Recall of Device Recall CADD Prizm

According to U.S. Food and Drug Administration, this recall involved a device in United States that was produced by Smiths Medical MD, Inc..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    37486
  • Event Risk Class
    Class 3
  • Event Number
    Z-0718-2007
  • Event Initiated Date
    2007-01-19
  • Event Date Posted
    2007-04-05
  • Event Status
    Terminated
  • Event Country
  • Event Terminated Date
    2007-05-25
  • Event Source
    USFDA
  • Event Source URL
  • Notes / Alerts
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Extra notes in the data
    ambulatory infusion pump - Product Code FRN
  • Reason
    The year of manufacture had been incorrectly entered as 1980 into the pumps software during the manufacture process. as a result the pump may go into an immediate 'clock battery is low/service immediately' or 'clock battery needs service soon' alarm message indicating that the pump requires servicing.
  • Action
    All consignees who have received the affected CADD-Prizm PCS II ambulatory infusion pumps and the CADD-Prizm VIP System ambulatory infusion pumps were contacted via telephone by February 1, 2007. The problem will be explained to the consignee and they will be asked to return the pumps to Smiths Medical MD in order for the problem to be corrected.

Device

  • Model / Serial
    Serial numbers affected:  691099, 691108, 691117, 691700, 691704, 691713, 691719, 691724, 691729, 691731, 691734, 691737, 691739, 691744, 691748, 691756, 691761, 691764, 691769, 691772, 691776, 691780, 691785, 691788, 691793, 691794, 691798.
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    Nationwide.
  • Product Description
    Deltec branded CADD-Prizm PCS II ambulatory infusion pump and the CADD-Prizm VIP system ambulatory infusion pumps; (Both) Model 6101; Order #21-8861-01 (PCS II) and 21-8821-01 (VIP). Smiths Medical MD, Inc., St. Paul, MN 55112 USA; RX ONLY.
  • Manufacturer

Manufacturer

  • Manufacturer Address
    Smiths Medical MD, Inc., 1265 Grey Fox Rd, Saint Paul MN 55112-6929
  • Source
    USFDA