532-0000572, 92532-0000734, 92532-0000735, 92532-0000556, 92532-0000718, 92532-0000620, 92532-0000578, 92532-0000555, 92532-0000700, 92532-0000733, 92532-0000617, 92532-0000716, 92532-0000717, 92532-0000581, 92532-0000577, 92532-0000618, 92532-0000616, 92532-0000600, 92532-0000619, 92532-0000713, 92532-0000612, 92532-0000615, 92532-0000553, 92532-0000576, 92532-0000558, 92532-0000573, 92532-0000611, 92532-0000609, 92532-0000730, 92532-0000726, 92532-0000711, 92532-0000554, 92532-0000552, 92532-0000728, 92532-0000729, 92532-0000706, 92532-0000708, 92532-0000704, 92532-0000727, 92532-0000731, 92532-0000563, 92532-0000506, 92532-0000703, 92532-0000623, 92532-0000562, 92532-0000701, 92532-0000582, 92532-0000705, 92532-0000736, 92532-0000720, 92532-0000707, 92532-0000725, 92532-0000710, 92532-0000568, 92532-0000564, 92532-0000567, 92532-0000569, 92532-0000719, 92532-0000721, 92532-0000722, 92532-0000709, 92532-0000565, 4035366, 92532-0001085, 92532-0000501, 92532-0000522, 92532-0001071, 92532-0001193, 92532-0001047, 92532-0000591, 92532-0000580, 92532-0000526, 92532-0000610, 92532-0000723, 92532-0000911, 92532-0001095, 92532-0000689, 92532-0000745, 92532-0000583, 92532-0000956, 92532-0000678, 92532-0000916, 92532-0000999, 92532-0000954, 92532-0000751, 92532-0000680, 92532-0000950, 92532-0000949, 92532-0001013, 92532-0000945, 92532-0000940, 92532-0000539, 92532-0001070, 92533-0001193, 92532-0000520, 92532-0000575, 92532-0000629, 92532-0000613, 92532-0000622, 92532-0001005, 92532-0000732, 92532-0000684, 92532-0000702, 376219, 4111502, 4111491, 4111492, 4111493, 4111494, 4111495, 4111497, 4111498, 4111499, 4111469, 4111501, 4111488, 4111503, 4111504, 4111505, 4111507, 4111508, 4111509, 4111510, 4111500, 4111479, 4111470, 4111471, 4111472, 4111473, 4111474, 4111475, 4111476, 4111490, 4111478, 4111489, 4111480, 4111481, 4111482, 4111483, 4111484, 4111485, 4111487, 4111513, 4111477, 4111545, 4111536, 4111537, 4111538, 4111539, 4111540, 4111541, 4111542, 4111511, 4111544, 4111533, 4111546, 4111547, 4111548, 4111549, 4111551, 4111468, 4111553, 4111552, 4111543, 4111525, 4111516, 4111514, 4111515, 4111517, 4111519, 4111520, 4111521, 4111522, 4111535, 4111524, 4111534, 4111526, 4111527, 4111528, 4111529, 4111530, 4111531, 4111532, 4111512, 4111523, 4111576, 4111564, 4111566, 4111568, 4111569, 4111570, 4111571, 4111572, 4111573, 4111586, 4111575, 4111561, 4111577, 4111578, 4111579, 4111580, 4111581, 4111582, 4111583, 4111584, 4111585, 4111574, 4111456, 4111466, 4111467, 4111465, 4111464, 4111463, 4111462, 4111461, 4111460, 4111459, 4111563, 4111457, 4111562, 4111455, 4111554, 4111555, 4111556, 4111557, 4111558, 4111559, 4111560, 4111567, 4111458, 4111598, 4111565, 4111604, 4111603, 4111602, 4111601, 4111599, 4111597, 4111596, 4111595, 4111589, 4111600, 4111594, 4111588, 4111587, 4111590, 4111591, 4111592, 4111593, 4037313, 4037358, 4037356, 4037355, 4037350, 4037349, 4037346, 4037344, 4037336, 4037333, 4037329, 4037328, 4037249, 4037323, 4037366, 4037310, 4037309, 4037297, 4037292, 4037282, 4037278, 4037277, 4037274, 4037259, 4037325, 4037427, 4035404, 4037447, 4037442, 4037438, 4037437, 4037436, 4037434, 4037433, 4037431, 4037430, 4037359, 4037428, 4037363, 4037411, 4037410, 4037408, 4037404, 4037401, 4037398, 4037381, 4037376, 4037371, 4037345, 4037429, 4044122, 4046399, 4044118, 4044126, 4046394, 4046392, 4046396, 4046390, 4044109, 4046397, 4044120, 4044127, 4046398, 4044125, 4044124, 4044123, 4046395, 4044121, 4044119, 4044128, 4044105, 4044092, 4044093, 4044094, 4044095, 4044096, 4044097, 4044098, 4044099, 4044100, 4044101, 4044102, 4044111, 4044104, 4044117, 4044106, 4044107, 4044108, 4046408, 4044110, 4046401, 4044112, 4044113, 4044114, 4044115, 4044116, 4044103, 4046466, 4046406, 4046476, 4046475, 4046473, 4046472, 4046471, 4046470, 4046478, 4046467, 4046479, 4046465, 4046463, 4046462, 4046461, 4046458, 4046455, 4046454, 4046468, 4044129, 4044138, 4044137, 4044136, 4044135, 4044134, 4044133, 4044132, 4046477, 4044130, 4046451, 4046488, 4046487, 4046486, 4046485, 4046483, 4046482, 4046481, 4044131, 4046411, 4046421, 4046419, 4046418, 4046417, 4046416, 4046415, 4046414, 4046453, 4046412, 4046424, 4046410, 4046409, 4046456, 4046407, 4046405, 4046404, 4046403, 4046413, 4046436, 4046402, 4046450, 4046449, 4046447, 4046446, 4046443, 4046442, 4046422, 4046438, 4046423, 4046435, 4046433, 4046432, 4046431, 4046429, 4046427, 4046425, 4046452, 4046441, 4044050, 4044063, 4044062, 4044061, 4044060, 4044059, 4044058, 4044057, 4044056, 4044055, 4044054, 4044053, 4044064, 4044051, 4044047, 4044049, 4044048, 4044046, 4044044, 4044043, 4044042, 4044041, 4044040, 4044039, 4046459, 4044091, 4044052, 4044081, 4044045, 4044065, 4044090, 4044089, 4044088, 4044087, 4044086, 4044085, 4044084, 4044083, 4044082, 4044080, 4044079, 4044072, 4044066, 4044067, 4044068, 4044069, 4044071, 4044078, 4044073, 4044074, 4044075, 4044076, 4044077, 4044070, 4138842, 4138843, 4138844, 4138845, 4138846, 4138847, 4138849, 4138848, 4138850, 4138841, 4138832, 4138851, 4138840, 4138839, 4138838, 4138837, 4138836, 4138835, 4138833, 4138831, 4138872, 4138829, 4138830, 4138834, 4138864, 4138828, 4138803, 4138876, 4138875, 4138874, 4138873, 4138871, 4138869, 4138868, 4138867, 4138870, 4138865, 4138852, 4138863, 4138862, 4138861, 4138860, 4138859, 4138858, 4138857, 4138856, 4138855, 4138854, 4138853, 4138866, 4138786, 4138799, 4138798, 4138797, 4138796, 4138795, 4138794, 4138793, 4138792, 4138778, 4138790, 4138800, 4138787, 4138791, 4138785, 4138784, 4138783, 4138782, 4138781, 4138780, 4138779, 4138805, 4138827, 4138777, 4138788, 4138818, 4138825, 4138824, 4138789, 4138801, 4138823, 4138822, 4138821, 4138819, 4138817, 4138816, 4138815, 4138814, 4138809, 4138804, 4138820, 4138826, 4138802, 4138808, 4138806, 4138810, 4138811, 4138812, 4138813, 4138807, 4147706, 4147707, 4147705, 4147694, 4147704, 4147703, 4147708, 4147702, 4147701, 4147700, 4147699, 4147698, 4147697, 4147709, 4147695, 4147719, 4147693, 4147692, 4147691, 4147696, 4147723, 4147736, 4147735, 4147734, 4147733, 4147732, 4147731, 4147730, 4147729, 4147728, 4147727, 4147726, 4147717, 4147724, 4147710, 4147722, 4147721, 4147720, 4146652, 4147718, 4147690, 4147716, 4147715, 4147714, 4147713, 4147712, 4147711, 4147725, 4146619, 4146654, 4146630, 4146629, 4146628, 4146627, 4146626, 4146625, 4146624, 4146623, 4146622, 4146632, 4146620, 4146633, 4146618, 4146617, 4146616, 4146615, 4146614, 4146613, 4146642, 4146643, 4147757, 4147737, 4146621, 4146646, 4147688, 4147687, 4146658, 4146657, 4146656, 4146655, 4146653, 4146651, 4146650, 4146649, 4146631, 4146647, 4147689, 4146645, 4146644, 4146641, 4146640, 4146639, 4146638, 4146637, 4146636, 4146635, 4146634, 4146648, 4147666, 4147672, 4147781, 4147780, 4147779, 4147778, 4147777, 4147663, 4147783, 4147665, 4147784, 4147667, 4147738, 4147668, 4147755, 4147669, 4147670, 4147771, 4147664, 4147792, 4146692, 4146693, 4146694, 4146695, 4147793, 4146696, 4146697, 4147782, 4146699, 4146700, 4147791, 4147790, 4147789, 4147788, 4147787, 4147786, 4147785, 4146698, 4147774, 4147671, 4147681, 4147682, 4147683, 4147684, 4147685, 4147686, 4147679, 4147775, 4147678, 4147773, 4147772, 4147766, 4147767, 4147768, 4147769, 4147770, 4147776, 4146708, 4147673, 4147674, 4146701.4146702, 4146703, 4146704, 4146705, 4147680, 4146707, 4146689, 4146709, 4146710, 4146711, 4146712, 4147675, 4147676, 4147677, 4146706, 4146662, 4146691, 4147761, 4147762, 4147763, 4147764, 4147765, 4146659, 4147759, 4146661, 4147758, 4146663, 4146664, 4146665, 4146666, 4146667, 4146668, 4146669, 4146660, 4147748, 4147739, 4147740, 4147741, 4147742, 4147743, 4147744, 4147745, 4147760, 4147747, 4146672, 4147749, 4147750, 4147751, 4147752, 4147753, 4147754, 4147756, 4147746, 4147798, 4147806, 4147805, 4147804, 4147803, 4147802, 4147801, 4147808, 4147799, 4147809, 4147797, 4147796, 4147795, 4147794, 4146686, 4146687, 4146688, 4147800, 4146681, 4146690, 4146673, 4146674, 4146675, 4146676, 4146677, 4146678, 4147807, 4146680, 4146671, 4146682, 4146683, 4146684, 4146685, 4147812, 4147811, 4147810, 4146679, 4155336, 4155331, 4155332, 4155333, 4155334, 4155305, 4155335, 4155310, 4155307, 4155339, 4155308, 4155309, 4155330, 4155322, 4155340, 4155337, 4155338, 4155306, 4155319, 4155341, 4155379, 4155311, 4155312, 4155314, 4155315, 4155316, 4155324, 4155318, 4155329, 4155320, 4155321, 4155323, 4155313, 4155325, 4155326, 4155327, 4155328, 4155317, 4155380, 4155367, 4155368, 4155369, 4155370, 4155371, 4155372, 4155373, 4155374, 4155375, 4155377, 4155378, 4155364, 4155381, 4155382, 4155383, 4155387, 4155386, 4155304, 4155385, 4155225, 4155384, 4155376, 4155354, 4155343, 4155344, 4155345, 4155346, 4155347, 4155348, 4155349, 4155350, 4155351, 4155366, 4155353, 4155365, 4155355, 4155356, 4155357, 4155358, 4155359, 4155360, 4155361, 4155362, 4155363, 4155342, 4155352, 4155248, 4155227, 4155239, 4155240, 4155241, 4155242, 4155243, 4155244, 4155245, 4155237, 4155247, 4155236, 4155249, 4155250, 4155251, 4155252, 4155253, 4155254, 4155255, 4155256, 4155257, 4155246, 4155226, 4155216, 4155217, 4155218, 4155219, 4155220, 4155221, 4155222, 4155223, 4155238, 4155469, 4155260, 4155388, 4155228, 4155229, 4155230, 4155231, 4155232, 4155233, 4155234, 4155235, 4155224, 4155293, 4155283, 4155284, 4155285, 4155286, 4155287, 4155288, 4155289, 4155290, 4155258, 4155292, 4155280, 4155294, 4155295, 4155296, 4155297, 4155298, 4155299, 4155300, 4155301, 4155302, 4155291, 4155271, 4155303, 4155261, 4155262, 4155263, 4155264, 4155265, 4155266, 4155267, 4155268, 4155282, 4155270, 4155281, 4155272, 4155273, 4155274, 4155275, 4155276, 4155277, 4155278, 4155279, 4155259, 4155269, 4155513, 4155503, 4155504, 4155505, 4155506, 4155507, 4155508, 4155509, 4155510, 4155480, 4155512, 4155500, 4155564, 4155514, 4155515, 4155516, 4155517, 4155518, 4155519, 4155520, 4155511, 4155491, 4155481, 4155482, 4155483, 4155484, 4155389, 4155486, 4155467, 4155488, 4155502, 4155490, 4155501, 4155492, 4155493, 4155494, 4155495, 4155496, 4155497, 4155498, 4155499, 4155523, 4155489, 4155557, 4155521, 4155546, 4155563, 4155562, 4155565, 4155561, 4155547, 4155560, 4155544, 4155558, 4155543, 4155556, 4155555, 4155554, 4155553, 4155548, 4155549, 4155550, 4155551, 4155552, 4155559, 4155533, 4155485, 4155524, 4155525, 4155526, 4155527, 4155528, 4155529, 4155530, 4155545, 4155532, 4155522, 4155534, 4155535, 4155536, 4155537, 4155538, 4155539, 4155540, 4155541, 4155542, 4155531, 4155422, 4155432, 4155413, 4155414, 4155415, 4155416, 4155417, 4155418, 4155419, 4155411, 4155421, 4155410, 4155423, 4155487, 4155425, 4155479, 4155427, 4155428, 4155429, 4155430, 4155431, 4155420, 4155400, 4155390, 4155391, 4155392, 4155393, 4155394, 4155395, 4155396, 4155397, 4155412, 4155399, 4155426, 4155401, 4155402, 4155403, 4155404, 4155405, 4155406, 4155407, 4155408, 4155409, 4155398, 4155468, 4155458, 4155459, 4155460, 4155461, 4155433, 4155463, 4155424, 4155464, 4155457, 4155466, 4155462, 4155470, 4155471, 4155472, 4155473, 4155474, 4155475, 4155476, 4155477, 4155478, 4155465, 4155440, 4155434, 4155435, 4155436, 4155439, 4155456, 4155441, 4155442, 4155443, 4155444, 4155445, 4155452, 4155437, 4155446, 4155454, 4155453, 4155455, 4155451, 4155450, 4155449, 4155448, 4155447, 4171103, 4171102, 4171119, 4171113, 4171123, 4171122, 4171121, 4171120, 4171087, 4171118, 4171116, 4171117, 4171114, 4171104, 4171112, 4171111, 4171110, 4171108, 4171107, 4171106, 4171105, 4171115, 4171071, 4171089, 4171079, 4171078, 4171077, 4171076, 4171075, 4171074, 4171081, 4171072, 4171082, 4171070, 4171069, 4171068, 4171067, 4171066, 4171124, 4171163, 4171073, 4171091, 4171100, 4171099, 4171098, 4171097, 4171096, 4171095, 4171094, 4171080, 4171092, 4171101, 4171090, 4171065, 4171088, 4171086, 4171085, 4171084, 4171083, 4171093, 4171171, 4171162, 4171179, 4171178, 4171177, 4171176, 4171175, 4171174, 4171181, 4171172, 4171182, 4171170, 4171169, 4171168, 4171167, 4171166, 4171165, 4171161, 4171173, 4171190, 4171200, 4171199, 4171197, 4171196, 4171195, 4171194, 4171193, 4171180, 4171191, 4171160, 4171189, 4171188, 4171187, 4171186, 4171185, 4171184, 4171183, 4171192, 4171133, 4171164, 4171141, 4171140, 4171139, 4171138, 4171137, 4171136, 4171143, 4171134, 4171144, 4171132, 4171131, 4171130, 4171129, 4171128, 4171127, 4171126, 4171135, 4171152, 4171159, 4171158, 4171157, 4171156, 4171064, 4171155, 4171198, 4171142, 4171153, 4171125, 4171151, 4171150, 4171149, 4171148, 4171147, 4171146, 4171145, 4171154, 4170989, 4170982, 4170983, 4170984, 4170985, 4170986, 4170965, 4170988, 4170980, 4170990, 4170992, 4170993, 4170994, 4170995, 4170996, 4170987, 4170974, 4170966, 4170967, 4170968, 4170969, 4170970, 4170971, 4171063, 4170973, 4170981, 4170975, 4170976, 4170977, 4170978, 4170979, 4171000, 4170972, 4170931, 4170997, 4171020, 4171021, 4171022, 4171023, 4171024, 4171018, 4171026, 4171017, 4170930, 4170929, 4170928, 4170927, 4170926, 4170933, 4171025, 4171008, 4171010, 4171001, 4171002, 4171003, 4171004, 4171005, 4171019, 4171007, 4170999, 4171009, 4171011, 4171013, 4171014, 4171015, 4171016, 4171006, 4171037, 4171030, 4171044, 4171043, 4171042, 4171041, 4171040, 4171046, 4171038, 4171047, 4171036, 4171035, 4171034, 4171033, 4171032, 4171031, 4171039, 4171056, 4171062, 4171012, 4170964, 4171061, 4171060, 4171059, 4171045, 4171057, 4171049, 4171055, 4171054, 4171053, 4171052, 4171050, 4171048, 4171058, 4170958, 4170948, 4170949, 4170950, 4170951, 4170952, 4170953, 4170947, 4170957, 4170955, 4170959, 4170960, 4171029, 4171051, 4170961, 4170962, 4170963, 4170956, 4170934, 4170936, 4171028, 4170935, 4170946, 4170937, 4170939, 4171027, 4170940, 4170945, 4170944, 4170942, 4170943, 4170941, 4170938, 4130935, 4130934, 4130933, 4130932, 4130931, 4130930, 4130936, 4130928, 4130940, 4130927, 4130929, 4130948, 4130937, 4130944, 4130939, 4130942, 4130947, 4130943, 4130946, 4130945, 4130926, 4130938, 4130913, 4130907, 4122652, 4131000, 4130949, 4122651, 4122649, 4122648, 4130908, 4130909, 4130910, 4122650, 4130912, 4130925, 4130914, 4130915, 4130916, 4130917, 4130918, 4130919, 4130920, 4130921, 4130922, 4130923, 4130924, 4130911, 4130992, 4130979, 4130980, 4130981, 4130982, 4130983, 4130984, 4130985, 4130987, 4130988, 4130989, 4130998, 4130991, 4130976, 4130993, 4130994, 4130995, 4130996, 4130997, 4130999, 4131001, 4131002, 4122653, 4130902, 4130990, 4130963, 4130951, 4130952, 4130953, 4130954, 4130955, 4130956, 4130957, 4130958, 4130959, 4130960, 4130978, 4130962, 4130977, 4130964, 4130965, 4130966, 4130967, 4130968, 4130969, 4130972, 4130973, 4130974, 4130975, 4130950, 4130961, 4122715, 4122727, 4122726, 4122725, 4122724, 4122723, 4122722, 4122721, 4122720, 4122719, 4122718, 4130904, 4122716, 4122730, 4122714, 4122713, 4122712, 4122711, 4122710, 4122709, 4122708, 4122707, 4122706, 4122746, 4122717, 4122741, 4130897, 4130898, 4130899, 4130900, 4130901, 4122647, 4130903, 4131003, 4122745, 4122744, 4122728, 4122742, 4122729, 4122740, 4122739, 4122738, 4122737, 4122736, 4122735, 4122734, 4122733, 4122732, 4122731, 4122704, 4122743, 4122664, 4122677, 4122676, 4122675, 4122674, 4122673, 4122672, 4122671, 4122669, 4122668, 4122667, 4122705, 4122665, 4122680, 4122663, 4122662, 4122661, 4122660, 4122659, 4122658, 4122657, 4122656, 4122655, 4130906, 4122666, 4122691, 4122703, 4122702, 4122701, 4122700, 4122699, 4122698, 4122697, 4122696, 4122695, 4122694, 4122678, 4122692, 4122679, 4122690, 4122689, 4122688, 4122687, 4122686, 4122685, 4130905, 4122683, 4122682, 4122681, 4122654, 4122693, 4131040, 4131024, 4131025, 4131026, 4131027, 4131028, 4131029, 4131031, 4131032, 4131033, 4131023, 4131039, 4131030, 4131041, 4131042, 4131044, 4131045, 4131046, 4131037, 4131036, 4131038, 4131035, 4131034, 4131004, 4131006, 4131007, 4131009, 4131022, 4131010, 4131005, 4131011, 4131012, 4131013, 4131014, 4131019, 4131008, 4131015, 4131021, 4131020, 4131018, 4131017, 4131016, 4105913, 4105909, 4105910, 4105911, 4105935, 4105912, 4105908, 4105928, 4105942, 4105941, 4105940, 4105939, 4105938, 4105937, 4105936, 4107838, 4105934, 4105907, 4105932, 4105931, 4105933, 4105929, 4105914, 4105927, 4105926, 4105925, 4105924, 4105923, 4105922, 4105921, 4105920, 4105919, 4105918, 4105917, 4105916, 4105915, 4105930, 4107819, 4107840, 4107827, 4107826, 4107825, 4107824, 4107823, 4107822, 4107829, 4107820, 4107830, 4107818, 4107817, 4107806, 4105943, 4106001, 4107851, 4107807, 4107821, 4107839, 4105905, 4107850, 4107846, 4107845, 4107805, 4107844, 4107843, 4107828, 4107841, 4105906, 4107837, 4107836, 4107835, 4107834, 4107833, 4107832, 4107831, 4107842, 4107895, 4107881, 4107882, 4107883, 4107884, 4107885, 4107849, 4107888, 4107890, 4107892, 4105999, 4107894, 4107878, 4106004, 4106003, 4106002, 4106000, 4105998, 4105997, 4105996, 4105995, 4105994, 4107893, 4107868, 4107889, 4107852, 4107860, 4107848, 4107861, 4107862, 4107863, 4107864, 4107865, 4107880, 4107867, 4107879, 4107869, 4107870, 4107871, 4107872, 4107873, 4107874, 4107875, 4107876, 4107877, 4105991, 4107866, 4105955, 4105993, 4105965, 4105964, 4105963, 4105962, 4105961, 4105960, 4105959, 4105958, 4105967, 4105956, 4105968, 4105954, 4105953, 4105952, 4105951, 4105950, 4105949, 4105948, 4105947, 4105946, 4105945, 4105957, 4105979, 4105944, 4105990, 4105989, 4105988, 4105987, 4105986, 4105985, 4105984, 4105983, 4105966, 4105981, 4105992, 4105978, 4105977, 4105976, 4105975, 4105974, 4105973, 4105972, 4105971, 4105970, 4105969, 4105982, 4104102, 4104113, 4104112, 4104111, 4104110, 4104109, 4104108, 4104107, 4104106, 4104105, 4104138, 4104103, 4104116, 4104100, 4104097, 4104096, 4104094, 4104093, 4104092, 4104091, 4104090, 4104089, 4104104, 4104126, 4104137, 4104136, 4104135, 4104134, 4104133, 4104132, 4104131, 4104130, 4104129, 4104114, 4104127, 4104115, 4104125, 4104124, 4104123, 4104122, 4104121, 4104120, 4104119, 4104118, 4104117, 4104086, 4104128, 4104047, 4104088, 4104058, 4104057, 4104056, 4104055, 4104054, 4104053, 4104052, 4104051, 4104060, 4104048, 4104061, 4104046, 4104045, 4104044, 4104043, 4104042, 4104041, 4104040, 4104039, 4107891, 4107853, 4104050, 4104072, 4104101, 4104084, 4104082, 4104081, 4104080, 4104079, 4104078, 4104076, 4104075, 4104059, 4104073, 4104087, 4104071, 4104070, 4104069, 4104068, 4104067, 4104066, 4104065, 4104064, 4104063, 4104062, 4104074, 4107854, 4107855, 4107856, 4107857, 4107858, 4107859, 4104098, 4114939, 4114982, 4114971, 4114995, 4114994, 4114993, 4114991, 4114990, 4114989, 4114986, 4114985, 4114997, 4114983, 4114998, 4114981, 4114980, 4114979, 4114978, 4114977, 4114976, 4114975, 4114974, 4114973, 4115021, 4114984, 4115009, 4115020, 4115019, 4115018, 4115017, 4115016, 4115015, 4115014, 4115013, 4115012, 4114996, 4115010, 4114970, 4115008, 4115007, 4115006, 4115005, 4115004, 4115003, 4115002, 4115001, 4115000, 4114999, 4115011, 4114928, 4114972, 4114940, 4114938, 4114937, 4114936, 4114934, 4114933, 4114932, 4114931, 4114944, 4114929, 4114945, 4114927, 4114926, 4114925, 4114924, 4114923, 4114922, 4114921, 4114920, 4115119, 4114942, 4114930, 4114958, 4114969, 4114968, 4114967, 4114966, 4114965, 4114964, 4114963, 4114962, 4114961, 4114943, 4114959, 4114949, 4114957, 4114956, 4114955, 4114954, 4114953, 4114952, 4114951, 4114950, 4114948, 4114946, 4114960, 4115082, 4115093, 4115092, 4115091, 4115090, 4115089, 4115088, 4115087, 4115086, 4115085, 4115071, 4115083, 4115096, 4115081, 4115080, 4115079, 4115078, 4115077, 4115076, 4115075, 4115074, 4115073, 4115072, 4115084, 4115107, 4115117, 4114947, 4115022, 4115116, 4115115, 4115114, 4115113, 4115112, 4115111, 4115110, 4115094, 4115108, 4115095, 4115106, 4115105, 4115104, 4115103, 4115102, 4115101, 4115100, 4115099, 4115098, 4115097, 4115109, 4115036, 4115050, 4115049, 4115048, 4115023, 4115047, 4115046, 4115045, 4115044, 4115042, 4115040, 4115039, 4115051, 4115037, 4115043, 4115035, 4115034, 4115033, 4115032, 4115031, 4115030, 4115029, 4115028, 4115027, 4115026, 4115070, 4115025, 4115038, 4115061, 4115053, 4115054, 4115055, 4115056, 4115058, 4115052, 4115060, 4115062, 4115063, 4115041, 4115066, 4115067, 4115068, 4115069, 4115059, 4115065, 4111518, 4111496, 4111506, 4111486, 4111550, 4022097, 4022226, 4022180, 4022191, 4035415, 4037307, 4035362, 4010572, 4010591, 4010587, 4018133, 4018067, 4010585, 4018062, 4018056, 4018099, 4018124, 4010561, 4018112, 4018024, 4018080, 4010568, 4010635, 4001399, 4001498, 4001433, 4006695, 4000787, 4002236, 4006721, 4006671, 4046440, 4046439, 4046389, 4046445, 4046400, 4046469, 4046448, 4046480, 4046474, 4046464, 4146670, 4170991, 4170954, 4131043, 4130986, 4122684, 4122670, 4130970, 92532-0001152, 92532-0001231, 92532-0001104, 4105980, 4104049, 4115118, 4114987, 4115024, 4114988, 4115057, 4114935, 4107901, 4107896, 4107897, 4107898, 4107887, 4107900, 4107902, 4107903, 4107899, 4107886, 4107847, 4107904, 4107815, 4107814, 4107813, 4107812, 4107811, 4107810, 4107809, 4107808 and 4107816.
  • Product Classification
  • Device Class
    3
  • Implanted device?
    No
  • Distribution
    Worldwide Distribution -- UNITED STATES, ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BAHAMAS, BAHRAIN, BELARUS, BELGIUM, BERMUDA, BOLIVIA, BRAZIL, BULGARIA, CANADA, CAYMAN ISLANDS, CHILE, CHINA, COLOMBIA, COSTA RICA, CROATIA, CYPRUS, CZECH REPUBLIC, CZECHOSLOVAKIA, DENMARK, DOMINICAN REPUBLIC, ECUADOR, EGYPT, EL SALVADOR, FINLAND, FRANCE, GERMANY, GRECE, GUAM, GUATEMALA, HOLLAND, HONDURAS, HONG KONG, HUNGARY, ICELAND, INDIA, INDONESIA, IRAQ, IRELAND, ISRAEL, ITALY, JAMAICA, JAPAN, JORDAN, KENYA, KOREA, KUWAIT, LEBANON, LIBYA, LITHUANIA, MALAYSIA, MALTA, MEXICO, MONACO, MORROCCO, NAMIBIA, NETHERLANDS, NEW CALEDONIA, NEW ZEALAND, NIGERIA, NORWAY, OMAN, PAKISTAN, PANAMA, PARAGUAY, PERU, PHILIPPINES, POLAND, PORTUGAL, PUERTO RICO, QATAR, ROMANIA, RUSSIA, SAUDI ARABIA, SCOTLAND, SERBIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TRINIDAD & TOBAGO, TURKEY, UAE, UKRAINE, UNITED KINGDOM, URUGUAY, VENEZUELA, VIETNAM, VIRGIN ISLANDS AND YEMEN.
  • Product Description
    Burdick Cardiovive 92532 automated external defibrillator. This semi-automatic model requires the user to press a shock button in order for the device to deliver defibrillation energy (when appropriate). || Intended for use in either in-hospital or out-of-hospital settings for emergency treatment of victims exhibiting symptoms of sudden cardiac arrest who are unresponsive and not breathing.
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Cardiac Science Corporation, 3303 Monte Villa Pkwy, Bothell WA 98021-8969
  • Manufacturer Parent Company (2017)
  • Source
    USFDA